← Product Code [JZO](/productcode/JZO) · K052407

# ARCHITECT ANTI-TPO IMMUNOASSAY, CALIBRATORS & CONTROLS (K052407)

_Fisher Diagnostics · JZO · Jul 18, 2006 · Immunology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K052407

## Device Facts

- **Applicant:** Fisher Diagnostics
- **Product Code:** [JZO](/productcode/JZO.md)
- **Decision Date:** Jul 18, 2006
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 866.5870
- **Device Class:** Class 2
- **Review Panel:** Immunology

## Indications for Use

Architect® Anti-TPO is a chemiluminescent microparticle Immunoassay (CMIA) for the quantitative determination of the IgG class of thyroid peroxidase autoantibodies (anti-TPO) in human serum and plasma (EDTA and Heparin) on the Architect® i System. The Architect® Anti-TPO assay is intended for use as an aid in the diagnosis of autoimmune thyroid disease. The Architect® Anti-TPO Calibrators are for the calibration of the Architect® i System when used for the quantitative determination of IgG class of thyroid peroxidase autoantibodies (anti-TPO) in human serum and plasma. The Architect® Anti-TPO controls are for the estimation of test precision and the detection systematic analytical deviations of the Architect® i System when used for the quantitative determination of IgG class of thyroid peroxidase autoantibodies (anti-TPO) in human serum and plasma.

## Device Story

Architect Anti-TPO is a quantitative chemiluminescent microparticle immunoassay (CMIA) for detecting IgG thyroid peroxidase autoantibodies in human serum/plasma. System uses TPO-coated paramagnetic particles and acridinium-labeled anti-human IgG conjugate. Operation: two-step assay on Architect i System (i2000/i2000SR); sample mixed with particles, incubated, washed, then conjugate added; chemiluminescent reaction measured as relative light units (RLU). RLU output directly proportional to anti-TPO concentration. Used in clinical laboratories; operated by trained technicians. Results aid clinicians in diagnosing autoimmune thyroid disease. System includes reagents, calibrators, controls, and assay files.

## Clinical Evidence

Clinical performance evaluated using 500 specimens: 236 healthy individuals, 125 Graves' disease, and 139 Hashimoto's thyroiditis patients. Comparison with AxSYM predicate showed 92.6% total agreement. Clinical sensitivity for Hashimoto's was 67.6% and for Graves' was 95.2%. Normal reference range established as < 5.61 IU/mL based on 97.5th percentile of healthy population. Analytical performance included precision (total CV ≤ 10% for samples ≥ 5.61 IU/mL), linearity (3.0–1000 IU/mL), and interference testing (no significant interference from bilirubin, hemoglobin, triglycerides, or total protein).

## Technological Characteristics

CMIA technology; TPO-coated paramagnetic microparticles; acridinium-labeled anti-human IgG conjugate; chemiluminescent detection system; automated platform (ARCHITECT i System).

## Regulatory Identification

A thyroid autoantibody immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the thyroid autoantibodies (antibodies produced against the body's own tissues). Measurement of thyroid autoantibodies may aid in the diagnosis of certain thyroid disorders, such as Hashimoto's disease (chronic lymphocytic thyroiditis), nontoxic goiter (enlargement of thyroid gland), Grave's disease (enlargement of the thyroid gland with protrusion of the eyeballs), and cancer of the thyroid.

## Predicate Devices

- Abbott's AxSYM anti-Thyroid Microparticle Enzyme Immunoassay (k020348)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

1

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY

A. 510(k) Number:
k052407

B. Purpose for Submission:
New Device

C. Measurand:
Thyroid peroxidase autoantibodies (Anti-TPO), IgG

D. Type of Test:
Quantitative Chemiluminescent Microparticle Immunoassay

E. Applicant:
Fisher Diagnostics

F. Proprietary and Established Names:
Architect® Anti-TPO Immunoassay Reagents
Architect® Anti-TPO Calibrators and controls

G. Regulatory Information:
1. Regulation section:
21 CFR 866.5870, Thyroid Autoantibody Immunological Test System
21 CFR 862.1660, Quality Control Material (Assayed and Unassayed)
21 CFR 862.1150, Calibrator
2. Classification:
Class II
3. Product code:
JZO - Thyroid Antibody Test System
JJX - Single (specified) Analyte Control (Assayed and Unassayed)
JIT - Calibrator, Secondary
4. Panel:
Immunology 82
Clinical chemistry 75

H. Intended Use:
1. Intended use(s):
Architect® Anti-TPO is a chemiluminescent microparticle Immunoassay (CMIA) for the quantitative determination of the IgG class of thyroid peroxidase autoantibodies (anti-TPO) in human serum and plasma (EDTA and Heparin) on the Architect® i System. The Architect® Anti-TPO assay is intended for use as an aid in the diagnosis of autoimmune thyroid disease.

