MYLAB30, 50 AND 70 SYSTEMS UPGRADE, MODELS 7300, 7350, 6150
Applicant
Esaote, S.p.A.
Product Code
IYN · Radiology
Decision Date
Oct 17, 2005
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1550
Device Class
Class 2
Attributes
Pediatric
Indications for Use
Esaote's Model 7300 (MyLab30) is a compact ultrasound system used to perform diagnostic general ultrasound studies including Cardiac, Transesophageal, Peripheral Vascular, Neonatal Cephalic, Adult Cephalic, Small organ, Musculoskeletal (Conventional and Superficial), Abdominal, Fetal, Transvaginal, Transrectal, Pediatric and Intraoperative Abdominal. Esaote's Model 7350 (MyLab50) is a mainframe ultrasound system used to perform diagnostic general ultrasound studies including Cardiac, Transesophageal, Peripheral Vascular, Neonatal Cephalic, Adult Cephalic, Small organ, Musculoskeletal (Conventional and Superficial), Abdominal, Fetal, Transvaginal, Transrectal, Pediatric and Intraoperative Abdominal. Esaote's Model 6150 (MyLab70) is a mainframe ultrasound system used to perform diagnostic general ultrasound studies including Cardiac, Transesophageal, Peripheral Vascular, Neonatal Cephalic, Adult Cephalic, Small organ, Musculoskeletal (Conventional and Superficial), Abdominal, Fetal, Transvaginal, Transrectal, Pediatic Other: Urological, and Intraoperative Abdominal.
Device Story
MyLab30, 50, 70 are compact/mainframe ultrasound systems for diagnostic imaging. Inputs: ultrasound signals from phased, convex, linear array, and Doppler probes. Operation: B-Mode, M-Mode, Doppler, Color Flow Mapping, and Tissue Enhancement Imaging (TEI). Output: real-time diagnostic images on LCD display. Used in clinical settings by healthcare professionals. Features include Compound Imaging, VPAN, and Tissue Velocity Mapping. Output assists clinicians in diagnostic assessment of various anatomical structures and fluid flow. Benefits: non-invasive diagnostic visualization of internal anatomy and hemodynamics.
Clinical Evidence
Bench testing only. No clinical data provided. Substantial equivalence is based on design similarity to predicate devices and performance testing of new transducers and indications.
Technological Characteristics
Ultrasound imaging systems (21 CFR 892.1550, 892.1560). Materials: standard medical-grade plastics/metals. Sensing: ultrasonic pulsed echo/Doppler transducers. Connectivity: LCD display, probe ports. Software: embedded firmware for image processing (B-Mode, M-Mode, Doppler, TEI, Compound Imaging). Quality System: ISO 9001:2000 and ISO 13485 certified.
Indications for Use
Indicated for diagnostic ultrasound imaging or fluid flow analysis in patients requiring Cardiac, Transesophageal, Peripheral Vascular, Neonatal/Adult Cephalic, Small organ (Thyroid, Breast, Testicles), Musculoskeletal, Abdominal, Fetal, Transvaginal, Transrectal, Pediatric, Urological, and Intraoperative Abdominal examinations.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
510(k) Summary MyLab30. 50. 70 Systems Esaole. S.p.A.
K 052805
0C7 1 7 2005
### 510(k) Summary
The following 510(k) summary has been prepared pursuant to requirements specified in 21CFR 1807.92(a).
807.92(a)(1)
#### Submitter Information
Carri Graham, Official Correspondent The Anson Group 7992 Castleway Drive Indianapolis, Indiana 46250 Phone: (317) 849-1916 x103 Facsimile: (317) 577-9070
| Contact Person: | Carri Graham |
|-------------------------|----------------------------------------------------------------------------------------------------|
| Date: | September 29, 2005 |
| 807.92(a)(2) | |
| Trade Name: | MyLab30, 50, 70 Systems |
| Common Name: | Ultrasound Imaging System |
| Classification Name(s): | Ultrasonic pulse doppler imaging system 892.1550<br>Ultrasonic pulsed echo imaging system 892.1560 |
| Classification Number: | 90IYN; 90IYO |
#### 807.92(a)(3)
#### Predicate Device(s)
| K040596 | 7300 (MyLab30) | Esaote, S.p.A. |
|---------|----------------|----------------|
| K050326 | 7350 (MyLab50) | Esaote, S.p.A. |
| K051308 | 6150 (MyLab70) | Esaote, S.p.A. |
| K051837 | 6100 (MyLab90) | Esaote, S.p.A. |
{1}------------------------------------------------
510(k) Summary Mr Lab30, 50, 70 Systems Esaote. S.p.A.
