K052794 · DiaSorin, Inc. · JJF · Mar 31, 2006 · Clinical Chemistry
Device Facts
Record ID
K052794
Device Name
ETI-MAX 3000
Applicant
DiaSorin, Inc.
Product Code
JJF · Clinical Chemistry
Decision Date
Mar 31, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.2170
Device Class
Class 1
Attributes
Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K052794 · Mar 31, 2006
ETI-MAX 3000
DiaSorin, Inc.
Retrospective clinical repository serum samples
Retrospective clinical repository samples were used to assess the accuracy and performance of the ETI-MAX 3000 instrument by comparing results against the manual reference method.
Accuracy study using repository samples; Retrospective comparison study
159 repository samples per assay (ANA Screen, ENA Screen, Anti-SS-A (Ro)); Sample Size: 159 per assay; Number of Sites: 3 (DiaSorin Inc. and two external US laboratories)
Manual assay method
Percent agreement (positive, negative, overall) with manual method
Indications for Use
The ETI-MAX 3000™ is a fully automated EIA microtiter plate analyzer designed to perform the complete sample processing of qualitative and semi-quantitative assays with respect to (sample dilutions, sample and reagent dispensing, incubations, wash processes, plate transports) as well as the photometric measurement and evaluation. The qualitative and semi-quantitative performance of the ETI-MAX™ instrument were assessed using the DiaSorin ANAScreen ELISA kit, the DiaSorin Anti-SS-A(Ro)ELISA kit and the EuroDiagnostica AB ENA Single Well Screen kit.
Device Story
ETI-MAX 3000™ is a fully automated EIA microtiter plate laboratory analyzer. It accepts serum or plasma samples in primary tubes or test tubes. The system automates sample distribution, dilution, reagent addition, incubation, microplate washing, and photometric absorbance measurement. It features a barcode scanner for sample/rack identification and supports LIS integration. The device is operated by laboratory personnel in a clinical setting. The system mimics manual EIA/ELISA procedures, providing automated scheduling and data reduction. Output includes photometric results and assay reports, which assist clinicians in diagnostic decision-making for autoimmune conditions. The device benefits patients by increasing throughput, reducing manual handling errors, and ensuring standardized processing of diagnostic assays.
Clinical Evidence
Bench testing only. Performance validated using 159 repository samples for each of three assays (ANA Screen, ENA Screen, Anti-SS-A). Compared ETI-MAX 3000™ results against manual method results. ANA Screen: 96.9% positive agreement, 100% negative agreement. ENA Screen: 93.7% positive agreement, 100% negative agreement. Anti-SS-A (Qualitative): 100% positive/negative agreement. Anti-SS-A (Quantitative): 100% positive agreement, 99% negative agreement. Precision/reproducibility studies conducted across three sites showed acceptable %CVs. Linearity, carryover, and analytical sensitivity (LOD/LOB) were verified per NCCLS EP17-A.
Indicated for clinical laboratory use as a fully automated EIA microtiter plate analyzer for processing qualitative and semi-quantitative assays using serum or plasma samples.
Regulatory Classification
Identification
A micro chemistry analyzer for clinical use is a device intended to duplicate manual analytical procedures by performing automatically various steps such as pipetting, preparing filtrates, heating, and measuring color intensity. The distinguishing characteristic of the device is that it requires only micro volume samples obtainable from pediatric patients. This device is intended for use in conjunction with certain materials to measure a variety of analytes.
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
A. 510(k) Number:
k052794
B. Purpose for Submission:
New Device
C. Manufacturer and Instrument Name:
DiaSorin Inc.
DiaSorin ETI-MAX 3000™
D. Type of Test or Tests Performed:
For qualitative and semi-quantitative assays. The instrument performance was validated using three 510(k) cleared devices (ENA Screen, ANA Screen and Anti-SS-A (Ro) from Euro-Diagnostica).
E. System Descriptions:
1. Device Description:
The ETI-MAX 3000™ is a fully automated, EIA microtiter plate laboratory analyzer performing the complete sample processing (barcode scanner, predilution station, pipetting station, plate transport, wash station, incubators and photometric measurement). The instrument is controlled by the Windows PC software ETI-MAX 3000™. This software allows the user to process the pre-defined assays of DiaSorin.
The ETI-MAX 3000™ consists of 3 main components; sample processing, management system and software.
The sample processing system automates the steps required to process microplate assays from barcode reading, sample distribution, sample dilution, incubation, washing, reagent addition and detection using absorbance technology. The plate loading is managed with time scheduling, the plate capacity is 4, but can be up to 7 in continuous loading mode. There is also the capability to run multiple assays per plate.
