← Product Code [CDZ](/productcode/CDZ) · K052649

# DRG SLV TESTOSTERONE ELISA TEST (K052649)

_Drg Intl., Inc. · CDZ · Jan 27, 2006 · Clinical Chemistry · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K052649

## Device Facts

- **Applicant:** Drg Intl., Inc.
- **Product Code:** [CDZ](/productcode/CDZ.md)
- **Decision Date:** Jan 27, 2006
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 862.1680
- **Device Class:** Class 1
- **Review Panel:** Clinical Chemistry

## Indications for Use

An Enzyme Immunoassay for the in vitro diagnostic quantitative measurement of free active testosterone in saliva. Measurement of testosterone is used in the diagnosis and treatment of disorders involving the male sex hormones (androgens), including primary and secondary hypogonadism, delayed or precocious puberty, impotence in males and, in females hirsutism (excessive hair) and virilization (masculinization) due to tumors, polycystic ovaries, and adrenogenital syndromes.

## Device Story

The DRG Salivary Testosterone ELISA Kit is an in vitro diagnostic assay for professional use. It utilizes a competitive enzyme immunoassay principle to measure free active testosterone in human saliva samples. The device consists of a microtiter plate coated with mouse anti-testosterone antiserum, enzyme-conjugated testosterone (horseradish peroxidase), substrate solution (TMB), and stop solution. During the assay, testosterone in the patient sample competes with the enzyme-conjugated testosterone for binding sites on the plate. After incubation and washing, a substrate is added, producing a colorimetric signal measured at 450nm via a calibrated EIA reader. The intensity of the color is inversely proportional to the testosterone concentration in the sample. Healthcare providers use the quantitative output to assist in diagnosing and monitoring androgen-related conditions such as hypogonadism, puberty disorders, and virilization. The assay provides a standardized method for assessing hormonal status in clinical settings.

## Clinical Evidence

Bench testing only. Performance validated via method comparison studies (n=99 and n=81) against a commercial LIA method, yielding correlation of 0.904 and R2=0.9866. Analytical sensitivity: 1.857 pg/mL (lowest detectable); functional sensitivity: 7.1 pg/mL. Specificity evaluated via cross-reactivity testing against 12 steroids; minimal cross-reactivity observed. Reproducibility (intra-assay, inter-assay, inter-lot) demonstrated with CVs generally <10%. Linearity established from 7.1 to 4500 pg/mL.

## Technological Characteristics

Competitive enzyme immunoassay; microtiter plate coated with mouse monoclonal anti-testosterone antiserum; horseradish peroxidase conjugate; TMB substrate; colorimetric detection at 450nm. Requires calibrated EIA reader. Standards prepared in artificial saliva matrix; traceability to GC-MS methods. No electronic connectivity; standalone manual assay.

## Regulatory Identification

A testosterone test system is a device intended to measure testosterone (a male sex hormone) in serum, plasma, and urine. Measurement of testosterone are used in the diagnosis and treatment of disorders involving the male sex hormones (androgens), including primary and secondary hypogonadism, delayed or precocious puberty, impotence in males and, in females hirsutism (excessive hair) and virilization (masculinization) due to tumors, polycystic ovaries, and adrenogenital syndromes.

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
ASSAY ONLY TEMPLATE

A. 510(k) Number:
k052649

B. Purpose for Submission:
Premarket Notification 510(k) of intention to manufacture and market the DRG Salivary Testosterone Elisa Kit.

C. Measurand:
Testosterone

D. Type of Test:
Enzyme Immunoassay

E. Applicant:
DRG International, Inc.

F. Proprietary and Established Names:
DGR Salivary Testosterone Elisa Kit

G. Regulatory Information:
1. Regulation section:
21 CFR §862.1680 Testosterone Test System
2. Classification:
Class 1 (reserved)
3. Product code:
CDZ
4. Panel:
75 (Chemistry)

{1}

2

H. Intended Use:

1. Intended use(s):

See Indications for use below.

2. Indication(s) for use:

An Enzyme Immunoassay for the *in vitro diagnostic* quantitative measurement of free active testosterone in saliva. Measurement of testosterone is used in the diagnosis and treatment of disorders involving the male sex hormones (androgens), including primary and secondary hypogonadism, delayed or precocious puberty, impotence in males and, in females hirsutism (excessive hair) and virilization (masculinization) due to tumors, polycystic ovaries, and adrenogenital syndromes.

3. Special conditions for use statement(s):

For Professional use only. For in-vitro diagnostic use only

4. Special instrument requirements:

Calibrated EIA reader adjusted to read at 450nm.

