JSZ-STAT-TREE, AND JSZ-STAT-OR, CLEANING, DISINFECTING AND STORING SOLUTIONS; JSZ-WET-TREE AND JSZ-WET-OR, WETTING,
Applicant
Szabocsik and Associates
Product Code
MRC · Ophthalmic
Decision Date
Sep 1, 2005
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.5918
Device Class
Class 2
Indications for Use
JSZ-Stat-Tree Cleaning, Disinfecting, Storage Solution is indicated for use in the cleaning, chemical disinfection and storage of fluoro/silicone acrylate and silicone acrylate (RGP) contact lenses. JSZ-Stat-Or Cleaning, Disinfecting, Storage Solution is indicated for use in the cleaning, chemical disinfection and storage of fluoro/silicone acrylate and silicone acrylate (RGP) contact lenses. JSZ-Wet-Tree Wetting. Lubricating, Rewetting Drops are indicated for use to wet fluoro/silicone acrylate and silicone acrylate rigid gas permeable (RGP) contact lenses prior to lens insertion and to rewet and lubricate the lens while on the eye. JSZ-Wet-Or Wetting. Lubricating, Rewetting Drops are indicated for use to wet fluoro/silicone acrylate and silicone acrylate rigid gas permeable (RGP) contact lenses prior to lens insertion and to rewet and lubricate the lens while on the eye.
Device Story
Chemical solutions for RGP contact lens care; includes cleaning, disinfecting, and storage solutions (Stat-Tree/Stat-Or) and wetting, lubricating, and rewetting drops (Wet-Tree/Wet-Or). Solutions applied by patient to RGP lenses; cleaning/disinfecting performed during storage; wetting drops applied prior to insertion or while lens is on eye. Formulations identical to previously marketed predicate devices. No clinical data required due to established safety of active ingredients. Benefits include maintenance of lens hygiene, comfort, and ocular lubrication.
Clinical Evidence
No clinical data included. Evidence based on bench testing: cytotoxicity, acute oral toxicity, ocular irritation, guinea pig maximization, corneal wound healing, corneal penetration, disinfection efficacy, and preservative efficacy. All tests passed.
Technological Characteristics
Chemical solutions for RGP lens care. Formulations contain active ingredients identical to predicate devices. Stability and sterility verified via real-time testing. No electronic or software components.
Indications for Use
Indicated for cleaning, chemical disinfection, and storage of fluoro/silicone acrylate and silicone acrylate (RGP) contact lenses; wetting, lubricating, and rewetting of RGP lenses prior to insertion and while on the eye.
Regulatory Classification
Identification
A rigid gas permeable contact lens care product is a device intended for use in the cleaning, conditioning, rinsing, lubricating/rewetting, or storing of a rigid gas permeable contact lens. This includes all solutions and tablets used together with rigid gas permeable contact lenses.
Special Controls
*Classification.* Class II (Special Controls) Guidance Document: “Guidance for Industry Premarket Notification (510(k)) Guidance Document for Contact Lens Care Products.”
Predicate Devices
DeStat-3
DeStat-4
Stay-Wet 3
Stay-Wet 4
Reference Devices
Optimum by Lobob Cleaning, Disinfecting and Storage Solution
Optimum by Lobob Wetting and Rewetting Drops
Claris Cleaning and Soaking Solution
Submission Summary (Full Text)
{0}------------------------------------------------
SEP 0 1 2005
K052074
JSZ-Stat-Tree, JSZ-Stat-Or, JSZ-Wet-Tree, JSZ-Wet Or 510(k)
### SUMMARY, FREEDOM OF INFORMATION
# JSZ-STAT-TREE, AND JSZ-STAT-OR, CLEANING, DISINFECTING AND STORING SOLUTIONS;
# AND JSZ-WET-TREE AND JSZ-WET-OR, WETTING, LUBRICATING AND REWETTING SOLUTIONS
### FOR RIGID GAS PERMEABLE CONTACT LENSES
| 1. Common/Usual Names | JSZ-Stat-Tree, and JSZ-Stat-Or, Cleaning, Disinfecting and Storing Solutions;<br>and JSZ-Wet-Tree and JSZ-Wet-Or, Wetting, Lubricating and Rewetting Solutions |
|--------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 2. Trade/Proprietary | JSZ-Stat-Tree, and JSZ-Stat-Or, Cleaning, Disinfecting and Storing Solutions;<br>and JSZ-Wet-Tree and JSZ-Wet-Or, Wetting, Lubricating and Rewetting Solutions |
| 3. FDA Classification | Class II<br>(Performance Standards)<br>21 CFR 886.5918<br>Rigid Gas Permeable (RGP) contact lens solutions |
| 4. Performance standards | None established under section 514 |
5. Substantial equivalence
These products are substantially equivalent to several currently marketed products, such as DeStat-3 and Destat 4 Cleaning, Disinfecting and Storage Solutions, Stay-Wet 3 and Stay-Wet 4 Wetting and Rewtting Drops. Optimum by Lobob Cleaning, Disinfecting and Storage Solution, Wetting and Rewetting Drops, and Claris Cleaning and Soaking Solution.
