The Lanx VBR System is indicated for use to replace a vertebral body that has been resected or excised due to tumor or trauma/fracture. The device is intended for use as a vertebral body replacement in the thoracolumbar spine (from TI to LS) The Lanx VBR System may also be used in the thoracolumbar spine (i.e., Ti- L5) for partial replacement (i.e., partial vertebrectomy) of a diseased vertebral body resected or excised for the treatment of tumors in order to achieve anterior decompression of the spinal cord and neural tissues, and to restore the height of a collapsed vertebral body. The Lanx VBR System is also indicated for treating fractures of the thoracic and lumbar spine. The Lanx VBR System is designed to restore the biomechanical integrity of the anterior, middle, and posterior spinal column. For either indication the system must be used with supplemental internal fixation. Supplemental internal fixation is required to properly utilize this system.
Device Story
Lanx VBR is a spinal intervertebral body fixation orthosis; consists of flat, trapezoidal, rectangular, and curved PEEK implants with titanium position markers; hollowed central area accommodates bone graft; upper/lower surfaces feature transverse slots/grooves for stability. Used in thoracolumbar spine (T1-L5) to replace resected/excised vertebral bodies due to tumor or trauma/fracture; restores height of collapsed vertebrae; achieves anterior decompression of spinal cord/neural tissues; restores biomechanical integrity of spinal column. Requires supplemental internal fixation. Implanted by surgeons in clinical/OR settings. Benefits include structural support and stabilization of the spinal column following resection or fracture.
Clinical Evidence
Bench testing only. Fatigue and static testing performed according to accepted engineering and scientific principles. Results demonstrate performance equivalent to comparison devices. Biocompatibility information provided via device material Master File (MAF).
Technological Characteristics
Materials: Implant-grade PEEK with titanium position markers; instruments made of stainless steel. Design: Hollowed central area for bone graft; transverse slots/grooves on surfaces for fixation. Sterilization: Non-sterile; requires high-temperature steam sterilization (270°F/134°C for 18 minutes).
Indications for Use
Indicated for patients requiring vertebral body replacement or partial vertebrectomy in the thoracolumbar spine (T1-L5) due to tumor, trauma, or fracture. Requires use with supplemental internal fixation.
Regulatory Classification
Identification
A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vertebral plates that are punched into each of a series of vertebral bodies. An eye-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fastened or crimped at each eye-type screw. The device is used to apply force to a series of vertebrae to correct “sway back,” scoliosis (lateral curvature of the spine), or other conditions.
{0}------------------------------------------------
| 510(k) Summary Information<br>Premarket Notification, Section 510(k) | Lanx LLC<br>SEPT 17, 2005 |
|----------------------------------------------------------------------|---------------------------|
|----------------------------------------------------------------------|---------------------------|
Regulatory Authority: Safe Medical Devices Act of 1990, 21 CFR 807.92
#### 1. Device Name:
| Trade Name: | Lanx VBR |
|-----------------|----------------------------------------------|
| Common: | |
| Name(s): | Vertebral body replacement |
| Classification: | |
| Name(s): | Spinal intervertebral body fixation orthosis |
# 2. Establishment Name & Registration Number:
| Name: | Lanx LLC |
|---------|------------|
| Number: | 3004485144 |
#### 3. Classification(s):
Sec. 888.3060 Spinal intervertebral body fixation orthosis.
(a) Identification. A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vetebral plates that are punched into each of a series of vertebral bodies. An eye-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fareaded or crimped at each eye-type screw. The device is used to apply force to a series of vertebrae to correct "smay back," scoliosis (lateral curvature of the spine), or other conditions,
| Device Class: | Class II for the requested indications |
|-----------------------|----------------------------------------------|
| Classification Panel: | Orthopaedic and Rehabilitation Devices Panel |
| Product Code(s): | MQP |
# Equivalent Predicate Device:
Lanx LLC believes that the Lanx VBR is substantially equivalent to the following:
1. K003709, Mesh Cage, Surgical Dynamics
2. K001340, Stackable Cage, DePuy
3. K040928, PEEK Cage, Interpore/Cross
4. K042268, EBI CAS Spine Spacer System, EBI. L.P.
Equivalence can be seen in the design, material composition, surgical technique and intended use.
## 5. Device Description:
4.
Implants: The Lanx VBR implants consists of a series of flat, trapezoidal, rectangular and curved shaped The device is offered in different size/configurations to better approximate the anatomical implants. variation observed in different vertebral levels and/or patient anatomy. The devices are manufactured with a hollowed out central area to accommodate bone graft material The upper and lower surfaces are manufactured in such a way that a series of transverse slots or grooves are formed to improve stability and fixation once the device is inserted.
