The Clini-Lase System is an infrared lamp that is intended to emit energy in the infrared spectrum to provide topical heating for the purpose of elevating tissue temperture for the temporary relief of minor muscle and joint pain and stiffness, for the temporary relief of minor joint pain associated with arthritis, for the temporary increase in local circulation where applied, and the relaxation of muscles. In addition, the Clini-Lase System may also help muscle spasms, minor sprains and strains, and minor muscular back pain.
Device Story
Clini-Lase System is an infrared lamp device comprising a control base unit and interchangeable infrared heat lamp probes. The base unit contains electronics, system controls, and an LCD display, powered by an AC-DC adapter. The device emits infrared radiation (wavelengths 650, 785, 808 nm) via handheld probes to provide topical heating. It is intended for clinical use to elevate tissue temperature, promoting local circulation and muscle relaxation. The device is used to provide temporary relief for minor musculoskeletal pain, stiffness, and spasms. Healthcare providers operate the system to deliver targeted thermal therapy. The output is thermal energy applied to the patient's skin, which facilitates symptomatic relief of pain and muscle tension.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on non-clinical data, design, and labeling mitigations.
Technological Characteristics
Infrared lamp system; 12V DC power; wavelengths 650, 785, 808 nm; power output 25-500 mW; handheld probes with replaceable cable assembly; LCD display; control base unit; non-sterile.
Indications for Use
Indicated for patients requiring topical heating for temporary relief of minor muscle/joint pain, stiffness, arthritis-related joint pain, muscle spasms, minor sprains/strains, and minor muscular back pain; also indicated for temporary increase in local circulation and muscle relaxation.
Regulatory Classification
Identification
An infrared lamp is a device intended for medical purposes that emits energy at infrared frequencies (approximately 700 nanometers to 50,000 nanometers) to provide topical heating.
Special Controls
*Classification.* Class II (special controls). The device, when it is an infrared therapeutic heating lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 890.9.
{0}------------------------------------------------
# 510(k) Premarket Notification for the Clini-Lase
2
# 510(k) Summary of Saftey and Effectiveness
## 1. Submitter's Name and Address
American Spine Center, Ltd. 100 Mascoutah Avenue Belleville, IL 62220-3801 Phone: 618-233-6824 Fax: 618-233-6825
Key Contact: James J. White, President
Date Prepared: 09/22/2005
## 2. Device Name
Proprietary Name: Clini-Lase Laser System Common / Usual Name: Infrared Lamp Classification Name: Infrared Lamp (21 CFR 890.5500) Product Code: ILY
Model #: CL-Base, CL-2000, CL-1200, CL-500, CL-300, CL-275, CL-155, CL-55, CLP-Base, CLP-275 and CLP-55.
## 3. Legally-Marketed Predicate Devices
The Clini-Lase System are substantially equivalent to other infrared sources (product code ILY) currently in commercial distribution such as:
K032231 MedX LCT100 & MedX LCS Portable Laser K033923 Thor DDII IR Lamp Systems K040662 Vectra Genisis Laser System
K043586 Laser Sys*Stim 540, Model ME 540
K050370 Palomar LuxIR Handpiece
## Description of the Device
The Clini-Lase System consists of a clinical and portable control base unit, and seperate attachment infrared heat lamp probes. The control base unit houses the electronics, LCD display and system controls and is powered by 12 Volts DC suplied by an AC-DC Power Adapter. The attachment probes produce infrared radiation to provide topical heating. Various single and cluster probes are available for the base units. The probes are conected by a readily replacable independent cable assembly.
{1}------------------------------------------------
052647 2 of 2
## 510(k) Premarket Notification for the Clini-Lase System
## 4. Intended Use of the Device
The Clini-Lase System is an infrared lamp that is intended to emit energy in the infrared spectrum to provide topical heating for the purpose of elevating tissue temperture for the temporary relief of minor muscle and joint pain and stiffness, for the temporary relief of minor joint pain associated with arthritis, for the temporary increase in local circulation where applied, and the relaxation of muscles. In addition, the Clini-Lase System may also help muscle spasms, minor sprains and strains, and minor muscular back pain.
