K052197 · W.H.P.M., Inc. · DJR · Jun 9, 2006 · Clinical Toxicology
Device Facts
Record ID
K052197
Device Name
FIRST SIGN DRUG OF ABUSE URINE SCREENING TEST
Applicant
W.H.P.M., Inc.
Product Code
DJR · Clinical Toxicology
Decision Date
Jun 9, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.3620
Device Class
Class 2
Indications for Use
The First Sign Drug of Abuse Tests are rapid, chromatographic immunoassays for the qualitative detection of drugs-of-abuse in human urine. The tests may be run singly of in combinations of up to six drugs simultaneously. The cut-off concentrations and specific analytes tested for are listed below. This assay provides only a preliminary test result. A more specific alternate chemical method must be used to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method. Clinical and professional judgment must be applied to any drug of abuse test result, particularly when preliminary positive results are obtained. For Professional Use Only.
Device Story
Rapid chromatographic immunoassay for qualitative detection of drugs of abuse in human urine; detects nortriptyline, secobarbital, MDMA, oxazepam, oxycodone, and methadone. Device used in professional clinical settings; operated by healthcare professionals. Input is human urine sample; principle of operation is immunoassay; output is visual qualitative result. Preliminary results require confirmation via GC/MS or HPLC. Assists clinicians in identifying potential drug use; supports clinical decision-making when combined with professional judgment.
Clinical Evidence
Clinical evaluation compared First Sign test results against GC/MS or HPLC. Sample size not specified. Results: Nortriptyline (97.5% pos agreement, >99% neg agreement), Secobarbital (97.4% pos, 97.6% neg), MDMA (92.5% pos, >99% neg), Oxazepam (95.7% pos, >99% neg), Methadone (93.7% pos, 97.9% neg), Oxycodone (95% pos, >99% neg).
Technological Characteristics
Rapid chromatographic immunoassay; lateral flow format. Qualitative detection based on competitive binding. No electronic components, software, or external energy source required.
Indications for Use
Indicated for professional use for the qualitative detection of drugs of abuse (nortriptyline, secobarbital, MDMA, oxazepam, oxycodone, methadone) in human urine. Provides preliminary results requiring confirmation by GC/MS or HPLC.
Regulatory Classification
Identification
A methadone test system is a device intended to measure methadone, an addictive narcotic pain-relieving drug, in serum and urine. Measurements obtained by this device are used in the diagnosis and treatment of methadone use or overdose and to determine compliance with regulations in methadone maintenance treatment.
Special Controls
*Classification.* Class II (special controls). A methadone test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).
Predicate Devices
ACON BZO One Step Benzodiazepine Test Strip (K012300)
{0}------------------------------------------------
K052/97
JUN - 9 2006
# 510(k) Summary
Date of Summary: 30 May, 2005
#### Product Name
First Sign™ Drug of Abuse Urine Screening Tests
#### Sponsor and Manufacturer
WHPM, Inc. 9662 Telstar Avenue El Monte, CA 91731
WHPM, Bioresearch and Technology Co. Ltd. 806 Taihong Mansion, No 44 Chongwai Street Chongwai District Beijing, China 100062
Product will be manufactured in both locations and distributed through the California site.
#### Correspondent
Fran White, President MDC Associates, LLC 163 Cabot Street Beverly, MA 01915
# Substantial Equivalency
The First Sign™ Drug of Abuse Urine Screening Test is substantially equivalent to other tests currently on the market.
| First Sign Test Analyte | Predicate Device Name | Predicate Device 510(k) # |
|-------------------------|---------------------------------------------|---------------------------|
| Oxazepam | ACON BZO One Step Benzodiazepine Test Strip | K012300 |
| Oxycodone | ACON OXY One Step Oxycodone Test Strip | K043507 |
| Secobarbital | ACON BAR One Step Barbiturates Test Strip | K050593 |
| Methadone | ACON MTD One Step Methadone test Strip | K012595 |
| Nortriptyline | ACON TCA One Step Tricyclic Antidepressants | K021526 |
| MDMA | ACON MDMA One Step Ecstasy Test Strip | K022589 |
#### Product Description
The First Sign Drug of Abuse Tests are rapid, chromatographic immunoassays for the qualitative detection of drugs-of-abuse in human urine.
