INNOVA CT OPTION FOR DIGITAL FLUOROSCOPIC IMAGING SYSTEMS INNOVA MODELS 4100, 4100IQ, 3100, 3100IQ & 2100IQ
K052157 · Ge Medical Systems, LLC · JAK · Aug 25, 2005 · Radiology
Device Facts
Record ID
K052157
Device Name
INNOVA CT OPTION FOR DIGITAL FLUOROSCOPIC IMAGING SYSTEMS INNOVA MODELS 4100, 4100IQ, 3100, 3100IQ & 2100IQ
Applicant
Ge Medical Systems, LLC
Product Code
JAK · Radiology
Decision Date
Aug 25, 2005
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1750
Device Class
Class 2
Indications for Use
The Innova systems are indicated for use in generating fluoroscopic images of human anatomy for vascular angiography, diagnostic and interventional procedures, and optionally, rotational imaging procedures. They are also indicated for generating fluoroscopic images of human anatomy for cardiology, diagnostic, and interventional procedures. They are intended to replace fluoroscopic images obtained through image intensifier technology. Those devices are not intended for mammography applications. Innova CT is a software option, which reconstructs 3D volumes from Rotational Fluoroscopy acquisition to provide images that assist the physician in diagnosis, surgical planning, interventional procedures and treatment follow-up. Innova CT is intended for imaging bone and soft tissues as well as other internal body structures. Innova CT is not intended for mammography applications.
Device Story
Innova CT is a software option for GE Digital Fluoroscopic Imaging Systems (Innova 4100/3100/2100 series). It takes rotational fluoroscopy acquisition data as input; reconstructs 3D volumes of bone, soft tissue, and internal structures. Used in clinical settings (OR/interventional suites) by physicians to assist in diagnosis, surgical planning, interventional procedures, and treatment follow-up. Output is 3D volumetric images viewed by clinicians to guide clinical decision-making and procedural navigation. System hardware includes monoplane positioner, vascular/cardiac table, X-ray source, and amorphous silicon/cesium iodine digital detector.
Clinical Evidence
Sample clinical data included in submission. Bench testing performed via FMEA risk management process. No specific sensitivity/specificity metrics provided.
Technological Characteristics
Amorphous silicon detector with cesium iodine scintillator. Energy source: 480 VAC 50/60Hz. Compliance: UL 187, IEC 60601-1, UL 2601. Software-based 3D reconstruction from rotational fluoroscopy.
Indications for Use
Indicated for patients requiring fluoroscopic imaging for vascular, cardiac, diagnostic, and interventional procedures. Includes rotational imaging for 3D volume reconstruction of bone, soft tissue, and internal structures. Not for mammography.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Predicate Devices
DynaCT option of Siemens Vascular Angiographic system (K042646)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" intertwined and enclosed within a circle. The logo is in black and white.
.....
: 上一篇:
.
. . . .
:
AUG 25 2005
Koszis7
GE Healthcare
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
P.O. Box 414, W-400 Milwaukee, WI 53201 USA
:
. .
1. 1. 1. 1.
### Executive Summary
## 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS
This 510(k) summary of safety and effectiveness information is submitted in accordance with the requirements of 21 CFR Part 807.87(h).
| <b>Identification of Submitter:</b> | Larry A. Kroger, Ph.D.<br>Senior Regulatory Programs Manager<br>GE Healthcare<br>Tel. (262) 544-3894<br>Summary prepared: July 22, 2005 |
|-------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <b>Identification of Product:</b> | Digital Fluoroscopic Imaging System |
| <b>Device Trade Name:</b> | Innova CT option for Digital Fluoroscopic Imaging Systems<br>Innova 4100, Innova 4100 <sup>IQ</sup> , Innova 3100, Innova 3100 <sup>IQ</sup> , Innova 2100 <sup>IQ</sup> |
| <b>Classification Name:</b> | Fluoroscopic X-ray System |
| <b>Manufacturer:</b> | GE Medical Systems SCS.<br>283, rue de la Minière<br>78530 Buc Cedex, France |
| <b>Distributed by:</b> | GE Medical Systems, LLC, Milwaukee, WI |
| <b>Marketed Devices:</b> | The GE Healthcare Innova CT option is substantially equivalent<br>to the currently marketed DynaCT option of Siemens Vascular<br>Angiographic system (K042646). The Innova CT is intended for<br>use with Innova 4100 (renamed Innova 4100 <sup>IQ</sup> and previously<br>cleared under K033244), Innova 3100 (renamed Innova 3100 <sup>IQ</sup><br>and previously cleared under K031637), and Innova 2100 <sup>IQ</sup> (or<br>2100-IQ previously cleared under K050489). This opinion is<br>based on the information included in this premarket notification. |
| <b>Device Description:</b> | The Innova CT imaging is offered as an option for Innova 4100<br>Innova 3100, Innova 2100 <sup>IQ</sup> (2100-IQ), 4100 <sup>IQ</sup> and Innova 3100 <sup>IQ</sup> .<br>The Digital Fluoroscopic Imaging Systems are designed to<br>perform fluoroscopic x-ray examinations. The detector is<br>comprised of amorphous silicon with a cesium iodine scintillator.<br>The resulting digital image can be sent through a Fiber Channel<br>link to an acquisition system then to network (in using DICOM) |
| Innova CT_SumSE revA.doc<br>1/3 | |
、サイズ、スタート、ブラック、アイテム、スタ
:
and the was and the states
{1}------------------------------------------------
for applications such as post-processing, printing, viewing and archiving. Digital Fluoroscopic Imaging System consists of a monoplane positioner, a vascular or cardiac table, an X-RAY system and a digital detector.