The management system (computer) is a Pentium III, 500MHz, 64 Mbytes RAM or equivalent with a 3.5" Floppy drive, a 6.4 Gbyte Hard Disk, an Alphanumeric keyboard, standard Mouse, 15" color monitor, and Standard Inkjet printer.
The software (Operating system) uses Microsoft Windows® 95, 98 or 2000 32-bit application, for all operations, including assay creation and worklist generation, time scheduling, data reduction and reporting of results, and has multi-language capabilities. The software also provides in-process control, validation and verification.
2. Principles of Operation:
The ETI MAX and its predicate devices share the same principle of operation. The device consists of pipetting, washing and incubation station and uses the same basic technology to provide specimen pipetting and dilution, reagent pipetting, microplate washing and incubation.
{1}
The instrument is programmed to mimic the steps of the manual assay procedures for EIA/ELISA microtiter plate assays.
3. Modes of Operation:
The ETI-MAX 3000™ Analyzer is an open tube, batch mode analyzer with a continuous load option. The reagent bottles used from the test kit are placed on the instrument with the caps removed. The sample tubes can be the primary tubes with stoppers removed or the serum/plasma can be poured off into identified test tubes.
4. Specimen Identification:
There are multiple ways to identify the patient. Each ETI-MAX 3000™ sample rack holds 20 patient specimens. The sample rack has a barcode label at each position which identifies the rack and the unique sample position (1-20). There are 12 spaces (channels) numbered 1-12 which holds 1 sample rack each. The samples are placed in the rack and then put onto the instrument where the barcode scanner reads the barcodes on the rack.
1. The customer can use patient barcode identifiers if so desired. Patients are placed into samples racks, put on the instrument and the barcode scanner scans the rack, the rack barcode identifiers and the patient sample barcode identifier in each position.
2. LIS system – patient ID’s and test requests can be sent to the instrument through the lab LIS.
3. Manual identification
Patient names/ID’s can be entered via Patient Editor
The ETI-MAX 3000™ Analyzer can be set to pipette the samples in ascending order, descending order or None (no order).
- Ascending order – the samples will be pipetted in ascending order whether they are in the same rack or spread throughout several racks.
- Descending order – the samples will be pipetted in descending order whether they are the same rack or spread throughout several racks.
- None – if this option is selected, the instrument will pipette the all the samples in positions 1-20, starting in channel 12 (which is always the 1st position loaded) then once all the samples are pipetted in that rack, it will move to the next rack (channel 11) and pipette those samples (1-20).
5. Specimen Sampling and Handling:
The DiaSorin assays programmed on the ETI-MAX 3000™ do not use whole blood. All the DiaSorin assays use either serum or plasma as the sample type. Samples can be loaded onto the instrument in their primary tubes, or they can be poured off into test tubes ranging in size from 10 -16 mm in diameter.
If a dilution step is required, the instrument dispenses sample into the specimen diluent. Once this is completed the samples is then aspirated and dispensed into
2
{2}
the dilution tube a couple of times to ensure a good mixing of the sample and diluent.
The ETI-MAX 3000™ also performs a shaking of the microtiter plate once the specimens and reagents are added to the wells.
The operator is responsible for placing samples in the sample racks and loading or unloading the racks from the ETI-MAX 3000™ analyzer.
6. Calibration:
Each of the instrument modules require calibration. The calibration procedure is described in the Service manual of the instrument. The calibration procedures do not use whole blood or commercial calibration materials.
The calibration procedures use a Dye Test that is performed by the Service Engineers at the time of Installation, whenever service to one of the modules is performed and at the 6 month PM intervals.
Each assay also has its own calibration (assay verification) to ensure the assay is valid. Each of these verification specifications that are used for the manual assay are programmed into the ETI-MAX™ 3000 protocol and are used to ensure the Validity of the assay on the ETI-MAX 3000™.
7. Quality Control:
The ETI-MAX 3000™ Analyzer in and of itself does not require quality control procedures or commercial quality control materials. Each individual assay that is programmed on the ETI-MAX 3000™ follows the quality control information as recommended in the Instructions for Use for that assay. The quality control used on the ETI-MAX 3000™ would be the same as required for running the assay by the manual method.
Additional commercial quality control materials are used by the operator for the assay as required by their laboratory or regulating agency.