### 6. Indications for Use
JSZ-Stat-Tree Cleaning, Disinfecting, Storage Solution is indicated for use in the cleaning, chemical disinfection and storage of fluoro/silicone acrylate and silicone acrylate (RGP) contact lenses.
JSZ-Stat-Or Cleaning, Disinfecting, Storage Solution is indicated for use in the cleaning, chemical disinfection and storage of fluoro/silicone acrylate and silicone acrylate (RGP) contact lenses.
JSZ-Wet-Tree Wetting. Lubricating, Rewetting Drops are indicated for use to wet fluoro/silicone acrylate and silicone acrylate rigid gas permeable (RGP) contact lenses prior to lens insertion and to rewet and lubricate the lens while on the eye.
JSZ-Wet-Or Wetting. Lubricating, Rewetting Drops are indicated for use to wet fluoro/silicone acrylate and silicone acrylate rigid gas permeable (RGP) contact lenses prior to lens insertion and to rewet and lubricate the lens while on the eye.
{1}------------------------------------------------
# SUMMARY
Contained in this submission are comparisons of the products to the predicate devices, information on the chemistry and manufacturing, toxicology and microbiology. Because the predicate devices, with identical formulations, have been previously approved for market, and have been marketed over the last decade, no clinical data is included in this submission.
- I. Chemistry
The JSZ-Stat-Tree/JSZ-Wet-Tree and the JSZ-Stat-Or/JSZ-Wet-Or systems were compatible with all groups of RGP lenses. Both ISZ-Stat-Tree and JSZ-Stat-Or were shown to be effective daily cleaners.
Toxicology II.
> The toxicological testing of the the JSZ-Stat-Tree/JSZ-Wet-Tree, and JSZ-Stat-Or/JSZ-Wet-Or systems for the cleaning, disinfecting and storing, and wetting, lubricating and rewetting of rigid gas permeable lenses is summarized below.
| Test | DeStat-3/DeStat-4 | Stay-Wet-3/Stay-Wet 4 |
|-------------------------------------|-------------------|-----------------------|
| Cytotoxicity | NAa | passed |
| Acute Oral Toxicity | passed | passed |
| Ocular Irritation (21d rabbit) | passedb | |
| Ocular Irritation (72hr rabbit) | passedb | |
| Anesthetic Effect | NAa | passed |
| Guinea Pig Maximization | passedc | |
| Corneal Epithelial Wound<br>Healing | passedc | |
| Corneal Penetration | passedc | |
a NA - test not applicable to cleaning solutions
b Test used JSZ-Stat-Tree/JSZ-Wet-Tree, or JSZ-Stat-Or/JSZ-Wet-Or
° Tests on benzyl alcohol
#### III. Microbiology
The microbiological testing of the the JSZ-Stat-Tree/JSZ-Wet-Tree, and JSZ-Stat-Or! JSZ-Wet-Or systems for the cleaning, disinfecting and storing, and wetting, lubricating and rewetting of rigid gas permeable lenses is summarized below.
