K052751 · Mediserve Information Systems, Inc. · MOD · Dec 23, 2005 · Anesthesiology
Device Facts
Record ID
K052751
Device Name
VENTLINK SYSTEM
Applicant
Mediserve Information Systems, Inc.
Product Code
MOD · Anesthesiology
Decision Date
Dec 23, 2005
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5895
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
The VentLink system is intended for use with specified or validated ventilator devices to obtain ventilator data and provide the user with the ability to display, store, print and otherwise process that data to other interfaced systems. VentLink is not connected directly to any patient nor does the VentLink device remotely control any connected ventilator device.
Device Story
VentLink is a server-based interface application for ventilator data management; it functions as a service within the MediLinks clinical information suite. The system connects via Ethernet to configured ventilators to collect data at set intervals; it does not connect directly to patients or control ventilators. Data is transmitted to workstations or wireless laptops for clinician review, validation, and storage. The system enables automatic data collection, retention settings, and event alerting (e.g., alarms). It acts as a central monitoring station for remote summary review. The device benefits patients by facilitating centralized, efficient clinical data management and reporting for healthcare providers in hospital environments.
Clinical Evidence
No clinical data. The device underwent alpha validation testing to confirm functionality and hazard mitigation, and beta testing in clinical environments to ensure compatibility with specified ventilator systems. Bench testing included EMI compatibility and performance testing of software and hardware components.
Technological Characteristics
Software-based interface application running on server/workstation hardware. Connectivity via Ethernet (100BaseT). Supports wireless laptops via Citrix MetaFrame. Compatible with Windows 2000/XP and SQL Server 2000. No proprietary hardware provided; utilizes off-the-shelf components. Data transmission is non-altering.
Indications for Use
Indicated for use in any clinical setting while interfaced to specified or validated ventilator devices. Provides secondary display of ventilator data to configured workstations and wireless laptops; data may be made available to interfaced hospital information systems (HIS).
Regulatory Classification
Identification
A continuous ventilator (respirator) is a device intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Adult, pediatric, and neonatal ventilators are included in this generic type of device.
{0}------------------------------------------------
K05275/
## DEC 2 3 2005
# PREMARKET NOTIFICATION 510(k) SUMMARY As required by §807.92
#### Device Name - as required by 807,92(a)(2):
| Trade Name: | VentLink™ System VentLink is a Trade Mark of<br>MediServe Information Systems, Tempe, AZ |
|-----------------------------|------------------------------------------------------------------------------------------|
| Common/Classification Name: | Ventilator Data Management System/ Accessory To<br>Continuous Ventilator (respirator) |
| Classification Regulation: | 868.5895 |
| Device Class: | Class II (accessory to) |
| Product Code: | MOD |
| Panel: | Anesthesiology |
Premarket Notification submitter: MediServe Information Systems, 3225 South Hardy Drive, Ste. 101, Tempe, AZ 85282
Contact: Mark Ofori-Kyei, Director of Quality Assurance and Change Management
Preparation Date: 9/28/2005
#### A. LEGALLY MARKETED PREDICATE DEVICE – as required by 807.92(a)(3)
The VentLink device is substantially equivalent to the Ventnet system, K963633, submitted by Puritan Bennett Corporation, Tustin, CA.
#### B. DEVICE DESCRIPTION - as required by 807.92(a)(4)
The VentLink system is a device composed of MediServe Information Systems' proprietary software and various off-the-shelf hardware, operating system software and third-party proprietary software intended to be interfaced with specified or validated ventilator devices to allow authorized users to move ventilator patient data to wireless laptops and workstations configured with MediServe Information Systems' MediLinks data storage software'. Additionally, the VentLink system allows users to move ventilator data to the
<sup>1</sup> MediLinks data storage software is a Class I proprietary standalone software application of MediServe Information Systems, the submitter.
{1}------------------------------------------------
user's hospital information system (HIS). It is intended to be used in an environmentally controlled hospital and hospital type environment.
The VentLink system software allows users to:
- Enable or disable automatic device data collection. .
- Set a default data collection time interval. .
- Specify how long unauthenticated data will be retained. .
- Alert the user when data collection is paused. .
After user review and validation of results received via the Ethernet the results are stored short-term on the VentLink system and transmitted to MediLinks for acceptance and storage.
Specified or validated ventilator devices are interfaced and added to the VentLink system through VentLink's devices' dictionary.
The VentLink system does not alter or change any patient data, but does allow the user to display the received patient data.
VentLink will function within a wide array of connectivity modules which allow VentLink to adapt to the evolving connectivity technology.
VentLink also has advanced device-management functionality.
#### C. DEVICE CLAIMS - as required by 807.92(a)(4)
The VentLink™ system is an interface application for ventilator devices that runs as a service on a server. VentLink works in concert with MediLinks®, an integrated point-of-care application suite intended for managing clinical departmental information activities and tasks.
