The Site~Rite® 5 Ultrasound System with associated probes and accessories provide ultrasound guidance for placement of needles and catheters in vascular structures. Ultrasound guidance may occur intraoperatively or perculaneously. Ultrasound imaging of vascular structures, various organs and structures of the body may also be performed. The Site~Rite® 5 Ultrasound System is intended to be used for diagnostic ultrasound imaging of fluid flow analysis of the human body in the following clinical applications: Abdominal, Intraoperative (epiaortic scanning), Small Organ (breast, testes, thyroid, etc.), Neonatal Cephalic, Adult Cephalic, Cardiac, Peripheral Vascular, Musculo-skeletal Conventional.
Device Story
Portable, real-time B-mode ultrasonic pulsed echo imaging system; assists physicians in vascular access (needle/catheter placement) and diagnostic imaging. Inputs: ultrasonic signals via linear array probe. Processing: real-time B-mode imaging; image depth-dependent operating parameters. Outputs: visual ultrasound images on large screen; freeze frame; image saving/printing. Used in bedside, ICU, ER, angiography suites, and cath labs by physicians/clinicians. Facilitates needle/catheter guidance; improves procedural accuracy; enhances patient safety during vascular access.
Clinical Evidence
Bench testing only. Performance data evaluated against IEC 60601-1, IEC 60601-1-1, IEC 60601-1-2, IEC 60601-1-4, IEC 60601-2-37, UL 60601-1, NEMA UD-2, and NEMA UD-3 standards to demonstrate substantial equivalence.
Technological Characteristics
Portable B-mode ultrasound scanner; linear array probe; max imaging depth 6.0 cm. Battery or line voltage power. Standards: IEC 60601-1, IEC 60601-1-1, IEC 60601-1-2, IEC 60601-1-4, IEC 60601-2-37, UL 60601-1, NEMA UD-2, NEMA UD-3. Features: image freeze, save, print. Updated electronics design compared to predicate.
Indications for Use
Indicated for ultrasound guidance for needle and catheter placement in vascular structures (intraoperative or percutaneous) and diagnostic ultrasound imaging of fluid flow in abdominal, intraoperative (epiaortic), small organ, neonatal/adult cephalic, cardiac, peripheral vascular, and musculoskeletal structures. Not indicated for ophthalmic applications.
Regulatory Classification
Identification
A diagnostic ultrasonic transducer is a device made of a piezoelectric material that converts electrical signals into acoustic signals and acoustic signals into electrical signals and intended for use in diagnostic ultrasonic medical devices. Accessories of this generic type of device may include transmission media for acoustically coupling the transducer to the body surface, such as acoustic gel, paste, or a flexible fluid container.
{0}------------------------------------------------
KO52517
Site~Rite® 5 Ultrasound System Traditional 510(k)
______________________________________________________________________________________________________________________________________________________________________________
# SEP 2 9 2005
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
.
.
Section 5: 510(k) Summary of Safety and Effectiveness (21 CFR 807.92(a))
{1}------------------------------------------------
#### 5.2 Device Name
| Device Name: | Site-Rite® 5 Ultrasound System |
|-----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade Name: | Site-Rite® 5 Ultrasound System |
| Common/Usual Name: | Ultrasonic Pulsed Echo Imaging System and Accessories<br>Diagnostic Ultrasonic Transducers |
| Classification Name: | 90-IYO - Ultrasonic Pulsed Echo Imaging System<br>21 CFR 892.1560 – Class II<br>Diagnostic Device<br>90-ITX - Diagnostic Ultrasonic Transducer<br>21 CFR 892.1570 - Class II<br>Diagnostic Device |
| Classification Panel: | Radiology Devices Panel |
#### ડ 3 Address and Registration
The addresses and registration numbers of the manufacturer for the Site® 5 Ultrasound System are:
### Manufacturer:
5.4
| Submitter Name: | Bard Access Systems, Inc. (BAS)<br>[Wholly owned Subsidiary of C. R. Bard, Inc.] |
|----------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Address: | 5425 West Amelia Earhart Drive<br>Salt Lake City, Utah 84116 |
| Telephone Number: | (801) 595-0700 ext. 5421 |
| Fax Number: | (801) 595-5425 |
| Contact Person: | Kimberly Geisler |
| Date of Preparation: | August 9, 2005 |
| Registration Number: | 1720496 |
| Additional Registration Numbers: | C. R. Bard<br>2212754 |
| Manufacturing Site: | |