A quality control analysis report (Levy-Jennings plot), are available for tracking the results of a selected quality control material over a period of time. The QA Analysis report has to be activated on the ETI-MAX 3000™ by defining QA Labels in the Lot specific Values dialogue box, while preparing a worklist.
The QA-Label must match the label which is used in the assay layout definition.
8. Software:
FDA has reviewed applicant’s Hazard Analysis and Software Development processes for this line of product types:
Yes ☑ or No ☐
F. Regulatory Information:
{3}
1. Regulation section: 21 CFR 862.2170 (Micro chemistry analyzer for clinical use)
2. Classification: Class I
3. Product code: JJF
4. Panel: Chemistry (75)
G. Intended Use:
1. Indication(s) for Use: The ETI-MAX 3000™ is a fully automated EIA microtiter plate analyzer designed to perform the complete sample processing of qualitative and semi-quantitative assays with respect to (sample dilutions, sample and reagent dispensing, incubations, wash processes, plate transports) as well as the photometric measurement and evaluation. The qualitative and semi-quantitative performance of the ETI-MAX™ instrument were assessed using the DiaSorin ANAScreen ELISA kit, the DiaSorin Anti-SS-A(Ro)ELISA kit and the EuroDiagnostica AB ENA Single Well Screen kit.
2. Special Conditions for Use Statement(s): For Prescription Use only
H. Substantial Equivalence Information:
1. Predicate Device Name(s) and 510(k) numbers: Labotech (a.k.a ETI-LAB) (K922081) Mago™ Automated EIA Processor (K973177)
2. Comparison with Predicate Device:
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate (Labotech) |
| Intended Use/Indications for Use | The DiaSorin ETI-MAX 3000™ is a fully automated microtiter plate analyzer designed to perform the complete sample processing (sample dilutions, sample and reagent dispensing, incubations, wash processes, plate transports) as well as the photometric measurement and evaluation. | The Labotech is a fully automated microtiter plate analyzer designed to perform the complete sample processing (sample dilutions, sample and reagent dispensing, incubations, wash processes, plate transports) as well as the photometric measurement and evaluation. |
| Barcode Reading | Yes | Yes |
| Internal | X | X |
| External Hand Held *Optional | X | X |
| Sample Identification | X | X |
| Batch Loading | X | X |
| Sample / Reagent Dispensing Unit | Yes | Yes |
{4}
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate (Labotech) |
| Disposable Tips | X | X |
| Metal Needle * optional | | X |
| Pipetting area capacity (# of plates) | 4 | 3 |
| Mixing | X | X |
| Multi-Dispensing | X | X |
| Liquid Level Detection | Yes | Yes |
| Clot Detection | Yes | Yes |
| Incubation Assembly | Yes | Yes |
| Temperature controlled chambers | X | X |
| Reading Assembly | Yes | Yes |
| # of Chambers | 1 | 3 |
| # of Channels (per chamber) | 8 | 8 |
| Vertical with Photodiodes | X | X |
| Absorbance Reading | X | X |
| Method | | |
| Single or Double Beam w/over range filter | X | X |
| Spectrum (400nm to 700 nm) | X | X |
| Filters (404, 450, 492, 550, 620) | X | X |
| Washer Assembly | Yes | Yes |
| Buffer Capacity | 4 | 2 |
| Wash Head (8 channel, dual needle) | X | X |
| Buffer Level Sensing | X | X |
| Waste Tank Level Sensing | X | X |
| Standards Met | IEC 61010
1:90+A1:92+A2:95
EN 61326 - 1:97 + A1:98
IEC 60027
ASTM E 1381 - 91
ASTM E 1394 - 91 | IEC 1010-1/1990
IEC 801/1988
IEC 825
EN 55011 |
| Differences | | |
| --- | --- | --- |
| Item | Device | Predicate (Labotech) |
| Barcode Reading
Reagent Identification | X | |
| Loading Random Access | X | |
| Liquid Level Detection | Electronic | Pneumatic |
| Clot Detection | Electronic | Pneumatic |
| (Incubation) # of Chambers | 4 | 3 |
| Shaking (longitudinal) | X | |
| (Reading) # of Chambers | 1 | 3 |
| Kinetic Reading | X | |
{5}
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate (Mago) |
| Description/ Intended Use/ Indications for Use | The DiaSorin ETI-MAX 3000™ is a fully automated microtiter plate analyzer designed to perform the complete sample processing (sample dilutions, sample and reagent dispensing, incubations, wash processes, plate transports) as well as the photometric measurement and evaluation. | The Diamedix Mago™ Automated EIA Processor is indicated for use as a general purpose automated EIA processor for use in clinical laboratories. |
I. Special Control/Guidance Document Referenced (if applicable):
NCCLS EP5-A2
NCCLS EP17-A
J. Performance Characteristics:
1. Analytical Performance:
a. Accuracy:
The clinical studies were conducted at DiaSorin Inc. and at two external US laboratories. Testing was performed on 159 repository samples in each of the three assays. The breakdown for the 159 ANA repository samples is as follows:
- 88 negative samples (negative result using manual assay)
- 64 positive samples (positive result using manual assay)
- 7 equivocal samples (equivocal result using manual assay)
The breakdown for the 159 ENA repository samples is as follows:
- 92 negative samples (negative result using manual assay)
- 63 positive samples (positive result using manual assay)
- 4 equivocal samples (equivocal result using manual assay)
The breakdown for the 159 Anti-SS-A (Ro) repository samples is as follows:
- 104 negative samples (negative result using manual assay)
- 55 positive samples (positive result using manual assay)
- 0 equivocal samples (equivocal result using manual assay)
Study Analysis
All assay results were expressed in OD and S/CO. In addition, the Anti-SS-A (Ro) included U/mL. A sample was considered negative, equivocal, or positive based on the criteria as specified in the currently released assay package inserts.
