CYBOW 11 Reagent Strips are for the rapid visual determination of urobilinogen, glucose, bilirubin, ketones (acetoacetic acid), specific gravity, blood, pH, protein, nitrite, leukocytes and ascorbic acid in urine.
Device Story
CYBOW 11 Series Reagent Strips are dip-and-read test strips for in vitro diagnostic use. The device consists of multiple chemically reactive pads on a plastic strip. The operator dips the strip into a urine sample and visually compares the color of each pad to a color chart provided on the vial label. The device provides qualitative or semi-quantitative results for 11 analytes: urobilinogen, glucose, bilirubin, ketones, specific gravity, blood, pH, protein, nitrite, leukocytes, and ascorbic acid. These results assist clinicians in assessing carbohydrate metabolism, kidney and liver function, acid-base balance, and urinary tract infections. The device is intended for professional use in point-of-care or laboratory settings.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Reagent strip for urinalysis; colorimetric chemical reaction principle; visual readout; single-use; professional use only.
Indications for Use
Indicated for single use in professional near-patient (point-of-care) and centralized laboratory locations for screening at-risk patients to assist diagnosis of kidney function, urinary tract infections, carbohydrate metabolism (e.g., diabetes mellitus), liver function, acid-base balance, and urine concentration. Results are used with other diagnostic information to rule out disease states and determine if microscopic analysis is needed. For prescription use only.
Regulatory Classification
Identification
A urinary glucose (nonquantitative) test system is a device intended to measure glucosuria (glucose in urine). Urinary glucose (nonquantitative) measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, hypoglycemia, and hyperglycemia.
Submission Summary (Full Text)
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE
A. 510(k) Number:
k052525
B. Purpose for Submission:
Clearance to market CYBOW 11 Series Reagent Strips for Urinalysis
C. Measurand:
Urobilinogen, bilirubin and its conjugates, ketones (acetoacetic acid), blood, glucose, protein, nitrite, leukocytes, glucose, specific gravity, pH, and ascorbic acid in urine.
D. Type of Test:
Qualitative and semi-quantitative urine tests
E. Applicant:
DFI Co., Ltd.
F. Proprietary and Established Names:
CYBOW 11 Series Reagent Strips for Urinalysis
G. Regulatory Information:
1. Regulation section:
21 CFR §864.6550: Occult blood test.
21 CFR §862.1340: Urinary glucose (nonquantitative) test system.
21 CFR §862.1785: Urinary urobilinogen (nonquantitative) test system.
21 CFR §862.1115: Urinary bilirubin and its conjugates (nonquantitative) test system.
21 CFR §862.1435: Ketones (nonquantitative) test system.
21 CFR §862.1645: Urinary protein or albumin (nonquantitative) test system.
21 CFR §862.1510: Nitrite (nonquantitative) test system.
21 CFR §864.7675: Leukocyte peroxidase test.
21 CFR §862.1550: Urinary pH (nonquantitative) test system.
21 CFR §862.1095: Ascorbic acid test system.
{1}
2. Classification:
Class II
3. Product code:
Occult blood test - JIO
Urinary glucose (nonquantitative) test system - JIL
Urinary urobilinogen (nonquantitative) test system - CDM
Urinary bilirubin and its conjugates (nonquantitative) test system - JJB
Ketones (non-quantitative) test system - JIN
Urinary protein or albumin (non-quantitative) test system - JIR
Nitrite (non-quantitative) test system - JMT
Leukocyte peroxidase test - LJX
Urinary pH (nonquantitative) - CEN
Ascorbic acid test system - JMA
4. Panel:
Chemistry (75)
Hematology (82)
H. Intended Use:
1. Intended use(s):
CYBOW 11 Reagent Strips are for the rapid visual determination of urobilinogen, glucose, bilirubin, ketones (acetoacetic acid), specific gravity, blood, pH, protein, nitrite, leukocytes and ascorbic acid in urine.
2. Indication(s) for use:
CYBOW 11 Reagent Strips are for single use in professional near-patient (point of-care) and centralized laboratory locations. The strips are intended for use in screening at-risk patients to assist diagnosis in the following areas:
- Kidney function
- Urinary tract infections
- Carbohydrate metabolism (e.g. diabetes mellitus)
- Liver function
- Acid-base balance
- Urine concentration
The results can be used along with other diagnostic information to rule out certain disease states and to determine if microscopic analysis is needed. The test is to be read visually.
2
{2}
3. Special conditions for use statement(s):
For prescription use only
4. Special instrument requirements:
Not applicable to this device.
I. Device Description:
CYBOW Reagent Strips are dip-and-read test strips for In Vitro Diagnostic Use only for testing analytes in urine. Test result may provide information regarding the status of carbohydrate metabolism, kidney and liver function, acid-base balance, and urinary tract infection. It is measured by comparison of test paper attached to a plastic strip to a color chart printed on the vial label.
J. Substantial Equivalence Information:
1. Predicate device name(s):
Bayer Corporation MULTISTIX 10SG Reagent Strips
2. Predicate 510(k) number(s):
k852611
3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Specimen | Urine | Same |
| Intended Use Audience | Professional use | Same |
| Test results | Color comparison | Color comparison |
K. Standard/Guidance Document Referenced (if applicable):
CLSI EP6-A: "Evaluation of Linearity of Quantitative Measurement Procedures: A Statistical Approach"; Approved Guideline, 2003
CLSI EP09-A2: "Method Comparison and Bias Estimation Using Patient Samples"; Approved Guideline, 2002
CLSI EP12-A: "User Protocol for Evaluation of Qualitative Test Performance"; Approved Guideline, 2002
{3}
ISO 2895-1: “Sampling Procedures for Inspection by Attributes - Part 1: Sampling Schemes Indexed by Acceptance Quality Limit (AQL) for Lot-by-Lot Inspection-Second Edition”
L. Test Principle:
The device is composed of multiple chemically reactive spots separate from each other on a plastic strip. Read-out is accomplished by visually matching the position and color of an exposed spot to a color coded chart provided with the device.
