K052077 · Esaote, S.p.A. · MOS · Aug 16, 2005 · Radiology
Device Facts
Record ID
K052077
Device Name
G-SCAN CERVICAL SPINE COIL
Applicant
Esaote, S.p.A.
Product Code
MOS · Radiology
Decision Date
Aug 16, 2005
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.1000
Device Class
Class 2
Indications for Use
The G-scan Cervical Spine Coil is to be used in MR imaging of the cervical section of the spine column. G-scan is a Magnetic Resonance (MR) system that produces transversal, sagittal and coronal and oblique cross-section images of the limbs, joints and spinal column. It is intended for imaging portions of the upper limb, including the hand, wrist, forearm, elbow, arm and shoulder, imaging portions of the lower limb, including the foot, ankle, calf, knee, thigh and imaging portions of the spinal column, including the cervical, thoracic and lumbo-sacral sections. G-scan images correspond to the spatial distribution of protons (hydrogen nuclei) that determine magnetic resonance properties and are dependent on the MR parameters, including spin-latice relaxation time (T1), spin-spin relaxation time (T2), nuclei density, flow velocity and "chemical shift". When interpreted by a medical expert trained in the use of MR equipment, the images can provide diagnostically useful information.
Device Story
The G-scan Cervical Spine Coil is a solenoidal linear receiving coil accessory for the G-scan MRI system. It captures RF signals from the cervical spine to improve Signal-to-Noise Ratio (SNR) during imaging. The device features a mechanical lever-controlled head height adjustment and an integrated cushion for patient positioning. Operation involves inserting the patient's neck into the coil, which opens via non-magnetic contacts. The coil connects to the patient table for electrical wiring and automatic model recognition. During RF transmission, passive PIN diode circuits decouple the receiving coil to prevent parasitic currents. The resulting MR images are interpreted by trained medical experts to assist in clinical diagnosis of cervical spine conditions.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Solenoidal linear receiving coil; 3 copper tubes with non-magnetic contacts; 3 tuning capacitor groups; passive PIN diode decoupling circuit; dimensions: 26.4 x 32.1 x 36.6 cm (external), 17.5 x 15.1 x 5.0 cm (internal); connects to G-scan MRI system for signal acquisition.
Indications for Use
Indicated for MR imaging of the cervical spine column in patients requiring diagnostic imaging of the cervical vertebrae.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
K052077
510(k) Sammary (1-sean Esaote. Sp A
AUG 1 6 2005
# 510(k) Summary
The following 510(k) summary has been prepared pursuant to requirements specified in 21CFR 807.92(a).
#### 807.92(a)(1)
## Submitter Information
Carri Graham, Official Correspondent
7992 Castleway Drive
Indianapolis, IN 46250
| Phone: | (317) 849-1916 x103 |
|-----------------|---------------------|
| Facsimile: | (317) 577-9070 |
| Contact Person: | Carri Graham |
| Date: | July 29, 2005 |
807.92(a)(2)
| Trade Name: | G-scan |
|-------------------------|--------------------------------------|
| Common Name: | Coil, Magnetic Resonance Specialty |
| Classification Name(s): | Magnetic Resonance Diagnostic Device |
| Classification Number: | 90MOS |
807.92(a)(3)
Predicate Device(s)
| Esaote | G-scan | K042236 |
|--------|--------|---------|
|--------|--------|---------|
{1}------------------------------------------------
S 10(k) Summar. ( r-scan Esacte. Sp A
#### 807.92(a)(4)
### Device Description
The Cervical Spine Coil is to be used with the MRI system G-scan, cleared via K042236.
This is a solenoidal linear receiving coil, shaped for suitability to the areas under examination and designed to be matched to the patient's cervical vertebrae; the coil is also morphologically adapted to the examined area to obtain a good Signal-to-Noise Ratio.
The patient's neck is inserted into the coil after the coil itself has been opened by means of nonmagnetic contacts inserted in the coil loops.
The height of the patient's head can be regulated by a suitable mechanism with a lever command. A special cushion is attached to the coil assuring that the patient maintains comfortably his position during the examination.
The mechanical connection between the coil and the patient table is provided through a mechanism that positions the coil at the proper position for a cervical examination.
The coil is equipped with a connecting cable, which must be fitted to the connector on the patient table for proper electrical wiring and for automatic coil model recognition.
The coil resonator is composed of 3 copper tubes each one with two couples of non-magnetic contacts, the three tubes are arranged in a circular path and connected in series and with 3 tuning capacitors groups.
De-coupling between the transmitting and the receiving coil is accomplished via a couple of PIN diodes in a passive circuit. During the RF transmitting pulse, the diodes placed in the receiving coil decoupling circuit are switched on by the pulse and enable a LC parallel resonating circuit, at the working frequency, so that the overall impedance of the coil becomes very high and the parasitic current is minimised.
807.92(a)(5)
### Intended Use(s)
The G-scan Cervical Spine Coil is to be used in MR imaging of the cervical section of the spine column.
G-scan is a Magnetic Resonance (MR) system that produces transversal, sagittal and coronal and oblique cross-section images of the limbs, joints and spinal column. It is intended for imaging portions of the upper limb, including the hand, wrist, forearm, elbow, arm and shoulder, imaging portions of the lower limb, including the foot, ankle, calf, knee, thigh and imaging portions of the spinal column, including the cervical, thoracic and lumbo-sacral sections.