I. Device Description:

The DRG Salivary Testosterone Elisa Kit consists of the following:

1. Microtiter plate, 8 well snap-off strips, 12 strips, coated with (mouse) anti-Testosterone antiserum.
2. Reference Standard Set, 1 ml each, 0.0; 10; 50; 100; 500; 1000; 5000 pg/ml.
3. Enzyme-Conjugate, 26 ml, Testosterone conjugated to horseradish peroxidase, ready to use.
4. Substrate Solution - TMB, 25 ml, ready to use.
5. Stop Solution, 0.5M H₂SO₄, 14 ml, ready to use.
6. Wash Solution, 30 ml, Concentrate for 1200 ml.

J. Substantial Equivalence Information:

1. Predicate device name(s):

IMMUNO BIOLOGICAL LABORATORIES, IBL Testosterone LIA

2. Predicate 510(k) number(s):

k033786

{2}

3. Comparison with predicate:

|  Similarities  |   |   |
| --- | --- | --- |
|  Item | Predicate Device | New Device  |
|  Device Name | IBL Testosterone LIA | DRG SLV Testosterone ELISA  |
|  Analyte | Free active Testosterone | Same  |
|  Specimen | Serum or Saliva | Saliva  |
|  Method | Luminescence Immunoassay | Enzyme Immunoassay  |
|  Test Principle | Competitive Immunoassay. Competition is between a labeled and non-labeled antigen for a fixed number of antibody binding sites. The amount of labeled analyte bound to the antibody is inversely proportional to the concentration of the analyte present in the sample. | Same  |
|  Detection | Luminescence detection | Colorimetric detection  |
|  Calculation | Quantitative determination with standard curve | Same  |
|  Quality Control | 2 Controls at different levels | Recommended separate external controls  |
|  Indications for Use | Measurements of testosterone are used in the diagnosis and treatment of disorders involving the male sex hormones (androgens), including primary and secondary hypogonadism, delayed or precocious puberty, impotence in males and , in females hirsutism (excessive hair) and virilization (masculinization) due to tumors, polycystic ovaries, and adrenogenital syndromes. | Same  |
|  Detection Limit | 1.757 pg/mL | 1.857 pg/mL  |

K. Standard/Guidance Document Referenced (if applicable):

Haeckel, R., R.F. Walker and D. Colic (1989): Reference ranges for mixed saliva collected from literature. J. Clin. Chem. Clin. Biochem. 27, 249-252

{3}

L. Test Principle:

The DRG Salivary Testosterone ELISA Kit is based on the competition principle and the microplate separation. An unknown amount of free testosterone present in the sample and a fixed amount of testosterone conjugated with horseradish peroxidase compete for the binding sites of mouse monoclonal testosterone antiserum coated onto the wells. After one-hour incubation the microplate is washed to stop the competition reaction. After addition of the substrate solution the concentration of testosterone is inversely proportional to the optical density measured.

M. Performance Characteristics (if/when applicable):

1. Analytical performance:

a. Precision/Reproducibility:

INTRA-ASSAY PRECISION

The intra-assay (within-run) variation of the DRG SLV Testosterone ELISA was determined by repeated measurements of four saliva samples.

|   | Saliva 1 |   | Saliva 2 |   | Saliva 3 |   | Saliva 4  |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  Measurement | OD450 | Conc. pg/ml | OD450 | Conc. pg/ml | OD450 | Conc. pg/ml | OD450 | Conc. pg/ml  |
|  1 | 1.789 | 2.94 | 1.665 | 13.82 | 1.438 | 44.07 | 1.539 | 6.75  |
|  2 | 1.785 | 3.22 | 1.683 | 11.99 | 1.429 | 45.58 | 1.561 | 5.19  |
|  3 | 1.816 | 1.18 | 1.700 | 10.35 | 1.430 | 45.41 | 1.542 | 9.50  |
|  4 | 1.778 | 3.73 | 1.711 | 9.32 | 1.471 | 38.76 | 1.579 | 4.95  |
|  5 | 1.791 | 2.80 | 1.654 | 14.97 | 1.462 | 40.17 | 1.540 | 9.41  |
|  6 | 1.802 | 2.05 | 1.641 | 16.37 | 1.455 | 41.29 | 1.560 | 6.50  |
|  7 | 1.789 | 2.94 | 1.690 | 11.31 | 1.472 | 38.60 | 1.538 | 9.22  |
|  8 | 1.795 | 2.52 | 1.658 | 14.55 | 1.449 | 42.26 | 1.557 | 8.50  |
|  9 | 1.769 | 4.41 | 1.647 | 15.72 | 1.431 | 45.28 | 1.562 | 5.59  |
|  10 | 1.791 | 2.80 | 1.689 | 11.40 | 1.479 | 37.52 | 1.563 | 6.67  |
|  11 | 1.785 | 3.22 | 1.662 | 14.13 | 1.438 | 44.07 | 1.574 | 8.41  |
|  12 | 1.815 | 1.24 | 1.665 | 13.82 | 1.446 | 42.75 | 1.550 | 5.11  |
|  13 | 1.786 | 3.15 | 1.691 | 11.21 | 1.471 | 38.76 | 1.156 | 8.68  |
|  14 | 1.792 | 2.73 | 1.670 | 13.30 | 1.450 | 42.10 | 1.202 | 5.92  |
|  15 | 1.775 | 3.95 | 1.667 | 13.61 | 1.435 | 44.57 | 1.159 | 7.52  |
|  16 | 1.772 | 4.18 | 1.681 | 12.19 | 1.428 | 45.74 | 1.156 | 7.18  |
|  17 | 1.771 | 4.26 | 1.668 | 13.51 | 1.440 | 43.73 | 1.144 | 5.84  |
|  18 | 1.801 | 2.12 | 1.673 | 13.00 | 1.456 | 41.13 | 1.173 | 5.92  |
|  19 | 1.810 | 1.54 | 1.682 | 12.09 | 1.467 | 39.38 | 1.161 | 7.61  |
|  20 | 1.794 | 2.59 | 1.681 | 12.19 | 1.440 | 43.73 | 1.202 | 6.67  |
|   | Saliva 1 | Saliva 2 | Saliva 3 | Saliva 4  |
| --- | --- | --- | --- | --- |
|  Mean (pg/ml) | 2.88 | 12.94 | 42.25 | 7.06  |
|  SD | 0.946 | 1.787 | 2.655 | 1.485  |
|  CV (%) | 32.87 | 13.81 | 6.28 | 21.04  |
|  n = | 20 | 20 | 20 | 20  |