The VentLink service communicates via Ethernet with the ventilator devices that have been configured within MediLinks. The purpose of the VentLink service is to collect data from the designated ventilators at specific time intervals and store the data in a database utilized by MediLinks.
The VentLink service cannot be used without MediLinks. MediLinks provides the user interface to the VentLink service. A MediLinks user can enter new ventilators, set data collection intervals, create Patient - Ventilator associations, and review and authenticate ventilator data collected by the VentLink service. MediLinks allows the user to generate reports of monitored ventilator information and also provides means for archiving data.
MediLinks will be used as a Central Monitoring Station that will allow users to remotely review summary information from many different ventilators at a
{2}------------------------------------------------
glance. The system will also be configurable to page an appropriate user when certain events (e.g. alarms) occur.
#### PRODUCT AND TECHNICAL SPECIFICATIONS - as required by D. 807.92(a)(4)
The purpose of this section is to show recommended minimum specifications for the VentLink system. MediServe Information Systems does not supply any hardware. While specific requirements may vary from one customer to another, the specifications listed below provide general guidance in determining hardware and system software needed for VentLink.
#### Client/Server Workstations
| | Recommended |
|--------------|----------------------------------------|
| Workstations | |
| | • Pentium4 2.8 GHz or higher |
| | • 80 GB Hard Disk or greater |
| | • 512 MB RAM or greater |
| | • Windows 2000 Professional SP4 or |
| | • Windows XP Pro SP2 |
| | • 32-bit SVGA Graphics Card, 32 MB RAM |
| | • 1024x768 Display Resolution |
| | • SVGA Monitor (17") or greater |
- 100BaseT 32-bit PCI Network Card .
#### Thin Client Workstations
| | Recommended Recommended |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Workstations |
| | • Pentium4 2 GHz or higher |
| | 40 GB Hard Disk or greater |
| | 512 MB RAM or greater |
| | Windows 2000 Professional SP4 or |
| | Windows XP Pro SP1 |
| | Internet Explorer 6.0 or greater (with 128-bit |
| | encryption) |
| | 32-bit SVGA Graphics Card, 32 MB RAM |
| | 1024x768 Display Resolution |
| | SVGA Monitor (17") or greater |
| | 100BaseT 32-bit PC1 Network Card |
| Career Station Company Control Concession Company Controller Company Controller Company Controller Company Company Company Company Company Comments of Children | 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1<br>t the country of the county of the county of the first of the first of the first of the first of the first of the first of the first of the first of the first for the finishe |
Note: Thin client devices can be used for thin client computing via Citrix MetaFrame and must include Internet Explorer 5.5 or greater and support Citrix ICA and NFuse clients.
{3}------------------------------------------------
Database Server (up to 1200-1500 procedures/day or up to 50 concurrent users)
| Recommended | |
|-----------------|---------------------------------------------------------|
| Database Server | |
| • | Dual (2) Pentium4 2 GHz or greater |
| • | 3 GB RAM or greater |
| • | 4(+) x 36 GB Ultra 3/4 SCSI Hard Drives<br>(RAID-5) |
| • | 32-bit Caching Array Controller 64 MB RAM or<br>greater |
| • | 2 x 100BaseT 32-bit Network Card |
| • | Windows 2000 Server SP4 or Windows Server<br>2003 |
- Microsoft SQL Server 2000 Enterprise Ed. SP3 ●
- . Monitor, CD-ROM
# Database Server (up to 2500+ procedures/day or up to 100 concurrent users)
1 Recommended Recommended
#### Database Server
- Dual (2) Intel Xeon 3 GHz or greater. 4 . processors for larger volume sites
- . 4 GB RAM or greater
- . 6(+) x 36 GB or 72 GB Ultra 3/4 SCSI Hard Drives (RAID-5 or RAID 10)
- . 32-bit Multi-Channel Caching Array Controller 128 MB RAM or greater (the array should be constructed using multiple I/O channels) with additional embedded 64MB controller for system drive
- 2 x 100BaseT 32-bit Network Card .
- Windows 2000 Advanced Server SP4 or . Windows Advanced Server 2003
- Microsoft SQL Server 2000 Enterprise Edition . SP3
- Monitor, CD-ROM .
{4}------------------------------------------------
Thin Client Server(s)
| Recommended |
|----------------------------------------------------------|
| Citrix MetaFrame Server (up to 30 users) |
| • Dual (2) Xeon 2.8 GHz or greater |
| • 3 GB RAM or greater (64 MB per concurrent user) |
| • 3 x 36 GB Ultra 3/4 SCSI Hard Drives (RAID-5) or |
| • 2 x 72 GB Ultra 3/4 SCSI Hard Drives (RAID-1) |
| • 32-bit Caching Array Controller 64 MB RAM or greater |
| • 2 x 100BaseT 32-bit Network Card |
| • Windows 2000 Server SP4 or Windows Server 2003 |
| • Citrix MetaFrame v1.8 or XP (XPs, XPa, XPe) |
| • Citirx Load Balancing is required for multiple servers |
- Monitor, CD-ROM .
