K052316 · Cardiac Science, Inc. · MKJ · Jan 3, 2006 · Cardiovascular
Device Facts
Record ID
K052316
Device Name
POWERHEART ECD
Applicant
Cardiac Science, Inc.
Product Code
MKJ · Cardiovascular
Decision Date
Jan 3, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.5310
Device Class
Class 3
Attributes
Therapeutic, Pediatric
Indications for Use
The Powerheart ECD defibrillator system is intended to be used by personnel who have been trained in its operation. The Powerheart ECD is indicated for the termination of certain fatal arrhythmias, such as ventricular fibrillation and symptomatic ventricular tachycardia. Delivery of energy in the synchronized mode is a method for treating atrial fibrillation, atrial flutter, paroxysmal supraventricular tachycardia, and in relatively stable patients, ventricular tachycardia. The semi-automatic advisory mode is for use in cardiac arrest in patients of at least 8 years of age. The patient must be unconscious, pulseless, and not breathing spontaneously before using the defibrillator to analyze the patient's ECG rhythm. The Powerheart ECD 3-lead and 5-lead ECG monitoring allows for identification or interpretation of cardiac rhythms or dysrhythmias and calculation of heart rate. The Powerheart ECD noninvasive pacing as a therapy is indicated for patients with symptomatic bradycardia or asystole. The Powerheart ECD pulse oximetry is intended for the continuous external monitoring of arterial oxygen saturation and pulse rate and is indicated for use in any patient who is at risk of developing hypoxemia.
Device Story
Powerheart ECD is a portable, battery/AC-powered defibrillator/monitor/pacemaker for use by trained personnel. Device inputs include ECG signals (3/5-lead), SpO2, and impedance. RHYTHMx software analyzes ECG rhythms; STAR Biphasic waveform delivers impedance-compensated energy (2-270 Joules). Therapies include manual/semi-automatic defibrillation, cardioversion, and fixed/demand pacing. Output displayed on 320x240 TFT color screen; event history stored internally; thermal printer provided. Clinicians use output to guide resuscitation and cardiac therapy; device facilitates rhythm identification, heart rate calculation, and oxygen saturation monitoring. Benefits include rapid intervention for life-threatening arrhythmias and continuous patient monitoring.
Clinical Evidence
No clinical data. Bench testing, system testing, and safety testing were performed on hardware and software components to demonstrate functional and performance characteristics.
Technological Characteristics
Portable defibrillator/monitor/pacemaker; 320x240 TFT color display; AC or Li-Ion battery power. Delivers impedance-compensated biphasic energy (2-270J). Features 3/5-lead ECG, SpO2 monitoring, and 60mm thermal printer. Software-based rhythm analysis (RHYTHMx).
Indications for Use
Indicated for termination of fatal arrhythmias (ventricular fibrillation, symptomatic ventricular tachycardia) in patients; synchronized mode for atrial fibrillation, atrial flutter, paroxysmal supraventricular tachycardia, and stable ventricular tachycardia. Semi-automatic advisory mode for cardiac arrest in patients ≥8 years old (unconscious, pulseless, apneic). Noninvasive pacing for symptomatic bradycardia or asystole. Pulse oximetry for patients at risk of hypoxemia.
Regulatory Classification
Identification
An automated external defibrillator (AED) system consists of an AED and those accessories necessary for the AED to detect and interpret an electrocardiogram and deliver an electrical shock (e.g., battery, pad electrode, adapter, and hardware key for pediatric use). An AED system analyzes the patient's electrocardiogram, interprets the cardiac rhythm, and automatically delivers an electrical shock (fully automated AED), or advises the user to deliver the shock (semi-automated or shock advisory AED) to treat ventricular fibrillation or pulseless ventricular tachycardia.
