K052797 · Medtek Devices, Inc. · FYD · Feb 13, 2006 · General, Plastic Surgery
Device Facts
Record ID
K052797
Device Name
LAPEVAC, FILTRATION DEVICE FOR THE PERIOTONEUM
Applicant
Medtek Devices, Inc.
Product Code
FYD · General, Plastic Surgery
Decision Date
Feb 13, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.5070
Device Class
Class 2
Indications for Use
To remove airborne particles generated by tissue combustion during laparoscopic surgery, via filtration of gaseous media contained within the distended pneumoperitoneum in order to significantly improve visualization. LapEvac may be used in any laparoscopic surgery, as appropriate. Sterile, single use only device
Device Story
LapEvac is a battery-operated, sterile, single-use smoke evacuation system for laparoscopic surgery. It connects to inlet/outlet trocars to create a closed-circuit loop for filtering pneumoperitoneum gas (typically CO2). The system uses activated charcoal and ULPA filtration media to remove smoke particles, recirculating gas at approximately 4 liters per minute. Operated by surgical staff in the OR, the device is controlled via a 3-amp slide switch and has a 4-hour operational life. By clearing smoke from the surgical field, the device improves visualization for the surgeon, facilitating the procedure without affecting pneumoperitoneal pressure, temperature, or humidity.
Clinical Evidence
Bench testing only. Three verification tests were conducted: 1) Clarity of View and Evacuation Efficiency (verified smoke clearance time comparable to existing devices); 2) Effect on Pneumoperitoneum (verified no adverse impact on pressure, temperature, or relative humidity); 3) Power Budget (verified operation exceeds 4 hours).
Indicated for patients undergoing laparoscopic surgery to remove airborne particles generated by tissue combustion via filtration of pneumoperitoneum gaseous media to improve visualization.
Regulatory Classification
Identification
Air-handling apparatus for a surgical operating room is a device intended to produce a directed, nonturbulent flow of air that has been filtered to remove particulate matter and microorganisms to provide an area free of contaminants to reduce the possibility of infection in the patient.
Special Controls
*Classification.* Class II (special controls). The device, when it is an air handling bench apparatus, an air handling room apparatus, or an air handling enclosure apparatus, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 878.9.
Predicate Devices
Sun Medical SFE-200 (91154/A)
Sun Medical Smoke Evacuator Circuit (91154/A)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image is a logo for Buffalo Filter. The logo features the words "BUFFALO" and "Filter" in bold, black letters. The word "BUFFALO" is on top of the word "Filter", and there is a registered trademark symbol next to the word "Filter". To the left of the words is a black square with white lines inside.
K052797
FEB 1 3 2006
#### ' Leader in V Smoke Plume Evacuation and Aerosol Management "
595 Commerce Drive Buffalo, NY 14228 (800) 343-2324 (716) 835-7000 (716) 835-3414 fax www.buffalofilter.com email to: Infɔ@buffalofilter.com
## 510(k) SUMMARY
Buffalo Filter 595 Commerce Drive Buffalo, NY 14228
| Phone: | (800) 343.2324 |
|--------|----------------|
| | (716) 835.700 |
| Fax: | (716) 835.3414 |
| Contact Person: | Robert O. Dean |
|----------------------|----------------------------------------------------------------------------------------------------------|
| Date Prepared: | September 30, 2005 |
| Device Name: | LapEvac, Filtration Device for the<br>Pneumoperitoneum |
| Common Name: | Smoke Evacuator |
| Classification Name: | Apparatus, exhaust, surgical<br>878.5070 |
| Predicate Device: | Sun Medical SFE-200<br>Sun Medical Smoke Evacuator Circuit<br>Reviewed and released under 510(k) 91154/A |
Image /page/0/Picture/9 description: The image shows the logo for the American Society for Quality (ASQ). The text "Member Of: American Society for Quality" is above the logo. The logo itself consists of a stylized globe with the letters "ASQ" in large, bold font to the right of the globe.
Image /page/0/Picture/10 description: The image shows two logos and some text. The first logo is a UL (Underwriters Laboratories) logo with the words "REGISTERED FIRM" below it. The second logo is a UKAS (United Kingdom Accreditation Service) logo with the number 062 below it. The text above the logos reads "ISO 9001 - ISO 13485", and the text below the logos reads "D. HALO FILTER" and "A12634".
{1}------------------------------------------------
# KO52797
### Device Description:
LapEvac is a battery operated smoke evacuation system that connects directly to the inlet and outlet cannula/ trocars and creates a continuous closed circuit for filtering pneumoperitoneum gaseous media (typically CO2). The system contains the LapEvac unit and connection tubing. Filtration is accomplished using activated charcoal filtration media and ULPA filtration media. The system filters and recirculates at an approximate rate of 4 liters per minute. Operating life is approximately 4 hours. A 3 amp, UL rated slide switch provides on/off control.
#### Intended Use:
To remove airborne particles generated by tissue combustion during laparoscopic surgery, via filtration of gaseous media contained within the distended pneumoperitoneum in order to significantly improve visualization. LapEvac may be used in any laparoscopic surgery, as appropriate.
Sterile, single use only device
### Technological Comparison
Similarities of LapEvac and the predicate include the intended/indications for use - both devices are used in laparoscopic surgeries to filter pneumoperitoneum gas for removing smoke particles and thereby significantly increasing visualization within the pneumoperitoneum. The operating principle of the two devices is the same - the active recirculation of pneumoperitoneum gas and filtration through similar sub-micron sized filters.
Power supplies differ. LapEvac is battery operated at 6 volts versus 120volt AC for the predicate device.
Performance Data: Three sets of bench verification tests were conducted.
> Clarity of View and Evacuation Efficiency verified LapEvac's ability to clear smoke from the chamber in a time comparable to smoke filtration devices commonly used during laparoscopic surgeries.
Effect on Pneumoperitoneum verified that LapEvac has no effect on the pneumoperitoneal pressure, temperature, and relative humidity.
Power Budget verifies power consumption and assures that LapEvac operation exceeds 4 hours.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle-like symbol with three lines representing wings or feathers. The symbol is positioned to the right of the text, which is arranged in a circular pattern around the symbol. The text reads "DEPARTMENT OF HEALTH AND HUMAN SERVICES, USA".
#### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
FEB 1 3 2006
Mr. Robert O. Dean Vice President Medtrek Devices, Incorporated 595 Commerce Drive Buffalo, New York 14228
Re: K052797
Trade/Device Name: LapEvac, Filtration Devices for the Pneumoperitoneum Regulation Number: 21 CFR 878.5070 Regulation Name: Air-Handling Apparatus for Surgical Operating Room Regulatory Class: II Product Code: FYD Dated: January 20, 2006 Received: January 25, 2006
Dear Mr. Dean:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
### Page 2 - Mr. Dean
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if requirely the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Senette Y. Michie Dm.D.
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known): K052797
Device Name: LapEvac, Filtration device for the pneumoperitoneum
Indications For Use:
To remove airborne particles generated by tissue combustion during laparoscopic surgery, via filtration of gaseous media contained within the distended pneumoperitoneum in order to significantly improve visualization.
LapEvac may be used in any laparoscopic surgery, as appropriate.
Sterile, single use only device
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Shula A Murphy, MD, 1/10/08
Page 1 of 1
K052 747
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.