ANA Screen:
- negative <0.7 S/CO.
- equivocal ≥0.7 and <1.0 S/CO.
- positive ≥1.0 S/CO.
{6}
7
ENA Screen:
- negative <0.95 S/CO.
- equivocal ≥0.95 and ≤1.0 S/CO.
- positive >1.0 S/CO.
Anti-SS-A (Ro) Qualitative:
- negative <0.95 S/CO.
- equivocal ≥0.95 and ≤1.0 S/CO.
- positive >1.0 S/CO.
Anti-SS-A (Ro) Quantitative:
- negative ≤2.0 U/mL.
- positive >2.0 U/mL.
The performance of the ETI-Max was determined by calculating the percent agreement among negative samples, percent agreement among positive samples, and overall percent agreement with the reference (manual) method for each assay. The relevant 95% confidence limits were computed by applying the exact method.
Results:
The tables below show the comparison of the ETI-Max and Manual assay results for the retrospective samples population. The results are summarized below as positive, negative, and overall percent agreement with the Manual assay results with 95% exact confidence intervals.
| ANA Screen | Manual Method | | | |
| --- | --- | --- | --- | --- |
| ETI-MAX 3000™ | Positive | Negative | Equivocal | Total |
| Positive ≥1.0 S/CO | 62 | 0 | 0 | 62 |
| Negative <0.7 S/CO | 0 | 88 | 3 | 91 |
| Equivocal ≥0.7 and <1.0 S/CO | 2 | 0 | 4 | 6 |
| Total | 64 | 88 | 7 | 159 |
Positive agreement = 96.9% (62/64) (Exact 95% CI: 89.2 – 99.6%)
Negative agreement = 100.0% (88/88) (Exact 95% CI: 95.9 – 100.0%)
Overall agreement = 96.9% (154/159) (Exact 95% CI: 92.8 – 99.0%)
| ENA Screen | Manual Method | | | |
| --- | --- | --- | --- | --- |
| ETI-MAX 3000™ | Positive | Negative | Equivocal | Total |
| Positive >1.0 S/CO | 59 | 0 | 0 | 59 |
{7}
8
| Negative <0.95 S/CO | 4 | 92 | 4 | 100 |
| --- | --- | --- | --- | --- |
| Equivocal ≥0.95 and ≤1.0 S/CO | 0 | 0 | 0 | 0 |
| Total | 63 | 92 | 4 | 159 |
Positive agreement = 93.7% (59/63) (Exact 95% CI: 87.6 – 99.7%)
Negative agreement = 100.0% (92/92) (Exact 95% CI: 96.1 – 100.0%)
Overall agreement = 95.0% (151/159) (Exact 95% CI: 90.3 – 97.8%)
| Anti-SS-A (Ro) Qualitative | Manual Method | | | |
| --- | --- | --- | --- | --- |
| ETI-MAX 3000™ | Positive | Negative | Equivocal | Total |
| Positive >1.0 S/CO | 55 | 0 | 0 | 55 |
| Negative <0.95 S/CO | 0 | 104 | 0 | 104 |
| Equivocal ≥0.95 and ≤1.0 S/CO | 0 | 0 | 0 | 0 |
| Total | 55 | 104 | 0 | 159 |
Positive agreement = 100.0% (55/55) (Exact 95% CI: 93.5 – 100.0%)
Negative agreement = 100.0% (104/104) (Exact 95% CI: 96.5 – 100.0%)
Overall agreement = 100.0% (159/159) (Exact 95% CI: 97.7 – 100.0%)
| Anti-SS-A (Ro) Quantitative | Manual Method | | |
| --- | --- | --- | --- |
| ETI-MAX 3000™ | Positive | Negative | Total |
| Positive >2.0 U/mL | 54 | 1 | 55 |
| Negative ≤2.0 U/mL | 0 | 104 | 104 |
| Total | 54 | 105 | 159 |
Positive agreement = 100.0% (54/54) (Exact 95% CI: 93.4 – 100.0%)