{4}

The functional sensitivity of the assay is  $7.1~\mathrm{pg / mL}$ . This assumes that the lowest concentration having a  $\mathrm{CV\%}$  of approximately  $20\%$  is considered the functional sensitivity.

# INTER ASSAY PRECISION: LOT TO LOT

The inter-assay (between-run) variation was determined by triplicate measurements of five saliva samples in three different kit lots. Results reported in the table below:

|   | Sample 1 | Sample 2 | Sample 3 | Sample 4 | Sample 5  |
| --- | --- | --- | --- | --- | --- |
|  Mean (pg/ml) | 64.5 | 352.89 | 517.65 | 44.00 | 116.54  |
|  SD (pg/ml) | 3.77 | 13.44 | 15.01 | 1.53 | 5.00  |
|  CV (%) | 5.85 | 3.81 | 2.90 | 3.47 | 4.29  |
|  n = | 9 | 9 | 9 | 9 | 9  |

# b. Linearity/assay reportable range:

Two saliva samples containing different amounts of analyte were serially diluted with zero standard and assayed using DRG SLV testosterone ELISA. Percentage recovery was calculated by comparing the expected and observed values for SLV testosterone.

|  Sample | Dilution | OD 450 nm | Observed conc. pg/ml | Expected Conc. pg/ml | Recovery %  |
| --- | --- | --- | --- | --- | --- |
|  1(spiked) | undiluted | 0.113 | >5000 | 9000 |   |
|   |  1:2 | 0.139 | 4410.07 | 4500.00 | 98.00  |
|   |  1:4 | 0.247 | 2285.54 | 2250.00 | 101.58  |
|   |  1:8 | 0.435 | 1108.65 | 1125.00 | 98.55  |
|   |  1:16 | 0.670 | 582.77 | 562.50 | 103.60  |
|   |  1:32 | 1.025 | 265.73 | 281.30 | 94.46  |
|   |  1:64 | 1.350 | 132.78 | 140.60 | 94.44  |
|   |  1:128 | 1.622 | 67.89 | 70.30 | 96.57  |
|  2 (spiked) | undiluted | 0.457 | 1033.58 | 1033.58 |   |
| --- | --- | --- | --- | --- | --- |
|   |  1:2 | 0.730 | 505.48 | 516.79 | 97.81  |
|   |  1:4 | 1.030 | 262.95 | 258.40 | 101.76  |
|   |  1:8 | 1.385 | 122.69 | 129.20 | 94.96  |
|   |  1:16 | 1.660 | 60.89 | 64.60 | 94.26  |
|   |  1:32 | 1.855 | 30.84 | 32.30 | 95.48  |
|   |  1:64 | 1.985 | 15.46 | 16.15 | 95.73  |
|   |  1:128 | 2.125 | 2.78 | 8.07 | 34.45  |

{5}

|   | Sample 1 | Sample 2  |
| --- | --- | --- |
|  Concentration pg/ml | 9000.00 | 1033.58  |
|  Average % recovery | 98.20 | 96.70  |
|  Range of from | 94.4 | 94.3  |
|  % recovery to | 103.6 | 101.7  |
|  Accepted from | 85% | 85%  |
|  recovery to | 115% | 115%  |

The upper end detectability of SLV testosterone is 4410 pg/mL. These results demonstrate that the test system is performing equivalent to original.

c. Traceability, Stability, Expected values (controls, calibrators, or methods):

The standards prepared for the SLV Testosterone kits are buffer based (artificial saliva matrix). The range of calibrators were prepared by appropriate dilution from the maximum standard (Smax: 5000 pg/mL). The testosterone for the standards is purchased from a commercially available source, and is weighed in to make the 5000 pg/mL.