{0}------------------------------------------------
#### JAN 3 2006
### SECTION 2. SUMMARY AND CERTIFICATION
| 2.1. 510(k) Summary | |
|------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | Cardiac Science, Inc.<br>5474 Feltl Road<br>Minnetonka, MN 55343 |
| Contact Person: | Kenneth Olson<br>Phone: (952) 939-2912<br>Fax: (952) 939-4191 |
| Date Prepared: | August 22, 2005 |
| Trade Name: | Powerheart ECD™ |
| Classification Name<br>and Number: | Automated External Defibrillator<br>Class III, 21CFR 870.5310 |
| Product Code: | MKJ |
| Predicate Device(s): | The Powerheart ECD™ manufactured by Cardiac Science, Inc<br>is substantially equivalent to Medtronic Physio-Control<br>LIFEPAK® 20 Defibrillator (K012274). |
| Device Description: | The Powerheart® ECD is a defibrillator/monitor/pacemaker<br>intended for use by personnel trained in its operation. The<br>device is lightweight, portable, easy to use and reliable. It<br>incorporates a 320 x 240 transmissive TFT color display for<br>wide viewing angles in all light conditions. The device operates<br>using either an AC power supply or internal rechargeable Li-Ion<br>battery. The device provides continuous ECG monitoring and<br>three types of therapies: defibrillation, cardioversion and<br>external pacing. Defibrillation can be applied manually or<br>semi-automatically. Pacing therapy can be either fixed or<br>demand. The device employs patented RHYTHMx® software<br>which provides ECG rhythm analysis. STAR® Biphasic<br>waveform delivers impedance-compensated energy ranging<br>from 2-270 Joules. Features and options include external<br>paddles, spoons, disposable pads, 3- and 5-lead ECG, pulse<br>oximetry (SpO2), built-in 60 mm thermal printer, internal<br>storage of event history and remote synchronization to bedside<br>monitor. |
| Indications For Use: | The Powerheart ECD defibrillator system is intended to be used<br>by personnel who have been trained in its operation.<br><br>The Powerheart ECD is indicated for the termination of certain<br>fatal arrhythmias, such as ventricular fibrillation and<br>symptomatic ventricular tachycardia. Delivery of energy in the<br>synchronized mode is a method for treating atrial fibrillation,<br>atrial flutter, paroxysmal supraventricular tachycardia and in |
| | |
| | relatively stable patients, ventricular tachycardia. |
| | The semi-automatic advisory mode is for use in cardiac arrest in<br>patients of at least 8 years of age. The patient must be<br>unconscious, pulseless, and not breathing spontaneously before<br>using the defibrillator to analyze the patient's ECG rhythm. |
| | The Powerheart ECD 3-lead and 5-lead ECG monitoring allows<br>for identification or interpretation of cardiac rhythms or<br>dysrhythmias and calculation of heart rate. |
| | The Powerheart ECD noninvasive pacing as a therapy is<br>indicated for patients with symptomatic bradycardia or asystole. |
| | The Powerheart ECD pulse oximetry is intended for the<br>continuous external monitoring of arterial oxygen saturation and<br>pulse rate and is indicated for use in any patient who is at risk of<br>developing hypoxemia. |
| Functional and<br>Safety Testing: | Representative samples of the device components underwent<br>system, safety and bench testing on both hardware and software<br>to demonstrate appropriate functional and performance<br>characteristics. |
| Conclusion: | Based on the results of the testing described above, it is<br>concluded that the Powerheart ECD does not raise any different<br>questions regarding the safety or effectiveness as compared<br>with the predicate device. The Cardiac Science, Inc. Powerheart<br>ECD is substantially equivalent to the Medtronic Physio-<br>Control LIFEPAK 20 in terms of indications for use, features<br>and functions. |
#### 2.1 510(k) Summarv
510(k) Premarket Submission Cardiac Science, Inc.