The Architect® Anti-TPO Calibrators are for the calibration of the Architect® i System when used for the quantitative determination of IgG class of thyroid peroxidase autoantibodies (anti-TPO) in human serum and plasma. The Architect® Anti-TPO controls are for the estimation of test precision and the detection systematic analytical deviations of the Architect® i System when used for the quantitative determination of IgG class of thyroid peroxidase autoantibodies (anti-TPO) in human serum and plasma.

2. Indication(s) for use:

{1}

Same as intended use.

3. Special conditions for use statement(s):

For prescription use only.

4. Special instrument requirements:

Architect i System (Architect® i2000 and Architect i 200SR).

# I. Device Description:

The device consists of (a) TPO coated paramagnetic particles, (b) sample diluent MES buffer with protein-goat, (c) Anti-human IgG (mouse monoclonal) acridinium labeled conjugate, (d) wash buffer, (e) Architect i pre-trigger and trigger solutions, (f) calibrators, (g) controls and (h) a CD-ROM containing the assay files to be installed in the Architect i System.

# J. Substantial Equivalence Information:

1. Predicate device name(s):

Abbott's AxSYM anti-Thyroid Microparticle Enzyme Immunoassay

2. Predicate  $510(\mathrm{k})$  number(s):

k020348

3. Comparison with predicate:

|  Similarities  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|  Intended use | aid in the diagnosis of autoimmune thyroid disease | Same  |
|  Assay format | Quantitative | Same  |
|  Number of Calibrators | 6 levels (A-F) | same  |
|  Calibrator Range concentrations | 0.0-1000 IU/mL | same  |
|  Standardization material | NIBSC MRC 66/387 Anti-Thyroid microsome serum standard | Same  |
|  Control component | Human plasma | same  |
|  Differences  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|  Thyroid peroxidase - microparticle coating | Recombinant | From human thyroid glands  |
|  Conjugate | Mouse anti-human IgG (monoclonal) acridinium labeled in MES buffer | Goat anti-human IgG alkaline phosphate-in TRIS buffer  |
|  Control concentrations and range | Negative 0.0IU/mL Range ≤ 2.15 IU/mL Positive 75 IU/mL Range 45-105 IU/mL | Negative 0.0 IU/mL Range ≤ 12.0 IU/mL Positive 75 IU/mL Range 40-110 IU/mL  |
|  Assay diluent | MES buffer | TRIS buffer  |
|  Assay method | Chemiluminescent microparticle (CMIA) | Microparticle enzyme (MEIA)  |
|  Analyzer | Architect i System | AxSYM System  |

{2}

# K. Standard/Guidance Document Referenced (if applicable):

CLSI (NCCLS) EP5-A, EP6-A, EP7-A, EP17-A.

# L. Test Principle:

Chemiluminescent microparticle immunoassay (CMIA) referred as Chemiflex. It is a two step assay. In the first step sample and assay diluent (MES buffer) and TPO coated paramagnetic particles are mixed and incubated. Anti-TPO in the sample binds to the microparticles. After a wash step anti-human acridinium labeled conjugate is added in the second step. Following an incubation and wash cycle pre-trigger and trigger solutions are added to the reaction mixture. The resulting chemiluminescent reaction is measured as relative light units(RLU). A direct relationship exists between the amount of anti-TPO in the sample and the RLUs detected by the Architect  $i$  system optics.

# M. Performance Characteristics (if/when applicable):

# 1. Analytical performance:

# a. Precision/Reproducibility:

Precision studies were performed using Positive Control and 3 human panels (serum pools with low, medium and high concentrations), 3 lots of reagents in replicates of two at 2 separate times per day for 20 days on 3 instruments (according to CLSI EP5-A).

ARCHITECT Anti-TPO assay is designed to have an assay precision of  $\leq 10\%$  total CV for samples  $\geq 5.61~\mathrm{IU / mL}$ . Each reagent lot used a single calibration curve throughout the study. Data from this study are summarized in the following table.*