Negative agreement = 99.0% (104/105) (Exact 95% CI: 94.8 – 100.0%)
Overall agreement = 99.4% (158/159) (Exact 95% CI: 96.6 – 100.0%)
| | R Square | Intercept | | X Variable 1 | |
| --- | --- | --- | --- | --- | --- |
| | | Lower 95.0% | Upper 95.0% | Lower 95.0% | Upper 95.0% |
| ANA Screen | 0.974 | -0.007 | 0.056 | 0.779 | 0.82 |
| ENA Screen | 0.979 | 0.039 | 0.122 | 0.754 | 0.79 |
| SSA (Semi-quantitative) | 0.953 | -0.544 | 0.345 | 0.816 | 0.876 |
| SSA (Qualitative) | 0.983 | -0.125 | -0.046 | 0.955 | 0.996 |
{8}


{9}


# b. Precision/Reproducibility:
The instrument reproducibility/precision study was conducted at two external US laboratories and at DiaSorin Inc A coded panel comprised of 8 frozen
{10}
repository serum samples was prepared at DiaSorin and provided to each site for testing by the DiaSorin ETI-MAX $3000^{\mathrm{TM}}$ instrument and the three autoimmune assays. The samples were at or near the cut-off. All panel members were divided into aliquots and stored frozen prior to testing. The same coded panel was tested at all three sites, using four replicates per run in one run per day during five operating days at each trial site on the ETI-MAX $3000^{\mathrm{TM}}$ and manually.
Table of Precision for ANA Screen ETI-MAX 3000™ Method:
| | | | mean | within run | between run | total | instrument to instrument |
| --- | --- | --- | --- | --- | --- | --- | --- |
| ID# | Site | N | (S/CO) | %CV | %CV | %CV | %CV |
| #1 | 1 | 20 | 0.98 | 6.6 | 0 | 6.3 | |
| | 2 | 20 | 0.86 | 8.5 | 25.7 | 27.1 | 6.8 |
| | 3 | 20 | 0.93 | 20.6 | 0 | 20.4 | |
| #2 | 1 | 20 | 1.05 | 3.1 | 2.6 | 4.0 | |
| | 2 | 20 | 1.12 | 1.9 | 1.1 | 2.2 | 3.2 |
| | 3 | 20 | 1.09 | 2.0 | 0.5 | 2.0 | |
| #3 | 1 | 20 | 0.73 | 4.0 | 2.6 | 4.8 | |
| | 2 | 20 | 0.51 | 4.2 | 50.4 | 50.6 | 18.6 |
| | 3 | 20 | 0.71 | 4.3 | 3.2 | 5.4 | |
| #4 | 1 | 20 | 1.21 | 2.3 | 3.3 | 4.1 | |
| | 2 | 20 | 1.29 | 1.5 | 3.3 | 3.6 | 3.1 |
| | 3 | 20 | 1.25 | 4.3 | 0 | 4.2 | |
| #5 | 1 | 20 | 1.17 | 4.0 | 2.6 | 4.8 | |
| | 2 | 20 | 1.22 | 2.4 | 4.0 | 4.6 | 2.7 |
| | 3 | 20 | 1.23 | 2.7 | 2.0 | 3.4 | |
| #6 | 1 | 20 | 1.26 | 3.0 | 3.0 | 4.3 | |
| | 2 | 20 | 1.08 | 10.9 | 36.4 | 37.9 | 8.4 |
| | 3 | 19 | 1.23 | 22.0 | 0 | 21.8 | |
| #7 | 1 | 20 | 1.33 | 2.0 | 3.6 | 4.2 | |
| | 2 | 20 | 1.34 | 2.7 | 7.0 | 7.5 | 2.8 |
| | 3 | 20 | 1.40 | 3.0 | 0 | 2.9 | |
| #8 | 1 | 20 | 1.29 | 3.0 | 0.3 | 3.0 | |
| | 2 | 20 | 1.28 | 2.2 | 16.0 | 16.1 | 3.2 |
| | 3 | 20 | 1.36 | 2.9 | 4.2 | 5.1 | |
{11}
Table of Precision for ENA 6 Screen ETI-MAX 3000™ Method:
| | | | mean | within run | between run | total | instrument to instrument |
| --- | --- | --- | --- | --- | --- | --- | --- |
| ID# | Site | N | (S/CO) | %CV | %CV | %CV | %CV |
| #1 | 1 | 20 | 0.52 | 7.1 | 5.1 | 8.7 | |