| Name: | Dymax Corporation.<br>[Wholly owned Subsidiary of C.R. Bard, Inc.] |
| Address: | 271 Kappa Drive<br>Pittsburgh, Pennsylvania 15238 |
| Registration Number: | 2523003 |
| Device Classification | |
| Classification Name: | 90-IYO - Ultrasonic Pulsed Echo Imaging System<br>21 CFR 892.1560 -- Class II<br>Diagnostic Device<br><br>90-ITX - Diagnostic Ultrasonic Transducer<br>21 CFR 892.1570 - Class II<br>Diagnostic Device |
{2}------------------------------------------------
#### રે રે Predicate Device Information
The predicate devices are:
| Device Name: | Site~Rite® IV Ultrasound System |
|-------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade Name: | Site~Rite® IV Ultrasound System |
| Common/Usual Name: | Ultrasonic Pulsed Echo Imaging System and Accessories<br>Diagnostic Ultrasonic Transducers |
| Classification Name: | 90-IYO -- Ultrasonic Pulsed Echo Imaging System<br>21 CFR 892.1560 - Class II<br>Diagnostic Device<br><br>90-ITX - Diagnostic Ultrasonic Transducer<br>21 CFR 892.1570 - Class II<br>Diagnostic Device |
| Classification Panel: | Radiology Devices Panel |
| Premarket Notification: | K032135, concurrence date July 21, 2003 |
#### 5.6 Device Description
The Site-Rite® 5 Ultrasound System is a lightweight, low-output, real-time B-mode ultrasonic pulsed echo imaging system designed primarily to assist physicians in gaining vascular access to major veins and arteries. It offers high resolution imaging to a depth of 6.0 cm. The Site® 5 Ultrasound System is portable and therefore easy to use at the bedside and in a variety of clinical scenarios, including intensive care units, cmergency rooms, angiography suites, catheterization laboratories, etc. In addition, the Site® 5 Ultrasound System is designed with simple operating controls to facilitate easy operation.
Key features of the Site~Rite® 5 Ultrasound System are:
- . Compact, portable, real-time ultrasound scanner with:
- Intuitive controls allowing for rapid and easy operation ■
- . Choice of battery or line voltage power
- 발 Operating parameters of scanner determined by image depth
- . Image freeze frame
- 트 Image printing and saving
- r Large screen
#### 5.7 Intended Use
The Site~Rite® 5 Ultrasound System with associated probes and accessories provide ultrasound guidance for placement of needles and catheters in vascular structures. Ultrasound guidance may occur intraoperatively or perculaneously. Ultrasound imaging of vascular structures, various organs and structures of the body may also be performed.
The Site~Rite® 5 Ultrasound System is intended to be used for diagnostic ultrasound imaging of fluid flow analysis of the human body in the following clinical applications.
- י Abdominal
- Intraoperative (epiaortic scanning)
- " Small Organ (breast, testes, thyroid, etc.)
- Neonatal Cephalic
{3}------------------------------------------------
- 포 Adult Cephalic
- 트 Cardiac
- 트 Peripheral Vascular
- 트 Musculo-skeletal Conventional
The Site~Rite® 5 Ultrasound System is not intended for opthalmic applications.
- 5.8 Substantial Equivalence Summary
#### New device is compared to marketed device?
Yes.
#### Does the new device have the same indication statement?
No. The indications for use are the same as the predicate Site~Rite® IV Ultrasound System, except that the indication for intraoperative neurological imaging has been removed. In addition, the Indications for Use statement has been reorganized to emphasize the indication for vascular access guidance:
The Site~Rite® 5 Ultrasound System with associated probe and accessories provide ultrasound guidance for placement of needles and catheters in vascular structures. Ultrasound guidance may occur intraoperatively or percutaneously. Ultrasound imaging of vascular structures, various organs and structures of the body may also be performed.
### Do the differences alter the intended therapeutic/diagnostic/etc. effect (in deciding, impact on safety and effectiveness may be considered)?
No. The differences do not alter the intended use of the device.
# Does the new device have the same technological characteristics, e.g. design, material, etc?
Not in all regards. The Site® 5 Ultrasound System has an updated design compared to the predicate device. However, both the subject and predicate devices transmit and receive an ultrasound signal that is used to create an image to assist the radiologist or other clinician in placing a needle in the desired location.
