← Product Code [KRZ](/productcode/KRZ) · K052101

# C.F.A.S. (CALIBRATOR FOR AUTOMATED SYSTEMS) HBA1C (K052101)

_Roche Diagnostics · KRZ · Aug 26, 2005 · Hematology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K052101

## Device Facts

- **Applicant:** Roche Diagnostics
- **Product Code:** [KRZ](/productcode/KRZ.md)
- **Decision Date:** Aug 26, 2005
- **Decision:** SESE
- **Submission Type:** Abbreviated
- **Regulation:** 21 CFR 864.8165
- **Device Class:** Class 2
- **Review Panel:** Hematology

## Indications for Use

The Roche Diagnostics C.f.a.s. (Calibrator for automated systems) HbA1c is for use in the calibration of quantitative Roche methods on Roche clinical chemistry analyzers as specified in the enclosed value sheet. For Prescription use only.

## Device Story

C.f.a.s. HbA1c is a single-level lyophilized calibrator composed of hemolyzed sheep and human blood; used to calibrate quantitative Roche HbA1c methods on Roche clinical chemistry analyzers. The product is reconstituted with 2.0 mL of distilled or deionized water by laboratory personnel. Once loaded onto the analyzer, the system automatically performs dilutions to generate six calibration levels. The resulting calibration curve allows the analyzer to accurately quantify HbA1c levels in patient samples. This process ensures the accuracy of clinical chemistry measurements, supporting healthcare providers in the diagnosis and monitoring of glycemic control in patients.

## Clinical Evidence

No clinical data. Evidence consists of analytical performance and stability testing. Traceability established to CRM 522 (Hb) and IFCC reference method (HbA1c) via multi-laboratory value assignment. Real-time shelf-life stability (up to 30 months) and accelerated stability (5 days at 35°C) confirmed recovery within ±10% of assigned values. Reconstituted stability confirmed for 8 hours at 30°C, 2 days at 4°C, and 3 months at -20°C.

## Technological Characteristics

Lyophilized calibrator; matrix: hemolyzed sheep blood with human blood additives and 0.9% TTAB. Reconstituted with distilled/deionized water. Traceable to CRM 522 and IFCC reference materials. Barcode-encoded for automated analyzer calibration. Stability: 30 months shelf-life at 2-8°C.

## Regulatory Identification

A calibrator for hemoglobin or hematocrit measurement is a device that approximates whole blood, red blood cells, or a hemoglobin derivative and that is used to set instruments intended to measure hemoglobin, the hematocrit, or both. It is a material whose characteristics have been precisely and accurately determined.

## Special Controls

*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 864.9.

## Predicate Devices

- Tina-Quant HbA1c reagent kit calibrator ([K934070](/device/K934070.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

1

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE

A. 510(k) Number:
k052101

B. Purpose for Submission:
New device

C. Measurand:
Calibrator for Glycosylated hemoglobin (HbA1c)

D. Type of Test:
Not applicable - calibrator

E. Applicant:
Roche Diagnostics Corp.

F. Proprietary and Established Names:
C.f.a.s (Calibrator for Automated Systems) HbA1c

G. Regulatory Information:

1. Regulation section:
21 CFR §864.8165, Calibrator for hemoglobin or hematocrit measurement

2. Classification:
Class II

3. Product code:
KRZ

4. Panel:
81 (Hematology)

{1}

H. Intended Use:

1. Intended use(s):
See Indications for use below.

2. Indication(s) for use:
The Roche Diagnostics C.f.a.s. (Calibrator for automated systems) HbA1c is for use in the calibration of quantitative Roche methods on Roche clinical chemistry analyzers as specified in the enclosed value sheet.

3. Special conditions for use statement(s):
For Prescription use only.

4. Special instrument requirements:
Roche Cobas Integra 800

I. Device Description:

The Roche Diagnostic C.f.a.s HbA1c calibrator kit contains three vials of HbA1c calibrator. The vials contain freeze-dried (lyophilized) calibrator based on hemolyzed sheep blood. The concentration of the calibrator components have been adjusted to ensure optimal calibration of the appropriate Roche methods on clinical chemistry analyzers.

Reactive components

Hemolyzed sheep blood, with chemical additives and material of biological origin as specified. The origin of the biological additives is as follows:

Analyte
Hemoglobin
HbA1c

Origin
Sheep blood
Human blood

Non-reactive components

Preservatives and stabilizers

The concentrations and activities of the calibrator components are lot-specific and specified in the labeling. The values are also encoded in the calibrator barcode sheets for COBAS INTEGRA analyzers. The calibrators are designed to allow calibration to either the American NGSP system or the International IFCC system. The product has been tested and found negative by FDA accepted methods for HBV, HCV and HIV 1 and 2.