|  Sample | Instrument | Reagent Lot | n | Mean Conc. (IU/mL) | Within Run |   | Total  |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
|   |   |   |   |   |  SD | % CV | SD | % CV  |
|  Positive Control | 1 | A | 80 | 74.85 | 2.11 | 2.8 | 2.16 | 2.9  |
|   |   |  B | 80 | 74.20 | 1.89 | 2.6 | 2.04 | 2.7  |
|   |   |  C | 80 | 74.63 | 2.01 | 2.7 | 2.17 | 2.9  |
|   |  2 | A | 80 | 77.42 | 2.11 | 2.7 | 3.25 | 4.2  |
|   |   |  B | 80 | 75.32 | 1.92 | 2.5 | 2.54 | 3.4  |
|   |   |  C | 80 | 74.59 | 1.73 | 2.3 | 2.57 | 3.4  |
|   |  3 | A | 80 | 75.41 | 2.13 | 2.8 | 2.52 | 3.3  |
|   |   |  B | 80 | 75.48 | 1.90 | 2.5 | 2.13 | 2.8  |
|   |   |  C | 80 | 76.66 | 2.42 | 3.2 | 2.87 | 3.7  |
|  Panel 1 | 1 | A | 80 | 1.57 | 0.08 | 4.8 | 0.10 | 6.5  |
|   |   |  B | 80 | 1.46 | 0.06 | 3.8 | 0.09 | 5.8  |
|   |   |  C | 80 | 1.64 | 0.09 | 5.6 | 0.10 | 6.1  |
|   |  2 | A | 80 | 1.60 | 0.09 | 5.3 | 0.12 | 7.6  |
|   |   |  B | 80 | 1.53 | 0.06 | 3.9 | 0.11 | 7.2  |
|   |   |  C | 80 | 1.52 | 0.10 | 6.7 | 0.12 | 7.7  |
|   |  3 | A | 80 | 1.47 | 0.08 | 5.3 | 0.11 | 7.8  |
|   |   |  B | 80 | 1.47 | 0.07 | 4.7 | 0.13 | 8.5  |
|   |   |  C | 80 | 1.52 | 0.14 | 9.5 | 0.15 | 9.8  |

{3}

|  Sample | Instrument | Reagent Lot | n | Mean Conc. (IU/mL) | Within Run |   | Total  |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
|   |   |   |   |   |  SD | % CV | SD | % CV  |
|  Panel 2 | 1 | A | 80 | 20.98 | 0.65 | 3.1 | 0.76 | 3.6  |
|   |   |  B | 80 | 21.14 | 0.61 | 2.9 | 0.66 | 3.1  |
|   |   |  C | 80 | 21.51 | 0.71 | 3.3 | 0.75 | 3.5  |
|   |  2 | A | 80 | 21.27 | 0.61 | 2.9 | 0.98 | 4.6  |
|   |   |  B | 80 | 21.62 | 0.66 | 3.0 | 0.90 | 4.2  |
|   |   |  C | 80 | 20.82 | 0.67 | 3.2 | 0.85 | 4.1  |
|   |  3 | A | 80 | 21.00 | 0.73 | 3.5 | 0.86 | 4.1  |
|   |   |  B | 80 | 21.77 | 0.60 | 2.7 | 0.84 | 3.8  |
|   |   |  C | 80 | 21.24 | 0.70 | 3.3 | 0.89 | 4.2  |
|  Panel 3 | 1 | A | 80 | 214.78 | 5.14 | 2.4 | 6.48 | 3.0  |
|   |   |  B | 80 | 221.79 | 4.73 | 2.1 | 5.82 | 2.6  |
|   |   |  C | 80 | 216.71 | 5.36 | 2.5 | 6.36 | 2.9  |
|   |  2 | A | 80 | 219.32 | 4.41 | 2.0 | 8.61 | 3.9  |
|   |   |  B | 80 | 224.54 | 4.04 | 1.8 | 13.37 | 6.0  |
|   |   |  C | 80 | 218.73 | 5.76 | 2.6 | 13.18 | 6.0  |
|   |  3 | A | 80 | 212.91 | 6.11 | 2.9 | 6.84 | 3.2  |
|   |   |  B | 80 | 225.46 | 5.15 | 2.3 | 5.67 | 2.5  |
|   |   |  C | 80 | 228.17 | 5.80 | 2.5 | 7.09 | 3.1  |

# b. Linearity/assay reportable range:

The assay was shown to be linear from 3.0 to  $1000.0\mathrm{IU / mL}$  by the following study: three high sample pools (1000, 300, and  $30~\mathrm{IU / mL}$ ) were combined with a low pool calibrator A to provide 9 test points. All test points were analyzed in replicates of 5 using a single reagent lot. Automated dilution was compared to a manual 1:2 dilution using 9 human specimens with anti-TPO levels that were greater than calibrator E  $(250~\mathrm{IU / mL})$ . The manual dilution was done with Anti-TPO calibrator A. Summary of results is shown in the following table.

