K052671 · Hoca Medical Microsystem, Inc. · DXN · Feb 3, 2006 · Cardiovascular
Device Facts
Record ID
K052671
Device Name
HOCA EZ-CARE BP108, TENDER-CARE BP109
Applicant
Hoca Medical Microsystem, Inc.
Product Code
DXN · Cardiovascular
Decision Date
Feb 3, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1130
Device Class
Class 2
Indications for Use
The device is a noninvasive blood pressure measurement system intended to measure the systolic and diastolic blood pressures and pulse rate of an adult individual, over age 18, at home by using a non-invasive technique in which an inflatable cuff is wrapped around the wrist. The cuff circumference is limited to be 5.25" - 7.75".
Device Story
Noninvasive blood pressure monitor; uses oscillometric method to measure systolic/diastolic pressure and pulse rate. Input: pressure oscillations sensed via wrist cuff and closed air pipe system. Processing: microcomputer automatically senses pulse signal characteristics; identifies systolic pressure as point where cuff oscillation amplitude begins to increase and diastolic pressure as point where amplitude stops decreasing. Output: digital display of blood pressure and pulse rate. Used at home by adult patients. Provides objective data for patient monitoring; assists in tracking blood pressure trends.
Clinical Evidence
Bench testing only. Compliance with EN 60601-1:1990 (electrical safety), EN 60601-1-2 (EMC), and ANSI/AAMI SP10-1992 (Electronic or Automated Sphygmomanometers).
Technological Characteristics
Oscillometric measurement principle; microcomputer-based signal processing; wrist-worn inflatable cuff (5.25"-7.75"); digital display. Standards: EN 60601-1, EN 60601-1-2, ANSI/AAMI SP10.
Indications for Use
Indicated for adults over age 18 for home-based noninvasive measurement of systolic/diastolic blood pressure and pulse rate via wrist cuff (5.25"-7.75" circumference).
Regulatory Classification
Identification
A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.
Predicate Devices
Automatic Digital Blood Pressure Monitor, APM BP108A (K040159)
Submission Summary (Full Text)
{0}------------------------------------------------
FEB S 2006
K052671
71/2
# HOCA Medical Microsystem, Inc.
No.336-2, Chien Kuo Road, Chu Nan Chen, Miaoli Shien, 35045, Taiwan Telephone (886) 37-481909 (886) 37- 484906 Fax # e-mail: neo@smarthoca.com Website : www.smarthoca.com
#### 510(k) SUMMARY " દા
HOCA Medical Microsystem, Inc. Submitter's Name:
> No.336-2, Chien Kuo Road, Miaoli Shien, 350, Taiwan Fax: 886-37-484906 Telephone: 886-37-481909 e-mail : neo@smarthoca.com
Date summary prepared:
Proprietary Name:
Device Name:
#### HOCA EZ-CARE BP108, TENDER-CARE BP109 NONINVASIVE BLOOD PRESSURE MEASUREMENT SYSTEM Common or Usual Name: Blood Pressure Monitor, Class II, Classification Name: 21 CFR 870.1130
September 19, 2005
Indications for Use:
The device is a noninvasive blood pressure measurement system intended to measure the systolic and diastolic blood pressures and pulse rate of an adult individual, over age 18, at home by using a non-invasive technique in which an inflatable cuff is wrapped around the wrist. The cuff circumference is limited to be 5.25" - 7.75".
### Description of the device:
HOCA EZ-CARE BP108 and TENDER-CARE BP109 use the Oscillometric method to measure the blood pressure. The Oscillometric method is adopted clinically to measure the blood pressure recently. It is not needed to use the stethoscope, as in the traditional measuring method, to monitor the Korotkov sound when deciding the systolic or diastolic pressure. The Oscillometric method senses the vibrating signal via the closed air pipe system and utilizes the microcomputer to automatically sense the characteristics of the pulse signal. Through simple calculation, clinically proven, the reading can reflect the accurate real blood pressure, and the systolic pressure is defined as the pressure when the cuff pressure oscillating amplitude begins to increase and the diastolic pressure as the pressure when the cuff pressure oscillating amplitude stops decreasing.
{1}------------------------------------------------
K052471
F2/2
Image /page/1/Picture/1 description: The image shows a logo with a stylized leaf shape in black and white. Inside the leaf, there is a white square with a black plus sign in the center. To the right of the leaf, there is a cursive word that appears to be "foca".
### HOCA Medical Microsystem, Inc.
No.336-2, Chien Kuo Road, Chu Nan Chen, Miaoli Shien, 35045, Taiwan (886) 37- 484906 Telephone (886) 37- 481909 Fax # e-mail: neo@smarthoca.com Website : www.smarthoca.com
Performance Testing:
Electric Safety Requirement Test Report of EN 60601-1:1990 and EMC test report of EN 60601-1-2 ( EN 55011:1991 and EN 61000-4-2:1995) ANSI/AAMI SP10-1992 Electronic or Automated Sphygmomanometers
Legally marketed device for substantial equivalence comparison:
Automatic Digital Blood Pressure Monitor, APM BP108A (K040159)
Summary for substantial equivalence comparison:
Same characteristics: intended use, technological characteristics, power supply, display, measuring range, accuracy, operating and storage environments. The new device HOCA BP108 are all the same as the predicate device APM BP108A.
Different characteristics: the new device HOCA BP109 has the different appearance for the "start key".
They are decided to be substantially equivalent.
{2}------------------------------------------------
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
FEB 3 2006
HOCA Medical Microsystem, Inc. c/o Dr. Jen Ke-Min ROC Chinese-European Industrial Research Society No. 58 Fu-Chiun St. Hsin-Chu City Taiwan ROC
Re: K052671
Trade Name: HOCA Ez-Care BP108 and Tender-Care BP109 Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II (two) Product Code: DXN Dated: January 12, 2006 Received: January 18, 2006
Dear Dr. Ke-Min.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Dr. Jen Ke-Min
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0210. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Bhimimor for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows a logo with a leaf-like shape on the left and the word "Coca" on the right. Inside the leaf shape, there is a plus sign. The leaf is shaded in black, and the word "Coca" is written in a cursive font.
# HOCA Medical Microsystem, Inc.
No.336-2, Chien Kuo Road, Chu Nan Chen, Miaoli Shien, 35045, Taiwan Telephone (886) 37-481909 (886) 37- 484906 Fax # e-mail: neo@smarthoca.com Website : www.smarthoca.com
### Indications for Use
510(k) Number:
K052671
Device Name: HOCA Medical Microsystem, Inc. HOCA EZ-CARE BP108 and TENDER-CARE BP109
. Indications for use:
The device is a noninvasive blood pressure measurement system intended to measure the systolic and diastolic blood pressures and pulse rate of an adult individual, over age 18, at home by using a non-invasive technique in which an inflatable cuff is wrapped around the wrist. The cuff circumference is limited to be 5.25" - 7.75".
Prescription Use
AND/OR
Over-The-Counter Use __
(Part 21 CFR 801 Subpart D)
(21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Blummmmas
(Division Sign-Off)
Division of Cardiovascular Devices
510(k) Number K052671
Cl
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.