# MODIFICATION TO:  QUANTUM VERTEBRAL BODY REPLACEMENT (K052641)

_Quantum Orthopedics, Inc. · MQP · Apr 12, 2006 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K052641

## Device Facts

- **Applicant:** Quantum Orthopedics, Inc.
- **Product Code:** [MQP](/productcode/MQP.md)
- **Decision Date:** Apr 12, 2006
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 888.3060
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The Quantum Vertebral Body Replacement is intended for use in the thoracic and/or thoracolumbar spine (T1-L5) to replace a collapsed, damaged, or unstable vertebral body resected or excised (i.e., partial or total vertebrectomy procedures) due to tumor or trauma (i.e., fracture). This device is intended to be used with supplemental spinal fixation systems that have been cleared for use in the thoracic and/or lumbar spine (i.e., posterior pedicle screw and rod systems, anterior plate systems, and anterior screw and rod systems). The interior of the spacer can be packed with allograft or autograft.

## Device Story

Quantum Vertebral Body Replacement is a box-shaped spinal implant; features exterior teeth to prevent migration; available in multiple sizes. Implanted during partial or total vertebrectomy procedures in thoracic/thoracolumbar spine (T1-L5). Requires use with cleared supplemental spinal fixation systems (pedicle screw/rod or anterior plate/screw systems). Interior cavity designed for packing with allograft or autograft to facilitate fusion. Used by surgeons in clinical/OR settings to restore spinal column stability following tumor resection or traumatic fracture.

## Clinical Evidence

Bench testing only. Mechanical testing performed to demonstrate the device is capable of performing in accordance with its intended use.

## Technological Characteristics

Box-shaped vertebral body replacement; materials: titanium or polyetheretherketone (PEEK); exterior surface features teeth for fixation; available in multiple sizes; intended for use with supplemental spinal fixation systems; interior cavity for bone graft material.

## Regulatory Identification

A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vertebral plates that are punched into each of a series of vertebral bodies. An eye-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fastened or crimped at each eye-type screw. The device is used to apply force to a series of vertebrae to correct “sway back,” scoliosis (lateral curvature of the spine), or other conditions.

## Predicate Devices

- Quantum Vertebral Body Replacement ([K052641](/device/K052641.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# APR 1 2 2006

## 510(k) Summary Quantum Vertebral Body Replacement

| Manufacturer Identification<br>Submitted by: | Quantum Orthopedics, Inc.<br>2744 Loker Ave. W., Suite 100<br>Carlsbad, CA 92010<br>760-607-0121                                                                             |
|----------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Information:                         | Jason Blain<br>Chief Technology Officer<br>Quantum Orthopedics, Inc.<br>2744 Loker Ave. W., Suite 100<br>Carlsbad, CA 92010<br>760-607-0121<br>iblain@quantumorthopedics.com |
| Date Prepared:                               | September 22, 2005                                                                                                                                                           |
| Device Indentification<br>Proprietary Name:  | Quantum Vertebral Body Replacement                                                                                                                                           |
| Common Name:                                 | Vertebral Body Replacement                                                                                                                                                   |
| Classification Name:                         | Spinal Vertebral Body Replacement                                                                                                                                            |

#### Predicate Devices

The subject device is substantially equivalent to the previously cleared Quantum Vertebral Body Replacement.

### Device Description

The Quantum Vertebral Body Replacement is a generally box-shaped device with various holes located throughout its geometry. The exterior surface of the device has teeth to help keep the device from migrating once placed in its desired location. It is available in a multitude of sizes to suit the individual pathology and anatomic condition of the patient. The device may be made from titanium or polyetheretherketone (PEEK).

{1}------------------------------------------------

#### Intended Use of the Device

The Quantum Vertebral Body Replacement is intended for use in the thoracic and/or thoracolumbar spine (T1-L5) to replace a collapsed, or unstable vertebral body resected or excised (i.e., partial or total vertebrectomy procedures) due to tumor or trauma (i.e., fracture).

This device is intended to be used with supplemental spinal fixation systems that have been cleared for use in the thoracic and/or lumbar spine (i.e., posterior pedicle screw and rod systems, anterior plate systems, and anterior screw and rod systems). The interior of the spacer can be packed with allograft or autograft.

#### Performance Data

Mechanical testing indicates that the Quantum Vertebral Body Replacement is capable of performing in accordance with its intended use.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus symbol, which is often associated with healthcare. The text "DEPARTMENT OF HEALTH AND HUMAN SERVICES, USA" is arranged in a circular pattern around the symbol. The logo is black and white.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

APR 1 2 2006

Quantum Orthopedics, Incorporated c/o Jason Blain Chief Technology Officer 2744 Loker Ave. W., Suite 100 Carlsbad, California 92010

Re: K052641

Trade/Device Name: Quantum Veretebral Body Replacement Regulation Number: 21 CFR 888.3060 Regulation Name: Spinal Intervertebral Body Fixation System Regulatory Class: II Product Code: MQP Dated: March 14, 2006 Received: March 15, 2006

Dear Mr. Blain:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Dr to and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, inting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 - Mr. Jason Blain

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours.

Sincerely yours,
elmee

Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

## Indications for Use

510(k) Number: K052641

Device Name: Quantum Vertebral Body Replacement

#### Indications For Use:

The Quantum Vertebral Body Replacement is intended for use in the thoracic and/or thoracolumbar spine (T1-L5) to replace a collapsed, damaged, or unstable vertebral body resected or excised (i.e., partial or total vertebrectomy procedures) due to tumor or trauma (i.e., fracture).

This device is intended to be used with supplemental spinal fixation systems that have been cleared for use in the thoracic and/or lumbar spine (i.e., posterior pedicle screw and rod systems, anterior plate systems, and anterior screw and rod systems). The interior of the spacer can be packed with allograft or autograft.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

DRH, Office of Device Evaluation (ODE) Concurrence

(Division Sign-Off) Division of General, Restorer and Neurological Devices

Page 1 of 1

510(k) Number_ko52641

---

**Source:** [https://fda-staging.innolitics.com/device/K052641](https://fda-staging.innolitics.com/device/K052641)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).