When intended to promote fusion of the occipitocervical spine, and the thoracic spine, (Occiput-T3), the VERTEX® Reconstruction System is indicated for the following: DDD (neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, spinal stenosis, fracture, dislocation, failed previous fusion and/or tumors. Occipitocervical Plate/Rod/Occipital Screws/Hooks The occipitocervical plate/rods, occipital screws (3.5mm, 4.0mm and 4.5mm cancellous), and hooks are intended to provide stabilization to promote fusion following reduction of fracture/dislocation or trauma in the occipitocervical junction and the cervical spine. When used to treat these occipitocervical and cervical conditions, these screws are limited to occipital fixation only. The screws are not intended to be placed in the cervical spine. The use of the occipitocervical plate/rod requires bilateral fixation to C2 and below. Note: segmental fixation is recommended for these constructs. Hooks and Rods The hooks and rods are also intended to provide stabilization to promote fusion following reduction of fracture/dislocation or trauma in the cervical/upper thoracic (C1-T3) spine. Multi-axial Screws/Connectors The use of multi-axial screws (3.5mm and 4.0mm cancellous, and 4.0mm cortical) are limited to placement in T1-T3. The screws are not intended to be placed in the cervical spine. Titanium ATLAS® Cable System used with the VERTEX® Reconstruction System allows for cable attachment to the posterior cervical or thoracic spine.
Device Story
Posterior spinal reconstruction system; consists of rods, hooks, screws, multi-axial screws, and connecting components; used to stabilize the occipitocervical, cervical, and thoracic spine (Occiput-T3) to promote fusion. Surgeon tailors construct configuration for individual patient cases. Multi-axial screws include Nitinol (NiTi) retaining rings. Titanium ATLAS® cable system may be used for attachment. System is implanted by surgeons in clinical settings. Provides rigid fixation to address degenerative disc disease, trauma, and other spinal pathologies. Benefits include spinal stabilization and fusion support.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Fabricated from medical grade titanium or titanium alloy; multi-axial screw retaining rings made of Shape Memory Alloy (Nitinol). System includes rods, hooks, screws, and lateral connectors. Designed for posterior spinal fixation. Not compatible with stainless steel.
Indications for Use
Indicated for patients requiring spinal fusion in the occipitocervical, cervical, and thoracic (Occiput-T3) regions due to DDD, spondylolisthesis, spinal stenosis, fracture, dislocation, failed fusion, or tumors. Contraindicated for use with stainless steel implants.
Regulatory Classification
Identification
A spinal interlaminal fixation orthosis is a device intended to be implanted made of an alloy, such as stainless steel, that consists of various hooks and a posteriorly placed compression or distraction rod. The device is implanted, usually across three adjacent vertebrae, to straighten and immobilize the spine to allow bone grafts to unite and fuse the vertebrae together. The device is used primarily in the treatment of scoliosis (a lateral curvature of the spine), but it also may be used in the treatment of fracture or dislocation of the spine, grades 3 and 4 of spondylolisthesis (a dislocation of the spinal column), and lower back syndrome.
{0}------------------------------------------------
SEP 23 2005
Ko52376
# VERTEX® Reconstruction System 510(k) Summary September 19, 2005
| I. | Company: | Medtronic Sofamor Danek USA, Inc.<br>1800 Pyramid Place<br>Memphis, Tennessee 38132<br>(901) 396-3133 | | |
|-----|----------------------|-------------------------------------------------------------------------------------------------------|--|--|
| | Contact: | Richard W. Treharne, Ph.D.<br>Senior Vice President, Regulatory Affairs | | |
| II. | Product Name: | VERTEX® Reconstruction System | | |
| | Classification Name: | Spinal Interlaminal Fixation Orthosis, Pedicle Screw<br>Spinal System | | |
| | Regulation Number: | 888.3050, 888.3070 | | |
| | Code: | KWP, MNI | | |
#### Description: III.
The VERTEX® Reconstruction System is a posterior system, which consists of a variety of shapes and sizes of rods, hooks, screws, multi-axial screws, and connecting components, which can be rigidly locked to the rod in a variety of configurations, with each construct being tailormade for the individual case. Titanium ATLAS® cable may be used with this system at the surgeon's discretion.