#### 807.92 (a)(4)
#### Device Description
The 7300 (MyLab30), 7350 (MyLab50) and 6150 (MyLab70) system designs remain the same as those previously cleared by FDA via K040596, K050326, and K051308, respectively. They are compact ultrasound systems used to perform diagnostic general ultrasound studies. Their primary modes of operation are: B-Mode, M-Mode, Doppler and Color Flow Mapping and, on lower frequency probes, Tissue Enhancement Imaging (TEI). The systems are equipped with an LCD Color Display and can drive phased (PA), convex (CA) and linear array (LA) and Doppler probes.
The 7300 (MyLab30), 7350 (MyLab50) and 6150 (MyLab70) models are manufactured under an ISO 9001:2000 and ISO 13485 certified quality system.
807.92(a)(5)
#### Intended Use(s)
Esaote's Model 7300 (MyLab30) is a compact ultrasound system used to perform diagnostic general ultrasound studies including Cardiac, Transesophageal, Peripheral Vascular, Neonatal Cephalic, Adult Cephalic, Small organ, Musculoskeletal (Conventional and Superficial), Abdominal, Fetal, Transvaginal, Transrectal, Pediatric and Intraoperative Abdominal.
Esaote's Model 7350 (MyLab50) is a mainframe ultrasound system used to perform diagnostic general ultrasound studies including Cardiac, Transesophageal, Peripheral Vascular, Neonatal Cephalic, Adult Cephalic, Small organ, Musculoskeletal (Conventional and Superficial), Abdominal, Fetal, Transvaginal, Transrectal, Pediatric and Intraoperative Abdominal.
Esaote's Model 6150 (MyLab70) is a mainframe ultrasound system used to perform diagnostic general ultrasound studies including Cardiac, Transesophageal, Peripheral Vascular, Neonatal Cephalic, Adult Cephalic, Small organ, Musculoskeletal (Conventional and Superficial), Abdominal, Fetal, Transvaginal, Transrectal, Pediatic Other: Urological, and Intraoperative Abdominal.
{2}------------------------------------------------
maisig or and y dry roger plynny
## 9)(1) 2017-12-2018
# อมสร้างอิง เหตุการสตว์
| To be added via this 510(k) | MyLab90 (6100)<br>K051837<br>(Predicate) | MyLab30CV (7300)<br>K040596 | MyLab50 (7350)<br>K050326 | MyLab70 (6150)<br>K051308 |
|------------------------------------------------|-------------------------------------------|-----------------------------|-------------------------------------------|-------------------------------------------|
| Doppler Probes<br>- 2.0 CW<br>- 5.0 CW | <span style="text-align:center;">✔</span> | YES | <span style="text-align:center;">✔</span> | <span style="text-align:center;">✔</span> |
| IOE Probe<br>- IOE323 | <span style="text-align:center;">✔</span> | YES | <span style="text-align:center;">✔</span> | <span style="text-align:center;">✔</span> |
| Biopsy Attachment<br>- ABS15 | <span style="text-align:center;">✔</span> | YES | YES | YES |
| VPAN | <span style="text-align:center;">✔</span> | YES | YES | YES |
| Compound Imaging | <span style="text-align:center;">✔</span> | YES | YES | YES |
| Intelligent Real Time<br>Processing | NO | NO | NO | YES |
| Intraoperative Abdominal<br>Indication for Use | <span style="text-align:center;">✔</span> | YES | YES | YES |
ବ əsindən filmlərin film bax və vəsilə ədəd vasitəsindən fəsiləsinin cinsinə aid bitk
YES
# Notice
It has come to our attention that certain individuals and entities are falsely claiming to be affiliated with or endorsed by Zeno Power Systems, Inc. These individuals and entities are misrepresenting themselves and their products or services as being associated with Zeno Power Systems, Inc., when in fact they are not.
Zeno Power Systems, Inc. is not responsible for any actions, representations, or omissions made by these individuals or entities. We urge you to exercise caution and due diligence when dealing with anyone claiming to be affiliated with Zeno Power Systems, Inc.
If you have any questions or concerns about the legitimacy of any individual or entity claiming to be affiliated with Zeno Power Systems, Inc., please contact us directly at [info@zenopower.com](mailto:info@zenopower.com).
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of the department's name encircling a symbol. The symbol is composed of three stylized lines that resemble a human figure.