For the detection of urobilinogen, the device employs a modified Ehrlich’s reaction. Urobilinogen reacts with Ehrlich’s reagent to form a red-colored compound. Color changes from light orange-pink to dark pink.
For the detection of glucose, the device employs glucose oxidase to catalyze the oxidation of glucose to form hydrogen peroxide. The hydrogen peroxide thus formed then oxidizes a chromogen on the reaction pad by the action of peroxidase.
For the detection of bilirubin, diazonium salts in an acidic matrix on the strip undergo an azo-coupling reaction with bilirubin to form an azodye. The spot color changes from light tan to beige or light pink.
The device uses Legal’s test-nitroprusside reaction for the detection of ketones. Acetoacetic acid in an alkaline medium reacts with nitroferricyanide to produce a color change from beige to purple.
The device uses a correlation between the concentration of ionic species and the sample’s specific gravity to report an estimated specific gravity. Ionic solutes present in the urine release protons from a polyelectrolyte. The released protons decrease the pH on that strip spot and produce a color change in bromothymol blue from blue-green to yellow-green.
To detect blood, the device exploits the pseudo-peroxidase activity of the haem moiety of hemoglobin and myoglobin. A chromogen is oxidized by a hydroperoxide in the presence of haem and changes color from yellow to blue.
The device employs a combination of methyl red and bromothymol blue indicators to give distinct color changes from orange to green to blue (pH 5.0 to 9.0).
To qualitatively estimate protein concentration, the device uses the “error of indicators” principle. Proteins interact with an ionizable electrolyte driving the release of protons which in turn interact with a spectator indicator. The color change in the indicator is correlated with the protein concentration.
The device detects nitrite through a reaction of the nitrite with an aromatic amine. The resulting diazo compound reacts with another aromatic to form a pink compound.
4
{4}
One location on the strip contains an indoxyl ester and diazonium salt. Leukocytes contain an esterase that hydrolyzes the indoxyl ester. The liberated compound reacts with the diazonium salt on the strip to generate a purple compound. The concentration of leukocytes is correlated with a color change from beige to violet.
The ascorbic acid test uses the decolorization of Tillmann's reagent. The presence of ascorbic acid causes the color of the test field to change from gray-blue to yellow.
## M. Performance Characteristics (if/when applicable):
### 1. Analytical performance:
#### a. Precision/Reproducibility:
Within-run and Within-day precisions were determined at the manufacture's site, using level 1 and level 2 urinalysis controls.
Reproducibility testing was done in accordance with CLSI EP12-A "User Protocol for Evaluation of Qualitative Test Performance". Two clinical pathologists were employed as readers for the precision testing.
For within-run precision testing, twenty replicates were run on each of two levels of liquid urine controls. Each of the twenty replicates was assayed consecutively, using strips obtained from each of 3 lots of vial. For within-day precision testing, two levels were analyzed in duplicate, one a day, for 20 days using strips obtained from 3 lots of vials. Percentage agreements were calculated using the concentrations specified in the control labeling and the concentration ranges specified for the strip.
The results of the within-day testing:
| | Negative Control | | | | | Positive Control | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Analyte | Expected Result | Lot 10205 | Lot 10721 | Lot 20108 | Total Agreement | Expected Result | Lot 10205 | Lot 10721 | Lot 20108 | Total Agreement |
| Urobilinogen | Negative | 20/20 | 20/20 | 59/60 | 59/60 (98.3%) | 4 mg/dl | 19/20 | 19/20 | 18/20 | 56/60 (93.3%) |
| Glucose | Negative | 20/20 | 20/20 | 20/20 | 60/60 (100%) | 1000 mg/dl | 19/20 | 19/20 | 20/20 | 58/60 (96.6%) |
| Bilirubin | Negative | 19/20 | 20/20 | 19/20 | 58/60 (96.7%) | +++ | 17/20 | 17/20 | 18/20 | 52/60 (86.7%) |
| Ketones (acetoacetic acid) | Negative | 20/20 | 20/20 | 20/20 | 60/60 (100%) | 40 mg/dl | 20/20 | 17/20 | 19/20 | 56/60 (93.3%) |
| Specific Gravity | 1.020 | 17/20 | 18/20 | 19/20 | 54/60 (90%) | 1.020 | 17/20 | 18/20 | 17/20 | 52/60 (86.7%) |
| Blood | Negative | 19/20 | 20/20 | 20/20 | 59/60 (98.3%) | 250 RBC/ul | 19/20 | 20/20 | 19/20 | 58/60 (96.6%) |
| pH | 6 | 18/20 | 20/20 | 19/20 | 57/60 (95%) | 7 | 17/20 | 17/20 | 18/20 | 52/60 (86.7%) |
{5}
The results of the within-run testing:
| | Negative Control | | | | | Positive Control | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Analyte | Result | Lot 10205 | Lot 10721 | Lot 20108 | Total Agreement | Expected Result | Lot 10205 | Lot 10721 | Lot 20108 | Total Agreement |
| Protein | Negative | 20/20 | 20/20 | 19/20 | 59/60 (98.3%) | 100 mg/dl | 19/20 | 20/20 | 19/20 | 58/60 (96.7%) |
| Nitrite | Negative | 20/20 | 19/20 | 20/20 | 59/60 (98.3%) | Positive | 20/20 | 20/20 | 20/20 | 60/60 (100%) |
| Leukocytes | Negative | 20/20 | 20/20 | 20/20 | 60/60 (100%) | 75 WBC/ul | 18/20 | 19/20 | 17/20 | 54/60 (90.0%) |
| Ascorbic | Negative | 20/20 | 20/20 | 20/20 | 60/60 (100%) | Negative | 20/20 | 20/20 | 20/20 | 60/60 (100%) |
b. Linearity/assay reportable range:
The company assessed the range of the device by repeated testing with urine containing known concentrations of the measured analytes.