#### 5. Performance Data
The differences in the specifications of the Clini-Lase System and the predicate devices do not result in different performance or raise new questions of safety or efficacy.
#### 6. Non-clinical Data
Potentail hazards are identified and have been mitigated by design and labeling.
#### 7. Conclusion
Based on the foregoing, the Clini-Lase System is substantially equivalent to the legally-marketed predicate devices.
{2}------------------------------------------------
## Name of Manufacturer: American Spine Center, LTD. Laser Model Name and Number: Clini-Lase System CL-2000, CL-1200, CL-155, CL-500. CL-300, CL-275, CL-55
Laser Type: (Circle all that apply)
Alexandrite, Argon, CO2, Copper-Vapor, Diode, Dye, Nd:YAG, Erbium, Hol: YAG, Krypton, Ruby, KTP/532, Excimer, HENE, Accessory, Other
CED's ________________________________________________________________________________________________________________________________________________________________________
Indications in this application: The Clini-Lase System is an infrared lamp that is intended to emit energy in the infrared spectrum to provide topical heating for the purpose of elevating tissue temperature for the temporary relief of minor muscle and joint pain associated with arthritis, for the temporary increase in local circulation where applied, and the relaxation of muscles, minor sprains and strains, and minor muscular back pain.
#### FDA Document Control Number: K052647
FDA Product Code: 79GEX
Reviewer Computer Initials: CYH
Date of Clearance Letter: 11/04/05
Basis of Approval: (Circle all that apply)
Predicate Device (PD), Clinical Data (CD), Animal Data (AD), Specifications (SPECS), Bench Test Data (BTD), Historical Information (HI), Other
Description of Laser: Operation Modes: (Circle all that apply) CW, Pulsed Q-Switched, Mode Locked, Contact, Free Beam, Other ________________________________________________________________________________________________________________
Wavelength in Nanometers: 808, 650, 785
Power/Energy Range (Watts/Joules): 25 mW - 500 mW
Pulse Width: n/a
Repitition Rate: n/a
Delivery System: hand held probes Comments:
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features a stylized eagle with three heads, representing the department's focus on health, human services, and science. The eagle is surrounded by a circular border with the text "DEPARTMENT OF HEALTH AND HUMAN SERVICES, USA" written around it.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
NOV - 9 2005
James J. White President American Spine Center, LTD. Clini-Lase 100 Mascoutah Avenue Belleville, Illinois 6220-3801
Re: K052647
Trade/Device Name: Clini-Lase Systems, with Models: CL-Base, CL-2000, CL-1200, CL-500, CL-300, CL-275, CL-155, and CL-55 CLP-Base, CLP-275 and CLP-55
Regulation Number: 21 CFR 890.5500 Regulation Name: Infrared lamp Regulatory Class: II Product Code: ILY Dated: October 31, 2005 Received: October 31, 2005
Dear Mr. White:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
#### Page 2- James J. White
This letter will allow you to begin marketing your device as described in your Section 510(k) I ms letter will and w your e FDA finding of substantial equivalence of your device to a legally prematted predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you attire Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html
Sincerely vours.
Sincerely yours,
Barbaze Buehup
Mark N. Melkerson Acting Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
K092647/5
510(k) Premarket Notification for the Clini-Lase System
## Indication for Use:
## K052647
510(k) Number (if known):
Device Name:
Clini-Lase System, with Models:
CL-Base, CL-2000, CL-1200, CL-500, CL-300, CL-275, CL-155, and CL-55.
CLP-Base, CLP-275 and CLP-55.
## Indication For Use:
The Clini-Lase System is an infrared lamp that is intended to emit energy in the infrared spectrum to provide topical heating for the purpose of elevating tissue temperture for the temporary relief of minor muscle and joint pain and stiffness, for the temporary relief of minor joint pain associated with arthritis, for the temporary increase in local circulation where applied, and the relaxation of muscles. In addition, the Clini-Lase System may also help muscle spasms, minor sprains and strains, and minor muscular back pain.
X Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-the Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, OFFICE of Device Evaluation (ODE)
Tarbayemohmd
Division of General, Restorative, and Neurological Devices
510(k) Number K052647
Page 1 of 1
v
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.