### Could the new characteristics affect safety or effectiveness?
Yes. The updated design could affect the safety or effectiveness of the technological characteristics are similar to the predicate devices except that the electronics design has been updated from the Site~Rite® IV Ultrasound System. The Site~Rite® 5 Ultrasound System has been designed for manufacturability and improved reliability. The Site-Rite® 5 Ultrasound System was designed to utilize a single linear array probe with a maximum imaging depth of 6.0cm in place of mechanical sector fluid filled probes of the predicate device.
### Do the new characteristics raise new types of safety or effectiveness questions?
No. Safety and effectiveness questions are the same for the predicate and subject devices. Refer to the Risk Analysis Document in Attachment 7.
{4}------------------------------------------------
### Do accepted scientific methods exist for assessing effects of the new characteristics?
Yes. Testing based on the following standards will be completed in order to evaluate the device's performance. Testing will be completed prior to placing the Site~Rite® 5 Ultrasound System on the market.
- . IEC 60601-1:1988, Medical Electrical Equipment - Part 1: General Requirements for Safety
- 피 IEC 60601-1-1:2000, Medical Electrical Equipment -- Part 1-1: General Requirements for Safety - Collateral Standard: Safety Requirements for Medical Electrical Systems
- IEC 60601-1-2:2004, Medical Electrical Equipment Part 1-2: General Requirements for 내 Safety - Collateral Standard: Electromagnetic Compatibility - Requirements and Tests
- * IEC 60601-1-4:2000, Medical Electrical Equipment - Part 1-4: General Requirements for Safety - Collateral Standard: Programmable Electrical Medical Systems
- . IEC 60601-2-37:2005, Medical Electrical Equipment - Part 2-37: Particular Reguirements for the Safety of Ultrasonic Medical Diagnostic and Monitoring Equipment
- 포 UL 60601-1:2003, Medical Electrical Equipment, Part 1: General Requirements for Safety
- NEMA UD-2:2004, Acoustic Output Measurement Standard for Diagnostic Ultrasound . Equipment
- NEMA UD-3:2004, Standard for Real-Time Display of Thermal and Mechanical Acoustic Output Indices on Diagnostic Ultrasound Equipment
### Are performance data available to assess effects of new characteristics?
Performance testing will be completed prior to placing the Site® 5 Ultrasound System on the market. Performance data must meet predetermined acceptance criteria prior to placing the Site~Rite® 5 Ultrasound System on the market.
### Performance data demonstrate equivalence?
Performance data must demonstrate that the Site~Rite® 5 Ultrasound System is substantially equivalent to the Site~Rite® IV Ultrasound System prior to placement of the Site~Rite® 5 Ultrasound System on the market.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is a stylized image of a human figure with three lines extending from the head, resembling a bird in flight.
SEP 2 9 2005
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Bard Access Systems, Inc. Division of C.R. Bard, Inc. c/o Mr. Robert Mosenkis President CITECH 5200 Butler Pike PLYMOUTH MEETING PA 19462-1298
Re: K052517
Trade Name: Site~Rite 5° Ultrasound System Regulation Number: 21 CFR §892.1560 Regulation Name: Ultrasonic pulsed ccho imaging system Product Code: IYO Regulation Number: 21 CFR §892.1570 Regulation Name: Diagnostic ultrasonic transducer Product Code: ITX Regulatory Class: II Dated: September 13, 2005 Received: September 14, 2005
Dear Mr. Mosenkis:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use surfal the May 28, 1976, the enactment date of the Medical Device Amendments, or to eoninero prof to the ridg " way" and " with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls Cosmetions of the Act. The general controls provisions of the Act include requirements for annual provision, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
This determination of substantial equivalence applies to the following transducers intended for use with the Site~Rite 5° Ultrasound System as described in your premarket notification:
Transducer Model Number
Site~Rite 5
{6}------------------------------------------------
# Page 2 - Mr. Robert Mosenkis
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it n your de fior is olabilional controls. Existing major regulations affecting your device can be may of babyer to early in the 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that I Tease of action that your device complies with other requirements of the Act or any I Dri has made a deceminations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality laboring (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This determination of substantial equivalence is granted on the condition that prior to shipping the first device, you submit a postclearance special report. This report should contain complete information, including acoustic output measurements based on production line devices, requested in Appendix G, (enclosed) of the Center's September 30, 1997, "Information for Manufacturers Appondin of (encreove) version Ultrasound Systems and Transducers." If the special report is incomplete or contains unacceptable values (e.g., acoustic output greater than approved levels), then the 510(k) clearance may not apply to the production units which as a result may be considered adulterated or misbranded.