Samples of the candidate device were randomly selected from routine production following ISO 2895-1: "Sampling procedures for inspection by attribute-sampling scheme indexed by acceptances quality limit (AQL) by lot-to-lot inspection". Strips were selected from three manufacturing lots.
{6}
Test material was created by spiking analytes into negative urine or by serial dilution with negative urine of a known high concentration of an analyte in urine. Test material was adjusted to match the mid-point of a concentration range for a particular color block on the strip – if a particular color indicated a concentration of $2 - 4\mathrm{mg / dL}$, the test urine was adjusted to $3\mathrm{mg / dL}$.
Three pathologists independently read each of the three lots of test strips. Tests were done 10 times at each test concentration with 3 different product lots. Sample application was done according to the device's product insert. Color comparisons were made against the color chart that accompanies the product.
The company illustrated the performance of their device by presenting the results of these tests as a scoring matrix. The results of this testing is summarized below.
For urobilinogen, the company demonstrated:
Applied Concentration, mg/dL
| | | 0.1 | 1 | 2 | 4 | 8 |
| --- | --- | --- | --- | --- | --- | --- |
| Concentration reported by Strip, mg/dL | 8 | 0 | 0 | 0 | 1 | 28 |
| | 4 | 0 | 0 | 2 | 25 | 2 |
| | 2 | 0 | 1 | 26 | 3 | 0 |
| | 1 | 0 | 29 | 2 | 1 | 0 |
| | 0.1 | 30 | 0 | 0 | 0 | 0 |
| Total Number of samples applied at concentration | | 30 | 30 | 30 | 30 | 30 |
For glucose, the company demonstrated:
Applied Concentration, mg/dL
| | | Neg. | 100 | 250 | 500 | 1000 |
| --- | --- | --- | --- | --- | --- | --- |
| Concentration reported by Strip, mg/dL | 1000 | 0 | 0 | 0 | 2 | 26 |
| | 500 | 0 | 0 | 2 | 27 | 4 |
| | 250 | 0 | 1 | 27 | 1 | 0 |
| | 100 | 0 | 29 | 1 | 0 | 0 |
| | Neg. | 30 | 0 | 0 | 0 | 0 |
| Total Number of samples applied at concentration | | 30 | 30 | 30 | 30 | 30 |
{7}
For bilirubin, the company demonstrated:
| | Neg. | + | ++ | +++ |
| --- | --- | --- | --- | --- |
| +++ | 0 | 0 | 3 | 29 |
| ++ | 0 | 4 | 27 | 1 |
| + | 0 | 25 | 0 | 0 |
| Neg. | 30 | 1 | 0 | 0 |
| Total Number of samples applied at concentration | 30 | 30 | 30 | 30 |
For ketones, the company demonstrated:
| | Applied Concentration, mg/dL | | | | | |
| --- | --- | --- | --- | --- | --- | --- |
| | | Neg. | 5 | 15 | 40 | 100 |
| Concentration reported by Strip, mg/dL | 100 | 0 | 0 | 0 | 2 | 28 |
| | 40 | 0 | 0 | 1 | 25 | 2 |
| | 15 | 0 | 0 | 28 | 3 | 0 |
| | 5 | 0 | 29 | 1 | 0 | 0 |
| | Neg. | 30 | 1 | 0 | 0 | 0 |
| Total Number of samples applied at concentration | | 30 | 30 | 30 | 30 | 30 |
For the specific gravity, the company demonstrated:
| | Applied Specific Gravity | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | 1.000 | 1.005 | 1.010 | 1.015 | 1.020 | 1.025 | 1.030 |
| Specific gravity reported by Strip | 1.030 | 0 | 0 | 0 | 0 | 0 | 2 | 29 |
| | 1.025 | 0 | 0 | 0 | 0 | 4 | 26 | 1 |
| | 1.020 | 0 | 0 | 0 | 1 | 24 | 2 | 0 |
| | 1.015 | 0 | 0 | 1 | 26 | 2 | 0 | 0 |
| | 1.010 | 0 | 2 | 26 | 3 | 0 | 0 | 0 |
| | 1.005 | 0 | 28 | 3 | 0 | 0 | 0 | 0 |
| | 1.000 | 30 | 0 | 0 | 0 | 0 | 0 | 0 |
| Total Number of samples applied at concentration | | 30 | 30 | 30 | 30 | 30 | 30 | 30 |
{8}
For hemolyzed blood, the company demonstrated:
Applied RBC/ $\mu$ l
| | | Neg. | 10 RBC/μl | 50 RBC/μl | 250 RBC/μl |
| --- | --- | --- | --- | --- | --- |
| RBC/μl reported by Strip | 250 | | | | |
| | RBC/μl | 0 | 0 | 3 | 30 |
| | 50 | | | | |
| | RBC/μl | 0 | 3 | 26 | 0 |
| | 10 | | | | |
| Total Number of samples applied at concentration | RBC/μl | 0 | 27 | 1 | 0 |
| | Neg. | 30 | 0 | 0 | 0 |
| | | 30 | 30 | 30 | 30 |
For non-hemolyzed blood, the company demonstrated:
| | | Applied Non-hemolyzed blood | | |
| --- | --- | --- | --- | --- |
| | | Neg. | + | ++ |
| Non-hemolyzed blood reported by Strip | ++ | 0 | 4 | 27 |
| | + | 0 | 26 | 3 |
| | Neg. | 30 | 0 | 0 |
