K052183 · Soring GmbH Medizintechnik · QPB · Oct 7, 2005 · General, Plastic Surgery
Device Facts
Record ID
K052183
Device Name
SORING GMBH SONOCA LIPO
Applicant
Soring GmbH Medizintechnik
Product Code
QPB · General, Plastic Surgery
Decision Date
Oct 7, 2005
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.5040
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The SONOCA Lipo is an instrument indicated for selected ultrasound dissection, liquefaction, emulsifying and aspiration of soft tissues in General Surgery, Plastic and Reconstructive Surgery, and Gynecological Surgery applications. It is also indicated for the liquefaction and aspiration of localized subcutaneous fatty tissue for aesthetic body contouring. Typical users of this system are trained medical professionals.
Device Story
SONOCA Lipo is an ultrasonic surgical system for soft tissue dissection and lipoplasty. Device consists of ultrasonic generator, infiltration pump, and aspiration pump; trolley-mounted or desktop configuration. Operates by transmitting power from a longitudinal vibrating probe tip to tissue; probe integrates irrigation and aspiration to collect cell/tissue fragments. Used by trained medical professionals in surgical settings. Provides selective tissue dissection based on ultrasonically induced cutting effect, targeting parenchyma-rich organs while sparing sustentacular tissue. Acts as a supplementary tool to conventional surgery for body contouring and soft tissue management.
Clinical Evidence
Bench testing only. Device tested for compliance with UL 2601-1, EN 60601-1, EN 60601-1-2, IEC 950, CISPR 22, IEC 801-2/3, IEEE 1003, IEC 601-1, IEC 601-2-2, and DIN EN 61847.
Technological Characteristics
Ultrasonic surgical system; includes ultrasonic generator, infiltration pump, and aspiration pump. Features longitudinal vibrating probe tip for tissue interaction. Trolley-mounted or desktop form factor. Complies with electrical safety standards (IEC 601-1, IEC 601-2-2) and electromagnetic compatibility standards (EN 60601-1-2, CISPR 22).
Indications for Use
Indicated for soft tissue dissection, liquefaction, emulsification, and aspiration in general, plastic, reconstructive, and gynecological surgery; also indicated for liquefaction and aspiration of subcutaneous fatty tissue for aesthetic body contouring in trained medical professionals.
Regulatory Classification
Identification
A suction lipoplasty system is a device intended for aesthetic body contouring. The device consists of a powered suction pump (containing a microbial filter on the exhaust and a microbial in-line filter in the connecting tubing between the collection bottle and the safety trap), collection bottle, cannula, and connecting tube. The microbial filters, tubing, collection bottle, and cannula must be capable of being changed between patients. The powered suction pump has a motor with a minimum of 1/3 horsepower, a variable vacuum range from 0 to 29.9 inches of mercury, vacuum control valves to regulate the vacuum with accompanying vacuum gauges, a single or double rotary vane (with or without oil), a single or double diaphragm, a single or double piston, and a safety trap.
Special Controls
*Classification.* Class II (special controls). Consensus standards and labeling restrictions.
MISONIX INC. LYSONIX 2000/3000 ULTRASONIC SURGICA (K041058)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Soring Gmbh Medizintechnik Carl Alletto Submission Correspondent 1600 Manchester Way Corinth, Texas 76210
June 8, 2021
Re: K052183
Trade/Device Name: Soring Gmbh Sonoca Lipo Regulation Number: 21 CFR 878.5040 Regulation Name: Suction lipoplasty system Regulatory Class: Class II Product Code: QPB
Dear Carl Alletto:
The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated October 7, 2005. Specifically, FDA is updating this SE Letter because FDA has created a new product code to better categorize your device technology.
Please note that the 510/k) submission was not re-reviewed. For questions regarding this letter please contact Cindy Chowdhury, OHT4: Office of Surgical and Infection Control Devices, 240-402-6647, Cindy.Chowdhury@fda.hhs.gov.
Sincerely,
Cindy Chowdhury -S
Cindy Chowdhury, Ph.D., M.B.A. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, representing medicine and healing. The text "DEPARTMENT OF HEALTH AND HUMAN SERVICES, USA" is arranged in a circular pattern around the caduceus symbol. The logo is black and white.
Public Health Service
OCT 7 - 2005
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Soring GmbH Medizintechnik c/o Mr. Carl Alletto 1600 Manchester Way Corinth, Texas 76210
Re: K052183
Trade/Device Name: Soring, SONOCA-Lipo Regulation Number: 21 CFR 878.5040 Regulation Name: Suction lipoplasty system Regulatory Class: II Product Code: MUU Dated: August 7, 2005 Received: August 15, 2005
Dear Mr. Alletto:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{2}------------------------------------------------
## Page 2-Mr. Carl Alletto
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours
Barbara Inelund
for
Mark N. Molkerson
Mark N. Melkerson Acting Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
#### Indications for Use
510(k) Number (if known):
Device Name: Soring, SONOCA-Lipo
Indications for Use:
The SONOCA Lipo is an instrument indicated for selected ultrasound dissection, liquefaction, emulsifying and aspiration of soft tissues in General Surgery, Plastic and Reconstructive Surgery, and Gynecological Surgery applications. It is also indicated for the liquefaction and aspiration of localized subcutaneous fatty tissue for aesthetic body contouring.
