K052223 · Beckman Coulter, Inc. · GGT · Oct 6, 2006 · Hematology
Device Facts
Record ID
K052223
Device Name
ACCESS EPO ASSAY
Applicant
Beckman Coulter, Inc.
Product Code
GGT · Hematology
Decision Date
Oct 6, 2006
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 864.7250
Device Class
Class 2
Indications for Use
The Access EPO assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of erythropoietin levels in human serum and plasma (heparin) using the Access Immunoassay Systems. This assay is intended as an aid in the diagnosis of anemias and polycythemias. With the advent of the administration of recombinant erythropoietin as a biologic therapy to increase red blood cell mass, an erythropoietin assay may be used also to aid in the prediction and monitoring of response to recombinant erythropoietin treatment in persons with anemias. The Access EPO calibrators are intended to calibrate the Access EPO assay for the quantitative determination of EPO levels in human serum and plasma (heparin) using the Access Immunoassay Systems.
Device Story
Access EPO assay; paramagnetic particle, chemiluminescent immunoassay. Input: human serum or heparinized plasma samples. Operation: two-site immunoenzymatic (sandwich) assay; sample incubated with paramagnetic particles coated with mouse monoclonal anti-EPO and alkaline phosphatase conjugate; magnetic field separates bound from unbound materials; chemiluminescent substrate (Lumi-Phos 530) added; light measured via luminometer. Output: quantitative EPO concentration derived from stored multi-point calibration curve. Used in clinical laboratories on Access family of immunoassay analyzers. Healthcare providers use results to diagnose anemias/polycythemias and monitor recombinant EPO therapy response.
Clinical Evidence
Bench testing only. Precision: total CV ≤10% at >3 mIU/mL. Linearity: recovery 100±15%. Detection limit: ≤0.6 mIU/mL. Interference: no significant interference from hemoglobin, triglycerides, bilirubin, albumin, heparin, or common medications. Method comparison: internal (n=103) and external (n=113) studies vs. Quantikine ELISA showed high correlation (r=0.988 and 0.995). Matrix comparison (n=80) confirmed equivalence between serum and heparinized plasma.
Technological Characteristics
Paramagnetic particle-based chemiluminescent immunoassay. Solid phase: paramagnetic particles coated with goat anti-mouse IgG and mouse anti-recombinant human EPO monoclonal antibody. Conjugate: alkaline phosphatase (bovine) conjugate. Reagents include BSA, sodium azide, ProClin 300, and TRIS saline buffer. Standardization: WHO 2nd IRP 67/343. Measuring range: 0.6 - 750 mIU/mL. System: Access Immunoassay Systems.
Indications for Use
Indicated for quantitative determination of erythropoietin (EPO) in human serum and plasma (heparin) to aid in diagnosis of anemias and polycythemias, and to aid in prediction and monitoring of response to recombinant EPO therapy in patients with anemias.
Regulatory Classification
Identification
A erythropoietin assay is a device that measures the concentration of erythropoietin (an enzyme that regulates the production of red blood cells) in serum or urine. This assay provides diagnostic information for the evaluation of erythrocytosis (increased total red cell mass) and anemia.
Special Controls
*Classification.* Class II. The special control for this device is FDA's “Document for Special Controls for Erythropoietin Assay Premarket Notification (510(k)s).”
Predicate Devices
Quantikine IVD Erythropoietin ELISA Kit - R & D Systems, Inc. (K936016)
Nichols Advantage Chemiluminescent Erythropoietin Immunoassay - Nichols Institute Diagnostics (K980737)
EPO [Erythropoietin] ELISA Assay - Sangui Bio Tech, Inc. (K992799)
Submission Summary (Full Text)
{0}
1
510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
ASSAY ONLY TEMPLATE
A. 510(k) Number:
K052223
B. Purpose for Submission:
New Device
C. Measurand:
Erythropoietin (EPO)
D. Type of Test:
Quantitative, Immunoenzymatic
E. Applicant:
Beckman Coulter, Inc.
F. Proprietary and Established Names:
Access Erythropoietin (EPO) Assay and Calibrator
G. Regulatory Information:
1. Regulation section:
21 CFR 864.7250
21 CFR 862.1150
2. Classification:
Class II
3. Product code:
GGT
JIT
{1}
4. Panel:
81 (Hematology)
H. Intended Use:
1. Intended use(s):
The Access EPO assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of erythropoietin levels in human serum and plasma (heparin) using the Access Immunoassay Systems. This assay is intended as an aid in the diagnosis of anemias and polycythemias. With the advent of the administration of recombinant erythropoietin as a biologic therapy to increase red blood cell mass, an erythropoietin assay may be used also to aid in the prediction and monitoring of response to recombinant erythropoietin treatment in persons with anemias.