| Total Number of samples applied at concentration | | 30 | 30 | 30 |
For pH, the company demonstrated:
| pH reported by Strip | Applied pH | | | | | |
| --- | --- | --- | --- | --- | --- | --- |
| | 5 | 6 | 6.5 | 7 | 8 | 9 |
| | 9 | 0 | 0 | 0 | 1 | 29 |
| | 8 | 0 | 0 | 0 | 2 | 28 |
| | 7 | 0 | 0 | 3 | 28 | 1 |
| | 6.5 | 0 | 0 | 23 | 0 | 0 |
| | 6 | 1 | 26 | 4 | 0 | 0 |
| | 5 | 29 | 4 | 0 | 0 | 0 |
| Total Number of samples applied at concentration | 30 | 30 | 30 | 30 | 30 | 30 |
{9}
For protein, the company demonstrated:
| | | Applied protein, mg/dL | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | | Neg. | 15 | 30 | 100 | 300 | 1000 |
| Protein reported by Strip, mg/dL | 1000 | 0 | 0 | 0 | 0 | 1 | 29 |
| | 300 | 0 | 0 | 0 | 2 | 28 | 1 |
| | 100 | 0 | 0 | 3 | 28 | 1 | 0 |
| | 30 | 0 | 0 | 23 | 0 | 0 | 0 |
| | 15 | 1 | 26 | 4 | 0 | 0 | 0 |
| | Neg. | 29 | 4 | 0 | 0 | 0 | 0 |
| Total Number of samples applied at concentration | | 30 | 30 | 30 | 30 | 30 | 30 |
For nitrite, the company demonstrated:
| | | Applied nitrite | | |
| --- | --- | --- | --- | --- |
| | | Neg. | Tr. | Pos. |
| Nitrite reported by Strip | Pos. | 0 | 1 | 30 |
| | Tr. | 1 | 28 | 0 |
| | Neg. | 29 | 1 | 0 |
| Total Number of samples applied at concentration | | 30 | 30 | 30 |
For leukocytes, the company demonstrated:
| Applied leukocytes/μl | | | | | |
| --- | --- | --- | --- | --- | --- |
| | | Neg. | 25 Leu/μl | 75 Leu/μl | 500 Leu/μl |
| Lue/μl reported by Strip | 500 Leu/μl | 0 | 0 | 1 | 29 |
| | 75 Leu/μl | 0 | 1 | 26 | 1 |
| | 25 Leu/μl | 0 | 25 | 3 | 0 |
| | Neg. | 30 | 4 | 0 | 0 |
| | Total Number of samples applied at concentration | 30 | 30 | 30 | 30 |
{10}
For ascorbic acid, the company demonstrated:
| | | Applied ascorbic acid | | |
| --- | --- | --- | --- | --- |
| | | 20 | 40 | |
| | | Neg. | mg/dL | mg/dL |
| Ascorbic acid reported by Strip | 40 mg/dL | 0 | 4 | 27 |
| | 20 mg/dL | 0 | 26 | 3 |
| | Neg. | 30 | 0 | 0 |
| Total Number of samples applied at concentration | | 30 | 30 | 30 |
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
The company used a combination of real-time and accelerated aging studies to demonstrate their claimed shelf life. The company used samples taken from three different manufacturing lots in assessing their shelf life.
The company conducted real time aging studies to empirically determine the shelf life of their device. Bottles from each manufacturing lot were stored in an environmental chamber at temperatures ranging from $2 - 30^{\circ}\mathrm{C}$ . The relative humidity was set to less than $30\%$ . Strips were withdrawn monthly and tested with negative and spiked positive urine samples. The company demonstrated that the performance of their device did not change over the course of this 2 year storage test.
In addition, the company conducted accelerated aging studies at storage temperatures of $40^{\circ}\mathrm{C}$ , $50^{\circ}\mathrm{C}$ , and $60^{\circ}\mathrm{C}$ . Data supplied by company demonstrated that the strips could be stored at $60^{\circ}\mathrm{C}$ for 3 weeks without a change in performance. Strips could be stored at $50^{\circ}\mathrm{C}$ for 4 weeks and $40^{\circ}\mathrm{C}$ for 6 weeks without a change in performance.
The company conducted tests at room temperature and high relative humidity to quantify the impact of ambient water vapor. Strips stored at less than $50\%$ relative humidity did not demonstrate a change in performance over a 2 year period.
The data provided by the company supports their claim that the CYBOW 11 Reagent strips for Urinalysis are stable for 24 months when stored between $15 - 30^{\circ}\mathrm{C}$ and a relative humidity of less than $50\%$ .
d. Detection limit:
The company demonstrated their detection limit by challenging the concentration cutoff with samples adjusted to concentrations $20\%$ above and $20\%$ below the cutoff concentration specified for the analyte. The company
{11}
conducted this study using 20 strips from each of 3 lots for a total of 60 measurements at each concentration.
