IGNITE® Putty is indicated only for bony voids or gaps that are not intrinsic to the stability of the bony structure. IGNITE® Putty is intended to be gently packed into bony voids or gaps of the skeletal system (i.e., the extremities and pelvis). These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. The Bone Graft Syringe is intended for use as a piston syringe for aspiration of bone marrow, autologous blood, plasma, or other body fluids. The syringe can be used to mix bone graft materials with aspirated fluids and deliver the composite graft material to the orthopedic surgical site.
Device Story
IGNITE® Kit combines ALLOMATRIX® Injectable Putty and a Bone Graft Syringe. Putty serves as bone void filler for non-structural skeletal defects; syringe facilitates aspiration of bone marrow, blood, or plasma and delivery of composite graft material to orthopedic surgical sites. Surgeons mix putty powder with sterile water or autologous bone marrow aspirate (BMA) using the syringe. Device used in clinical orthopedic settings. Osteoinductivity verified via in vitro human osteosarcoma cell bioassay and in vivo athymic rat muscle pouch model. Viral inactivation validated for processing method. Benefits include providing a convenient, osteoinductive graft delivery system for bone repair.
Clinical Evidence
No clinical trials presented. Evidence consists of bench testing and animal models. Osteoinductivity validated via in vitro human cell bioassay (correlation coefficient 0.850, p<0.0005) and in vivo athymic rat muscle pouch model. Viral inactivation validated using a panel of model human viruses.
Technological Characteristics
Kit contains demineralized bone matrix (DBM) and a piston-style syringe. Osteoinductive potential verified by bioassay. Viral inactivation processing applied. Syringe is a manual delivery tool. No electronic components or software.
Indications for Use
Indicated for patients with bony voids or gaps in the skeletal system (extremities and pelvis) not intrinsic to structural stability, resulting from surgery or trauma. Syringe indicated for aspiration of bone marrow, blood, or plasma and delivery of graft material.
Regulatory Classification
Identification
A resorbable calcium salt bone void filler device is a resorbable implant intended to fill bony voids or gaps of the extremities, spine, and pelvis that are caused by trauma or surgery and are not intrinsic to the stability of the bony structure.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance: Resorbable Calcium Salt Bone Void Filler Device; Guidance for Industry and FDA.” See § 888.1(e) of this chapter for the availability of this guidance.
{0}------------------------------------------------
NOV - 7 2005
Image /page/0/Picture/1 description: The image shows the logo for Wright Medical Technology. The logo consists of a stylized "W" symbol on the left, followed by the text "WRIGHT MEDICAL TECHNOLOGY" in a serif font. There is also a handwritten "K052" on the right side of the logo.
A Wright Medical Group Company
# 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS
In accordance with the Food and Drug Administration Rule to implement provisions of the Safe Medical Devices Act of 1990 and in conformance with 21 CRF 807, this information serves as a Summary of Safety and Effectiveness for the IGNITE® Kit.
| Submitted By: | Wright Medical Technology, Inc. |
|-------------------------------------|---------------------------------------------------------------------------------------|
| Date: | October 14, 2005 |
| Contact Person: | Ehab M Esmail<br>Director, Regulatory Affairs<br>Phone: 901-867-120 Fax: 901-867 4630 |
| Proprietary Name: | IGNITE® Kit |
| Common Name: | Bone Void Filler |
| Classification Name and Reference: | Filler, calcium sulfate preformed pellets |
| Device Product Code and Panel Code: | Orthopedics/87/MOV |
# DEVICE INFORMATION
#### INTENDED USES/ INDICATIONS A.
IGNITE® Putty is indicated only for bony voids or gaps that are not intrinsic to the stability of the bony structure. IGNITE® Putty is intended to be gently packed into bony voids or gaps of the skeletal system (i.e., the extremities and pelvis). These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone.
The Bone Graft Syringe is intended for use as a piston syringe for aspiration of bone marrow, autologous blood, plasma, or other body fluids. The syringe can be used to mix bone graft materials with aspirated fluids and deliver the composite graft material to the orthopedic surgical site
#### B. DEVICE DESCRIPTION
The IGNITE® product combines existing Wright Medical products into a kit configuration: ALLOMATRIX® Injectable Putty (K020895, K041168), and accessory
Image /page/0/Picture/12 description: The image shows the logo for Wright Medical Technology. The logo consists of a stylized "W" on the left, followed by the words "WRIGHT MEDICAL TECHNOLOGY" stacked on top of each other. Below the logo is the text "A Wright Medical Group Company".
Image /page/0/Picture/13 description: The image shows a black and white drawing of a planet. The planet has a rough, textured surface with dark and light areas. There is a ring around the planet, and a small object is visible below it. The drawing is simple and lacks detail, giving it a somewhat abstract appearance.