## Interface Computers
| | Recommended |
|--------------------|--------------------------------------------------------|
| Interface Computer | |
| | • Dual (2) Intel Xeon 2.4 GHz or greater |
| | • 3 GB RAM or greater |
| | • 2 x 72 GB Ultra 3/4 SCSI Hard Drives (RAID-1) |
| | • 32-bit Caching Array Controller 16 MB RAM or greater |
| | • 100BaseT 32-bit Network Card |
| | • Windows 2000 Server SP4 or Windows Server 2003 |
| | • Microsoft SQL Server 2000 Personal Edition SP3 |
- Monitor, CD-ROM .
{5}------------------------------------------------
#### Network Environment
# 科技术学院 - Recommended - Commended - Commended
## Report Writing
- . Crystal Reports v11 Pro
### Backup and Printing
- Laser Quality Printer .
- . Uninterruptible Power Supply w/Monitoring Software (all servers)
- . Tape Backup Unit (DAT, DLT, etc.) w/Backup Software (e.g. ArcServe)
## Cabling and Connectivity
- 100BaseT (100Mbps) Ethernet Multiport Hub . or Switch
- . Category 5e Patch cabling (CAT5e)
- Category 5e Facility cabling (CAT5e) .
## Remote Support
- Terminal Services or Symantec pcAnywhere to . Database Server and Interface Computers
- VPN Client connection via the Internet. .
## LAN/WAN Bandwidth Requirements
- . 128Kbps per concurrent user (for Citrix: 32Kbps per concurrent user)
#### Wireless Requirements
- MediLinks must connect using Citrix . MetalFrame.
#### ASP Requirements
- TCP/IP protocol support required ●
- . Host-to-Host or Client-Host VPN connections between MediServe ASP and facility are supported. Contact MediServe for details.
{6}------------------------------------------------
#### INTENDED USE - as required by 807.92(a)(5) E.
#### Intended use:
The VentLink system is intended for use with specified or validated ventilator devices to obtain ventilator data and provide the user with the ability to display, store, print and otherwise process that data to other interfaced systems.
VentLink is not connected directly to any patient nor does the VentLink device remotely control any connected ventilator device.
## Indications For Use
The VentLink system is intended for use in any clinical setting while interfaced to any specified or validated ventilator device. The VentLink system is intended to provide a secondary display of data received from an interfaced ventilator device to appropriately configured workstations and wireless laptop computers and to be made available to any interfaced hospital information system (HIS).
#### TECHNOLOGICAL CHARACTERISTICS SUMMARY - as required by 下. 807.92(a)(6)
The VentLink system has the same indications for use as the Ventnet system, K963633. The VentLink system has the same technological characteristics as the Ventnet system. Section III Substantial Equivalence of this submission provides a detailed COMPARISON MATRIX of the VentLink system to the predicate Ventnet system.
The submitter claims that the VentLink system is substantially equivalent to the predicate device, the Ventnet system.
The technological characteristics of the VentLink system are very similar to those of the Ventnet system. The differences include:
| Characteristic | VentLink system | Ventnet system |
|---------------------------|----------------------------------------|-------------------------------------------------|
| Hardware | No hardware provided. | Solution includes hardware. |
| Computer hardware | Specified but not provided by<br>user. | Solution includes computer<br>hardware. |
| Computer Operating System | Specified but not provided by<br>user. | Solution includes computer<br>operating system. |
| Color Monitor | Specified but not provided by<br>user. | Solution includes color monitor. |
## TABLE OF DIFFERENCES BETWEEN THE VENTLINK SYSTEM AND THE VENTNET SYSTEM
{7}------------------------------------------------
| Radio Transmitter Interface<br>(wireless) | Specified but not provided by<br>user. | Solution includes radio<br>transmitter interface. |
|-------------------------------------------|----------------------------------------|---------------------------------------------------|
| Remote Transceiver | Not specified. | Solution includes remote<br>transceiver. |
| Central Transceiver | Not specified. | Solution includes central<br>transceiver. |
The submitter concludes that the VentLink system employs the same type of technological characteristics including computer hardware, operating system(s), cabling, networking and similar functionality to the Ventnet system. The majority of differences relate to evolutionary changes in technology that has occurred since the release of the Ventnet system.
#### NON-CLINICAL PERFORMANCE DATA TESTING AND REVIEW - as G. required by 807.92(b)(1)
Since the submitter uses only off-the-shelf hardware components, only certifications provided by the manufacturers of those components were obtained to document that EMI compatibility and susceptibility meets acceptable industry levels. Those certifications are attached in Exhibit 10 Component/Accessory Validation.