|  Sample ID | Automated Dilution (IU/mL) | Manual Dilution (IU/mL) | % Recovery  |
| --- | --- | --- | --- |
|  1 | 861.25 | 859.27 | 100.2  |
|  2 | 684.49 | 703.64 | 97.3  |
|  3 | 844.36 | 847.62 | 99.6  |
|  4 | 724.55 | 757.09 | 95.7  |
|  5 | 709.46 | 688.49 | 103.1  |
|  6 | 1105.65 | 1106.18 | 100.0  |
|  7 | 948.43 | 931.80 | 101.8  |
|  8 | 840.77 | 851.72 | 98.7  |
|  9 | 966.48 | 998.44 | 96.8  |

High dose hook - Ten samples with anti-TPO concentrations of 3458 IU/mL to 17815 IU/mL were analyzed and results were reported as  $&gt;1000$  IU/mL. No high dose

{4}

hook effect was observed.

c. Traceability, Stability, Expected values (controls, calibrators, or methods): The anti-TPO reference calibrator is traceable to the NIBSC 66/387 reference preparation. The anti-TPO reference calibrators are produced from the NIBSC 66/387 reference preparation with individual levels at 30.0, 15.0, 5.0, 1.5, 0.75, and 0.0 IU/mL. The proper weight fractions of the stock concentrate and calibrator diluent were mixed to form the individual levels anti-TPO master calibrator at 0.00, 5.0, 20.0, 62.5, 250.0, and 1000.0 IU/mL. The calibrators and controls are manufactured from human anti-TPO positive plasma pools.

Stability – Stability was established based on accelerated and real time studies on three lots of Calibrators at various temperatures (5, 25, 30, 35, -20 and -50°C). Stability test results and graphs are in the submission. The results indicate storage stability for 12 months when stored at -10°C. On-board stability is 30 days. Stability of reagents stored at 2-8° is shown as 9 months.

d. Limit of Blank: Limit of Blank was tested using 10 replicates of calibrator A and 4 replicates of calibrator B per run. The acceptance criterion was ≤ 1.0 IU/mL. 36 total runs were made across 3 instruments (i 2000) using 3 different reagent lots. The Limit of Blank was calculated as mean of all calculations plus 2 SDs for type 1 error of 2.5%. The Limit of Blank is 0.159 IU/mL for all 3 lots.

e. Analytical specificity: Interference from elevated levels of bilirubin (20mg/dl), hemoglobin (1000 mg/dl) triglycerides (1000mg/dl) and total protein 10g/dl) were tested. A study was performed based on CLSI EP7-A. Specimens with anti-TPO levels between 45.07 and 361.64 IU/mL were used. Specimens were spiked between 133.44 and 568.78 IU/mL. The average interference observed ranged from -3.6% to +3.7% (within ≤15% acceptable range).

f. Assay cut-off: Values greater than 5.6 IU/mL are considered positive.

2. Comparison studies:

a. Method comparison with predicate device: The performance of the Architect Anti-TPO was compared to AxSYM Anti-TPO assay. A total of 500 specimens (236 samples from healthy individuals and 264 samples from patients with autoimmune thyroid disease- 125 Graves' disease and 139 Hashimoto's Disease) were evaluated. Specimens were tested in replicates of one using the Architect Anti-TPO assay with three reagent lots. 446 samples were &lt; 1000 IU/mL. The remaining 54 samples were not included in the correlation since their anti-TPO concentrations were greater than the assay range (&gt;1000 IU/mL). Data from this study are summarized in the following table.

|  ARCHITECT Anti-TPO | AxSYM Anti-TPO  |   |   |
| --- | --- | --- | --- |
|   |  Negative | Positive | Total  |
|  Negative | 242 | 32 | 274  |
|  Positive | 5 | 221 | 226  |
|  Total | 247 | 253 | 500  |

{5}

Negative Agreement  $= 98.0\%$  (242/247) with  $95\%$  CI:  $95.3\%$  to  $99.3\%$

Positive Agreement  $= 87.4\%$  (221/253) with  $95\%$  CI:  $82.6\%$  to  $91.2\%$

Total Agreement  $= 92.6\%$

Sample Range (ARCHITECT) = 0.0 to 27430.8 IU/mL

Sample Range (AxSYM Anti-TPO) = 1.5 to 50390 IU/mL

|  Linear Regression Analysis  |   |   |   |   |
| --- | --- | --- | --- | --- |
|  Specimen Type | n | Slope (95% CI) | Intercept (95% CI) | Correlation Coefficient  |
|  Autoimmune Thyroid Disease | 211 | 1.198 (1.138 to 1.259) | 15.34 (-3.25 to 33.93) | 0.937  |
|  Apparently Healthy Individuals | 235 | 0.924 (0.909 to 0.938) | -3.34 (-4.35 to -2.34) | 0.993  |

