← Product Code [KTT](/productcode/KTT) · K052065

# ALLIANCE MEDICAL CORPORATION REPROCESSED EXTERNAL FIXATION DEVICES (K052065)

_Alliance Medical Corp. · KTT · Aug 22, 2005 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K052065

## Device Facts

- **Applicant:** Alliance Medical Corp.
- **Product Code:** [KTT](/productcode/KTT.md)
- **Decision Date:** Aug 22, 2005
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 888.3030
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

Reprocessed external fixation devices are indicated for use in patients requiring external skeletal fixation and treatment of fractures, osteotomy, arthrodesis, correction of deformities, fracture revision, bone reconstruction procedures, limb lengthening, correction of bony or soft tissue deformities and segmental bony or soft tissue defects.

## Device Story

Reprocessed external fixation devices consist of frames, clamps, rods, couplings, pins, posts, fasteners, washers, nuts, hinges, sockets, and screws. Used for management of bone fractures, reconstructive, and corrective orthopedic surgery. Devices provide skeletal stability or traction. Operated by orthopedic surgeons in clinical/surgical settings. Reprocessed devices are provided non-sterile for hospital-based sterilization prior to use. Functionality and mechanical performance are identical to original predicate devices.

## Clinical Evidence

Bench testing only. Validation of reprocessing and function testing demonstrated that the reprocessed devices perform as originally intended.

## Technological Characteristics

Mechanical external fixation system comprising metal alloys, plastics, and composites. Components include frames, clamps, rods, pins, and fasteners. Design, dimensions, and materials are identical to predicate devices. Non-sterile; requires hospital-based sterilization.

## Regulatory Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- Reprocessed Orthofix® External Fixation Devices ([K023714](/device/K023714.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K052065

AUG 2 2 2005

PART B: 510(k) SUMMARY

| <b>Submitter:</b> | Alliance Medical Corporation |
|-------------------|------------------------------|
|                   | 10232 South 51st Street      |
|                   | Phoenix, Arizona 85044       |

Contact: Moira Barton Regulatory Affairs Manager (480) 763-5300 (o) (480) 763-5310 (f) mbarton@alliance-medical.com

Date of preparation: July 20, 2005

Trade/Proprietary Name: Reprocessed External Fixation Name of device: Devices Common or Usual Name: External Fixation Devices, Fixation Appliance. Single/Multiple Component Classification Name: Single/Multiple Component Metallic Bone Fixation Appliances and Accessories and Smooth or Threaded Metallic Bone Fixation Fastener

## Predicate device(s):

- K023714 Reprocessed Orthofix® External Fixation Devices
Device description: External fixation devices are specially designed frames, clamps, rods, rod-to-rod couplings, pins, posts, fasteners, wire fixations, fixation bolts, washers, nuts, hinges, sockets, connecting bars and screws used for the management of bone fractures and reconstructive, as well as corrective, orthopedic surgery. Materials used include metal alloys, plastic and composites. These materials are chosen to address a wide range of fractures and applications as well as to allow for the appropriate amount of rigidity and stability.

Intended use: External Fixation Devices are intended to be used for the fixation of supracondylar, or condylar fractures of the femur; for fusion of a joint; for surgical procedures that involve cutting the bone, for fixation of bone fractures; bone reconstruction; as a quide pin for insertion of other implants; or may be implanted through the skin so that a pulling force or traction may be applied to the skeletal system; and others may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

> Alliance Medical Corporation Modified Reprocessed External Fixation Devices Special 510(k)

{1}------------------------------------------------

| Indications<br>statement:         | Reprocessed external fixation devices are indicated for use in<br>patients requiring external skeletal fixation and treatment of<br>fractures, osteotomy, arthrodesis, correction of deformities,<br>fracture revision, bone reconstruction procedures, limb<br>lengthening, correction of bony or soft tissue deformities and<br>segmental bony or soft tissue defects.                                                                                                                                                                                                                                                                                                |
|-----------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Technological<br>characteristics: | The design, materials, and intended use of the Reprocessed<br>External Fixation Devices are identical to the predicate devices.<br>The mechanism of action of the Reprocessed External Fixation<br>Device is identical to the predicate devices in that the same<br>standard mechanical design, materials, shapes and sizes are<br>utilized. There are no changes to the claims, intended use,<br>clinical applications, patient population, performance<br>specifications, or method of operation. The only change is the<br>modified external fixation devices will be provided non-sterile.<br>Sterilization of the devices will occur in the hospital prior to use. |
| Performance data:                 | Bench and laboratory testing was conducted to demonstrate<br>performance (safety and effectiveness) of the Reprocessed<br>External Fixation Devices.<br><br>Validation of reprocessing Function Testing<br>Performance testing demonstrates that Reprocessed External<br>Fixation Devices perform as originally intended.                                                                                                                                                                                                                                                                                                                                               |
| Conclusion:                       | Alliance Medical Corporation concludes that the modified device<br>(the Reprocessed External Fixation Device) is safe, effective<br>and substantially equivalent to the predicate devices, as<br>described herein.                                                                                                                                                                                                                                                                                                                                                                                                                                                      |

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows a circular seal with the words "DEPARTMENT OF HEALTH AND HUMAN SERVICES, USA" written around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread. The eagle is depicted with thick, black lines, giving it a bold and graphic appearance. 

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

AUG 2 2 2005

Ms. Moira Barton Regulatory Affairs Manager Alliance Medical Corporation 10232 South 51st Street Phoenix, Arizona 85044

Re: K052065

Trade/Device Name: Alliance Medical Corporation Reprocessed External Fixation Devices Regulation Numbers: 21 CFR 888.3030, 21 CFR 888.3040 Regulation Names: Single/multiple component metallic bone fixation appliances and accessories, Smooth or threaded metallic bone fixation fastener Regulatory Class: II Product Codes: KTT, KTW, JEC Dated: July 19, 2005 Received: July 29, 2005

Dear Ms. Barton:

We have reviewed your Section 510(k) premarket notification of intent to market the device w o nave roviewed above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

## Page 2 - Ms. Moira Barton

This letter will allow you to begin marketing your device as described in your Section 510(k) I mis letter will and by of substantial equivalence of your device of your device to a legally promated predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you atent office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small other general international and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Barbara Buehus

Mark N. Melkerson Acting Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

## Indications for Use

510(k) Number (if known): K052065

Device Name: Alliance Medical Corporation Reprocessed External Fixation Devices

Indications for Use: Reprocessed external fixation devices are indicated for use in patients requiring external skeletal fixation and treatment of fractures, osteotomy, arthrodesis, correction of deformities, fracture revision, bone reconstruction procedures, limb lengthening, correction of bony or soft tissue deformities and segmental bony or soft tissue defects.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Barbara Buehrle for MXM

(Division Sig Division of General, Restorati and Neurological Devices

510(k) Number K052065

---

**Source:** [https://fda-staging.innolitics.com/device/K052065](https://fda-staging.innolitics.com/device/K052065)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