The special report should reference the manufacturer's 510(k) number. It should be clearly and prominently marked "ADD-TO-FILE" and should be submitted in duplicate to:
> Food and Drug Administration Center for Devices and Radiological Health Document Mail Center (HFZ-401) 9200 Corporate Boulevard Rockville, Maryland 20850
This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html
{7}------------------------------------------------
Page 3 – Mr. Robert Mosenkis
If you have any questions regarding the content of this letter, please contact Mr. Rodrigo C. Perez at (301) 594-1212.
Sincerely yours,
- Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosures
{8}------------------------------------------------
#### 1.2 Statement of Indications for Use
510(k) Number (if known):
K052 SI7
Device Name:
·
Site~Rite® 5 Ultrasound System
Indications for Use:
The Site® 5 Ultrasound System with associated probe and accessories provide ultrasound guidance for placement of needles and catheters in vascular structures. Ultrasound guidance may occur intraoperatively or percutaneously. Ultrasound imaging of vascular structures, various organs and structures of the body may also be performed
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________ (21 CFR 801 Subpart C)
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) _ .___
_..... ........
Concurrence of CDRH, Office of Device Evaluation (ODE)
Marney C brigdon
(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number K052517
.0005
. . . ...
{9}------------------------------------------------
#### Site~Rite® 5 Ultrasound System
#### Diagnostic Ultrasound Indications for Use Form
#### Fill out one form for each ultrasound system and each transducer.
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | |
|-------------------------------|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| Clinical Applications | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(Specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal | | | | | | | | | | |
| Abdominal | | P | | | | | | | | |
| Intraoperative (specify) | | P | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small Organ (specify) | | P | | | | | | | | |
| Neonatal Cephalic | | P | | | | | | | | |
| Adult Cephalic | | P | | | | | | | | |
| Cardiac | | P | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral Vascular | | P | | | | | | | | |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal Conventional | | P | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | | | |
| Other (specify | | | | | | | | | | |
N = new indication; P = previously cleared by FDA; E = added under Appendix E
Additional Comments:
Intraoperative (epiaortic scanning)
Small organ (breast, testes, thyroid, etc.)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) (Concurrence of CDRH, Office of Device Evaluation (ODE))
Prescription Use (Pcr 21 CFR 801.109)
Nancy C. Langdon
(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number K052517
000511
{10}------------------------------------------------
# Site~Rite®) 5 Ultrasound Transducer
Diagnostic Ultrasound Indications for Use Form
# Fill out one form for each ultrasound system and each transducer.
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | | Mode of Operation | | | | | | | | | |
|-------------------------------|---|-------------------|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|--|
| Clinical Applications | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(Specify) | |
| Ophthalmic | | | | | | | | | | | |
| Fetal | | | | | | | | | | | |
| Abdominal | | P | | | | | | | | | |
| Intraoperative (specify) | | P | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | | |
| Pediatric | | | | | | | | | | | |
| Small Organ (specify) | | P | | | | | | | | | |
| Neonatal Cephalic | | P | | | | | | | | | |
| Adult Cephalic | | P | | | | | | | | | |
| Cardiac | | P | | | | | | | | | |
| Transesophageal | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | |
| Transurethral | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | |
| Peripheral Vascular | | P | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | |
| Musculo-skeletal Conventional | | P | | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | | | | |
| Other (specify | | | | | | | | | | | |
N = new indication; P = previously cleared by FDA; E = added under Appendix E
______________________________________________________________________________________________________________________________________________________________________________
Additional Comments:
Intraoperative (epiaortic scanning) __________________________________________________________________________________________________________________________________________
Small organ (breast, testes, thyroid, etc.) . . ______________________________________________________________________________________________________________________________
(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED). __________________________________________________________________________________________________ (Concurrence of CDRH, Office of Device Evaluation (ODE))
· · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · ·
Jancyc brogdonPrescription Use (Per 21 CFR 801.109)
(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number K052517
000512
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.