Materials: all implants are made from implant grade PEEK with titanium position markers, Instruments: The instruments are made from instrument grade stainlent post
Indications for Use. The Lanx VBR System is indicated for use to replace a vertebral body that has been resected or excised due to tumor or trauma/fracture. The device is intended for use as a vertebral body replacement in the thoracolumbar spire (from TI to LS) The Lanx VBR System may also be used in the thoracolumbar spine (i.e., Ti- L5) for partial replacement (i.e., partial vertebrectomy) of diseased
{1}------------------------------------------------
vertebral body resected or excised for the treatment of tumors in order to achieve anterior decompression of the spinal cord and neural tissues, and to restore the height of a collapsed vertebral body. The Lanx VBR System is also indicated for treating fractures of the thoracic and lumbar spine. The Lanx VBR System is designed to restore the biomechanical integrity of the anterior, middle, and posterior spinal column. For fixstion is secured to may be most be used with supplemental internal fixation. Supplemental internal fixation is required to properly utilize this system.
Testing Summary. Fatigue and static testing is complete. Samples were tested according to accepted engineering and scientific principals. Test results demonstrate that the system can be expected to perform in a manner equivalent to the comparison device material Master File (MAF) recorded with the FDA provided additional biocompatibility information.
#### 6. Applicant Name & Address:
Lanx, LLC 1155 Alpine Avenue, Suite 320 Boulder, CO 80304 303-443-7500 Fax: 303-443-7501
## 7. Company Contact:
Michael Fulton, M.D. Lanx, LLC 1155 Alpine Avenue, Suite 320 Boulder, CO 80304 303-443-7500 Fax: 303-443-7501
## 8. Submission Correspondent:
Mr. David W. Schlerf Buckman Company, Inc. 200 Gregory Lane, Suite C-100 Pleasant Hill, CA 94523-3389 925.356.2640 - 925.356.2654 - fax david(@fda-help.com
#### 9. Performance Standards:
United States Food and Drug Administration mandated performance standards for this device do not exist. Various voluntary performance standards are utilized of tills device do not exist Standard Operating Process certification and qualification and qualification procedures, Quality, Systems Regulations, ISO materials standards and ISO 9000 series quality regulations, LANX, LLC, also meets appropriate general controls authorized under Sections 501, 502, 516, 516, 516, 518, 519, and 520 of the Food, Drug, and Cosmetic Act,
## I O. Sterilization Information:
The Lanx Spinal Fixation System is supplied "NON-STERILE" and must be sterilized prior to use. The recommended sterlization process is high temperature sterilization. The referenced sterilization cycle produces a Sterility Assurance Level (SAL) of att least 106
The validated cycle is: Method: Steam Cycle: Gravity Temperature: 270°F (134°C) Exposure Time: 18 minutes
{2}------------------------------------------------
| 510(k) Summary Information | Lanx LLC |
|----------------------------------------|---------------|
| Premarket Notification, Section 510(k) | SEPT 17, 2005 |
Regulatory Authority: Safe Medical Devices Act of 1990, 21 CFR 807.92
#### 1. Device Name:
Lanx VBR Trade Name: Common Name(s): Vertebral body replacement Classification Spinal intervertebral body fixation orthosis Name(s):
## 2. Establishment Name & Registration Number:
| Name: | Lanx LLC |
|---------|------------|
| Number: | 3004485144 |
## 3. Classification(s):
Sec. 888.3060 Spinal intervertebral body fixation orthosis.
(a) Identification. A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vertebral plates that are punched into each of a series of vertebral odiaes. An eye-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fastened or rimed at each eye-type screw. The device is used to apply force to a series of vertebrae to correct "sway back," scoliosis (lateral curvature of the spine), or other conditions.
| Device Class: | Class II for the requested indications |
|-----------------------|----------------------------------------------|
| Classification Panel: | Orthopaedic and Rehabilitation Devices Panel |
| Product Code(s): | MQP |
## 4. Equivalent Predicate Device:
Lanx LLC believes that the Lanx VBR is substantially equivalent to the following:
- 1. K003709, Mesh Cage, Surgical Dynamics
- 2. K001340, Stackable Cage, DePuy
- 3. K040928, PEEK Cage, Interpore/Cross
- 4. K042268, EBI CAS Spine Spacer System, EBI. L.P.
Equivalence can be seen in the design, material composition, surgical technique and intended use
## 5. Device Description:
Implants: The Lance VBR implants consists of a series of flat, trapezoidal, rectangular and curved shaped The device is offered in different size/configurations to better approximate the anatomical implants. variation observed in different vertebral levels and/or patient anatomy. The devices are manufactured with a hollowed out central area to accommodate bone graft material. The upper and lower surfaces are manufactured in such a way that a series of transverse slots or grooves are formed to improve stability and fixation once the device is inserted.
Materials: all implants are made from implant grade PEEK with titanium position markers, Instruments: The instruments are made from instrument grade stainlens steel.
