K052218 · Cane S.R.L. · FRN · Nov 23, 2005 · General Hospital
Device Facts
Record ID
K052218
Device Name
CRONO FIVE AND CRONO PCA 50
Applicant
Cane S.R.L.
Product Code
FRN · General Hospital
Decision Date
Nov 23, 2005
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5725
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The portable Crono Five and Crono PCA 50 infusion devices have been designed for use in subcutaneous infusion of prescribed liquid medicines. The Crono Five and Crono PCA 50 has been designed for use in subcutaneous, intravenous, epidural and intrathecal infusions of prescribed liquid medicines.
Device Story
Crono Five and Crono PCA 50 are ambulatory infusion pumps for controlled drug delivery; used in hospital or home settings by clinicians or patients. Devices accept dedicated 10, 20, or 50 ml syringes. Operation modes include continuous, bolus on request (PCA), clinician bolus, and combined. Pumps feature keypad lock-out to prevent unauthorized reprogramming and an LCD for displaying delivery status, programmed values, and remaining time. Devices provide precise drug administration; benefit patients by enabling portable, controlled therapy. Healthcare providers use output data to monitor therapy progress and ensure correct dosage delivery.
Clinical Evidence
Bench testing, EMC testing, and user testing were performed. No clinical trial data provided. Testing confirmed flow rate accuracy (+/-2% for Crono Five, +/-3% for Crono PCA 50) and occlusion pressure performance (2.2 bar +/- 0.5 bar).
Technological Characteristics
Ambulatory infusion pump; battery-powered (3V Lithium 123A). Infusion modalities: continuous, PCA, clinician bolus, combined. Features LCD interface and keypad lock-out. Dedicated syringe sizes: 10, 20, 50 ml. Flow rate accuracy +/-2-3%. Max occlusion pressure 2.2 bar +/- 0.5 bar.
Indications for Use
Indicated for subcutaneous, intravenous, epidural, and intrathecal infusion of prescribed liquid medicines in patients requiring controlled drug administration in hospital or home settings.
Regulatory Classification
Identification
An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.
{0}------------------------------------------------
NOV 2 3 2005
MEDICAL TECHNOLOGY
Canė S.r.l. società unipersonale
P.IVA - Reg. Impr. TO: 04384410017
R.Ε.Α. ΤΟ Ν. 629783
Cap. Soc. Euro 10.400 i.v.
Image /page/0/Picture/4 description: The image shows a logo for CERMET, which is a certification organization. The logo includes the text "CERMET" in bold letters, along with the phrase "SISTEMA DI GESTIONE CERTIFICATO". Below this, it indicates compliance with standards REG. N. 3506, UNI EN ISO 9001:2000, and ISO 13485:2003.
## EXHIBIT 2
CANÈ S.r.l. Via Pavia, 105/I 10090 Rivoli-Cascine Vica (Torino) Italy Tel.: ++39-011-957.48.72 Fax ++39-011-959.88.80 Contact: Mario Cané. President July 4, 2005
## 510(k) Summary
1. Identification of the Device: Proprietary - Trade Name: Crono Five and Crono PCA 50. Classification Name: 80 FRN. Common/Usual Name: Ambulatory Infusion Pump.
## 2. Equivalent legally marketed devices
This product is similar in function and design to the Crono cleared under 510(k) number K041414.
## 3. Indications for Use (intended use)
The portable Crono Five and Crono PCA 50 infusion devices have been designed for use in subcutaneous infusion of prescribed liquid medicines.
## 4. Description of the Device
Crono Five is a new, innovative ambulatory infusion pump that is very appreciated by health care professionals and patients due to its small size and technical features.
Crono Five is especially suited for controlled drug administration on patients in hospital or undergoing a therapy at home. The pump is suitable for subcutaneous, intra-venous, epidural and intrathecal infusions allowing 4 different administration modalities:
## 1. Continuous
- 2. Bolus dose upon request (PCA)
- 3. Clinician bolus (managed by clinician)
- 4. Combined (continuous + bolus upon request + clinician bolus)
Crono Five pump uses dedicated syringes which are available in two sizes: 10 and 20 ml.
## SINGERI
{1}------------------------------------------------
# SINGERT
Image /page/1/Picture/2 description: The image shows a logo or wordmark that includes a stylized cross-like symbol on the left, followed by the word "CANE" in bold, sans-serif font. The cross symbol is formed by thick, black bars arranged to create a symmetrical shape with a white space in the center. The letters in "CANE" are also constructed from thick, black bars, giving them a blocky and modern appearance. The overall design is simple, bold, and visually striking.
