UROK ROCKING WHEELCHAIR, MODEL 700-4XX

K052548 · Amg Medical, Inc. · IOR · Jan 6, 2006 · Physical Medicine

Device Facts

Record IDK052548
Device NameUROK ROCKING WHEELCHAIR, MODEL 700-4XX
ApplicantAmg Medical, Inc.
Product CodeIOR · Physical Medicine
Decision DateJan 6, 2006
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 890.3850
Device ClassClass 1
AttributesTherapeutic

Indications for Use

The URok Rocking Wheelchair is intended to provide mobility to persons limited to a sitting position while the rocking mechanism is locked in an upright position. The wheelchair is intended to redistribute pressures exerted on the buttocks and lower back when rocking or tilted at an angle. The wheelchair is intended to provide comfort adjustment while rocking or tilted at an angle.

Device Story

Manually operated, attendant or user-propelled mechanical wheelchair; aluminum TIG-welded frame; padded sling seat. Features: rocking motion via two fiberglass leaf springs under seat center of gravity; tilt-in-space capability. Rocking initiated by user head movement; arc limited by anti-tippers and rock arrestors. Rocking/tilt locked via cable-controlled pins operated by bicycle-style brake levers on push handles. Used in indoor/outdoor settings for mobility, pressure relief, and positioning. Benefits: dynamic pressure management, comfort, and head control. Healthcare providers (therapists) evaluate patients for appropriate use.

Clinical Evidence

Bench testing only. Device performance verified against ANSI/RESNA WC series requirements. Pressure redistribution efficacy during rocking observed and recorded using adult human volunteers on a pressure recording mat.

Technological Characteristics

Manual, mechanical wheelchair. Materials: aluminum tubing frame, fiberglass leaf springs, padded sling upholstery (ISO 7176-16 compliant). Sensing/Actuation: mechanical leaf springs for rocking; cable-actuated pins for locking/tilt. Dimensions: rigid, non-folding. Connectivity: none. Sterilization: N/A.

Indications for Use

Indicated for persons limited to a sitting position requiring mobility, pressure redistribution on buttocks/lower back, and comfort adjustment. Recommended for evaluation by accredited rehabilitation or occupational therapist.