| | 2 | 20 | 0.48 | 3.0 | 8.9 | 9.4 | 4.5 |
| | 3 | 20 | 0.50 | 6.0 | 5.7 | 8.3 | |
| #2 | 1 | 20 | 1.61 | 4.5 | 4.3 | 6.2 | |
| | 2 | 20 | 1.82 | 3.5 | 2.4 | 4.2 | 6.2 |
| | 3 | 20 | 1.70 | 2.8 | 4.4 | 5.2 | |
| #3 | 1 | 20 | 1.60 | 3.4 | 6.4 | 7.3 | |
| | 2 | 20 | 1.63 | 6.8 | 5.9 | 8.9 | 1.6 |
| | 3 | 20 | 1.66 | 4.0 | 3.0 | 5.0 | |
| #4 | 1 | 20 | 1.81 | 5.3 | 7.5 | 9.1 | |
| | 2 | 20 | 1.76 | 5.3 | 19.6 | 20.3 | 3.0 |
| | 3 | 20 | 1.87 | 2.8 | 4.2 | 5.0 | |
| #5 | 1 | 20 | 1.25 | 5.0 | 4.9 | 7.0 | |
| | 2 | 20 | 1.20 | 10.9 | 20.8 | 23.4 | 3.1 |
| | 3 | 20 | 1.27 | 5.4 | 3.2 | 6.3 | |
| #6 | 1 | 20 | 1.45 | 4.2 | 3.7 | 5.6 | |
| | 2 | 20 | 1.51 | 6.0 | 8.8 | 10.7 | 2.0 |
| | 3 | 20 | 1.50 | 4.4 | 4.6 | 6.3 | |
| #7 | 1 | 20 | 1.30 | 3.2 | 2.9 | 4.3 | |
| | 2 | 20 | 1.50 | 13.7 | 10.2 | 17.1 | 8.3 |
| | 3 | 20 | 1.30 | 9.9 | 5.1 | 11.1 | |
| #8 | 1 | 20 | 1.47 | 5.8 | 3.9 | 7.0 | |
| | 2 | 20 | 1.59 | 7.9 | 10.2 | 12.9 | 5.0 |
| | 3 | 20 | 1.45 | 3.4 | 6.5 | 7.4 | |
{12}
Table of Precision for Anti-SS-A (Ro) ETI-MAX 3000™ Method:
| | | | mean | within run | between run | total | instrument to instrument |
| --- | --- | --- | --- | --- | --- | --- | --- |
| ID# | Site | N | (S/CO) | %CV | %CV | %CV | %CV |
| #1 | 1 | 20 | 0.81 | 4.2 | 6.1 | 7.4 | |
| | 2 | 20 | 0.80 | 8.1 | 5.9 | 10.0 | 3.7 |
| | 3 | 20 | 0.86 | 7.1 | 0 | 7.1 | |
| #2 | 1 | 20 | 0.93 | 3.1 | 4.8 | 5.7 | |
| | 2 | 20 | 0.88 | 4.7 | 2.8 | 5.5 | 2.9 |
| | 3 | 20 | 0.91 | 2.6 | 3.1 | 4.1 | |
| #3 | 1 | 20 | 1.00 | 3.1 | 5.0 | 5.9 | |
| | 2 | 20 | 0.84 | 8.4 | 42.6 | 43.4 | 12.1 |
| | 3 | 20 | 1.07 | 5.1 | 0 | 4.9 | |
| #4 | 1 | 20 | 1.00 | 4.3 | 3.0 | 5.2 | |
| | 2 | 20 | 0.80 | 4.8 | 42.0 | 42.3 | 12.2 |
| | 3 | 20 | 1.00 | 3.6 | 4.7 | 5.9 | |
| #5 | 1 | 20 | 0.99 | 4.3 | 3.7 | 5.6 | |
| | 2 | 20 | 1.04 | 4.3 | 2.7 | 5.1 | 4.7 |
| | 3 | 20 | 1.08 | 7.0 | 7.7 | 10.4 | |
| #6 | 1 | 20 | 1.01 | 3.4 | 3.9 | 5.2 | |
| | 2 | 20 | 0.99 | 3.7 | 3.8 | 5.3 | 2.5 |
| | 3 | 20 | 1.04 | 3.8 | 2.6 | 4.6 | |
| #7 | 1 | 20 | 1.10 | 4.2 | 3.7 | 5.5 | |
| | 2 | 20 | 1.05 | 12.7 | 6.5 | 14.3 | 8.3 |
| | 3 | 20 | 1.23 | 6.6 | 3.9 | 7.6 | |
| #8 | 1 | 20 | 1.00 | 3.9 | 3.4 | 5.2 | |
| | 2 | 20 | 1.03 | 4.3 | 1.4 | 4.5 | 4.7 |
| | 3 | 20 | 1.09 | 4.6 | 7.4 | 8.7 | |
c. Linearity:
Four samples were serially diluted across the assay range to study the linearity of the ANA Screen, ENA 6 Screen and Anti-SS-A (Ro) semiquantitative assays using the manual and ETI-MAX methods. Samples were tested in singlicate. Acceptance Criteria: The ETI-MAX $3000^{\mathrm{TM}}$ result within $+/-1$ dilution of the manual result.
Results:
{13}
| | R Square | Intercept | | X Variable 1 | |
| --- | --- | --- | --- | --- | --- |
| | | Lower 95.0% | Upper 95.0% | Lower 95.0% | Upper 95.0% |