All construction and materials are compliant with UL 187 and IEC Materials: 60601-1 for the existing parts of the product and with UL 2601 and IEC 60601-1 for the new parts.
The design is validated through Failures Modes Effects Analysis Design: (FMEA) process, which allows managing the risks.
Energy Source: 480 VAC 50/60Hz.
Indications for Use: For Innova 4100, Innova 410019, Innova 3100, Innova 310010, Innova 2100'0 with Innova 3T option:
> The Innova systems are indicated for use in generating fluoroscopic images of human anatomy for vascular angiography, diagnostic and interventional procedures, and optionally, rotational imaging procedures. They are also indicated for generating fluoroscopic images of human anatomy for cardiology, diagnostic, and interventional procedures.
> They are intended to replace fluoroscopic images obtained through image intensifier technology. Those devices are not intended for mammography applications.
> Innova CT is a software option, which reconstructs 3D volumes from Rotational Fluoroscopy acquisition to provide images that assist the physician in diagnosis, surgical planning, interventional procedures and treatment follow-up.
> Innova CT is intended for imaging bone and soft tissues as well as other internal body structures.
Innova CT is not intended for mammography applications
#### Comparison with
The GE Healthcare option Innova CT is substantially equivalent to the currently marketed option DynaCT of Siemens Vascular Angiographic system cleared under K042646 .
- Dyna CT indications for use: DynaCT is an x-ray imaging software option, which allows the reconstruction of twodimensional images acquired with a standard angiographic Carm device into a three-dimensional image format.
:
Innova CT_SumSE revA.doc 2/3
{2}------------------------------------------------
DynaCT is intended for imaging both hard and soft tissues as well as other internal body structures for diagnosis, surgical planning, interventional procedures and treatment follow-up.
This opinion is based on the information contained in the comparison table and the product data sheets.
Summary of the Studies:
Sample clinical data for the Innova CT option are included in this submission.
Conclusions:
GE Healthcare considers that the Innova CT option for Digital Fluoroscopic Imaging Systems Innova 2100, Innova 4100, Innova 4100",
Innova 3100, Innova 3100 " and Innova 2100 " to be equivalent with the DynaCT option The potential hazards, related to the use of Innova CT option are controlled by a risk management plan including:
- A hazard identification .
- A risk evaluation ●
- A Software Development and Validation Process
Innova CT_SumSE revA.doc ર્સ્ડ
2017 2017
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol consisting of three curved lines that resemble a human figure.
AUG 25 2005
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. Larry A. Kroger, Ph.D. Senior Regulatory Programs Manager GE Healthcare P.O. Box 414, W-400 MILWAUKEE WI 53201
#### Re: K052157
Trade/Device Name: Innova CT option for Digital Fluoroscopic Imaging Systems, Innova 4100, Innova 410010, Innova 3100, Innova 3100 10, Innova 210010 Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: II
Product Code: IZI, JAA, and LLZ Dated: August 5, 2005 Received: August 8, 2005
Dear Mr. Kroger:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) This letter will anow you to begin maning of substantial equivalence of your device to a legally prematication: "The PDF maing or our device and thus, permits your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific access your of the following numbers, based on the regulation number at the top of this letter:
| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|-----------------|----------------------------------|--------------|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology) | 240-276-0120 |
| Other | 1 | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Also, prease note the regulation on your responsibilities under the Act from the 807.77). I ou may ovain one generational and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address
http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# C. Indications for Use
510(k) Number (if known): x 052157
Device Name: Innova CT option for Digital Fluoroscopic Imaging Systems Innova 4100, Innova 410010, Innova 3100, Innova 310019, Innova 210010
Indications for Use:
- Indications for USe.
The Innova systems are indicated for use in generating fluoroscopic images of The Innova systems are indicated for doo in genostic and interventional procedures, human analorny for vascular anglography, diagnosticated for generating and optionally, fotational maging procedures. They ar disgnostic, and interventional
fluoroscopic images of human anatomy for cardiology, diagnos abtained through fluoroscopic Infages of numan anatonly for sarensisy).
procedures. They are intended to replace fluoroscopic images obtained through procedures: They are intended is topice is not intended for mammography applications.
- applications.
Innova CT is a software option which reconstructs 3D volumes from in diagnosi ପ innova CT IS a soltware option which roosnet assist the physician in diagnosis,
Fluoroscopy acquisition to provide images that assist the physician wa Fluoroscopy acquisition to procedures and treatment follow-up.
- surgical planning, interventional procession in the many as well as other internal ロ body structures.
- body Structures:
Innova CT is not intended for mammography applications ם
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use _ (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF
NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Nancyl
(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number K052157
AttC_Indications_for_use_revA.doc
1 / 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.