MEDICAL TECHNOLOGY
Canė S.r.l. società unipersonale
P.IVA - Reg. Impr. TO: 04384410017
R.Ε.Α. ΤΟ Ν. 629783
Cap. Soc. Euro 10.400 i.v.
Image /page/1/Picture/4 description: The image shows a logo for CERMET, which is a certification organization. The logo includes the text "CERMET" in a bold font, along with the phrase "SISTEMA DI GESTIONE CERTIFICATO". Below this, it states "REG. N. 3506" and "UNI EN ISO 9001:2000". The logo is contained within a rounded rectangular shape.
The pump is provided with key-pad lock-out functions in order to avoid un-authorized reprogramming of the drug administration protocol.
The liquid crystal display (LCD) can show relevant information to caregivers and patients like programmed values, amount of drug delivered, remaining delivery time etc.
Crono PCA 50 is essentially the same as the Crono Five, but has a 50 ml infusion capacity.
## 5. Safety and Effectiveness, comparison to predicate device.
The results of bench, EMC, and user testing indicates that the new device is as safe and effective as the predicate device.
| Characteristic | Crono PCA - K013822 | Crono Five<br>Crono PCA 50 |
|--------------------------|---------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------|
| Intended Use | subcutaneous, intra-venous,<br>epidural and intrathecal<br>infusions of prescribed<br>liquid medicines. | SAME |
| Physical characteristics | Crono go - K013840 | Crono Five<br>Crono PCA 50 |
| Size | 3" x 1.85" x 1.14" (77 x 48<br>x 29 mm). | SAME<br>For Crono PCA 50: 84.5 x<br>55 x 42 mm. |
| Weight | 4.0 oz (115 g) (battery<br>included). | SAME<br>For Crono PCA 50: 140 g<br>(battery included). |
| Battery | Power Source Lithium<br>battery (3V) of the 123 A<br>type. | SAME |
| Infusion per impulse | 22 μl. | Crono Five: 5 μl.<br>Crono PCA 50: 20 μl. |
| Flow rate accuracy | +/-2%. | Crono Five: +/-2%.<br>Crono PCA 50: +/-3%. |
| Max. Occlusion pressure | 4.5 bar +/-1 bar. | Crono Five: 2.2 bar +/-0.5<br>bar.<br>Crono PCA 50: 2.2 bar +/-<br>0.5 bar. |
| Capacity | 10 or 20 ml. | 10, 20 ml or 50 ml. |
## 6. Substantial Equivalence Chart
## 7. Conclusion
After analyzing both bench and user testing data, it is the conclusion of CANÈ S.r.I. that Crono pumps are as safe and effective as the predicate device, has few software differences, and has no new indications for use, thus rendering it substantially equivalent the predicated device.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the circle is an abstract symbol that resembles an eagle or a bird in flight, composed of three curved lines.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
NOV 2 3 2005
Cane S.R. L. Mr. Carlo Musso Quality Manager Via Pavia, 105/I Rivoli-Cascine Vica (Torino), ITALY 10090
Re: K052218
Trade/Device Name: CRONO FIVE AND CRONO PCA 50 Regulation Number: 21 CFR 880.5725 Regulation Name: Infusion pump Regulatory Class: II Product Code: FRN and MEA Dated: September 21, 2005 Received: September 26, 2005
Dear Mr. Musso:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
### Page 2 - Mr. Musso
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Sulta y. Mckin Omd
Chiu Lin. Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "CANE" in bold, black letters. To the left of the word is a symbol that looks like two mirrored "C" shapes, also in bold black. The letters are sans-serif and appear to be part of a logo or title. The background is plain and white, which makes the black text stand out.
VED CAL TECHNOLOGY
Canė S.r.l. società unipersonale
P.IVA - Reg. Impr. TO: 04384410017
R.Ε.Α. ΤΟ Ν. 629783
Cap. Soc. Euro 10.400 i.v.
Image /page/4/Picture/2 description: The image shows a logo with the word "CERMET" in bold letters. Below the word CERMET are the words "SISTEMA DI GESTIONE CERTIFICATO" stacked on top of each other. At the bottom of the logo are the words "REG. N. 3506" followed by "UNI EN ISO 9001:2000" and "ISO 13485:2002".
SINGERT
j) Indications for Use
Koszug 510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
Device Name: Crono Five and Crono PCA 50.
Indications For Use: the Crono Five and Crono PCA 50 has been designed for use in subcutaneous, intravenous, epidural and intrathecal infusions of prescribed liquid medicines.
Prescription Use X . (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Joon Lir. For ADW al
(Division Sign-Off) (Division Sign-Off)
Division of Anesthesiology, General Hospital, Division of Control, Dental Devices
Koszus 510(k) Number:_
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.