# Indications for Use
The First Sign Drug of Abuse Tests are rapid, chromatographic immunoassays for the qualitative detection of drugs-of-abuse in human urine. The tests may be run singly or in combinations of up to six drugs simultaneously. The cut-off concentrations for these drugs are as follows: Nortriptyline 1,000ng/mL, Secobarbital 300ng/mL, MDMA 500ng/mL, Oxazepam 300ng/mL, Methadone 300ng/mL, Oxycodone 100ng/mL. For Professional Use Only.
{1}------------------------------------------------
# Performance Characteristics
A clinical evaluation compared test results between the First Sign Drugs of Abuse Tests and GC/MS or HPLC results. The results are summarized below.
| First Sign Test | Positive Agreement | Negative Agreement | Overall Agreement |
|-----------------|--------------------|--------------------|-------------------|
| Nortriptyline | 97.5% | >99% | 98.7% |
| Secobarbital | 97.4% | 97.6% | 97.5% |
| MDMA | 92.5% | >99% | 96.2% |
| Oxazepam | 95.7% | >99% | 97.5% |
| Methadone | 93.7% | 97.9% | 96.2% |
| Oxycodone | 95% | >99% | 97.5% |
# Conclusion
Clinical studies demonstrate the substantial equivalence between the First Sign Drugs of Abuse Tests [Nortriptyline, Secobarbital, MDMA, Oxazepam, Methadone, and Oxycodone] and commercially available FDA-cleared drugs of abuse tests. The studies also demonstrated that the First Sign tests are safe and effective in detecting drugs of abuse at or above their stated cut-off concentrations.
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its body and wing. The eagle is positioned to the right of a circular emblem that contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter.
Public Health Service
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
JUN - 9 2006
WHPM, Inc c/o Ms. Fran White President MDC Associates, LLC. 163 Cabot Street Beverly, MA 01915
Re: k052197
> Trade/Device Name: First Sign Drug of Abuse Urine Screening Test Regulation Number: 21 CFR 862.3620 Regulation Name: Methadone test system Regulatory Class: Class II Product Code: DJR, LFG, DIS, LAF, JXM, DJG Dated: May 11, 2006 Received: May 12, 2006
Dear Ms. White:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
{3}------------------------------------------------
Page 2 --
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Alberto Gutierrez, Ph.D.
Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
#### 510(k) Number (if known): K052197
Device Name: First Sign Drug of Abuse Urine Screening Test
Indications For Use:
The First Sign Drug of Abuse Tests are rapid, chromatographic immunoassays for the qualitative detection of drugs-of-abuse in human urine. The tests may be run singly of in combinations of up to six drugs simultaneously. The cut-off concentrations and specific analytes tested for are listed below.
This assay provides only a preliminary test result. A more specific alternate chemical method must be used to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method. Clinical and professional judgment must be applied to any drug of abuse test result, particularly when preliminary positive results are obtained.
For Professional Use Only.
| Test Name | Analyte Tested | Cut-off Concentration |
|--------------------------|----------------|-----------------------|
| First Sign Nortriptyline | nortriptyline | 1,000ng/mL |
| First Sign Secobarbital | secobarbital | 300ng/mL |
| First Sign MDMA | mdma | 500ng/mL |
| First Sign Oxazepam | oxazepam | 300ng/mL |
| First Sign Oxycodone | oxycodone | 100ng/mL |
| First Sign Methadone | methadone | 300ng/mL |
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CORH. Office of In Vitro Diagnostic Devices (OIVD)
Division Sign-Off
Office of In Vitro Diagnostic Device Evaluation and Safety
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.