Regulatory Classification

Identification

A mechanical wheelchair is a manually operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K052548 JAN 6 2006 ## 510(k) SUMMARY # AMG Medical Inc.'s URok Rocking Wheelchair Submitter,s Name, address, Telephone Number, Contact Person and Date Prepared AMG Medical Inc. 8505 Dalton Montreal, Canada H4T IV5 514-737-5251 Phone: Fax: 514-737-6572 Kenneth Levesque Contact Person: Director, RA/QA Date Prepared: September 07, 2005 #### Name of Device and Name/Address of Sponsor URok Rocking Wheelchair AMG Medical Inc. 8505 Dalton Montreal, Canada H4T 1V5 514-737-5251 Phonc: 514-737-6572 Fax: Common or Usual Name: Manual Wheelchair Classification Name: Wheelchair, Mechanical 89IOR ### Predicate Devices: The URok Rocking Wheelchair is substantially equivalent to the RxRocker (K902438) wheelchair for the rocking feature. The URok Rocking Wheelchair is substantially equivalent to the Invacare Solara wheelchair for the pressure redistribution feature. #### Background In 2001 AMG Medical Inc. purchased all rights to the RxRocker Corp. with the exception of the following: AMG granted RxRocker Corp. the right to continue to sell the RxRocker only to the US Government (including the Veteran's Administration and US military hospitals under the patents and existing technology and know how found the RxRocker wheelchair. {1}------------------------------------------------ In consideration of the above sales agreement and continuing sales of the original 510k device, AMG is not submitting this 510k for a change to an existing device but rather submitting for a new 510k premarket approval referencing the RxRocker as a predicate device. # Intended Use: The URok Rocking Wheelchair is intended to provide mobility to persons limited to a sitting position while the rocking mechanism is locked in an upright position. The wheelchair is intended to redistribute pressures exerted on the buttocks and lower back when rocking or tilted at an angle. The wheelchair is intended to provide comfort adjustment while rocking or tilted at an angle. # Technological Characteristics The AMG Medical Inc. URok Rocking Wheelchair is a manually operated, attendant or user propelled, mechanical rocking wheelchair. The device is a lightweight wheelchair, suitable for both indoor and outdoor use. The device consists of an aluminum frame, one size of rear wheel, and smaller front pivoting casters for steering and turning. The frame is constructed from aluminum tubing that is TIG-welded. This device is a rigid, non-folding type of wheelchair that incorporates a padded, sling style seat. The upholstery meets the requirements of ISO 7176-16, flammability standards. The AMG URok Rocking Wheelchair includes a rocking feature which permits dynamic pressure management of the seating pressure upon the buttocks. The rocking feature is realized by the placement of two vertically positioned fibrogless Ieaf springs, which are connected to the seat and frame, under the seat's centre of gravity. The balance of the seat upon the springs is precise enough so that the rocking motion may be initiated by the user simply by gently moving the head backward and forward. Safety features such as anti tippers and rock arrestors limit the arc of the rocking motion. The rocking feature can be "locked out" by an attendant so that the device can be used as a wheelchair with no rocking motion. The AMG URok wheelchair also has a "tilts in space" feature. The rocking mcchanism can be locked in multiple positions to provide "tilt in space" that permits receing meemainst backwards thereby allowing gravity to hold them in position. This is used to help in positioning, comfort, or head control. The tilt in space is initiated by the attendant pushing down on the push handles to tilt the scat and back and locking out the rocking action when the desired angle of tilt is reached. The rocking feature is released or is locked out by two cable controlled pins that insert into holes within two vertically mounted metal plates located under the seat support tubes. The cables are controlled using bicycle brake type levers in onto bed on the rear push tube or handle. 5-9 {2}------------------------------------------------ # Substantial equivalence: The AMG URok Rocking Wheelchair is substantially equivalent to numerous manual, mechanical wheelchairs currently in the marketplace, for which FDA has granted marketing clearance through the 510(k) premarket notification process. More specifically the AMG rocking wheelchair is substantially equivalent to the RxRocker (K902438) wheelchair for it's rocking feature and to the Invacare Solara wheelchair for it's pressure redistribution feature. Each of the products is manually operated, attendant or user propelled with the same intended functions and uses which are to provide mobility to persons that may be limited to a seated position and to provide pressure relief. Additionally, in one wheelchair pressure relief is achieved either by rocking or by "tilt in space" (URok Rocking Wheelchair) and in the other wheelchair pressure redistribution is achieved solely by "tilt in space" (Invacare Solara). # Performance data: The AMG URok Rocking Wheelchair meets the performance requirements specified in the ANSI/RESNA WC series of wheelchair requirements. Pressure redistribution obtained while rocking was observed and recorded using adult human volunteers seated on a pressure recording mat placed upon the seat cushion. {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the seal of the Department of Health and Human Services (HHS) of the United States. The seal features a stylized eagle with three stripes forming its body and wings. The text "DEPARTMENT OF HEALTH AND HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 JAN 6 2006 Mr. Kenneth Levesque A.M.G. Medical, Inc. 8505 Dalton Montreal, Quebec H4T 1V5 CANADA Re: K052548 Trade/Device Name: URok rocking wheelchair, Model 700-4xx Regulation Number: 21 CFR 890.3850 Regulation Name: Mechanical wheelchair Regulatory Class: II Product Code: IOR Dated: December 21, 2005 Received: December 22, 2005 Dear Mr. Levesque: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {4}------------------------------------------------ Page 2 - Mr. Levesque This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally premained modicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html Sincerely yours, Mark N. Melkerson Acting Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ # Indications for Use Statement 510(k) Number (if known): Device Name: URok Rocking Wheelchair # Indications for Use: The URok Rocking Wheelchair, by AMG Medical Inc., is intended to provide mobility to persons limited to a sitting position while the rocking mechanism is locked in an upright position. The wheelchair is intended to redistribute pressures exerted on the buttocks and lower back when rocking or tilted at an angle. The wheelchair is intended to provide comfort adjustment while rocking or tilted at an angle. AMG Medical inc. recommends that an accredited rehabilitation or occupational therapist and supplier evaluate all customers of its products. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) Division of General, Restorative, and Neurological Devices | 510(k) Number | K052548 | |---------------|---------| |---------------|---------| | Page | 1 of 1 | |------|--------| |------|--------| 4-1
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...