Public Health Service
OCT 17 2005
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Esaote, S.p.A. % Ms. Carri Graham Consultant The Anson Group 7992 Castleway Drive INDIANAPOLIS IN 46250
Re: K052805
Trade Name: Model 7300 (MyLab30), Model 7350 (MyLab50), and Model 6150 (MyLab70) Ultrasound Imaging Systems Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulation Number: 21 CFR 892.1560 Regulation Name: Ultrasonic pulsed echo imaging system Regulation Number: 21 CFR 892.1570 Regulation Name: Diagnostic ultrasonic transducer Regulatory Class: II Product Code: IYN, IYO, and ITX Dated: September 29, 2005 Received: October 4, 2005
#### Dear Ms. Graham:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device mendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to I cactar Food, Drag, and sions of the Act. The general controls provisions of the Act include ane general only provents for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
This determination of substantial equivalence applies to the following transducers intended for r ms determination 7300 (MyLab30), Model 7350 (MyLab50), and Model 6150 (MyLab70) Ultrasound Imaging Systems, as described in your premarket notification:
{4}------------------------------------------------
#### Transducer Model Number
$$\frac{\text{IOE323}}{\underline{\text{CW}}}$$
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean r hat FDA has made a determination that your device complies with other requirements of the Act that I Drinas Intact and regulations administered by other Federal agencies. You must or any I vith all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic for an as a tion control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This determination of substantial equivalence is granted on the condition that prior to shipping the first device, you submit a postclearance special report. This report should contain complete information, including acoustic output measurements based on production line devices, requested in Appendix G, (enclosed) of the Center's September 30, 1997 "Information for Manufacturers Sceking Marketing Clearance of Diagnostic Ultrasound Systems and Transducers." If the special report is incomplete or contains unacceptable values (e.g., acoustic output greater than approved levels), then the 510(k) clearance may not apply to the production units which as a result may be considered adulterated or misbranded.
The special report should reference the manufacturer's 510(k) number. It should be clearly and prominently marked "ADD-TO-FILE" and should be submitted in duplicate to:
> Food and Drug Administration Center for Devices and Radiological Health Document Mail Center (HFZ-401) 9200 Corporate Boulevard Rockville, Maryland 20850
This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small
{5}------------------------------------------------
Page 3 – Ms. Graham
Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html
If you have any questions regarding the content of this letter, please contact Rodrigo C. Perez at (301) 594-1212.
Sincerely yours,
Nancy C. Burgdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure(s)
{6}------------------------------------------------
#### Diagnostic Ultrasound Indications for Use Form
#### Model 7300
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Intended Use: Diagnostic | | | | | | | | | | |
|----------------------------------|---|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal | | P | P | P | | P | P | | P [2] | P [3] |
| Abdominal | | P | P | P | | P | P | | P [2] | P [3] |
| Intraoperative<br>(Abdominal) | | N | N | N | | N | N | | N [2] | N [3] |
| Intraoperative<br>Neurological | | | | | | | | | | |
| Pediatric | | P | P | P | | P | P | | P [2] | P [3] |
| Small Organ (specify) [1] | | P | P | P | | P | P | | P [2] | P [3] |
| Neonatal Cephalic | | P | P | P | | P | P | | P [2] | |
| Adult Cephalic | | P | P | P | | P | P | | P [2] | |
| Cardiac | | P | P | P | | P | P | | P [2] | P [3] |
| Transesophageal | | P | P | P | | P | P | | P [2] | |
| Transrectal | | P | P | P | | P | P | | P [2] | |
| Transvaginal | | P | P | P | | P | P | | P [2] | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral Vascular | | P | P | P | | P | P | | P [2] | |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | P | P | P | | P | P | | P [2] | |
| Musculo-skeletal<br>Superficial | | P | P | P | | P | P | | P [2] | |
| Other (specify) | | | | | | | | | | |
N=new indication; P=previously cleared by FDA; E= added under Appendix E Additional Comments:
[1] Small organs include Thyroid, Breast and Testicles.
