← Product Code [KTT](/productcode/KTT) · K052918

# REPROCESSED EXTERNAL FIXATION DEVICES (K052918)

_Alliance Medical Corp. · KTT · Jan 27, 2006 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K052918

## Device Facts

- **Applicant:** Alliance Medical Corp.
- **Product Code:** [KTT](/productcode/KTT.md)
- **Decision Date:** Jan 27, 2006
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3030
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

External Fixation Devices are intended to be used for the fixation of supracondylar, or condylar fractures of the femur; for fusion of a joint; for surgical procedures that involve cutting the bone, for fixation of bone fractures; bone reconstruction; as a guide pin for insertion of other implants; or may be implanted through the skin so that a pulling force or traction may be applied to the skeletal system; and others may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that pulling force (traction) may be applied to the skeletal system.

## Device Story

Reprocessed external fixation devices consist of frames, clamps, rods, couplings, pins, bolts, and hinges used for orthopedic fracture management and reconstruction. Devices are reprocessed from previously used components to restore functionality. Used by surgeons in clinical/OR settings to stabilize bone segments, provide traction, or facilitate bone healing. Mechanical assembly provides rigidity and stability to skeletal system. Benefits include stabilization of fractures and support for reconstructive procedures. Output is physical skeletal fixation.

## Clinical Evidence

Bench testing only. Testing included validation of reprocessing, sterilization validation, and functional testing to demonstrate performance equivalent to original devices.

## Technological Characteristics

Constructed from metal alloys, plastic, and composites. Mechanical design includes frames, clamps, rods, and fasteners. Provided sterile or non-sterile. No software or electronic components.

## Regulatory Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- TempFix™ External Fixation System ([K021933](/device/K021933.md))
- TempFix™ External Fixation System, MR Safe ([K875022](/device/K875022.md))
- DePuy ACE AIM® Titanium Supracondylar Nail ([K875012](/device/K875012.md))
- Hoffmann Dynamic Wrist Fixator ([K910986](/device/K910986.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K052918 (pg 1 of 2)
JAN 27 2006

## PART B: 510(k) SUMMARY

| Submitter:           | Alliance Medical Corporation<br>10232 South 51st Street<br>Phoenix, Arizona 85044                                                                                                                                                                                                                                                      |
|----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact:             | Moira Barton<br>Regulatory Affairs Manager<br>(480) 763-5300 (o)<br>(480) 763-5310 (f)<br>mbarton@alliance-medical.com                                                                                                                                                                                                                 |
| Date of preparation: | August 2, 2005                                                                                                                                                                                                                                                                                                                         |
| Name of device:      | Trade/Proprietary Name: Reprocessed External Fixation<br>Devices<br>Common or Usual Name: External Fixation Devices, Fixation<br>Appliance, Single/Multiple Component<br>Classification Name: Single/Multiple Component Metallic Bone<br>Fixation Appliances and Accessories and Smooth or Threaded<br>Metallic Bone Fixation Fastener |

| Predicate Device | 510(k) Title | Manufacturer                               |
|------------------|--------------|--------------------------------------------|
|                  | K021933      | TempFix™ External Fixation System          |
|                  | K875022      | TempFix™ External Fixation System, MR Safe |
|                  | K875012      | DePuy ACE AIM® Titanium Supracondylar Nail |
|                  | K910986      | Hoffmann Dynamic Wrist Fixator             |

Device description: External fixation devices are specially designed frames, clamps, rods, rod-to-rod couplings, pins, posts, fasteners, wire fixations, fixation bolts, washers, nuts, hinges, sockets, connecting bars and screws used for the management of bone fractures and reconstructive, as well as corrective, orthopedic surgery. Materials used include metal alloys, plastic and composites. These materials are chosen to address a wide range of fractures and applications as well as to allow for the appropriate amount of rigidity and stability.

Intended use: External Fixation Devices are intended to be used for the fixation of supracondylar, or condylar fractures of the femur; for fusion of a joint; for surgical procedures that involve cutting the bone, for fixation of bone fractures; bone reconstruction; as a guide pin for insertion of other implants; or may be implanted through the skin so that a pulling force or traction may be applied to the skeletal system; and others may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that pulling force (traction) may be applied to the skeletal system.

{1}------------------------------------------------

K052918 (pg 2 of 2)

- Indications Reprocessed external fixation devices are indicated for use in statement: patients requiring external skeletal fixation and treatment of fractures, osteotomy, arthrodesis, correction of deformities, fracture revision, bone reconstruction procedures, limb lengthening, correction of bony or soft tissue deformities and segmental bony or soft tissue defects.
- Technological The design, materials, and intended use of the Reprocessed External Fixation Devices are identical to the predicate devices. characteristics: The mechanism of action of the Reprocessed External Fixation. Device is identical to the predicate devices in that the same standard mechanical design, materials, shapes and sizes are utilized. There are no changes to the claims, intended use, clinical applications, patient population, performance specifications, or method of operation.
- Performance data: These devices will be provided sterile and non-sterile depending on customer preference.