The reference values (calibrators/controls) were established using (Gas chromatography-mass spectrophotometry) GC-MS methods, as per the guidelines for quality assurance in medical laboratories, Instand E.V. Germany (L.D. Dikkesche. Et al. 1988: De toepassing kwaliteitcontroleprogramma's voor progesterone-, cortisol-, testosterone- en oestradiolbe[alingen in serums. Tijdschr NVKC 13: 148-155).

WHO standard is not available.

The functional quality of the kit lots were tested using the Lyphochek controls from BioRad, but controls are not included in the kit. These controls are commercially available and can be purchased by the customers.

Real time stability, Accelerated stability, and Saliva sample stability were validated for this assay.

d. Detection limit:

The analytical sensitivity of SLV Testosterone ELISA was calculated from the mean minus 2SD of 20 replicate analyses of the zero standard.

|  Mean | 1.718  |
| --- | --- |
|  SD | 0.018  |
|  2 x SD | 0.035  |
|  Mean – 2SD | 1.683 corresponds to 1.857 pg/ml  |
|  N | 20  |

{6}

e. Analytical specificity:

Cross reactivity was tested with the following compounds whose chemical structure could potentially cause interference with the SLV Testosterone ELISA. The specificity of the antiserum used for the ELISA was evaluated by determination of the cross-reactivity at 50% displacement of various compounds listed in the table below.

The cross-reactivity is defined as:

Concentration of testosterone at 50% B/BO x 100
Concentration of cross-reactant giving 50% B/BO

|  Steroid | % Cross reaction  |
| --- | --- |
|  Testosterone | 100%  |
|  5α-Dihydrotestosterone | 0.80%  |
|  Androstenedione | 0.90%  |
|  11β-hydroxysterone | 3.30%  |
|  17α-methyltestosterone | 0.10%  |
|  19-Nortestosterone | 3.30%  |
|  Epitestosterone | 0.10%  |
|  Estradiol | 0.10%  |
|  Progesterone | < 0, 10%  |
|  Cortisol | < 0, 10%  |
|  Estrone | < 0, 10%  |
|  Danazol | < 0, 10%  |

f. Assay cut-off:

Not applicable

2. Comparison studies:

a. Method comparison with predicate device:

COMPARISON TO IBL LIA

Concentration of testosterone in 81 saliva samples collected from 40 - 65 year old men and women using DRG SLV testosterone kit. The results were compared with those obtained from IBL-LIA method.

Correlation Coefficient = 0.99328

{7}

![img-0.jpeg](img-0.jpeg)

Number of XY Pairs = 81

Pearson  $r = 0.9933$

95% confidence interval = 0.9895 to 0.9957

R squared = 0.9866

b. Matrix comparison:

Not applicable

3. Clinical studies:

a. Clinical Sensitivity:

Not applicable

b. Clinical specificity:

Not applicable

c. Other clinical supportive data (when a. and b. are not applicable):

4. Clinical cut-off:

Not applicable

{8}

9

5. Expected values/Reference range:

In order to determine the normal range of SLV Testosterone, saliva samples from 187 adult male and 188 adult female apparently healthy subjects, age 21 to 75 years, were collected in the morning and analyzed using the DRG SLV Testosterone ELISA kit. The following range was calculated from this study.

|   | Men ♂ |   |   | Women ♀  |   |   |
| --- | --- | --- | --- | --- | --- | --- |
|  Age Group Years | Range (5-95%) | Median | n | Range (5-95%) | Median | n  |
|  21-30 | 47.2-136.2 | 92.8 | 42 | 7.9-50.4 | 20.8 | 40  |
|  31-40 | 46.8-106.8 | 73.6 | 37 | <7.0-44.8 | 17.1 | 40  |
|  41-50 | 36.5-82.7 | 58.8 | 34 | <7.0-39.4 | 18.3 | 38  |
|  51-60 | 19.2-89.0 | 44.5 | 36 | <7.0-29.8 | 19.2 | 38  |
|  61-75 | 12.2-68.6 | 38.9 | 38 | <7.0-29.3 | 16.0 | 32  |

N. Proposed Labeling:

The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.

O. Conclusion:

The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

{9}

10

---

**Source:** [https://fda-staging.innolitics.com/device/K052649](https://fda-staging.innolitics.com/device/K052649)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
