SCANDIUS TRITIS TIBIAL ACL RECONSTRUCTION SYSTEM

K052810 · Scandius Biomedical, Inc. · MBI · Dec 20, 2005 · Orthopedic

Device Facts

Record IDK052810
Device NameSCANDIUS TRITIS TIBIAL ACL RECONSTRUCTION SYSTEM
ApplicantScandius Biomedical, Inc.
Product CodeMBI · Orthopedic
Decision DateDec 20, 2005
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3040
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The intended use of the Scandius TriTis Tibial ACL Reconstruction System is for fixation of ligament and tendon grafts in cruciate ligament reconstructions of the knee.

Device Story

TriTis Tibial ACL Reconstruction System is a three-piece implant system for tibial ACL reconstruction. Components include a graft cleat to secure soft tissue graft in the bone tunnel, plus a washer and fixation screw to secure graft ends and cleat to bone. System includes specialized instrumentation for bone tunnel preparation and device placement, along with a sterilization tray. Used by surgeons in clinical/OR settings for knee ligament reconstruction. Provides mechanical fixation of grafts to bone to facilitate healing and joint stability.

Clinical Evidence

Bench testing only; mechanical ultimate failure strength data provided to support substantial equivalence.

Technological Characteristics

Three-piece non-degradable soft tissue fixation fastener system. Includes graft cleat, washer, and fixation screw. Mechanical fixation principle. Supplied with instrumentation and sterilization tray.

Indications for Use

Indicated for fixation of ligament and tendon grafts in cruciate ligament reconstructions of the knee.

Regulatory Classification

Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ### 510(k) SUMMARY #### 1. SUBMITTER: Scandius Biomedical, Inc. 11A Beaver Brook Road Littleton, MA 01460 Contact: Eric Bannon, Regulatory Consultant Date Prepared: December 19, 2005 #### 2. DEVICE: Trade Name: Scandius TriTis Tibial ACL Reconstruction System Classification Name: Fastener, Fixation, Non-Degradable, Soft Tissue The Product Code: MBI ### 3. PREDICATE DEVICE: The predicate device used to determine substantial equivalence for the Scandius ACL Reconstruction System was the Innovasive Devices Intratunnel Tibial Fixation Fastener. ## 4. DEVICE DESCRIPTION: The TriTis Tibial ACL Reconstruction System consists of a three piece implant designed to fixation soft tissue for tibial ACL reconstruction. The graft cleat secures the soft tissue graft in the tunnel. The washer and fixation screw secures the ends of the graft and cleat to the bone. The System includes instrumentation to prepare the bone tunnel and place the device as well as a sterilization tray. #### 5. INTENDED USE: The intended use of the Scandius TriTis Tibial ACL Reconstruction System is for fixation of ligament and tendon grafts in cruciate ligament reconstructions of the knee. ## 6. COMPARISON OF CHARACTERISTICS: - The devices have the same intended and indication for use; have similar technical characteristics and principles of operation. - 트 The devices use similar implant materials {1}------------------------------------------------ - Bench testing demonstrates that any minor technological differences do not .. raise any new questions of safety and effectiveness. ## 7. PERFORMANCE DATA: The following performance data was provided in support of the substantial equivalence determination: - Mechanical ultimate failure strength . {2}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three lines representing its wings and a circular border. The text "DEPARTMENT OF HEALTH AND HUMAN SERVICES USA" is arranged around the top half of the circle. DEC 2 0 2005 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Mr. Eric Bannon Regulatory Representative Scandius Biomedical, Inc. 11A Beaver Brook Road Littleton, Massachusetts 01460 Re: K052810 Trade/Device Name: TriTis Tibial ACL Reconstruction System Regulation Number: 21 CFR 888.3040 Regulation Name: Fastener, Fixation, Non-Degradable, Soft Tissue Regulatory Class: II Product Code: MBI Dated: November 28, 2005 Received: November 29, 2005 Dear Mr. Bannon: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set {3}------------------------------------------------ Page 2 - Mr. Eric Bannon forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic forth in the quality systems (QD) roggistions 531-542 of the Act); 21 CFR 1000-1050. product radiation control provisions (Declients on 1 evice as described in your Section 5 10(k) I his letter will anow you w ough mailing of substantial equivalence of your device to a legally premarket notification: "The PDF intains sification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific acrice for your as (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small other general informational and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, ( محمد Mark N. Melkerson Acting Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # Indications for Use 510(k) Number (if known): K052810 Device Name: Scandius TriTis Tibial ACL Reconstruction System Indications for Use: The Scandius TriTis ACL Reconstruction System is intended for use indications for Ober The Beardon grafts in cruciate ligament reconstructions of the knee. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) Division of General, Restorative, and Neurological Devices Page _1_of__1_ 510(k) Number_Ko528|0 (Posted November 13, 2003)
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...