The Access EPO calibrators are intended to calibrate the Access EPO assay for the quantitative determination of EPO levels in human serum and plasma (heparin) using the Access Immunoassay Systems.
2. Indication(s) for use:
Same as above.
3. Special conditions for use statement(s):
Not applicable.
4. Special instrument requirements:
For use with the Access family of Immunoassay analyzers (Access, Access 2, Synchron LX®i 725, UniCel DxI 800, UniCel DxC 600i)
I. Device Description:
The Access EPO assay consists of the reagent pack and calibrators. The reagent pack consist of three specific reagents; (1) Paramagnetic particles coated with goat anti-mouse IgG: mouse anti-recombinant human EPO monoclonal antibody, BSA, sodium azide and ProClin 300, (2) Chicken anti-recumbent mouse alkaline phosphatase (bovine) conjugate, BSA, 0.1% sodium aide and 0.7% ProClin 300, (3) TRIS saline buffer containing BSA, proteins (chicken, bovine, mouse), <0.1% sodium azide and 0.17% ProClin 300. Two reagent packs containing 50 tests per pack are provided for a total of 100 assay determinations. The calibrator kit provides calibrators at 6 levels: zero (S0) and approximately 5, 25, 125, 375, 750ml mIU/ml (S1 - S5) and one Calibration Card. The S0 calibrator vial contains 10 ml and each S1 - S5 vial contains
2
{2}
2.5mL. Other items needed to perform the assay include the Access substrate and wash buffers for use with the Access family of Immunoassay Analyzers in the clinical laboratory.
## J. Substantial Equivalence Information:
1. Predicate device name(s):
Quantikine IVD Erythropoietin ELISA Kit - R & D Systems, Inc.
IMMULITE EPO - Diagnostic Products Corp.
Nichols Advantage Chemiluminescent Erythropoietin Immunoassay - Nichols Institute Diagnostics
EPO [Erythropoietin] ELISA Assay - Sangui Bio Tech, Inc.
2. Predicate 510(k) number(s):
K936016, K983203, K980737, K992799
3. Comparison with predicate:
| Similarities | | | | | |
| --- | --- | --- | --- | --- | --- |
| Characteristic | BCI Access EPO | RDS Quantikine | DPC Immulite | Nichols Advantage | Sangui Bio EPO ELISA Kit |
| Indications for use | For the quantitative determination of erythropoietin levels in human serum and plasma (heparin) using the Access Immunoassay Systems. This assay is intended as an aid in the diagnosis of anemias and polycythemias. With the advent of the administration of recombinant erythropoietin as a biologic therapy to increase red blood cell mass, an erythropoietin assay may be used also to aid in the prediction and monitoring of response to recombinant erythropoietin treatment in persons with anemias. | For the quantitative measurement of EPO levels in serum or plasma by enzyme linked immunosorbent assay (ELISA). The quantitative determination of EPO aids in the diagnosis of anemias and polycythemias. | For the quantitative measurement of EPO in serum or heparinized plasma. It is intended strictly for the in vitro diagnostic use as an aid in the diagnosis of anemias and polycythemias. | For the quantitative determination of EPO concentration in human serum as an adjunct in the diagnosis of anemia and polycythemias. | For the quantitative determination of erythropoietin levels in human serum. This assay is intended as an aid in the diagnosis of anemias and polycythemias. With the advent of the administration of recombinant erythropoietin as a biologic therapy to increase red blood cell mass, an erythropoietin assay may be used also to aid in the prediction and monitoring of response to recombinant erythropoietin treatment in persons with anemias. |
{3}
| Similarities | | | | | |
| --- | --- | --- | --- | --- | --- |
| Characteristic | BCI Access EPO | RDS Quantikine | DPC Immulite | Nichols Advantage | Sangui Bio EPO ELISA Kit |
| Assay type/format | 2-site simultaneous immunometric (sandwich) chemiluminescent | ELISA | 2-site sequential immunometric (sandwich) chemiluminescent | Same | 2-site simultaneous immunometric (sandwich) chemiluminescent (ELISA) |
| Composition | Paramagnetic particles coated with goat anti-mouse IgG: mouse anti-recombinant human EPO monoclonal antibody, BSA, sodium azide and ProClin 300.