(2) Applicable combined modes: B+M+PW+CW+CFM+PD
(3) Tissue Enhancement Imaging (TEI)
Compound Imaging VPAN Tissue Velocity Mapping (TVM) CMM CnTl
(PLEASE DO ROT WRITE BELOW THIS LINE. CONTINUE ON ANOTHER PAGE IF
concurrence of CDRH, Office of Device Evaluation (t
Prescription Use (Per 21 CFR 801.109)
Nancy C. Roardon
Divisi n of Reproductive and logical Devic 510(k) Number
{7}------------------------------------------------
#### Diagnostic Ultrasound Indications for Use Form
#### Model 7350
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | | | | | | Mode of Operation | | | | |
|----------------------------------|---|---|---|-----|-----|-------------------|----------------------|------------------------------|-----------------------|--------------------|
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal | | P | P | P | | P | P | | P [2] | P [3] |
| Abdominal | | P | P | P | | P | P | | P [2] | P [3] |
| Intraoperative<br>(Abdominal) | | N | N | N | | N | N | | N [2] | N [3] |
| Intraoperative<br>Neurological | | | | | | | | | | |
| Pediatric | | P | P | P | P | P | P | | P [2] | P [3] |
| Small Organ (specify) [1] | | P | P | P | P | P | P | | P [2] | P [3] |
| Neonatal Cephalic | | P | P | P | P | P | P | | P [2] | |
| Adult Cephalic | | P | P | P | P | P | P | | P [2] | P [3] |
| Cardiac | | P | P | P | P | P | P | | P [2] | P [3] |
| Transesophageal | | P | P | P | P | P | P | | P [2] | |
| Transrectal | | P | P | P | | P | P | | P [2] | |
| Transvaginal | | P | P | P | | P | P | | P [2] | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral Vascular | | P | P | P | P | P | P | | P [2] | |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | P | P | P | P | P | P | | P [2] | |
| Musculo-skeletal<br>Superficial | | P | P | P | P | P | P | | P [2] | |
| Other (specify) | | | | | | | | | | |
N=new indication; P=previously cleared by FDA; E= added under Appendix E Additional Comments:
[1] Small organs include Thyroid, Breast and Testicles.
[2] Applicable combined modes: B+M+PW+CW+CFM+PD
- [3] Tissue Harmonic Imaging (TEI)
CMM
Tissue Velocity Mapping (TVM)
VPAN
Compound Imaging
- CnTI
(PLEASE DO NOT WRITE BELOW THIS LINE. CONTINUE ON ANOTHER PAGE IF NEEDED
concurrence of CDRH, Office of Device Evaluation (ODE)
Nancy Burden
(Division Sign-Off)
Division of Reproductive, Abdi and Radiological Devices 510(k) Number
{8}------------------------------------------------
#### Diagnostic Ultrasound Indications for Use Form
#### Model 6150
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | | Mode of Operation | | | | | | | | |
|----------------------------------|---|-------------------|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal | P | P | P | P | P | P | P | | P [2] | P [3] |
| Abdominal | P | P | P | P | P | P | P | | P [2] | P [3] |
| Intraoperative<br>(Abdominal) | N | N | N | N | N | N | N | | N [2] | N [3] |
| Intraoperative<br>Neurological | | | | | | | | | | |
| Pediatric | P | P | P | P | P | P | P | | P [2] | P [3] |
| Small Organ (specify) [1] | P | P | P | P | P | P | P | | P [2] | P [3] |
| Neonatal Cephalic | P | P | P | P | P | P | P | | P [2] | P [3] |
| Adult Cephalic | P | P | P | P | P | P | P | | P [2] | P [3] |
| Cardiac | P | P | P | P | P | P | P | | P [2] | P [3] |
| Transesophageal | P | P | P | P | P | P | P | | P [2] | P [3] |
| Transrectal | P | P | P | P | P | P | P | | P [2] | P [3] |
| Transvaginal | P | P | P | P | P | P | P | | P [2] | P [3] |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral Vascular | P | P | P | P | P | P | P | | P [2] | P [3] |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | P | P | P | P | P | P | P | | P [2] | P [3] |
| Musculo-skeletal<br>Superficial | P | P | P | P | P | P | P | | P [2] | P [3] |
| Other (Urological) | P | P | P | P | P | P | P | | P [2] | P [3] |
N=new indication; P=previously cleared by FDA; E= added under Appendix E Additional Comments:
[1] Small organs include Thyroid, Breast and Testicles.