# b. Matrix comparison:

Twenty donors were evaluated in each of the nine tubes – serum separator tube plus vs. serum, sodium heparin glass vs. serum, sodium heparin plastic vs. serum, lithium heparin glass vs. serum, lithium plastic vs. serum, plasma separator lithium heparin glass vs. serum, plasma separator lithium heparin plastic vs. serum, EDTA glass vs. serum and EDTA plastic vs. serum. For each donor, samples from each tube type were assayed in replicates of five. Samples from two donors with anti-TPO values greater than the assay range were excluded from the analysis. The mean percentage interference ranged from 1.16 to  $1.8\%$ . Linear regression analysis was also performed and results are summarized below.

Serum separator tube plus vs. serum,  $y = 1.0178x - 2.7939$ ,  $R^2 = 0.9995$

Sodium Heparin glass vs. serum,  $y = 1.0036x + 1.1448$ ,  $R^2 = 0.9995$

Sodium Heparin plastic vs. serum,  $y = 1.012x - 1.3539$ ,  $R^2 = 0.9995$

Lithium Heparin glass vs. serum,  $y = 1.0097x - 0.8132$ ,  $R^2 = 0.9996$

Lithium Heparin plastic vs. serum,  $y = 1.0163x - 2.0565$ ,  $R^2 = 0.9995$

Plastic separator Lithium Heparin glass vs. serum,  $y = 1.007x - 0.964$ ,  $R^2 = 0.9996$

Plastic separator Lithium Heparin plastic vs. serum,  $y = 1.0104x - 1.994$ ,  $R^2 = 0.9996$

EDTA glass vs. serum,  $y = 0.997x - 1.3241$ ,  $R^2 = 0.9998$

EDTA plastic vs. serum,  $y = 1.0088x - 1.2826$ ,  $R^2 = 0.9993$

# 3. Clinical studies:

# a. Clinical Sensitivity:

Clinical sensitivity was evaluated by testing 139 clinically defined Hashimoto's

Thyroiditis and 125 Grave's Disease specimens. Anti-TPO levels  $\geq 5.6$  IU/mL is considered positive for Architect assay and  $\geq 12$  IU/mL for comparison assay.

The samples used in the method comparison study were from multiple collection sites

{6}

and tested at the sponsor's facility at two different timings noted as study 1 and study 2. The results are summarized below:

|  ARCHITECT Anti-TPO  |   |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- | --- |
|  Hashimoto’s Thyroiditis |   |   |   | Graves’ Disease  |   |   |
|  Study # | n | No. of Positive | % Positive (95% CI) | n | No. of Positive | % Positive (95% CI)  |
|  1 | 89 | 57 | 64.0 (53.2 to 73.9) | 75 | 69 | 92.0 (83.4 to 97.0)  |
|  2 | 50 | 37 | 74.0 (59.7 to 85.4) | 50 | 50 | 100.0 (92.9 to 100.0)  |
|  Total | 139 | 94 | 67.6 (59.2 to 75.3) | 125 | 119 | 95.2 (89.8 to 98.2)  |
|  Comparison Assay-AxSYM Anti-TPO  |   |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- | --- |
|  1 | 89 | 76 | 85.4 (76.3 to 92.0) | 75 | 71 | 94.7 (86.9 to 98.5)  |
|  2 | 50 | 47 | 94.0 (83.5 to 98.7) | 50 | 50 | 100.0 (92.9 to 100.0)  |
|  Total | 139 | 123 | 88.5 (82.0 to 93.3) | 125 | 121 | 96.8 (92.0 to 99.1)  |

There was a decrease in clinical sensitivity for Hashimoto’s Thyroiditis of 20.9%.

b. Clinical specificity:
Besides Grave’s and Hashimoto’s disease no other autoimmune diseases were tested.

c. Other clinical supportive data (when a and b are not applicable):
None

4. Clinical cut-off:
5.61 IU/mL

5. Expected values/Reference range:
Human serum specimens were collected from a population of 236 apparently healthy individuals. All specimens had TSH values within the normal reference range. Of this study population, 9 specimens delivered positive results on a commercially available anti-TPO assay device and were excluded from further normal range analysis. The 97.5 percentile concentration of the remaining population was 5.61 IU/mL. In this study population, the normal range is &lt; 5.61 IU/mL. A total of 97.8% (222/227) of the population had values within this normal range.

N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10

O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

---

**Source:** [https://fda-staging.innolitics.com/device/K052407](https://fda-staging.innolitics.com/device/K052407)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
