K052063 · A & A Medical Supplies, Inc. · ILY · Sep 30, 2005 · Physical Medicine
Device Facts
Record ID
K052063
Device Name
HUANQIU HEATING LAMP
Applicant
A & A Medical Supplies, Inc.
Product Code
ILY · Physical Medicine
Decision Date
Sep 30, 2005
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5500
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
HUANQIU Heating lamp, including CQ-27, CQ-29 and CQ-67, is an infrared lamp that emits the infrared spectrum to provide topical heating for the purpose of elevating tissue temperature, to temporarily increase local blood circulation, and to temporarily relieve minor muscle and joint pain and stiffness. The lamps may also help to relieve minor pain associated with muscle spasms, minor sprains and strains, and minor muscular back pain.
Device Story
HUANQIU Heating Lamps (models CQ-27, CQ-29, CQ-67) are infrared heating devices consisting of a heating head, swing arm, control box, and pedestal. The device utilizes a rare earth ceramic plate to emit infrared radiation (2 to 50 microns) to provide topical heat. The device is powered by 110V AC. Users operate the device by positioning the heating head 8 to 12 inches from the treatment area to achieve skin temperatures of 40-45°C. The device includes a timer (up to 60 minutes) and either standard or digital display controls. It is intended to elevate tissue temperature, increase local blood circulation, and provide temporary relief for minor musculoskeletal pain and stiffness. The heating plate requires replacement after 1,000 to 1,500 hours of use.
Clinical Evidence
Bench testing only. No clinical data provided. Equivalence is based on technical specifications, material composition, and performance testing (skin temperature output at varying distances) compared to the predicate device.
Technological Characteristics
Infrared lamp; rare earth ceramic heating plate; 110V AC power; 250W max power; 2-50 micron emission spectrum; 1,000-1,500 hour plate life; 4-5 casters; 2 fuses; standard or digital display control options. Tested to IEC 60601-1 (GB 9706.1-1995).
Indications for Use
Indicated for patients requiring topical heating to elevate tissue temperature, increase local blood circulation, and temporarily relieve minor muscle and joint pain, stiffness, muscle spasms, minor sprains, strains, and minor muscular back pain.
Regulatory Classification
Identification
An infrared lamp is a device intended for medical purposes that emits energy at infrared frequencies (approximately 700 nanometers to 50,000 nanometers) to provide topical heating.
Special Controls
*Classification.* Class II (special controls). The device, when it is an infrared therapeutic heating lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 890.9.
{0}------------------------------------------------
Image /page/0/Picture/2 description: The image shows the text "Crosslinks International, Inc." in a serif font. The text is arranged in two lines, with "Crosslinks" on the first line and "International, Inc." on the second line. The text is black and the background is white. The image appears to be a logo or heading from a document.
# 510 (k) Premarket Notification Summaries
### Submission Correspondent
George Su, Ph.D. Crosslinks International 1800 Century Park East, Suite 600 Los Angeles, CA 90067 USA
Tel: 310-229-5748 Fax: 310-388-1067 Email: crosslinks2000/a aol.com
## Applicant:
A & A MEDICAL SUPPLIES INC 5288 valley blvd #1 Los Angeles, CA 90032
Tel: 323-223-6318
# Sponsor of the 510(k) Submission:
Same as above.
# Device Name:
Proprietary Name: HUANQIU Heating Lamp Common/Usual Name: Infrared Lamp Device Name: Infrared Lamp Product Code: ILY Classification: Class II
# Predicate Device:
TDP CQ-27 Heat Lamp Establishment: LHASA Medical Inc. Regulation Number: 890.5500 Product Code: ILY 510(k) Number: K003538 Registration Number: 1222811 Owner/Operator Number: 9003816
{1}------------------------------------------------
Crosslinks International, Inc.
### Description of the HUANQIU Heating Lamps:
The device consists of heating head, swing arm, control box and pedestal with extension. It is used to provide topical heating to the body and is specially engineered using a rare earth ceramic plate. Emission spectrum ranges from 2 to 50 microns. The emission heating plate should be replaced after 1,000 to 1500 hours of usage. 110 volt power, 250 watts. There are two different types of control: standard control and digital display control.
### Indications for use
HUANQIU Heating lamp, including CQ-27, CQ-29 and CQ-67, is an infrared lamp that emits the infrared spectrum to provide topical heating for the purpose of elevating tissue temperature, to temporarily increase local blood circulation, and to temporarily relieve minor muscle and joint pain and stiffness. The lamps may also help to relieve minor pain associated with muscle spasms, minor sprains and strains, and minor muscular back pain.
#### Substantial Equivalence Information
HUANQIU CQ-27, CQ-29 and CQ-67 Heating Lamps are generally equivalent to CQ-27 Heat Lamp (K003538) manufactured by Lhasa Medical, Inc. on label and labeling, materials used, specifications and intended use. All the technical data are tested according to GB 9706.1-1995 which is identical to IEC 60601-1
The equivalence information is focused on intended use, materials and specifications. HUANQIU Heating lamps are powered by 110V AC, They have 4 or 5 casters and 2 fuses. The working life of the heating plate is 1000 to 1500 hours. The skin temperature of the treatment area is at 40~45℃ when treating distances are from 8" to 12" from heating unit to treatment surface.
{2}------------------------------------------------
Crosslinks
International, Inc.
Table 1.