Chicken anti-recumbent mouse alkaline phosphatase (bovine) conjugate, BSA, 0.1% sodium aide and 0.7% ProClin 300,
TRIS saline buffer containing BSA, proteins (chicken, bovine, mouse), <0.1% sodium azide and 0.17% ProClin 300 | Microplate - polystyrene microplate coated with mouse monoclonal antibody against recombinant human EPO.
Conjugate - anti EPO polyclonal (rabbit) antibody: horseradish peroxidase conjugate w/ thimerosal as a preservative | Solid phase - Polystyrene bead coated with an anti-ligand derived from streptavidin.
Reagent – ligand labeled murine monoclonal anti-EPO antibody with preservative. Alkaline phosphatase conjugated to goat polyclonal anti-EPO antibody in buffer, with preservatives. | Streptavidin coated magnetic particles in a buffer containing goat, rabbit and mouse gamma globulin with sodium azide and ProClin 300.
Acridinium ester-labeled mouse monoclonal antibody to human EPO in a buffered protein solution with antimicrobial agents. Biotin labeled mouse monoclonal antibody to human EPO in a buffered protein solution, with sodium azide and ProClin 300. | Carboxyl-terminal mouse monoclonal antibody & an affinity purified region-restricted amino-terminal sheep antibody. Solid phase coated with avidin. Capture antibodies are coupled with biotin. Horseradish peroxidase labeled Tag antibody. |
| Measuring/Reportable Range | 0.6 - 750 mIU/mL | | | 5-700 mIU/mL | Unknown |
| Expected Values | 2.59-18.50 mIU/mL | 3.3 – 16.6 mIU/mL | 3.7 – 29.5 mIU/mL | 5.0 – 25.1 mIU/ml | Unknown |
| Analytical Sensitivity | ≤ 0.6 mIU/mL | ~ <0.6 mIU/mL | | | Unknown |
| Standardization | WHO 2^{nd} IRP 67/343 | WHO 2^{nd} IRP 67/343 | WHO 2^{nd} IRP 67/343 | WHO 1^{st} IS 87/684 | Unknown |
| Sample Types | Serum, heparinized plasma | Serum & EDTA plasma | Serum, heparinized plasma | Serum | Serum |
| Sample Size | 85μl | | 100 μl | | Unknown |
{4}
| Differences | | | | |
| --- | --- | --- | --- | --- |
| Characteristic | BCI Access EPO | RDS Quantikine | DPC Immulite | Nichols Advantage |
| Calibrator Stability | 28 days | Each run | 2 weeks | 1 week |
| Sample Size | 85μl | ≥ 0.25 mL | | 200 μl |
| Analytical Sensitivity | ≤ 0.6 mIU/mL | | 0.24 mIU/mL | 1.2 mIU/mL |
| Measuring/ Reportable Range | 0.6 - 750 mIU/mL | 0 - 200 mIU/mL | 0 - 200 mIU/mL | |
# K. Standard/Guidance Document Referenced (if applicable):
Document for Special Controls for Erythropoietin Assay Premarket Notification [510(k)s] - April 1999
Abbreviated 510(k) Submission for In Vito Diagnostic Calibrators – February 1999
EP5A Evaluation of Precision Performance of Clinical Chemistry Device Approved Guideline, CLSI/NCCLS
EP7-P Interference Testing in Clinical Chemistry, CLSI/NCCLS
EP9-A2 Method Comparison and Bias Estimation Using Patient Samples, Approved Standard-Second Edition, CLSI/NCCLS
EP14A2 Evaluation of Matrix Effects, CLSI/NCCLS
C28-A2 How to Define and Determine Reference Intervals in the Clinical Laboratory, CLSI/CLSI/NCCLS
# L. Test Principle:
The Access EPO assay is a two-site immunoenzymatic ("sandwich") assay. A sample is added to a reaction vessel along with the paramagnetic particles coated with mouse monoclonal anti-EPO blocking reagent and the alkaline phosphatase conjugate. After incubation in a reaction vessel, materials bound to the solid phase are held in a magnetic field while unbound materials are washed away. Then, the chemiluminescent substrate Lumi-Phos 530 is added to the vessel and light generated by the reaction is measured with a luminometer. The light production is directly proportion to the concentration of EPO in the sample. The amount of analyte in the sample is determined from a stored, multi-point calibration curve.
