NON-STERILE POWDER FREE NATURAL/GREEN LANO-E LATEX EXAM GLOVES, WITH OR WIHOUT CITRUS/PEPPERMINT SCENT PLUS A PROTEIN
K052666 · Ascend Eagle, Inc. · LYY · Jan 25, 2006 · General Hospital
Device Facts
Record ID
K052666
Device Name
NON-STERILE POWDER FREE NATURAL/GREEN LANO-E LATEX EXAM GLOVES, WITH OR WIHOUT CITRUS/PEPPERMINT SCENT PLUS A PROTEIN
Applicant
Ascend Eagle, Inc.
Product Code
LYY · General Hospital
Decision Date
Jan 25, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6250
Device Class
Class 1
Indications for Use
These patient examination gloves are disposable devices intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between the patient and the examiner.
Device Story
Non-sterile, powder-free natural or green latex examination gloves; coated with pharmaceutical-grade lanolin and Vitamin E; optional citrus/peppermint scent. Designed for use by healthcare providers in clinical settings to prevent cross-contamination between patient and examiner. Device acts as a physical barrier. Performance verified via ASTM D3578-01a standards, including tensile strength, elongation, and water-tight testing. Features protein labeling claim of <50 micrograms of total water-extractable protein per gram of glove to reduce potential for latex-related allergic reactions.
Clinical Evidence
Bench testing only. Device meets ASTM D3578-01a requirements for physical properties (tensile strength, elongation) and FDA 1000ml water-tight test. Biocompatibility testing passed. Residual protein levels confirmed at <50 micrograms/gram.
Technological Characteristics
Material: Latex. Cuff: Beaded. Powder residue: <2mg/glove. Standards: ASTM D3578-01a, EN 455-1/2/3:2000, ISO 2859-1:1999, ISO 9001:2000. Coating: Lanolin and Vitamin E (USP/Ph.Eur./DAB/BP compliant). Dimensions: 240mm length, 0.13-0.18mm palm thickness. Tensile strength: 21 MPa (before aging), 16 MPa (after aging).
Indications for Use
Indicated for use as a disposable medical glove worn on the examiner's hand or finger to prevent cross-contamination between patient and examiner. Suitable for over-the-counter use.
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
{0}------------------------------------------------
#### JAN 2 5 2006
#### SUMMARY PREMARKET 510(k) NOTIFICATION
Non-Sterile Powder-Free Natural/Green Lano-E Latex Exam Gloves, with or without Citrus/Peppermint scent plus a protein labeling claim.
510(k) Number (if known): K052666
Submission Applicant/Official Correspondent:
Jarilyn Lim, President Ascend Eagle Inc. 140-B, Dodd Court American Canyon, CA 94503 707-648-1526 Tel: Fax: 707-648-1534
Submitted: September 26, 2005
#### Description of the Device:
Trade and Proprietary Name: Non-Sterile Powder-Free Natural/Green Lano-E Latex Exam Gloves, with or without Citrus/Peppermint scent plus a protein labeling claim. (Multiple Private Labels) Common Name: Latex Examination Gloves Classification Name: Patient Examination Glove (per 21 CFR 880.6251) Class I: Powder-Free Latex examination glove 80LYY that meets all of the requirements of ASTM Standard D 3578-01aE2. Predicative Devices: Latex Powder-Free Examination Gloves
#### Intended Use of the Device:
These patient examination gloves are disposable devices intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between the patient and the examiner.
#### Substantial Equivalence:
Both in its intended use and/or physical characteristics, this device is equivalent to devices currently marketed by US companies. It is Substantially Equivalent to the devices manufactured by SGMP Co., LTD, except by coating, K032293, and by N.S. Uni-Gloves Sdn. Bhd., except by scent, K013163.
#### Summary of Technological Characteristics:
Material: Latex Cuff: Beaded Powder Residue: Maximum 2mg/glove Quality Assurance: In compliance with ASTM D3578-01a, EN 455-1: 2000, EN 455-2: 2000, EN 455-3: 2000, ISO 2859-1:1999 and manufactured under ISO 9001:2000.