| ANA Screen | 0.982 | -0.112 | -0.067 | 1.196 | 1.307 |
| ENA Screen | 0.983 | -0.037 | 0.059 | 1.197 | 1.324 |
| Anti-SS-A (Ro) | 0.996 | 0.036 | 0.066 | 1.169 | 1.216 |


{14}

# d. Carryover:
# Study Design
A Reference Plate consisting of 5 strips of Negative samples $(n = 40)$ was tested in each assay (ANA Screen, ENA 6 Screen, and Anti-SS-A (Ro) to establish the baseline for comparison during the carry over tests. The mean, SD and $+ / - 3$ SD range of the negative sample on the reference plate were calculated.
# Pipettor and Washer Carryover Tests
One high dose sample was tested in replicates of four, followed by a negative sample tested in replicates of eight (in 3 separate sequences) to evaluate Pipettor carryover.
Two high dose sample strips $(n = 16)$ followed by 2 negative sample strips $(n = 16)$ were tested to evaluate washer carryover.
Acceptance Criteria: The mean of the negative samples on the Pipettor and Washer Carry Over Test plates within the 3SD range of the Negative Sample on the Reference Plate.
# Results
ANA Screen
| | Reference Plate Negative Sample | Pipettor Carry Over Negative Sample | Washer Carry Over Negative Sample |
| --- | --- | --- | --- |
| Mean OD | 0.141 | 0.143 | 0.141 |
| SD | 0.014 | | |
| +/- 3SD Range | 0.099 – 0.183 | Pass | Pass |
ENA 6 Screen
{15}
| | Reference Plate Negative Sample | Pipettor Carry Over Negative Sample | Washer Carry Over Negative Sample |
| --- | --- | --- | --- |
| Mean OD | 0.160 | 0.141 | 0.128 |
| SD | 0.009 | | |
| +/- 3SD Range | 0.133 – 0.187 | Pass | Pass |
Anti-SS-A (Ro)
| | Reference Plate Negative Sample | Pipettor Carry Over Negative Sample | Washer Carry Over Negative Sample |
| --- | --- | --- | --- |
| Mean OD | 0.105 | 0.098 | 0.093 |
| SD | 0.012 | | |
| +/- 3SD Range | 0.069 – 0.141 | Pass | Pass |
e. Interfering Substances:
Not applicable
2. Other Supportive Instrument Performance Data Not Covered Above:
Analytical Sensitivity:
Analytical sensitivity was established using parametric methods for the determination of Limit of Detection (LOD) according to NCCLS EP17-A. Limit of Blank (LOB) was estimated from a minimum of 60 measurements of multiple blank (analyte-free) samples. The Blank standard deviation was multiplied by 1.645 ( $95^{\text{th}}$ percent of the standard Gaussian distribution) and added to the Blank average. In order to establish LOD, a sample standard deviation was obtained from repeated measurement of several low-concentration samples; a minimum of 60 results were then used to calculate a pooled standard deviation. The pooled standard deviation was multiplied by a term $C_{\beta}$ , representing the $95^{\text{th}}$ percentile of the standard Gaussian distribution and a correction factor, and then added to LOB to establish LOD.