The VERTEX® Reconstruction System is fabricated from medical grade titanium or titanium alloy. The VERTEX® Reconstruction System also includes a retaining ring for the multi-axial screw made of Shape Memory Alloy (Nitinol – NiTi). Shape Memory Alloy is compatible with medical grade titanium or titanium alloy implants only. Do not use with stainless steel.
The purpose of this submission was to add modified lateral connectors to the VERTEX® Reconstruction System.
#### IV Indications
When intended to promote fusion of the occipitocervical spine, cervical spine, and the thoracic spine, (Occiput-T3), the VERTEX® Reconstruction System is indicated for the following: DDD (neck pain of discogenic origin with degeneration of the disc confirmed by history and
$$
\hat{\urcorner} \urcorner \urcorner
$$
{1}------------------------------------------------
radiographic studies), spondylolisthesis, spinal stenosis, fracture, dislocation, failed previous fusion and/or tumors.
# Occipitocervical Plate/Rod/Occipital Screws/Hooks
The occipitocervical plate/rods, occipital screws (3.5mm, 4.0mm and 4.5mm cancellous), and hooks are intended to provide stabilization to promote fusion following reduction of fracture/dislocation or trauma in the occipitocervical junction and the cervical spine. When used to treat these occipitocervical and cervical conditions, these screws are limited to occipital fixation only. The screws are not intended to be placed in the cervical spine.
The use of the occipitocervical plate/rod requires bilateral fixation to C2 and below. Note: segmental fixation is recommended for these constructs.
### Hooks and Rods
The hooks and rods are also intended to provide stabilization to promote fusion following reduction of fracture/dislocation or trauma in the cervical/upper thoracic (C1-T3) spine.
### Multi-axial Screws/Connectors
The use of multi-axial screws (3.5mm and 4.0mm cancellous, and 4.0mm cortical) are limited to placement in T1-T3. The screws are not intended to be placed in the cervical spine.
Titanium ATLAS® Cable System used with the VERTEX® Reconstruction System allows for cable attachment to the posterior cervical or thoracic spine.
#### V. Substantial Equivalence:
Documentation has been provided which demonstrates that the VERTEX® Reconstruction System is substantially equivalent to VERTEX® Reconstruction System that was previously cleared in K042524 and K003780.
page 2 of 2
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image is a black and white logo for the Department of Health and Human Services, USA. The logo features a circular border with the text "DEPARTMENT OF HEALTH AND HUMAN SERVICES, USA" arranged around the perimeter. Inside the circle is a stylized emblem resembling an eagle or bird in flight, composed of three curved lines.
Public Health Service
SEP 2 3 2005
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Richard W. Treharne, Ph.D. Senior Vice President, Regulatory Affairs Medtronic Sofamor Danek 1800 Pyramid Place Memphis, Tennessee 38132
Re: K052376
K052370
Trade/Device Name: VERTEX® Reconstruction System Regulation Number: 21 CFR 888.3050 Regulation Name: Spinal interlaminal fixation orthosis Regulatory Class: II Product Code: KWP, MNI Dated: August 29, 2005 Received: August 30, 2005
Dear Dr. Treharne:
We have reviewed your Section 510(k) premarket notification of intent to market the device We have reviewed your Section 310(x) premained in substantially equivalent (for the indications
referenced above and have determined the device is substantially interstate referenced above and nave determined the arreted predicate devices marketed in interstate for use stated in the enclosure) to legally manced producal Device Amendments, or to
commerce prior to May 28, 1976, the enactment answerings of the Federal Food. Drug commerce prior in May 28, 1970, the cities with the provisions of the Federal Food. Drug, devices that have been reciassified in accordance who we proval application (PMA).
and Cosmetic Act (Act) that do not require approval of a provisions of the Act . The and Cosmetic Act (Act) that do not require approvine of the general controls provisions of the Act. The
You may, therefore, market the device, subject to the general negat You may, therefore, market the device, sucject to urements for annual registration, listing of
general controls provisions of the Act include required michrending and general controls provisions of the rect mendo requirements.
devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is classified (see above) more more regulations affecting your device can
may be subject to such additional controls. Title 01 - 808 - 808 - 808 - 808 - 104 m may be subject to such additions. Existing major regaleto 898. In addition, FDA may
be found in the Code of Federal Regulations, Title 21, Parts 800 to 899 be found in the Code of I cacial regarations, your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean Please be advised that FDA \$ issualice of a subscribe with other requirements of the Act
that FDA has made a determination that your device complies. You must that FDA has made a determination that your as res by other Federal agencies. You must or any Federal statules and regulations administered of registration and listing (21 l
comply with all the Act's requirements, including, but not limited to: requirements as comply with all the Act Stequirements, more and manufacturing practice requirements as set CFR Part 807); labeling (21 CFR Pactor); good and if applicable, the electronic
forth in the quality systems (QS) regulation (21 CFR Pat 820); and if applicable, the electron forth in the quality systems (Q5) regulation (21-51-542 of the Act); 21 CFR 1000-1050.
product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2- Richard W. Treharne, Ph.D.
This letter will allow you to begin marketing your device as described in your Section 510(k)
The success of the same of a local at the samination of your device to a legal This letter will allow you to begin marketing your avince of your device of your device to a legally premarket notification. The PDA inding of substantial equice and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please
rs to and the many of the 110 - All and 1400 - All any desse note the regul If you desire specific auvice for your do no no no a come note the regulation entitled, the regulation entitled, would obtain contact the Office of Compliance at (240) 270 - 1210 - 121 Part 807.97). You may obtain
"Misbranding by reference to premarket notification" (21) - 1 the Division of Small "Misbranding by reference to prematics notification (the Act from the Division of Small)
other general information on your responsibilities under the Act from worker (800) 63 other general information on your responsion.com its toll-free number (800) 638-2041 or
Manufacturers, International and Consumer Assistance at their unnortlinges btml Manufacturers, International and Consumer Pisolation around consection of the new learner in the support/index.html.
Sincerely yours,
Mark N. Melkerson
Mark N. Metkerson Acting Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Page 1 _ of __ 1 __ 1 __
August 2005
510(k) Number (if known): K052376
VERTEX® Reconstruction System Device Name: _________________________________________________________________________________________________________________________________________________________________
Indications for Use:
When intended to promote fusion of the occipitocervical spine, and the thoracic spine, (Occiput-T3), the VERTEX® Reconstruction System is indicated for the following: DDD (neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, spinal stenosis, fracture, dislocation, failed previous fusion and/or tumors.
# Occipitocervical Plate/Rod/Occipital Screws/Hooks
The occipitocervical plate/rods, occipital screws (3.5mm, 4.0mm and 4.5mm cancellous), and hooks are intended to provide stabilization to promote fusion following reduction of fracture/dislocation or trauma in the occipitocervical junction and the cervical spinc. When used to treat these occipitocervical and cervical conditions, these screws are limited to occipital fixation only. The screws are not intended to be placed in the cervical spine.
The use of the occipitocervical plate/rod requires bilateral fixation to C2 and below. Note: segmental fixation is recommended for these constructs.
### Hooks and Rods
The hooks and rods are also intended to provide stabilization to promote fusion following reduction of fracture/dislocation or trauma in the cervical/upper thoracic (C1-T3) spine.
# Multi-axial Screws/Connectors
The use of multi-axial screws (3.5mm and 4.0mm cancellous, and 4.0mm cortical) are limited to placement in T1-T3. The screws are not intended to be placed in the cervical spine.
Titanium ATLAS® Cable System used with the VERTEX® Reconstruction System allows for cable attachment to the posterior cervical or thoracic spine.
| Prescription Use (Part 21 CFR 801 Subpart D) | X | AND/OR | Over-The-Counter Use (21 CFR 807 Subpart C) | |
|----------------------------------------------|---|--------|---------------------------------------------|--|
|----------------------------------------------|---|--------|---------------------------------------------|--|
(PLEASE SIGN BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
(Division Sign-Off)
Concurrence of CDRH, Division of General, Restorative, and Neurological Devices
510(k) Number K052376
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.