। (
{2}------------------------------------------------
# JSZ-Stat-Tree, JSZ-Stat-Or, JSZ-Wet-Tree, JSZ-Wet Or 510(k)
| Test | JSZ-Stat-Tree | JSZ-Stat-Or | JSZ-Wet-Tree | JSZ-Wet-Or |
|---------------------------------------------|---------------|-------------|--------------|------------|
| Disinfection Efficacy<br>(microbial, viral) | passed | passed | NAa | NA |
| Neutralizer Efficacy | passed | passed | NA | NA |
| Preservative Efficacy | passed | passed | passed | passed |
| Sterility/Stabilityb | passed | passed | passed | passed |
ª NA - Not applicable to rewetting solutions
6 Protocol for real-time stability of JSZ products included
# IV. Clinical Studies
Clinical data is not required for solutions using the same active ingredients at the same concentrations as currently marketed products.
V. Substantial Equivalence
The solutions are identical to the DeStat-3, DeStat-4, Stay-Wet 3, and Stay-Wet-4 solutions previously approved, and substantially equivalent to the Optimum by Lobob Cleaning, Disinfecting and Storage Solution, Optimum by Lobob Wetting and Rewetting Drops, and Claris by Menicon Cleaning and Soaking Solution.
VI. Manufacturing
Manufacturing details are included in the manufacturing section.
VII. Shelf Life
The shelf life is assumed to be the same as in the predicate filings and a protocol for on-going real time shelf-life is included.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three stripes forming its wing, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion around the eagle. The text is in all caps and appears to be in a sans-serif font. The logo is black and white.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
SEP 0 1 2005
John Szabocsik, Ph.D. President SZABOCSIK AND ASSOCIATES 203 North Wabash Avenue Chicago, Illinois 60601
Re: K052074
Trade/Device Name: JSZ-Stat-Tree, and JSZ-Stat-Or, Cleaning, Disinfecting and Storing Solutions; and JSZ-Wet-Tree and JSZ-Wet-Or, Wetting, Lubricating and Rewetting Solutions
Regulation Number: 21 CFR 886.5918 Regulation Name: Rigid Gas Permeable (RGP) contact lens solutions Regulatory Class: Class II Product Code: MRC Dated: July 29, 2005 Received: August 1, 2005
Dear Dr. Szabocsik :
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
# Page 2 - John Szabocsik, Ph.D.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 827-8910. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
David M. Whipple
David M. Whipple Acting Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
JSZ-Stat-Tree, JSZ-Stat-Or, JSZ-Wet-Tree, JSZ-Wet Or 510(k)
510(k) NUMBER (IF KNOWN) K052074
DEVICE NAME
JSZ-Stat-Tree Cleaning, Disinfecting, Storage Solution JSZ-Stat-Or Cleaning, Disinfecting, Storage Solution
JSZ-Wet-Tree Wetting. Lubricating, Rewetting Drops JSZ-Wet-Or Wetting. Lubricating, Rewetting Drops.
# INDICATIONS FOR USE
JSZ-Stat-Tree Cleaning, Disinfecting, Storage Solution is indicated for use in the JSZ-Stat-Tree Cleaning, Disameeting, Sessings of Stages
cleaning, chemical disinfection and storage of fluoro/silicone acrylate and silicone acrylate (RGP) contact lenses.
JSZ-Stat-Or Cleaning, Disinfecting, Storage Solution is indicated for use in the JSZNal-Of Cleaning, Districting, Stellige Beer ago Bearther and silicone acrylate (RGP) contact lenses.
JSZ-Wet-Tree Wetting. Lubricating, Rewetting Drops are indicated for use to wet JSZ-Wel-Tree Wetting. Dubrieading acrylate rigid gas permeable (RGP) contact lenses Indord/sincolle act yiate and barrest and lubricate the lens while on the eye.
JSZ-Wet-Or Wetting. Lubricating, Rewetting Drops are indicated for use to wet JSZ-Wel-Or Welting: Edwrieauing, songlate rigid gas permeable (RGP) contact lenses prior to lens insertion and to rewet and lubricate the lens while on the eye.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Amilham
Concurrence of CDRH, Office of Device Evaluation (ODE)
Over-The-Counter-Use U (Per 21 CFR 801.109) OR Prescription Use _ (Optional Format 1-2-96)
stmlest
ision of Ophthalm
10(k) Number K052074
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.