The submitted device has undergone significant verification and validation testing. Alpha validation testing included testing of all functionality and confirmation that all identified hazards have been adequately addressed by software functionality, the user interface or documentation.
Alpha validation activities included specified system software and operating software performance and environmental testing within the specified environment.
Additionally, each specified ventilator system has been or will be fully validated by users in beta testing environments PRIOR to implementation to ensure that each specified or validated ventilator operating environment meets all appropriate specifications.
The submitter requires that any user of an un-validated specified or unspecified ventilator device perform a documented, successful, ventilator/VentLink validation prior to authorizing the user operate their system in a live environment. See Section IV, for a description of the User's Validation Protocol and Exhibit 12 Users' Manual, Implementation Manual and Users' Validation Protocol for a User's Validation Protocol provided to users of un-validated specified or unspecified ventilator devices.
The submitter's validation testing fully documents that specified ventilator device's data is transmitted from the ventilator through the VentLink system into
{8}------------------------------------------------
the submitter's MediLinks device2 with user control of data review and acceptance. No changes were made to transmitted ventilator data.
In these regards, the VentLink system is similar to and substantially equivalent to the capabilities of the predicate device, the Ventnet system.
The predicate device, Ventnet system, did not provide or reference any clinical tests submitted in compliance with 807.92(b)(2), therefore the submitter believes such clinical testing is not appropriate or required by FDA and has not made or provided any summary of such testing.
<sup>2</sup> MediLinks is a Trade Mark of MediServe Information Systems, Tempe, AZ
{9}------------------------------------------------
#### CONCLUSIONS AND SUMMARY- as required by 807.92(b)(3) and (c) H.
The nonclinical testing, in the form of alpha and beta validation studies document that the VentLink system is
- . Safe, and
- As effective as the predicate device, the Ventnet system, and .
- Performs as well or better than the predicate device, the Ventnet system, . and, is therefore,
- Substantially equivalent to the identified predicate device, the Ventnet . system.
{10}------------------------------------------------
Image /page/10/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles an abstract human figure or a bird in flight, composed of three curved lines.
DEC 2 3 2005
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. Mark Ofori-Kyei Official Correspondent Mediserve Information Systems, Incorporated 3225 Suoth Hardy Drive, Suite 101 Tempe, Arizona 85282
Re: K052751
Trade/Device Name: VentLinkTM System Regulation Number: 868.5895 Regulation Name: Continuous Ventilator Regulatory Class: II Product Code: MOD Dated: September 30, 2005 Received: September 30, 2005
Dear Mr. Kyei:
We have reviewed your Section 510(k) premarket notification of intent to market the device we have reviewed your becater mined the device is substantially equivalent (for the referenced above and have actoring/osure) to legally marketed predicate devices marketed in indications for use stated in the encreations of the enactment date of the Medical Device interstate commerce pror to fray 30, 20, 20, 2017 10:00 accordance with the provisions of Ameliuments, of to devices that nave o Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general approvin uppreation (the Act. The general controls provisions of the Act include controls provibions of the or annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III II your device is classified (500 additional controls. Existing major regulations affecting (I MA), it may oc subject to back adam f Federal Regulations, Title 21, Parts 800 to 898. In your device can be roundsh further announcements concerning your device in the Federal Register.
{11}------------------------------------------------
Page 2 - Mr. Kyei
Please be advised that FDA's issuance of a substantial equivalence determination does not Please be advised that IDA S Issualled of a substance ice complies with other requirements mean that FDA has made a decemination administered by other Federal agencies. of the Act of ally rederal statutes and regulations, but not limited to: registration You must colliply with an the Fee 3 requirement 801); good manufacturing practice and ifsing (21 CFR I art 607), laboling (21 CFR Part 820); and if (21 CFR Part 820); and if requirements as set forth in the quality bytellis (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) This letter will anow you to begin maileding your substantial equivalence of your device to a premarket notification. The PDA miams of sassification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), If you desire specific advice for your device on our career on and the regulation in entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the may obtain other general missimations and Consumer Assistance at its toll-free Division (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Smith Y. Michael Davis
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{12}------------------------------------------------
# INDICATIONS FOR USE
interfaced hospital information system (HIS).
K052751 [510(k)] Number: VentLink™ system Device Name: The VentLink System is intended to be used in any Indications For Use: THC VentEining while interfaced to any specified or validated ventilator device. The VentLink system is intended to provide a secondary display of data received from an interfaced ventilator device to appropriately configured workstations and wireless laptop computers and to be made available to any
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-the-Counter Use (21 CFR 807 Subpart C)
(Please DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Cure Suleom
K052751
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.