Indications for Use. The Lanx VBR System is indicated for use to replace a vertebral body that has been resected or excised due to tumor or trauma/fracture. The device is intended for use as a vertebral body themely when with the thoracolumbar spine (from TI to LS) The Lanx VBR System may also be used in the thoracolumbar spine (i.e., Ti- L5) for partial replacement (i.e., partial vertebrectomy) of diseased
{3}------------------------------------------------
vertebral body resected or excised for the treatment of tumors in order to achieve anterior decompression of the spinal cord and neural tissues, and to restore the height of a collapsed vertebral body. The Land VRR System is also indicated for treating fractures of the thoracic and lumbar spine. The Lanx VBR System is designed to restore the biomechanical integrity of the anterior, middle, and posterior spinal column. For either indication the system must be used with supplemental fixation. Supplemental internal fixation is required to properly utilize this system.
Testing Summary. Fatigue and static testing is complete. Samples were tested according to accepted engineering and scientific principals. Test results demonstrate that the system can be expected to perform in a manner equivalent to the comparison device. Additionally, a device material Master Five (MAF) recorded with the FDA provided additional biocompatibility information.
#### 6. Applicant Name & Address:
Lanx、LLC 1155 Alpine Avenue, Suite 320 Boulder, CO 80304 303-443-7500 Fax: 303-443-7501
## 7. Company Contact:
Michael Fulton, M.D. Lanx, LLC 1155 Alpine Avenue, Suite 320 Boulder, CO 80304 303-443-7500 Fax: 303-443-7501
## 8. Submission Correspondent:
Mr. David W. Schlerf Buckman Company, Inc. 200 Gregory Lane, Suite C-100 Pleasant Hill, CA 94523-3389 925.356.2640 - 925.356.2654 - fax david@fda-help.com
#### 9. Performance Standards:
United States Food and Drug Administration mandated performance standards for this device do not exist. Various voluntary performance standards are utilized overaliands of tilliged include ASTM, Surtame Rocedures, vendor & process certification and qualification procedures, Qualification procedures, Qualification procedures, Qualification procedures, Qualification pr Systems Regulations, ISO materials standards and ISO 9000 series quality regulations. LANX, LLC, also meets appropriate general controls authorized under Sections 501, 516, 516, 518, 519, and 520 of the Food, Drug, and Cosmetic Act.
## 1 0. Sterilization Information:
The Lanx Spinal Fixation System is supplied "NON-STERILE" and must be sterilized prior to use. The recommended sterilization process is high temperature sterilization of sterilization. The referenced sterilization cycle produces a Sterility Assurance Level (SAL) of at least 106
The validated cycle is: Method: Steam Cycle: Gravitv Temperature: 270°F (134°C) Exposure Time: 18 minutes
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized caduceus symbol, which features a staff entwined with three snakes, representing healing and medicine. The logo is surrounded by the text "DEPARTMENT OF HEALTH AND HUMAN SERVICES, USA" in a circular arrangement.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
OCT 7 - 2005
Lanx, LLC c/o David W. Schlerf Buckman Company, Inc. 200 Gregory Lane, Suite C-100 Pleasant Hill. California 94523
Re: K052384
Trade/Device Name: Lanx VBR Regulation Number: 21 CFR 888.3060 Regulation Name: Spinal intervertebral body fixation orthosis Regulatory Class: II Product Code: MQP Dated: August 26, 2005 Received: August 30, 2005
Dear Mr. Schlerf:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{5}------------------------------------------------
Page 2- David W. Schlerf
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Signature: Mark McMullen
Mark N. Melkerson Acting Director Division of General, Restorative, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Page 1 of 1
510(k) Number: K052384
LanxVBR Device Name(s):
Indications for Use:
The Lanx VBR System is indicated for use to replace a vertebral body that has been resected or excised due to tumor or trauma/fracture. The device is intended for use as a vertebral body replacement in the thoracolumbar spine (from TI to LS) The Lanx VBR System may also be used in the thoracolumbar spine (i.e., Ti- L5) for partial replacement (i.e., partial vertebrectomy) of a diseased vertebral body resected or excised for the treatment of tumors in order to achieve anterior decompression of the spinal cord and neural tissues, and to restore the height of a collapsed vertebral body. The Lanx VBR System is also indicated for treating fractures of the thoracic and lumbar spine. The Lanx VBR System is designed to restore the biomechanical integrity of the anterior, middle, and posterior spinal column. For either indication the system must be used with supplemental internal fixation. Supplemental internal fixation is required to properly utilize this system.
Over-The-Counter Use Prescription Use X OR PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NECESSARY Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off) Division of General, Restorative, and Neurological Devices
510(k) Number K052384
(Per 21 CFR 801.109)
(Optional format 1-2-96)
ાં ર
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.