{1}------------------------------------------------
Inspection Parameters:
| Criteria | Inspection Level | AQL |
|-------------------------|------------------|-----|
| Dimensions | S-2 | 4.0 |
| Physical Properties | S-2 | 4.0 |
| Water Tight Test 1000ml | G-1 | 1.5 |
| Visual Major Defects | G-1 | 1.5 |
| Visual Minor Defects | G-1 | 2.5 |
## Physical Properties:
| Dimensions: | |
|-------------------|-----------------------------------------------------|
| Overall Length: | 240 mm minimum |
| Width: | 95 mm minimum (for medium glove) |
| Palm Thickness: | 0.13 to 0.18 mm (at center of palm) |
| Finger Thickness: | 0.15 to 0.20 mm (at 15mm from tip of center finger) |
| Cuff Thickness: | 0.10 to 0.15 mm (at 40mm from the beaded end) |
### BEFORE AGING
Tensile Strength: 21. Mpa minimum Ultimate Elongation:700% minimum Pinhole AQL 1.5 minimum
AFTER AGING
16.0 Mpa minimum 500% minimum 1.5 minimum
#### Biocompatibility:
The biocompatibility test results are as per attached in Appendix B1 and B2 and show that the gloves passed the tests for examination gloves.
#### Residual Protein Level:
The extractable protein content test result is a per attached in Appendix B5 and show that the gloves meets the protein labeling claim level at < 50ugg.
Special Properties: Processed with pharmaceutical quality lanolin as the emollient and conditioning agent. Also contains Vitamin E which complies with the current USP, Ph.Eur., DAB and BP monographs.
Packaging: 100 pcs per dispenser box, 10 boxes per case, 1,000 gloves per case
Conclusion: These Powder-Free Latex Examination Gloves meet the physical property requirements of ASTM D 3578-01 and the FDA 1000 ml water test both before and after aging. It also meets the protein labeling claim level at < 50uggl
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the seal of the Department of Health and Human Services (HHS). The seal features an abstract eagle design with three stylized lines representing the eagle's wings. The eagle is positioned above a wavy line, likely representing water or a horizon. Encircling the eagle is the text "DEPARTMENT OF HEALTH AND HUMAN SERVICES, USA" in a circular arrangement.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JAN 2 5 2006
Ms. Jarilyn Lim President Ascend Eagle, Incorporated 140-B. Dodd Court American Canyon, California 94503
Re: K052666
Trade/Device Name: Non-Sterile, Powder-Free, Latex Examination Gloves, with Lanolin and Vitamin E Coating, Natural or Green Color, with or Without Scent (Combined Citrus Peppermint), and with Protein Labeling Claim (50 Micrograms or Less of total water Soluble Protein Per Gram of Glove) Regulation Number: 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LYY Dated: January 3, 2006 Received: January 9, 2006
Dear Ms. Lim:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially cquivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Fxisting major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
#### Page 2 -- Ms. Lim
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820): and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510/k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97), You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Carl
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# INDICATIONS FOR USE
## APPLICANT:
Jarilyn Lim, President Ascend Eagle Inc. 140-B, Dodd Court American Canyon, CA 94503
#### 510(k) NUMBER: K052666
## DEVICE NAME:
Non-Sterile, Powder-Free, Latex Examination Gloves, with Lanolin and Vitamin E Coating, Natural or Green Color, With or Without Scent (combined Citrus-Peppermint), and with Protein Labeling Claim (50 Micrograms or less of total water soluble protein per Gram of glove)
## Indications For Use:
The Non-Sterile, Powder-Free, Latex Examination Gloves, with Lanolin and Vitamin E Coating, Natural or Green Colonese &acmination (Hores, With Lanoin and Vitamin F
and with Protein Labelian Claim (Combined Citrus-Peppermint), and with Protein Labeling Claim (50 micrograms or less total water extractable protein per gram of glove) is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Prescription Use .............................. AND/ OR Per 21 CFR 801.109
Over-The -Counter Use ...... X ....... (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Shula P. Murphy 12/25/06
U.S. Army General Hospital,
Medical Control, Dental Devices
K052666
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.