Predicate device: TDP CQ-27 Heat Lamp manufactured by Lhasa Medical, Inc.
| ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | The new device: HUANQIU Heating Lamp CQ-27 |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------|
| | and and and the commend the production of the first of the first of the first of the first of the first of the may |
| Item | The predicate device | The new device | | |
|----------------------|------------------------------------------------------------------|------------------------------------------------------------------|--------------|------------|
| Components | heating head, swing arm, control box and pedestal with extension | heating head, swing arm, control box and pedestal with extension | | |
| Power Frequency | 50/60 Hz | 50/60 Hz | | |
| Power | $\leq 250 W$ | $\leq 250 W$ | | |
| Spectrum Ranges | 2 to 50 microns | 2 to 50 microns | | |
| Warming Up Time | 15 minutes | 15 minutes | | |
| Total Working Hours | 1,200 to 1500 hours | 1,000 to 1500 hours | | |
| Operating Timer | Up to 60 minutes | Up to 60 minutes | | |
| Number of Fuse | 2 | 2 | | |
| Inner Cover Diameter | 12 cm | 12 cm | | |
| Number of Casters | 4 | 4 | | |
| Skin Temperature | The distance from the lamp head and the result | | | |
| | 8"<br>45°C | 10"<br>43 °C | 12"<br>41 °C | 8"<br>45°C |
Table 2.
Predicate device: TDP CQ-27 Heat Lamp manufactured by Lhasa Medical, Inc. The new device: HUANQIU Heating Lamp CQ-29
| Item | The predicate device | The new device | | |
|----------------------|------------------------------------------------------------------|------------------------------------------------------------------|-------------|------------|
| Components | heating head, swing arm, control box and pedestal with extension | heating head, swing arm, control box and pedestal with extension | | |
| Power Frequency | 50/60 Hz | 50/60 Hz | | |
| Power | < 250 W | < 250 W | | |
| Spectrum Ranges | 2 to 50 microns | 2 to 50 microns | | |
| Warming Up Time | 15 minutes | 15 minutes | | |
| Total Working Hours | 1,200 to 1500 hours | 1,000 to 1500 hours | | |
| Operating Timer | Up to 60 minutes | Up to 60 minutes | | |
| Number of Fuse | 2 | 2 | | |
| Inner Cover Diameter | 12 cm | 20 cm | | |
| Number of Casters | 4 | 4 | | |
| Skin Temperature | The distance from the lamp head and the result | | | |
| | 8"<br>45°C | 10"<br>43°C | 12"<br>41°C | 8"<br>45°C |
{3}------------------------------------------------
Crosslinks
International, Inc.
Table 3.
Predicate device: TDP CQ-27 Heat Lamp manufactured by Lhasa Medical, Inc. The new device: HUANQIU Heating Lamp CQ-67
| Item | The predicate device | The new device |
|----------------------------|------------------------------------------------------------------------------|------------------------------------------------------------------------------------------|
| Components | heating head<br>swing arm<br>standard control box<br>pedestal with extension | heating head<br>swing arm<br>control box with digital display<br>pedestal with extension |
| Power Frequency | 50/60 Hz | 50/60 Hz |
| Power | < 250 W | < 250 W |
| Spectrum Ranges | 2 to 50 microns | 2 to 50 microns |
| Warming Up Time | 15 minutes | 15 minutes |
| Total Working Hours | 1,200 to 1500 hours | 1,000 to 1500 hours |
| Operating Timer | Up to 60 minutes | Up to 60 minutes |
| Number of Fuse | 2 | 2 |
| Inner Cover Diameter/width | 12 cm | 30 cm |
| Number of Casters | 4 | 5 |
| Skin Temperature | The distance from the lamp head and the result | |
| | 8"<br>45 | 10"<br>43°C |
| | 12"<br>41°C | 8"<br>46 °C |
| | 10"<br>42°C | 12"<br>40°C |
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the seal of the Department of Health and Human Services (HHS). The seal features the department's name encircling a symbol. The symbol consists of a stylized caduceus, a staff with two snakes entwined around it, which is a common symbol associated with medicine and healthcare. The seal is in black and white.
SEP 3 0 2005
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
A & A Medical Supplies Inc. c/o George Su, Ph.D. Crosslinks International 1800 Century Park East, Suite 600 Los Angeles, California 90067
Re: K052063
Trade/Device Name: Infrared Lamp Regulation Number: 21 CFR 890.5500 Regulation Name: Infrared lamp Regulatory Class: II Product Code: ILY Dated: May 4, 2005 Received: July 29, 2005
Dear Dr. Su:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{5}------------------------------------------------
Page 2 - George Su, Ph.D.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Mark A. Milliman
Mark N. Melkerson Acting Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Premarket Notification 510(k) Report for HUANQIU HEATING LAMP
# Indications for Use
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
Device Name: _________________________________________________________________________________________________________________________________________________________________
Indications for Use:
HUANQIU Heating lamp, including CQ-27, CQ-29 and CQ-67, is an infrared lamp that emits the infrared spectrum to provide topical heating for the purpose of elevating tissue temperature, to temporarily increase local blood circulation, and to temporarily relieve minor muscle and joint pain and stiffness. The lamps may also help to relieve minor pain associated with muscle spasms, minor sprains and strains, and minor muscular back pain.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Mark A. Millkerson
(Division Sign-Off) Division of General, Restorative, and Neurological Devices
**510(k) Number** K052063
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.