The Mesa Spinal System is a non-cervical spinal fixation device intended for posterior, non-pedicle fixation for the following indications: degenerative disc disease ( DDD ) (defined as back pain of discogenic origin with degencration of the disc confirmed by history and radiographic studies ); spondylolisthesis; trauma ( i.e. fracture or dislocation ); spinal stenosis; curvatures ( i.e. scoliosis, kyphosis; and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion. The Mesa Spinal System is also intended for non-cervical pedicle screw fixation for the following indications: trauma ( i.e. fracture or dislocation ); spinal stenosis; curvatures ( i.e. scoliosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion. It is also indicated for the treatment of severe spondylolisthesis ( grades 3 and 4 ) of the L5-S1 vertebra in skelctally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine ( L3 to sacrum ) with removal of the attainment of a solid fusion.
Device Story
Mesa Spinal System is top-loading, posterior spinal fixation system; components include pedicle screws, rods, hooks, transverse connectors. Used as adjunct to fusion to immobilize thoracic, lumbar, sacral spinal segments. Surgeon implants hardware to stabilize spine; provides structural support during bone healing. System components available in various sizes to match patient anatomy. Device used in surgical setting by orthopedic or neurosurgeons. Clinical benefit includes stabilization of spinal segments to facilitate fusion in patients with degenerative or traumatic spinal conditions.
Clinical Evidence
Bench testing only. Biomechanical testing performed per ASTM F1717 to compare performance against predicate systems.
Technological Characteristics
Top-loading posterior spinal fixation system. Materials: Ti6Al-4V ELI alloy per ASTM and ISO standards. Components: pedicle screws, rods, hooks, transverse connectors. Non-sterile (implied).
Indications for Use
Indicated for non-cervical spinal fixation in patients with degenerative disc disease, spondylolisthesis, trauma (fracture/dislocation), spinal stenosis, scoliosis, kyphosis, lordosis, tumor, pseudoarthrosis, or failed fusion. Includes pedicle screw fixation for severe L5-S1 spondylolisthesis (grades 3-4) in skeletally mature patients.
Regulatory Classification
Identification
(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.
Special Controls
*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
“Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.
Predicate Devices
Denali Spine System (K2M, LLC)
Versalok Low Back Fixation System (Wright Medical)
Ovation Polyaxial System (Osteotech)
Submission Summary (Full Text)
{0}------------------------------------------------
DEC 7 2005
K052398
Page lot 1
510(k) Summary
## for the
### Mesa Spinal System
s safety and effectiveness summary for the Mesa Spinal System is provided as required per Section 513(i)(3) of the Food, Drug and Cosmetic Act.
#### 1. Submitter :
K2M, LLC 751 Miller Drive SE, Suite Fi Leesburg, VA 20175
August 29, 2005 Date Prepared:
Contact Person : Richard W. Woods K2M, LLC 751 Miller Drive SE, Suite F1 Leesburg, VA 20175 Telephone: 703-777-3155
| 2. Tradename: | Mesa Spinal System |
|----------------------|----------------------------------------------------------------------------------------------------------------|
| Common Name: | Spine Fixation System |
| Classification Name: | Pedicle Screw Spinal System (21 CFR 888.3070(b)(1))<br>Spinal Interlaminal fixation orthosis (21 CFR 888.3050) |
## 3. Predicate or legally marketed devices which are substantially equivalent :
- Denali Spine System ( K2M, LLC ) .
- Versalok Low Back Fixation System ( Wright Medica! ) .
- Ovation Polyaxial System ( Osteotech ) .
#### 4. Description of the device :
The Mesa Spine System is a top-loading, multiple component, posterior spinal fixation system which consists of pedicle screws, rods, hooks, and transverse connectors. All of the components are available in a variety of sizes to match more closely the patient's anatomy.
Materials: The devices are manufactured from Ti6Al-4V ELI alloy per ASTM and ISO standards.
Function: The system functions as an adjunct to fusion to provide immobilization of spinal segments of the thoracic, lumbar and / or sacral spine.
#### 5. Intended Use:
The Mesa Spinal System is a non-cervical spinal fixation device intended for posterior, non-pedicle fixation for the following indications: degenerative disc disease ( DDD ) (defined as back pain of discogenic origin with degencration of the disc confirmed by history and radiographic studies ); spondylolisthesis; trauma ( i.e. fracture or dislocation ); spinal stenosis; curvatures ( i.e. scoliosis, kyphosis; and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion.
The Mesa Spinal System is also intended for non-cervical pedicle screw fixation for the following indications: trauma ( i.e. fracture or dislocation ); spinal stenosis; curvatures ( i.e. scoliosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion. It is also indicated for the treatment of severe spondylolisthesis ( grades 3 and 4 ) of the L5-S1 vertebra in skelctally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine ( L3 to sacrum ) with removal of the attainment of a solid fusion.
## 6. Comparison of the technological characteristics of the device to predicate and legally marketed devices :
The Mesa Spine System was biomechanically tested and compared to the predicate systems and other currently marketed systems and performed equal to or better than these systems in ASTM testing to F1717. The design features and sizing of the components were also compared and the Mesa System found to be substantially the same as these systems. It is manufactured from the same FDA recognized materials and is indicated for the same intended uses as these systems. There are no significant differences between the Mesa Spinal Systems currently being marketed which would adversely affect the use of the product. It is substantially equivalent to these other devices in design, function, material and intended use.