For urobilinogen, the company demonstrated:
| Sample | 0.8*Cutoff Conc. | Cutoff Conc. | 1.2*Cutoff Conc. |
| --- | --- | --- | --- |
| Normal | 58 | 0 | 0 |
| Positive | 2 | 60 | 60 |
| Percent Correct | 96.7% | 100% | 100% |
| Cutoff Used | 2 Ehrlich units/dL | | |
For glucose, the company demonstrated:
| Sample | 0.8*Cutoff Conc. | Cutoff Conc. | 1.2*Cutoff Conc. |
| --- | --- | --- | --- |
| Normal | 57 | 1 | |
| Positive | 3 | 59 | 60 |
| Percent Correct | 95% | 98.3% | 100% |
| Cutoff Used | 50 mg/dL | | |
For bilirubin, the company demonstrated:
| Sample | 0.8*Cutoff Conc. | Cutoff Conc. | 1.2*Cutoff Conc. |
| --- | --- | --- | --- |
| Normal | 59 | 0 | 0 |
| Positive | 1 | 60 | 60 |
| Percent Correct | 98.3% | 100% | 100% |
| Cutoff Used | 0.5 mg/dL | | |
For ketones, the company demonstrated:
| Sample | 0.8*Cutoff Conc. | Cutoff Conc. | 1.2*Cutoff Conc. |
| --- | --- | --- | --- |
| Normal | 57 | 0 | 0 |
| Positive | 3 | 60 | 60 |
| Percent Correct | 95% | 100% | 100% |
| Cutoff Used | 5 mg/dL | | |
For hemolyzed blood, the company demonstrated:
| Sample | 0.8*Cutoff Conc. | Cutoff Conc. | 1.2*Cutoff Conc. |
| --- | --- | --- | --- |
| Normal | 52 | 1 | 0 |
| Positive | 8 | 59 | 60 |
| Percent Correct | 86.7% | 98.3% | 100% |
| Cutoff Used | 10 RBC/μL | | |
{12}
For non-hemolyzed blood, the company demonstrated:
| Sample | 0.8*Cutoff Conc. | Cutoff Conc. | 1.2*Cutoff Conc. |
| --- | --- | --- | --- |
| Normal | 50 | 2 | 0 |
| Positive | 10 | 58 | 60 |
| Percent Correct | 83.3% | 96.7% | 100% |
| Cutoff Used | 10 RBC/μL | | |
For protein, the company demonstrated:
| Sample | 0.8*Cutoff Conc. | Cutoff Conc. | 1.2*Cutoff Conc. |
| --- | --- | --- | --- |
| Normal | 42 | 8 | 0 |
| Positive | 18 | 52 | 60 |
| Percent Correct | 70.0% | 86.7% | 100% |
| Cutoff Used | 12 mg/dL | | |
For nitrite, the company demonstrated:
| Sample | 0.8*Cutoff Conc. | Cutoff Conc. | 1.2*Cutoff Conc. |
| --- | --- | --- | --- |
| Normal | 55 | 0 | 0 |
| Positive | 5 | 60 | 60 |
| Percent Correct | 91.7% | 100% | 100% |
| Cutoff Used | 0.05 mg/dL | | |
For leukocytes, the company demonstrated:
| Sample | 0.8*Cutoff Conc. | Cutoff Conc. | 1.2*Cutoff Conc. |
| --- | --- | --- | --- |
| Normal | 41 | 5 | 0 |
| Positive | 19 | 55 | 60 |
| Percent Correct | 68.3% | 91.7% | 100% |
| Cutoff Used | 23 WBC/μL | | |
For ascorbic acid, the company demonstrated:
| Sample | 0.8*Cutoff Conc. | Cutoff Conc. | 1.2*Cutoff Conc. |
| --- | --- | --- | --- |
| Normal | 58 | 0 | 0 |
| Positive | 2 | 60 | 60 |
| Percent Correct | 96.7% | 100% | 100% |
| Cutoff Used | 20 mg/dL | | |
e. Analytical specificity:
The company conducted tests using analytes suspected of interfering with their device. Fresh negative urine was spiked with the analyte of interest and the interfering analyte. The company documented the impact of the interfering analyte in their product insert. A summary of the company's findings:
{13}
| Test | Interfering Analyte | Impact on test |
| --- | --- | --- |
| Urobilinogen | p-amino salicylic acid | False positive |
| | Azo gantrinsin | Test inoperative – unable to read |
| Glucose | High specific gravity (>1.2) | False negative |
| | High pH (>8.5) | False negative |
| | Ascorbic acid (>40 mg/dL) | False negative |
| | Elevated ketones (>40 mg/dL) | False negative |
| Bilirubin | Selenium | False positive |
| | Pyrindum (drug metabolite) | False positive |
| | Indoxyl sulfate | Test inoperative – unable to read |
| | Therapeutic dyes | False positive |
| Ketones | Colored contaminates | False positive |
| | L-dopa and metabolites | False positive |
| | High specific gravity (>1.025) | False positive |
| | Low pH | False positive |
| | Phenosulfonphthalein(>0.05 mg/dL) | False positive |
| Blood (hemolyzed and non-hemolyzed) | Microbial peroxidase | False positive |
| | Ascorbic acid (>40 mg/dL) | False negative |
| Specific Gravity | Elevated proteins (>300 mg/dL) | High specific gravity not indicative of high urine electrolytes |
| | Buffered alkaline samples (pH > 8.5) | Low reported specific gravity |
| | Buffered acidic samples (pH < 5) | Low reported specific gravity |
| pH | Cross-contamination from adjacent reagent pads | Low reported pH |
| Protein | Turbidity/ Lipemic samples | Test inoperative – unable to read |
| | High pH (>9) | False positive |
| Nitrite | | |
| | Ascorbic acid | False negative |
| Leukocyte | High glucose (>1000 mg/dL) | Low reported WBC/μL |
| | High albumin (>1000 mg/dL) | Low reported WBC/μL |
| | Formaldehyde (>5%) | Low reported WBC/μL |
| | Oxalic acid (>50 mg/dL) | Low reported WBC/μL |
| | Oxidizing agents (permanganate) | Low reported WBC/μL |
# f. Assay cut-off:
Not applicable for devices of this type. The performance of the device around
{14}
the "not present"/"present" cutoff is discussed under d) Detection Limit above.