G-scan images correspond to the spatial distribution of protons (hydrogen nuclei) that determine magnetic resonance properties and are dependent on the MR parameters, including spin-latice relaxation time (T1), spin-spin relaxation time (T2), nuclei density, flow velocity and "chemical shift". When interpreted by a medical expert trained in the use of MR equipment, the images can provide diagnostically useful information.
The indications for use of the G-scan MRJ system with the Cervical Spine Coil are the same as those for the G-scan previously cleared via K 042236.
{2}------------------------------------------------
## Technological Characteristics
The technological characteristics of the Cervical Spine Coil are similar to the characterisities of the predicate device linear receiving coils.
presided device intear received ally adapted to the examined area for increasing the Signal-to-Noise Ratio.
| Characteristics | G-scan<br>K042236 | Modified<br>G-scan |
|---------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Cervical Spine Coil | Linear Receiving Coils:<br>- Shoulder coil 1: 22.4 x 21.6 x 12.6 cm<br>(h x w x d) external; 14.5 x 17.5 x 12.6 cm (h x w x d) internal<br>- Flexible coil 6: 4.0 x 33.0 x 28.5 cm (h x w x d) maximum external<br>- Thoracic 8 - Lumbar Spine coil: 4.5 x 32.2 x 28.4 cm (h x w x d) maximum external<br><br>DPA Receiving Coils:<br>- Knee coil 2: 22.5 x 21.0 x 18.3 cm (h x w x d) external; 14.3 x 16.0 x 18.3 cm (h x w x d) internal<br>- Hand coil 3: 17.8 x 17.5 x 20 cm (h x w x d) external; 11.9 x 7.2 x 20 cm (h x w x d) internal<br>- Foot/Ankle coil 4: 22.0 x 19.2 x 28.5 cm (h x w x d) external; 14.6 x 10.0 x 28.5 cm (h x w x d) internal<br>- Shoulder coil 7: 21.5 x 30.0 x 21.8 cm (h x w x d) maximum external | Linear Receiving Coils:<br>- Shoulder coil 1: 22.4 x 21.6 x 12.6 cm<br>(h x w x d) external; 14.5 x 17.5 x 12.6 cm (h x w x d) internal<br>- Flexible coil 6: 4.0 x 33.0 x 28.5 cm (h x w x d) maximum external<br>- Thoracic 8 – Lumbar Spine coil: 4.5 x 32.2 x 28.4 cm (h x w x d) maximum external<br>- Cervical Spine Coil 9: 26.4 x 32.1 x 36.6 cm (h x w x d) maximum external; 17.5 x 15.1 x 5.0 cm (h x w x d) internal<br><br>DPA Receiving Coils:<br>- Knee coil 2: 22.5 x 21.0 x 18.3 cm (h x w x d) external; 14.3 x 16.0 x 18.3 cm (h x w x d) internal<br>- Hand coil 3: 17.8 x 17.5 x 20 cm (h x w x d) external; 11.9 x 7.2 x 20 cm (h x w x d) internal<br>- Foot/Ankle coil 4: 22.0 x 19.2 x 28.5 cm (h x w x d) external; 14.6 x 10.0 x 28.5 cm (h x w x d) internal<br>- Shoulder coil 7: 21.5 x 30.0 x 21.8 cm (h x w x d) maximum external |
## Substantial Equivalence Comparison Table
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows a circular logo for the Department of Health & Human Services - USA. The logo features a stylized depiction of a bird or eagle in flight, composed of three curved lines. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the circumference of the circle, framing the bird symbol. The logo is black and white.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Esaote, S.p.A. % Ms. Carri Graham Official Correspondent The Anson Group 7992 Castleway Drive INDIANAPOLIS IN 46250
### AUG 1 6 2005
Re: K052077
Trade/Device Name: G-scan Cervical Spine Coil Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device
Regulatory Class: II Product Code: MOS Dated: July 29, 2005 Received: August 1, 2005
Dear Ms. Graham:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) 240-276-0115 |
|-----------------|-----------------------------------------------|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology) 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology) 240-276-0120 |
| Other | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
#### Indications for Use
510(k) Number (if known): Kos 20 77
Device Name: G-scan Cervical Spine Coil
Indications for Use:
The G-scan Cervical Spine Coil is to be used in MR imaging of the cervical section of the spine column.
G-scan is a Magnetic Resonance (MR) system that produces transversal, sagittal and coronal and oblique cross-section images of the limbs, joints and spinal column. It is intended for imaging portions of the upper limb, including the hand, wrist, forearm, elbow, arm and shoulder, imaging portions of the lower limb, including the foot, ankle, calf, knee, thigh and hip and imaging portions of the spinal column, including the cervical, thoracic and lumbo-sacral sections.
G-scan images correspond to the spatial distribution of protons (hydrogen nuclei) that determine magnetic resonance properties and are dependent on the MR parameters, including spin-lattice relaxation time (T1), spin-spin relaxation time (T2), nuclei density, flow velocity and "chemical shift". When interpreted by a medical expert trained in the use of MR equipment, the images can provide diagnostically useful information.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Nancy C. Brogdon
Page 1 of 1
(Division Sigh Off) Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number
12
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.