{1}------------------------------------------------
:
12
{2}------------------------------------------------
#### 2.2. 510(k) Checklist
# Traditional 510(k) Premarket Notification Checklist for Class III Device
| No. | Item | Included | Location |
|-----|------------------------------------------------------------------------|----------|--------------|
| 1 | Cover letter clearly identifying submission as<br>"Traditional 510(k)" | Yes | Cover Letter |
| 2 | Table of Contents | Yes | TOC |
| 3 | Indication for Use | Yes | IFU page |
| 4 | Manufacturer name and address | Yes | Section I |
| 5 | Contact person, phone number, fax number | Yes | Section 1 |
| 6 | Device common name and trade name | Yes | Section 1 |
| 7 | Establishment registration number | Yes | Section 1 |
| 8 | Device classification and product code | Yes | Section 1 |
| 9 | 510(k) Summary | Yes | Section 2.1 |
| 10 | Identification of predicate device | Yes | Section 2.1 |
| 11 | Truthful and Accurate Statement | Yes | Section 2.4 |
| 12 | Class III Certification and Summary | Yes | Section 2.5 |
| 13 | Proposed labeling | Yes | Section 3 |
| 14 | Description of the device | Yes | Section 4 |
| 15 | Substantial equivalence comparison | Yes | Section 5 |
| 16 | Performance testing | Yes | Section 6 |
| 17 | System Hazard Analysis | Yes | Section 6.1 |
| 18 | Clinical evaluation | Yes | Section 6.6 |
| 19 | Biocompatibility data | Yes | Section 7 |
| 20 | Sterilization information | Yes | Section 8 |
| 21 | Software documentation | Yes | Section 9 |
| 22 | Compliance with performance standards | Yes | Section 10 |
{3}------------------------------------------------
#### 2.3. Premarket Notification Statement
:
. '
:
In lieu of a Pre Market Notification Statement, a 510(k) Summary is provided in Section 2.1. of this 510(k) Premarket Notification.
:
{4}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES. USA
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JAN 3 2006
Cardiac Science, Inc. c/o Mr. Kenneth F. Olson Chief Technical Officer 5474 Feltl Road Minnetonka, MN 55343-7982
Re: K052316 Trade Name: Powerheart ECD Regulation Number: 21 CFR 870.5310 Regulation Name: Automated External Defibrillator Regulatory Class: Class III Product Code: MKJ Dated: December 20, 2005 Received: December 21, 2005
Dear Mr. Olson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 - Mr. Kenneth F. Olson
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050, This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely vours,
Bheminar for
Bram D. Zuckerman, M.D. Director Division of Cardiovacsular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
## Indications for Use
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
Device Name: Powerheart ECD TM
The Powerheart ECD defibrillator system is intended to be used by personnel who have been trained in its operation.
The Powerheart ECD is indicated for the termination of certain fatal arrhythmias, such as ventricular fibrillation and symptomatic ventricular tachycardia. Delivery of energy in the synchronized mode is a method for treating atrial fibrillation, atrial flutter, paroxysmal supraventricular tachycardia, and in relatively stable patients, ventricular tachycardia.
The semi-automatic advisory mode is for use in cardiac arrest in patients of at least 8 years of age. The patient must be unconscious, pulseless, and not breathing spontaneously before using the defibrillator to analyze the patient's ECG rhythm.
The Powerheart ECD 3-lead and 5-lead ECG monitoring allows for identification or interpretation of cardiac rhythms or dysrhythmias and calculation of heart rate.
The Powerheart ECD noninvasive pacing as a therapy is indicated for patients with symptomatic bradycardia or asystole.
The Powerheart ECD pulse oximetry is intended for the continuous external monitoring of arterial oxygen saturation and pulse rate and is indicated for use in any patient who is at risk of developing hypoxemia.
Prescription Use _____________________________________________________________________________________________________________________________________________________________ AND/OR (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page I of l
Blammymer
(Division Sign-Off
Division of Cardiovascular Devices
510(k) Number K052316
510(k) Premarket Submission Cardiac Science, Inc.
Powerheart ECD CONFIDENTIAL
August 23, 2005
9
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.