# 2. Comparison studies:
# a. Method comparison with predicate device:
The company conducted a method comparison of the submitted device to the legally marketed predicate, the Bayer Multistix 10 SG. Using fresh negative and spiked urine samples, 3 pathologists recorded the results of measurements on 300 samples made with the proposed device and to those made with the Bayer predicate. Strips from 3 manufacturing lots of the propose device were used in the comparison. Measurements were made according to the respective product inserts.
For the specific gravity, the company the company demonstrated the following correlation between the proposed and predicate device:
| Proposed Device | 1.030 | 0 | 0 | 0 | 0 | 0 | 0 | 50 |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | 1.025 | 0 | 0 | 0 | 0 | 10 | 54 | 18 |
| | 1.020 | 0 | 0 | 0 | 8 | 56 | 0 | 0 |
| | 1.015 | 0 | 0 | 15 | 48 | 1 | 0 | 0 |
| | 1.010 | 0 | 0 | 33 | 1 | 0 | 0 | 0 |
| | 1.005 | 0 | 6 | 0 | 0 | 0 | 0 | 0 |
| | 1.000 | 0 | 0 | 0 | 0 | 0 | 0 | 0 |
| | SG | 1.000 | 1.005 | 1.010 | 1.015 | 1.020 | 1.025 | 1.030 |
| | Predicate Device | | | | | | | |
For pH, the company the company demonstrated the following correlation between the proposed and predicate device:
| Proposed Device | 9 | 0 | 0 | 0 | 0 | 3 | 2 | 7 |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | 8 | 0 | 0 | 0 | 0 | 15 | 8 | 3 |
| | 7 | 0 | 0 | 2 | 43 | 3 | 2 | 1 |
| | 6.5 | 2 | 5 | 21 | 12 | 0 | 0 | 0 |
| | 6 | 12 | 84 | 17 | 2 | 0 | 0 | 0 |
| | 5 | 47 | 9 | 0 | 0 | 0 | 0 | 0 |
| | pH | 5 | 6 | 6.5 | 7 | 7.5 | 8 | 8.5 |
| | Predicate Device | | | | | | | |
{15}
For urobilinogen, the company demonstrated the following correlation between the proposed and predicate device:
| Proposed Device (mg/dL) | 8 | 0 | 0 | 0 | 3 | 55 |
| --- | --- | --- | --- | --- | --- | --- |
| | 4 | 0 | 0 | 3 | 53 | 4 |
| | 2 | 0 | 3 | 54 | 3 | 1 |
| | 1 | 0 | 53 | 3 | 1 | 0 |
| | 0.1 | 60 | 4 | 0 | 0 | 0 |
| | Uro | 0.1 | 1 | 2 | 4 | 8 |
| | | Predicate Device (mg/dL) | | | | |
For glucose, the company demonstrated the following correlation between the proposed and predicate device:
| Proposed Device (mg/dL) | >1000 | 0 | 0 | 1 | 3 | 33 | 19 |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | 500 | 0 | 0 | 3 | 54 | 5 | 1 |
| | 250 | 0 | 4 | 53 | 2 | 2 | 0 |
| | 100 | 0 | 54 | 3 | 1 | 0 | 0 |
| | Negative | 60 | 2 | 0 | 0 | 0 | 0 |
| | Glucose | Negative | 100 | 250 | 500 | 1000 | 2000 |
| | Predicate Device (mg/dL) | | | | | | |
For urinary bilirubin, the company demonstrated the following correlation between the proposed and predicate device:
| Proposed Device | 3+ | 0 | 2 | 2 | 72 |
| --- | --- | --- | --- | --- | --- |
| | 2+ | 0 | 4 | 71 | 3 |
| | 1+ | 4 | 70 | 5 | 5 |
| | Negative | 56 | 4 | 2 | 0 |
| | Bilirubin | Negative | 1+ | 2+ | 3+ |
| | Predicate Device | | | | |
For urinary ketones, the company demonstrated the following correlation between the proposed and predicate device:
| Proposed Device (mg/dL) | >100 | 0 | 0 | 1 | 3 | 34 | 18 |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | 40 | 0 | 0 | 2 | 50 | 4 | 2 |
| | 15 | 0 | 4 | 52 | 4 | 2 | 0 |
| | 5 | 2 | 51 | 4 | 3 | 0 | 0 |
| | Negative | 58 | 5 | 1 | 0 | 0 | 0 |
| | Ketones | Negative | 5 | 15 | 40 | 80 | 160 |
| | Predicate Device (mg/dL) | | | | | | |
{16}
For hemolyzed urinary blood, the company demonstrated the following correlation between the proposed and predicate device:
| Proposed Device (RBC/μl) | 250 | 0 | 0 | 0 | 5 | 43 |
| --- | --- | --- | --- | --- | --- | --- |
| | 50 | 0 | 0 | 5 | 41 | 5 |
| | 10 | 4 | 45 | 45 | 4 | 2 |
| | Negative | 56 | 5 | 0 | 0 | 0 |
| | Hemolyzed Blood | Negative | 10 | 25 | 80 | 250 |
| | Predicate device (RBC/μl) | | | | | |
For non-hemolyzed blood in the urine, the company demonstrated the following correlation between the proposed and predicate device:
| Proposed Device (RBC/μl) | 80 | 0 | 1 | 17 |
| --- | --- | --- | --- | --- |
| | 10 | 4 | 17 | 3 |
| | Negative | 56 | 2 | 0 |
| | Non-hemolyzed blood | Negative | 10 | 80 |