{2}

J. Substantial Equivalence Information:

1. Predicate device name(s):
Tina-Quant HbA1c reagent kit

2. Predicate 510(k) number(s):
k934070

3. Comparison with predicate:
The table below indicates the similarities and differences between the modified C.f.a.s. HbA1c and its predicate device Tina-Quant HbA1c test system (k934070)

|  Similarities  |   |   |
| --- | --- | --- |
|  Characteristic | New Device Roche C.f.a.s HbA1c (k052101) | Predicate Device Tina-Quant HbA1c (k934070)  |
|  Intended Use | C.f.a.s. HbA1c is for the use in the calibration of quantitative Roche methods on Roche clinical chemistry analyzers as specified in the enclosed value sheet | Calibrator included in kit for calibration of the quantitative Tina-Quant HbA1c method.  |
|  Format | Lyophilized | Lyophilized  |
|  Matrix and Composition | Hemolysate derived from human and sheep blood; 0.9% TTAB (tetradecyl trimethylammonium bromide) | Hemolysate derived from human and sheep blood; 0.9% TTAB (tetradecyl trimethylammonium bromide)  |
|  Differences  |   |   |
| --- | --- | --- |
|  Item | New Device Roche C.f.a.s HbA1c (k052101) | Predicate Device Tina-Quant HbA1c (k934070)  |
|  Handling | Reconstituted with 2.0 mL distilled or deionized water. | Reconstituted with 1.0 mL distilled or deionized water.  |
|  Levels | Four levels | Single level  |
|  Stability | Unopened: stable up to the expiration date
Reconstituted: 2 days @ 2-8° C
8 hours @ 15-25° C
3 months @ (-15 to -25) ° C | Unopened: stable up to the expiration date
Reconstituted: 2 days @ 2-8° C
8 hours @ 20-25° C
3 months @ 20 ° C  |

{3}

4

K. Standard/Guidance Document Referenced (if applicable):

FDA Guidance for Industry “Abbreviated 510(k) submissions for In Vitro Diagnostic Calibrators; Final”.

Deleted:
Formatted: Indent: Left: 18 pt
Deleted: See Traceability, Stability, Expected values (controls, calibrators, or methods) below.

L. Test Principle:
Not applicable.

M. Performance Characteristics (if/when applicable):

1. Analytical performance:

a. Precision/Reproducibility:
Not applicable.

b. Linearity/assay reportable range:
Not applicable.

c. Traceability, Stability, Expected values (controls, calibrators, or methods):

Traceability: The sponsor states that the absolute Hemoglobin (Hb) values for the C.f.a.s. HbA1c are traceable to the CRM 522, through the use of the cyanomethemoglobin (CNMetHb) reference method. For more information see the following references:

(1) International Committee for standardization in hematology expert panel on hemoglobinometry. Recommendations for reference method for hemoglobinometry in human blood and specifications for international haemoglobin cyanide reference preparation. Clinical &amp; Laboratory Hematology, 9, 78-79, 1987.
(2) Preparation and certification of a reference material of haemoglobin cyanide for standardization of blood hemoglobin measurement CRM 522; Report EUR 16101 EN European Commission bcr information reference material 1995.

The sponsor also states, that the absolute HbA1c values are traceable to the IFCC reference method through the use of the IFCC calibrator set “Los Angeles 2004”, Levels 1-8.

Value assignment – Hb values: The value assignment process for the Hb values was performed in five laboratories. Each laboratory conducted three independent runs where 30 whole blood samples were measured:

{4}

- in duplicate using CNMetHb reference method and
- in singleton using a commercially available method

Value assignment and transfer process - HbA1c values: Assigned values were verified using a commercially available assay against the IFCC Calibrator set (Levels 1-8) in five laboratories, with each lab performing three independent runs on Integra 800 analyzers.

Real-Time Shelf life stability was established by taking a set of stored calibrators (2-8°C) and assaying them after 12, 24, and 30 months. The acceptance criterion was established at ±10% of the initial assigned value.

Accelerated stability was established by taking C.f.a.s. HbA1c calibrator that was stored at 2-8°C and then stored for 5 days at 35°C. Five replicates of this on-test material was analyzed and compared to a reference material (fresh C.f.a.s. HbA1c). The average of the five replicates was calculated as a percentage of the reference value. The acceptance criterion is recovery of ±10% of the reference value.

Stability of reconstituted calibrator was established by reconstituting the C.f.a.s. calibrator according to package insert directions and then store a 30°C for 8 hours; at 4°C for 2 days; and at -20°C for 3 months. Five replicates of each on-test material were analyzed and compared to a reference material (freshly reconstituted C.f.a.s. HbA1c). The average of the five replicates was calculated as a percentage of the reference value. The acceptance criterion is recovery of ±10% of the reference value.

d. Detection limit:
Not applicable.

e. Analytical specificity:
Not applicable.

f. Assay cut-off:
Not applicable.

2. Comparison studies:

a. Method comparison with predicate device:
Not applicable.

{5}

b. Matrix comparison:
Not applicable.

3. Clinical studies:
a. Clinical Sensitivity:
Not applicable.
b. Clinical specificity:
Not applicable.
c. Other clinical supportive data (when a. and b. are not applicable):
Not applicable.

4. Clinical cut-off:
Not applicable.

5. Expected values/Reference range:
Not applicable.

N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.

O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

---

**Source:** [https://fda-staging.innolitics.com/device/K052101](https://fda-staging.innolitics.com/device/K052101)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