{1}------------------------------------------------
Bone Graft Syringe (K023088) with accessory components which are exempt from 510(k) requirements pursuant to 21 CFR 878.4800. The IGNITE® convenience kit provides surgeons the option of mixing the ALLOMATRIX® powder with sterile water or mixing with autologous bone marrow aspirate (BMA).
#### C. MATERIALS
The implant materials used for the IGNITE® Kit are substantially equivalent to the previously submitted and cleared ALLOMATRIX® implant materials.
#### D. SUBSTANTIAL EQUIVALENCE INFORMATION
The intended use, material composition, and design features of the IGNITE® Kit are substantially equivalent to the previously submitted and cleared ALLOMATRIX® product. The safety and effectiveness of the IGNITE® Kit is adequately supported by the substantial equivalence information, materials data, and testing results provided within this Premarket Notification.
## Osteoinductivity Potential
The DBM incorporated into IGNITE® Kits is assayed in vitro for its osteoinductive potential. The bioassay measures the proliferation of Saos human osteosarcoma cells in the presence of human DBM compared to positive and negative controls (osteoinductivity index). Results from the bioassay were correlated with results from implantation of DBM into athymic rat muscle", which demonstrated a correlation coefficient of 0.850 (p<0.0005) and accurately predicted the in vivo osteoinductivity of 99 of 101 donor lots. Additionally, clinical results using DBM with >0.20 and <0.20 demonstrated a significant difference in healing as evaluated by radiography, 92% and 33% healing, respectively.3
IGNITE® Kit were assayed in vivo in the rat muscle pouch model- and found to be osteoinductive. Each lot of IGNITE® Kit is assayed in vivo in the athymic rat muscle pouch to ensure the osteoinductivity potential of the final product.
Adkisson HD, Strauss-Schoenberger J, Gillis M, Wilkins R, Jackson M, and Hruska KA. Rapid Quantitative 1 Bioassay of Osteoinduction, J Ortho Res, 2000, 18:503-511.
Lindholm TS, Urist MR. A quantitative analysis of new bone formation by induction in composite grafts of 2 bone marrow and bone matrix, Clin Orthop 1980 Jul-Aug;(150):288-300. Note: The product is considered osteoinductive if one specimen (explant) contains new bone (i.e. bone occupied with lamellae), cartilage, and/or chondrocytes.
3 Wilkins RM, Clinical Effectiveness of Demineralized Bone Matrix Assayed in Human Cell Culture, Advances in Tissue Banking. 1999 3:113-124
## Viral Inactivation Validation
The method for processing the DBM and CBM contained in IGNITE® Kits was evaluated for its viral inactivation potential. A panel of model potential human viruses representing various virus types, sizes, shapes, and genomes were evaluated. The viral inactivation testing demonstrated suitable viral inactivation potential of the processing method for a wide spectrum of potential human viruses.
Image /page/1/Picture/14 description: The image shows the logo for Wright Medical Technology. The logo consists of a stylized "W" on the left, followed by the words "WRIGHT MEDICAL TECHNOLOGY" stacked on top of each other. Below the words is a horizontal line, and below that is the text "A Wright Medical Group Company".
Image /page/1/Picture/15 description: The image shows a black and white illustration of a planet, possibly Earth, with visible landmasses and cloud cover. A curved line above the planet suggests an atmospheric layer or a celestial object in the background. The overall style is reminiscent of a vintage or scientific illustration.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image is a black and white logo for the Department of Health and Human Services (HHS) in the USA. The logo features the department's emblem, which is a stylized representation of a human figure. The words "DEPARTMENT OF HEALTH AND HUMAN SERVICES - USA" are arranged in a circular pattern around the emblem.
NOV - 7 2005
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Ehab M. Esmail Director, Regulatory Affairs Wright Medical Technology, Inc. 5677 Airline Road Arlington, Tennessee 38002
Re: K052913
Trade/Device Name: IGNITE® KIT Regulation Number: 21 CFR 888.3045 Regulation Name: Resorbable calcium salt bone void filler device Regulatory Class: II Product Code: MQV Dated: October 14, 2005 Received: October 17, 2005
Dear Mr. Esmail:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2 –Ehab. M. Esmail
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120 . Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Signature
Mark N. Melkerson, Acting Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number: K052913
IGNITE® KIT Device Name:
Indications For Use:
IGNITE® Putty is indicated only for bony voids or gaps that are not intrinsic to the stability of the bony structure. IGNITE® Putty is intended to be gently packed into bony voids or gaps of the skeletal system (i.e., the extremities and pelvis). These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone.
The Bone Graft Syringe is intended for use as a piston syringe for aspiration of bone marrow, autologous blood, plasma, or other body fluids. The syringe can be used to mix bone graft materials with aspirated fluids and deliver the composite graft material to the orthopedic surgical site.
Prescription Use V (Per21 CFR 801.109)
OR
Over-The Counter Use (Optional Format 1-2-96)
## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of General Restorative Devices
510(k) Number K052913
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.