[2] Applicable combined modes: B+M+PW+CW+CW+CFM+PD
- |3| Tissue Harmonic Imaging (TEI)
CMM VPAN Compound Imaging CriTI Tissue Velocity Mapping (TVM)
(PLEASE DO NOT WRITE BELOW THIS LINE. CONTINUE ON ANOTHER PAGE IF NEEDED
concurrence of CDRH, Off
Prescription Use (Per 21 CFR 801.109)
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number K052805
{9}------------------------------------------------
#### 7300, 7350, and 6150 Systems
#### IOE323
| Clinical Application | Mode of Operation | | | | | | | | | | |
|----------------------------------|-------------------|---|---|-------------|-------------|---------------------------|------------------------------|------------------------------|-----------------------|--------------------|-------|
| | A | B | M | PWD<br>(PW) | CWD<br>(CW) | Color<br>Doppler<br>(CFM) | Amplitude<br>Doppler<br>(PD) | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify) | |
| Ophthalmic | | | | | | | | | | | |
| Fetal | | | | | | | | | | | |
| Abdominal | | | | | | | | | | | |
| Intraoperative (Adominal) | | N | N | N | | | N | N | | N (2) | N (3) |
| Intraoperative Neurological | | | | | | | | | | | |
| Pediatric | | N | N | N | | | N | N | | N [2] | N [3] |
| Small Organ (specify) [1] | | N | N | N | | | N | N | | N [2] | N [3] |
| Neonatal Cephalic | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | |
| Cardiac | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | |
| Transurethral | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | |
| Peripheral Vascular | | N | N | N | | | N | N | | N [2] | N [3] |
| Laparoscopic | | | | | | | | | | | |
| Muscolo-skeletal<br>Conventional | | N | N | N | | | N | N | | N [2] | N [3] |
| Muscolo-skeletal<br>Superficial | | N | N | N | | | N | N | | N [2] | N [3] |
| Other | | | | | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
[1] Small organs include Thyroid, Breast and Testicles.
[2] Applicable combined modes: B+M+PW+CFM+PD
- [3] TEI (Tissue Enhanced Imaging)
VPAN
Compound Imaging CnTI
(PLEASE DO NOT WRITE BELOW THIS LINE. CONTINUE ON ANOTHER PAGE IF HELDED)
concurrence of CDRH, Office of Device Evaluation (ODE)
Nancy C. Brogdon
(Division Sign-Off)
Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number ________________________________________________________________________________________________________________________________________________________________
{10}------------------------------------------------
#### 7300 System
#### 2 CW
| | Mode of Operation | | | | | | | | | |
|-------------------------------------------------|-------------------|---|---|-------------|-------------|---------------------------|------------------------------|------------------------------|-----------------------|--------------------|
| Clinical Application | A | B | M | PWD<br>(PW) | CWD<br>(CW) | Color<br>Doppler<br>(CFM) | Amplitude<br>Doppler<br>(PD) | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal | | | | | | | | | | |
| Abdominal | | | | | | | | | | |
| Intraoperative (specify) | | | | | | | | | | |
| Intraoperative<br>Neurological | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small Organ (specify) | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | | | | N | | | | | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | | |
| Laparoscopic<br>Muscolo-skeletal | | | | | | | | | | |
| Conventional<br>Muscolo-skeletal<br>Superficial | | | | | | | | | | |
| Other | | | | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
#### (PLEASE DO NOT WRITE BELOW THIS LINE. COETIEUE OR ANOTHER PAGE IF NEEDED)
511(x) Number ________________________________________________________________________________________________________________________________________________________________
concurrence of CDRH, Office of Device Evaluation (ODE)
Nancy C. Brigdon
(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
K052205
{11}------------------------------------------------
#### 7300 System
#### 5 CW
| | Mode of Operation | | | | | | | | | |
|----------------------------------|-------------------|---|---|-------------|-------------|---------------------------|------------------------------|------------------------------|-----------------------|--------------------|
| Clinical Application | A | B | M | PWD<br>(PW) | CWD<br>(CW) | Color<br>Doppler<br>(CFM) | Amplitude<br>Doppler<br>(PD) | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal | | | | | | | | | | |
| Abdominal | | | | | | | | | | |
| Intraoperative (specify) | | | | | | | | | | |
| Intraoperative<br>Neurological | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small Organ (specify) | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral Vascular | | | | | N | | | | | |
| Laparoscopic | | | | | | | | | | |
| Muscolo-skeletal<br>Conventional | | | | | | | | | | |
| Muscolo-skeletal<br>Superficial | | | | | | | | | | |
| Other (Urological) | | | | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
#### (PLEASE DO NOT WRITE BELOW THIS LINE. CONTINUE ON ANOTHER PAGE IF BERDED)
concurrence of CDRH, Office of Device Evaluation (ODE)
Nancy C. Broughton
(Division Sign-Off) Entision of Reproductive, Abdominal, and Rediological Devices 510(k) Number_
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.