| Assay | Manual | | | | ETI-MAX 3000™ | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | n | LOB | n | LOD | n | LOB | n | LOD |
| ANAScreen | 68 | 0.087 | 70 | 0.144 | 68 | 0.116 | 71 | 0.133 |
| ENAScreen | 90 | 0.070 | 88 | 0.122 | 90 | 0.106 | 88 | 0.136 |
| Anti-SSA(Ro) | 82 | 0.084 | 80 | 0.133 | 74 | 0.109 | 104 | 0.138 |
Assay Drift:
Eight replicates of a low positive sample were tested at the beginning of the plate in strip 2 (just after the reference sample and assay controls) and again at the end of the plate in strip 12 to evaluate assay drift. The mean, SD and $+/-2$ SD range for S/CO and OD were calculated. OD and U/mL were calculated for the Quantitative Anti-SSA assay.
{16}
Acceptance Criteria: The low positive S/CO, U/mL and OD results from strip 12 must fall within the 2 SD range (S/CO, U/mL and OD) calculated for strip 2. If the results from strip 12 do not fall within the 2SD ranges the assay will need to be repeated.
## Results
### ANA Screen
| | Strip 2 Low Positive Sample | | Strip 12 Low Positive Sample | |
| --- | --- | --- | --- | --- |
| | OD | S/CO | OD | S/CO |
| Mean | 0.315 | 0.757 | 0.303 | 0.728 |
| SD | 0.010 | 0.023 | | |
| +/- 2SD Range | 0.295-0.335 | 0.711-0.803 | Pass | Pass |
### ENA 6 Screen
| | Strip 2 Low Positive Sample | | Strip 12 Low Positive Sample | |
| --- | --- | --- | --- | --- |
| | OD | S/CO | OD | S/CO |
| Mean | 0.418 | 1.502 | 0.379 | 1.363 |
| SD | 0.024 | 0.085 | | |
| +/- 2SD Range | 0.370-0.466 | 1.332-1.672 | Pass | Pass |
### Anti-SS-A (Ro) Semi-quantitative
| | Strip 2 Low Positive Sample | | Strip 12 Low Positive Sample | |
| --- | --- | --- | --- | --- |
| | OD | U/mL | OD | U/mL |
| Mean | 0.796 | 7.31 | 0.773 | 7.04 |
| SD | 0.023 | 0.27 | | |
| +/- 2SD Range | 0.751-0.841 | 6.77-7.86 | Pass | Pass |
### Dye test data to determine the accuracy of pipettor and washer:
Upon request, the sponsor provided the dye tests data which is used to determine the accuracy of the pipettor and washer. A summary of the results of the pipettor and washer dye test along with the Acceptance/Validation Criteria is listed below. These tests were performed on the ETI-MAX 3000™ Analyzer asset number 10142 at the DiaSorin site. A printout from the instrument for each test is attached for your review.
### Verification of Residual Volume (Washer)
| Assay Name | Criteria | Result |
| --- | --- | --- |
| Val-Wash.Asy | Max(W1) < 1.2% | 0.675 < 1.2 |
| | W1 < 0.4% | 0.310313 < 0.4 |
### Verification of Dilution 1:100 and 10μL+200μL
| Assay Name | Criteria | Result |
| --- | --- | --- |
| Dilution.asy | CV(W1) <= 10% | 1.66693 <= 10 |
| | CV(D) <= 5% | 2.56137 <= 5 |
{17}
Verification of Multi-Dispensing 8x100μL
| Assay Name | Criteria | Result |
| --- | --- | --- |
| 8x100.asy | CV(W1) <= 2.5% | 1.47316 <= 2.5 |
Verification of Single Pipette $100\mu \mathrm{L}$ and $50\mu \mathrm{L}$
| Assay Name | Criteria | Result |
| --- | --- | --- |
| 100x50.asy | CV(W1) <= 2.5% | 0.784174 <= 2.5 |
| | CV(W2) <= 5% | 3.30849 <=5 |
These tests are performed at the time of installation of the instrument, every 6 months for Preventative Maintenance and whenever service on the pipettor is performed.
# K. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
# L. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.