Bench and laboratory testing was conducted to demonstrate performance (safety and effectiveness) of Reprocessed External Fixation Devices. This included the following tests:

- Validation of reprocessing ●
- Sterilization Validation (for sterile devices) .
- . Function test(s)

Performance testing demonstrates that Reprocessed External Fixation Devices perform as originally intended.

- Conclusion: Alliance Medical Corporation concludes that the modified devices (Reprocessed External Fixation Device) are safe, effective, and substantially equivalent to the predicate devices as described herein.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the seal of the Department of Health and Human Services (HHS) of the United States. The seal features the department's name encircling a symbol. The symbol is a stylized eagle with three lines representing its wings, and the eagle's head is facing to the left.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JAN 2 7 2006

Ms. Moira Barton Manager, Regulatory Affairs Alliance Medical Corporation 10232 South 51st Street Phoenix, Arizona 85044

Re: K052918

Trade/Device Name: Alliance Medical Corporation Reprocessed External Fixation Devices (See enclosed list) Regulation Number: 21 CFR 888.3030 and 21 CFR 888.3040 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories, Smooth or threaded metallic bone fixation fastener Regulatory Class: II Product Code: KTT, KTW, JEC Dated: December 21, 2005 Received: January 5, 2006

Dear Ms. Barton:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, I'DA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA 's issuance of a substantial equivalence determination docs not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set

{3}------------------------------------------------

## Page 2 - Ms. Barton

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0210. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely vours.

Sincerely yours,

Barbara BuchmuD
for

Mark N. Melkerson Acting Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

| System                               | Part        | Description                                                    |
|--------------------------------------|-------------|----------------------------------------------------------------|
| Ace-Fisher® Fixator<br>System        | FC-10027    | Composite 2/3 Ring, Small (6" diam.)                           |
|                                      | FC-10028    | Composite 1/3 Ring, Small (6" diam.)                           |
|                                      | FC-10029    | Composite 2/3 Ring, Medium (8" diam.)                          |
|                                      | FC-10030    | Composite 1/3 Ring, Medium (8" diam.)                          |
|                                      | FC-10031    | Composite 2/3 Ring, Large (10" diam.)                          |
|                                      | FC-10032    | Composite 1/3 Ring, Large (10" diam.)                          |
|                                      | 8180-04-008 | Ring Foot Medium                                               |
|                                      | 8180-04-011 | Medium Foot Frame                                              |
|                                      | 8180-04-013 | Foot Frame Crossbar                                            |
|                                      | FA-10000-3  | Long Rod                                                       |
|                                      | FA-10000-2  | Medium Rod                                                     |
|                                      | FA-10000-1  | Short Rod                                                      |
|                                      | FA-10000-0  | Extra Short Rod                                                |
|                                      | FA-10355    | 1 Pin Holder                                                   |
|                                      | FA-10360    | 2 Pin Holder                                                   |
|                                      | FA-10365    | 3 Pin Holder                                                   |
|                                      | FA-10054-2  | 90° Pin Holder Adapter                                         |
|                                      | FF-10605    | 90° Adapter Rod                                                |
|                                      | FF-10023    | Ring Connector                                                 |
|                                      | FF-10610    | Articulation                                                   |
| Hoffman Classic<br>Sterile Wrist Kit | A13         | Articulation 8/8 mm                                            |
|                                      | A14         | Articulation 8/5 mm                                            |
|                                      | A16         | Compression/Distraction Bar                                    |
|                                      | A18/150     | Connecting Rod 150mm                                           |
|                                      | A18/200     | Connecting Rod 200mm                                           |
|                                      | A18/250     | Connecting Rod 250mm                                           |
|                                      | A18/300     | Connecting Rod 300mm                                           |
|                                      | A18/350     | Connecting Rod 350mm                                           |
|                                      | A18/400     | Connecting Rod 400mm                                           |
|                                      | AC11        | Ball Joint w/ Rod                                              |
|                                      | AC10        | Ball Joint                                                     |
|                                      | AC16        | Washer                                                         |
|                                      | 8082-00-000 | Classic Sterile Wrist Kit<br>(AMC does not reprocess the pins) |

Reprocessed External Fixation Device Models found to be Substantially Equivalent:

: :

{5}------------------------------------------------

## 2. Indications for Use Statement

## 510(k) Number (if known): K052918

Device Name: Alliance Medical Corporation Reprocessed External Fixation Devices

Indications For Use: Reprocessed external fixation devices are indicated for use in patients requiring external skeletal fixation and treatment of fractures, osteotomy, arthrodesis, correction of deformities, fracture revision, bone reconstruction procedures, limb lengthening, correction of bony or soft tissue deformities and segmental bory or soft tissue defects. ·

Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR

Over-the-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Tartian Meehim

(Division Sign-Of Division of General, Restorative, and Neurological Devices

Alliance Medical Corporation Reprocessed External Fixation Device Traditional 510(k)

**510(k) Number** K052918

14

---

**Source:** [https://fda-staging.innolitics.com/device/K052918](https://fda-staging.innolitics.com/device/K052918)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).