Typical users of this system are trained medical professionals.
Prescription Use VV (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
(Posted November 13, 2003)
Oevbara Bacheupformm
(Division Sign Off)
Division of General, Restorative, and Neurological Devices
**510(k) Number** K052183
{4}------------------------------------------------
0C1 7 - 2005
K052183 Page 1/2
## 510(k) Summary of Safety & Effectiveness
(as required by 21 CFR 807.92c)
Date Prepared:
August 7, 2005
#### Submitter's Information:
Soring GmbH Medizintechnik Justus-v.Liebig 10 25451 Quickborn Germany
Telephone: 49 4106-5055 Fax: 49 4106-5271 Email: info@soering.com
#### Trade Name, Common Name, Classification:
| Trade name: | Soring GmbH, Sonoca™-Lipo |
|----------------------|---------------------------------|
| Common name: | Instrument, Ultrasonic Surgical |
| Classification name: | General and Plastic Surgery |
## Predicate Device:
| DEVICE CLASSIFICATION<br>NAME | PUMP, PORTABLE,<br>ASPIRATION (MANUAL OR<br>POWERED) | SYSTEM, SUCTION,<br>LIPOPLASTY |
|-------------------------------|------------------------------------------------------|---------------------------------------------------------|
| REGULATION NUMBER | 878.4780 | 878.5040 |
| 510(K) NUMBER | K992026 | K041058 |
| DEVICE NAME | SORING GMBH, SONOCA<br>300 | MISONIX INC. LYSONIX<br>2000/3000 ULTRASONIC<br>SURGICA |
| APPLICANT | SORING GMBH | MISONIX, INC. |
| PRODUCT CODE | BTA AND LFL | MUU |
| DECISION DATE | 09/23/1999 | 05/17/2004 |
#### Device Description:
SONOCA Lipo™ is a modified version of the SONOCA 300 cleared by FDA under K992026 and can be used in two ways:
- 1. As an ultrasonic dissector / aspirator as in the SONOCA 300 to cut, irrigate, and suction at the surgical site and
- 2. As an alternative to conventional Lipectomy because it is a combination of a standard Lipectomy suction device and an Ultrasound assisted Lipectomy for the selective dissection of human fatty tissue.
{5}------------------------------------------------
K052183 page 2/2
# 510(k) Summary of Safety & Effectiveness
(as required by 21 CFR 807.92c)
The SONOCA Lipo™ is not an alternative to conventional surgery but a supplementary tool which provides an advantage for the selective dissection of human tissue.
The advantage of the SONOCA is that the tissue selectivity of the ultrasonically induced cutting effect limits the spectrum to organs with a large proportion of parenchyma in which sustentacular tissue is only minimally developed
During the use of an ultrasonic dissector, power is transmitted from a longitudinal vibrating probe tip in the contact zone to tissue. The probe with the integrated aspiration/irrigation function collects the cell and tissue fragments.
The SONOCA Lipo™ can be trolley mounted or a desktop unit consisting of an ultrasonic generator, an infiltration pump and an aspiration pump. The nodular build of the electronic components allow the unit to function at a high capacity, totally free of electronic maintenance.
## Indications for Use:
The SONOCA Lipo™ is an instrument indicated for selected ultrasound dissection, liquefaction, emulsifying and aspiration of soft tissues in General Surgery, Plastic and Reconstructive Surgery, and Gynecological Surgery applications. It is also indicated for the liquefaction and aspiration of localized subcutaneous fatty tissue for aesthetic body contouring. Typical users of this system are trained medical professionals.
## Performance Data:
The subject and predicate devices have been designed and tested to pass the following Voluntary Standards: UL 260 1-1 Medical Electrical Equipment, Part 1: General Requirements for Safety EN 60601-1 Medical Electrical Equipment, Part 1: General Requirements for Safety EN 60601-1-2:1993 Electromagnetic Compatibility and FGC Part 18 EMC Requirement. The subject device complies with IEC 950 - Safety of Information Technology Equipment, CISPR 22, class A -- Electromagnetic Compatibility, IEC-801-2, IEC-801-3 - Electromagnetic Compatibility, IEEE 1003 . - General Pelectrical Safety for medical devices, IEC 601-1 -Electrical Safety for medical devices using RFpower, IEC 601-2-2- Ultrasonic surgical devices, DIN EN 61847
## Conclusion:
Similar to the predicate devices, the SONOCA Lipo™ does not control any life sustaining functions or services. The new device and the predicate devices share the same conformance to performance standards and both function as Ultrasonic Dissectors. Based on the information supplied in this 510(k), we conclude that the subject device is safe, effective, and substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.