| | Predicate device (RBC/μl) | | | |
For urinary protein, the company demonstrated the following correlation between the proposed and predicate device:
| Proposed Device (mg/dL) | 1000 | 0 | 0 | 0 | 0 | 2 | 18 |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | 300 | 0 | 0 | 0 | 3 | 35 | 2 |
| | 100 | 0 | 0 | 1 | 55 | 3 | 0 |
| | 30 | 0 | 3 | 54 | 2 | 0 | 0 |
| | 15 | 3 | 53 | 4 | 0 | 0 | 0 |
| | Negative | 57 | 4 | 1 | 0 | 0 | 0 |
| | Protein | Negative | 15 | 30 | 100 | 300 | 1000 |
| | | Predicate device (mg/dL) | | | | | |
For nitrite, the company demonstrated the following correlation between the proposed and predicate device:
| Proposed Device | Positive | 0 | 11 | 95 |
| --- | --- | --- | --- | --- |
| | +/- | 0 | 87 | 5 |
| | Negative | 100 | 2 | 0 |
| | Nitrite | Negative | +/- | Positive |
| | Predicate device | | | |
For leukocytes in the urine, the company demonstrated the following correlation between the proposed and predicate device:
| Proposed Device (Leu/μL) | 500 | 0 | 0 | 3 | 4 | 55 |
| --- | --- | --- | --- | --- | --- | --- |
| | 75 | 0 | 3 | 54 | 54 | 3 |
| | 25 | 2 | 44 | 2 | 2 | 2 |
| | Negative | 58 | 13 | 1 | 0 | 0 |
| | Leukocytes | Negative | 15 | 70 | 125 | 500 |
| | Predicate device (Leu/μL) | | | | | |
{17}
b. Matrix comparison:
The CYBOW 11 Reagent Strips for Urinalysis are only for use with urine.
# 3. Clinical studies:
a. Clinical Sensitivity:
Not applicable for a device of this type.
b. Clinical specificity:
Not applicable for a device of this type.
c. Other clinical supportive data (when a. and b. are not applicable):
The company demonstrated the equivalence of their proposed device to their predicate using patient samples analyzed in the environment in which the device will be used. The company undertook a clinical comparison of the device to the predicate at 2 hospitals' clinical laboratories. A total of 909 patient samples were compared over 30 days.
The clinical comparison followed CLSI EP9A "Method Comparison and Bias Estimation Using Patient Samples". In the study, the company used 3 different manufacturing lots for the proposed device and 3 different lots of strips from the cleared, on-market predicate.
Fresh urine samples obtained during routine processing at the medical facility were analyzed within 4 hours of collection. The fresh urine samples were detergent-free and were not centrifuged or spiked. Strip readings were conducted by laboratory personnel according to the respective device instructions.
For urobilinogen, the company demonstrated the following correlation between the proposed and predicate device using patient samples in a clinical lab setting:
| Proposed Device (mg/dL) | 8 | 0 | 0 | 0 | 3 | 6 |
| --- | --- | --- | --- | --- | --- | --- |
| | 4 | 4 | 0 | 2 | 16 | 5 |
| | 2 | 9 | 6 | 17 | 4 | 2 |
| | 1 | 10 | 22 | 6 | 1 | 0 |
| | Normal | 771 | 24 | 1 | 0 | 0 |
| | Urobilinogen | Normal | 1 | 2 | 4 | 8 |
| | Predicate Device (mg/dL) | | | | | |
For glucose, the company demonstrated the following correlation between the proposed and predicate device using patient samples in a clinical lab setting:
{18}
| Proposed Device (mg/dL) | 1000 | 0 | 0 | 2 | 6 | 36 | 34 |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | 500 | 0 | 1 | 15 | 40 | 2 | 3 |
| | 250 | 1 | 14 | 59 | 6 | 0 | 0 |
| | 100 | 23 | 53 | 7 | 5 | 0 | 0 |
| | Negative | 575 | 24 | 3 | 0 | 0 | 0 |
| | Glucose | Negative | 100 | 250 | 500 | 1000 | 2000 |
| | Predicate Device (mg/dL) | | | | | | |
For bilirubin, the company demonstrated the following correlation between the proposed and predicate device using patient samples in a clinical lab setting:
| Proposed Device (mg/dL) | 3+ | 0 | 1 | 4 | 21 |
| --- | --- | --- | --- | --- | --- |
| | 2+ | 0 | 4 | 24 | 3 |
| | 1+ | 30 | 29 | 10 | 7 |
| | Negative | 750 | 21 | 5 | 0 |
| | Bilirubin | Negative | 1+ | 2+ | 3+ |
| | Predicate Device (mg/dL) | | | | |
For urinary ketones, the company demonstrated the following correlation between the proposed and predicate device using patient samples in a clinical lab setting:
| Proposed Device (mg/dL) | >100 | 0 | 0 | 0 | 3 | 33 | 0 |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | 40 | 0 | 0 | 4 | 47 | 9 | 0 |
| | 15 | 7 | 8 | 32 | 10 | 4 | 0 |
| | 5 | 17 | 45 | 8 | 2 | 0 | 0 |
| | Negative | 658 | 20 | 2 | 0 | 0 | 0 |
| | Ketones | Negative | 5 | 15 | 40 | 80 | 160 |
| | Predicate Device (mg/dL) | | | | | | |
For specific gravity, the company demonstrated the following correlation between the proposed and predicate device using patient samples in a clinical lab setting:
| Proposed Device | 1.030 | 0 | 0 | 0 | 0 | 11 | 33 | 62 |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | 1.025 | 0 | 0 | 0 | 16 | 37 | 63 | 13 |
| | 1.020 | 0 | 0 | 20 | 62 | 91 | 30 | 9 |
| | 1.015 | 0 | 11 | 49 | 102 | 42 | 3 | 0 |
| | 1.010 | 4 | 58 | 83 | 9 | 2 | 0 | 0 |
| | 1.005 | 6 | 56 | 16 | 4 | 0 | 0 | 0 |
| | 1.000 | 14 | 4 | 0 | 0 | 0 | 0 | 0 |
| | SG | 1.000 | 1.005 | 1.010 | 1.015 | 1.020 | 1.025 | 1.030 |
| | Predicate Device | | | | | | | |
For urinary blood, the company demonstrated the following correlation between the proposed and predicate device using patient samples in a clinical lab setting:
{19}
| Proposed Device (RBC/μl) | 250 | 0 | 0 | 7 | 18 | 128 |
| --- | --- | --- | --- | --- | --- | --- |
| | 50 | 3 | 22 | 28 | 120 | 20 |
| | 10 | 12 | 98 | 101 | 26 | 8 |
| | Negative | 280 | 28 | 9 | 1 | 0 |
| | Hemolyzed Blood | Negative | 10 | 25 | 80 | 250 |
| | Predicate device (RBC/μl) | | | | | |
For urinary pH, the company demonstrated the following correlation between the proposed and predicate device using patient samples in a clinical lab setting:
| Proposed Device | 9 | 0 | 0 | 0 | 0 | 0 | 0 | 9 |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | 8 | 0 | 0 | 0 | 12 | 21 | 67 | 10 |
| | 7 | 0 | 2 | 15 | 110 | 92 | 16 | 0 |
| | 6.5 | 5 | 21 | 75 | 32 | 7 | 2 | 0 |
| | 6 | 61 | 159 | 37 | 7 | 1 | 0 | 0 |
| | 5 | 92 | 50 | 6 | 0 | 0 | 0 | 0 |
| | pH | 5 | 6 | 6.5 | 7 | 7.5 | 8 | 8.5 |
| | Predicate Device | | | | | | | |
For urinary protein, the company demonstrated the following correlation between the proposed and predicate device using patient samples in a clinical lab setting:
| Proposed Device (mg/dL) | 4+ | 0 | 0 | 0 | 0 | 2 | 18 |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | 3+ | 0 | 0 | 0 | 3 | 35 | 2 |
| | 2+ | 0 | 0 | 1 | 55 | 3 | 0 |
| | 1+ | 0 | 3 | 54 | 2 | 0 | 0 |
| | +/- | 3 | 53 | 4 | 0 | 0 | 0 |
| | Negative | 57 | 4 | 1 | 0 | 0 | 0 |
| | Protein | Negative | +/- | 1+ | 2+ | 3+ | 4+ |
| | Predicate device (mg/dL) | | | | | | |
For urinary nitrite, the company demonstrated the following correlation between the proposed and predicate device using patient samples in a clinical lab setting:
| Proposed Device | Positive | 6 | 4 | 108 |
| --- | --- | --- | --- | --- |
| | +/- | 10 | 43 | 8 |
| | Negative | 711 | 10 | 9 |
| | Nitrite | Negative | +/- | Positive |
| | Predicate device | | | |
For urinary leukocytes, the company demonstrated the following correlation between the proposed and predicate device using patient samples in a clinical lab setting:
{20}
| Proposed Device (Leu/μL) | 500 | 1 | 0 | 2 | 10 | 100 |
| --- | --- | --- | --- | --- | --- | --- |
| | 75 | 9 | 6 | 95 | 79 | 20 |
| | 25 | 3 | 66 | 21 | 24 | 5 |
| | Negative | 378 | 79 | 10 | 0 | 0 |
| | Leukocytes | Negative | 15 | 70 | 125 | 500 |
| | Predicate device (Leu/μL) | | | | | |
# 4. Clinical cut-off:
Not applicable for a device of this type.
# 5. Expected values/Reference range:
The company indicated the following published reference ranges for their analytes:
Urobilinogen: 0.1 to 1.0 Ehrlich unit /dL
Glucose: $< 100\mathrm{mg / dL}$
Bilirubin: $<$ Trace
Ketones: $0\mathrm{mg / dL}$ in urine with this test
pH: 5 - 9.
Blood: $< 3$ RBC/μL
Specific Gravity (SG): 1.001 to 1.035.
Protein: $< 20\mathrm{mg / dL}$
Nitrite: $0\mathrm{mg / dL}$
Leukocyte: Negative
Ascorbic acid: urinary output of $20 - 30\mathrm{mg / day}$
# N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
# O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.