# M. Performance Characteristics (if/when applicable):
# 1. Analytical performance:
# a. Precision/Reproducibility:
Precision testing was performed using patient samples and three levels of controls (0-10, 10-100, 100-500 mIU/mL), in duplicate, for 20 runs, completing 2 runs per day, for at least 10 day. The assay exhibits total precision of $\leq 10\%$ CV at EPO concentrations greater than $3\mathrm{mIU / mL}$ . Assay precision was tested at concentrations from approximately 9 to $475\mathrm{mIU / mL}$ . The within-run imprecision ranged from $1.8\%$ CV to $8.7\%$ CV. Total imprecision ranged from $2.6\%$ CV to $8.7\%$ CV.
{5}
b. Linearity/assay reportable range:
Dilution recovery studies were performed by diluting multiple human serum and plasma (heparin) samples at various levels with Access EPO Calibrator S0. Sample mean recovery values for all serum and plasma samples were within the range of $100 \pm 15\%$.
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
EPO reagents are stable for 28 days after opening and calibrators are stable for 90 days after opening. The calibration curve is stable for 28 days.
The measurand in the Access EPO Calibrators is traceable to the WHO Second IRP (67/343) a urine-derived form of human erythropoietin. Tractability process is based on EN ISO 17511.
d. Detection limit:
The detection limit was determined by processing a six point calibration curve, controls, and ten replicates of zero calibrator in multiple assays. The lowest detectable level of EPO distinguishable from zero (Access EPO Calibrator S0) with $95\%$ confidence is $\leq 0.6 \mathrm{mIU} / \mathrm{mL}$.
e. Analytical specificity:
The Access EPO assay was tested using spiked samples for interference of normal human blood constituents [hemoglobin, triglycerides (triolein), bilirubin and human serum albumin], heparin, and commonly encountered medications (acetaminophen, acetylsalicylic acid, ibuprofen, multi vitamins). A cross-reactivity study was also performed of compounds with similar molecular structure. The assay exhibited no significant interference or cross reactivity with these substances or compounds.
f. Assay cut-off:
Not applicable.
2. Comparison studies:
a. Method comparison with predicate device:
An internal and external study was performed with the Access EPO and RDS Quantikine ELISA assays. The internal study included apparently healthy adults and adults with elevated EPO concentration levels. The external study included normal subjects, subjects diagnosed with anemia or polycythemias, subjects on erythropoietin therapy, and procured or spiked samples. Samples
{6}
for both tests were selected to cover the predicate assay range (2.5 – 200 mIU/mL). The results are as follows:
Internal Site Study: Slope of 1.0511, intercept of –1.3595, and correlation coefficient (r) of 0.988. For this study N=103 with an EPO concentration range of approximately 3 to 182 mIU/mL.
External Site Study: Slope of 1.1216, intercept of –2.4168, and correlation coefficient (r) of 0.995. For this study N=113 with an EPO concentration range of approximately 3 to 193 mIU/mL.
b. Matrix comparison:
A comparison study was performed using matched serum and plasma (heparin) samples and three matched serum and plasma samples, each spiked with four different concentrations of erythropoietin to obtain samples with concentrations throughout the analytical range of the assay. The results as follows met the design specification.
| n | Slope
(95% Confidence Interval) | Intercept (mIU/mL)
(95% Confidence Interval) | Correlation Coefficient (r) |
| --- | --- | --- | --- |
| 80 | 1.0153 | 0.1827 | 0.998 |
3. Clinical studies:
a. Clinical Sensitivity:
Not applicable.
b. Clinical specificity:
Not applicable.
c. Other clinical supportive data (when a. and b. are not applicable):
4. Clinical cut-off:
No Applicable.
5. Expected values/Reference range:
Serum and plasma (heparin) samples were obtained from 122 apparently healthy adults in the Minneapolis/St. Paul metropolitan area in Minnesota and analyzed. The result below was based on 95% non parametric analysis of normal samples:
{7}
| n | Range of Observation (mIU/mL) | Expected Range (mIU/mL) |
| --- | --- | --- |
| 122 | 1.48 – 31.88 | 2.59 – 18.5 |
It is noted in the package insert that “the EPO concentration is in normal individuals can be affected by altitude, pregnancy and other factors” and recommended that each laboratory establish its own reference ranges to assure proper representation of specific populations.
# N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
# O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.