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized caduceus symbol, which features a staff with a snake winding around it, and three horizontal bars above it. The text "DEPARTMENT OF HEALTH AND HUMAN SERVICES . USA" is arranged in a circular pattern around the symbol.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
DEC 7 2005
Mr. Richard W. Woods Senior Vice President K2M, LLC 751 Miller Drive, S.E., Suite Fl Leesburg, Virginia 20175
Re: K052398
Trade/Device Name: Mesa Spinal System TraderDon Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: II Product Code: MNI, MNH, KWP Dated: November 16, 2005 Received: November 17, 2005
Dear Mr. Woods:
We have reviewed your Section 510(k) premarket notification of intent to market the device We have reviewed your Section > IV(x) premaince is substantialy equivalent (for the indications
referenced above and have determined the device is substanted in interstate referenced above and have determined the actives marketed in interstate
for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the enclosure) to legally manect prodical Device Amendments, or to
commerce prior to May 28, 1976, the enactment date of the Federal Food. Drug, commerce prior to May 28, 1976, the enactinent use of the Federal Food, Drug, devices that have been reclassified in accordance with the proval application (PMA).
and Cosmetic Act (Act) that do not require approval of a provisions of the Act. The and Cosmetic Act (Act) that do not require approval or a policions of the Act. The
You may, therefore, market the device, subject to the general contracts of listing of You may, therefore, market the device, subject to the general vonmul registration, listing of
general controls provisions of the Act include requirements for annual mishrandi general controls provisions of the Act mende requirements to the more of the management misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is classified (see above) into entitle tass in topedia ions affecting your device can
may be subject to such additional controls. Existing major regulations a may be subject to such additional controlist Listing introl organto 898. In addition, FDA may
be found in the Code of Federal Regulations, Title 21, Parts 800 to 898 may be found in the Code of Federal Regulations, Fride 24, 11, 2017, 11:15 Pederal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean Please be advised that FDA's issuance of a substance of the requirements of the Act
that FDA has made a determination that your device complies with other requirements You that FDA has made a determination hia your devres on per Federal agencies. You must
or any Federal statutes and regulations administed to: registration and listi or any Federal statutes and regulations and limited to registration and listing (21
comply with all the Act's requirements, including, but not mements as sct comply with all the Act s requrements, mendenting practice requirements as sch
CFR Part 807); labeling (21 CFR Part 801); good manafacturing practice requirements as sc CFR Part 807); labeling (21 CFR Part of ); good manakeraning philifapplicable, the electronic
forth in the quality systems (QS) regulation (21 CFR Part 820); and 1f applicabl forth in the quality systems (QS) regulation (Sections 531-542 of the Act); 21 CFR 1000-1050.
product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000
{2}------------------------------------------------
Page 2 -- Mr. Richard W. Woods
This letter will allow you to begin marketing your device as described in your Section 510(k)
The same of the first of a first of a location to be mixelers of your device to This letter will anow you to begin marksmig your antial equivalence of your device to a legally
premarket notification. The FDA finding of sybstantial equivales and thus, pe premarket notification. The FDA multing of substantal equive and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific advice for your ac on on on on on on on on the regulation entitled,
contact the Office of Compliance at (240) 276-0120. Also, please note the regulatio contact the Office of Compliance at (211) 216 other (21 CFR Part 807.97). You may obtain
"Misbranding by reference to premarket not he the of first the Division of Small "Misbranding by relefence to premation international the Act from the Division of Small
other general information on your responsibilities under the Act from pumber (800) 638 other general information on your responsionner Assistance at its toll-free number (800) 638-2041 or Manufacturers, International and Consumer Assistance at to restoration of the industry/support/index.html.
Sincerely yours,
Mark N. Melkerson
Mark N. Melkerson તેમ Acting Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known): K052398
Device Name: Mesa Spinal System
Indications for Use:
The Mesa Spinal System is a non-cervical spinal fixation device intended for posterior, non-pedicle fixation for the following indications: degenerative disc disease ( DDD ) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies ); spondylolisthesis; trauma ( i.e. fracture or dislocation ); spinal stenosis; curvatures (i.e. scoliosis; and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion.
The Mesa Spinal System is also intended for non-cervical pedicle screw fixation for the following indications: trauma ( i.e. fracture or dislocation ); spinal stenosis; curvatures ( i.e. scoliosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion. It is also indicated for the treatment of severe 'pondylolisthesis ( grades 3 and 4 ) of the L5-S1 vertebra in skeletally mature patients receiving fusion by utogenous bone graft having implants attached to the lumbar and sacral spine ( L3 to sacrum ) with removal of the implants after the attainment of a solid fusion.
Prescription Use X ... . . (Part 21 CFR 801 Subpart D)
AND/OR
Over-the-counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS-LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
signature
(Division Sigu-Off) Division of General, Restorative, and Neurological Devices
510(k) Number_ K 052 39 8 __
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.