K052059 · Inntec, Inc. · MQF · Nov 22, 2005 · Obstetrics/Gynecology
Device Facts
Record ID
K052059
Device Name
IUI CATHETER, MODEL 9057
Applicant
Inntec, Inc.
Product Code
MQF · Obstetrics/Gynecology
Decision Date
Nov 22, 2005
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 884.6110
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The InnTec, Inc. Intrauterine Insemination (IUI) Catheter is used for intrauterine insemination.
Device Story
InnTec IUI Catheter is a sterile, single-use device designed for intrauterine insemination. The catheter measures 17.5cm in length, tapering from 7 Fr at the hub to 5 Fr at the distal tip. The tip is closed and rounded with opposing eyelets; internal volume is 0.07cc. Centimeter markings (6, 7, 8, 9, 10cm) are printed on the shaft. The device is used by clinicians in a clinical setting to facilitate the delivery of sperm into the uterus. It functions as a conduit for insemination, aiding in assisted reproduction. The device is manually operated by a physician or trained healthcare professional.
Clinical Evidence
No clinical data; bench testing only.
Technological Characteristics
Catheter length 17.5cm; tapers 7 Fr to 5 Fr; closed rounded tip with 180-degree eyelets; 0.07cc internal volume; non-toxic ink markings. Sterilization per ANSI/AAMI/ISO 11135; biocompatibility per ANSI/AAMI/ISO 10993.
Indications for Use
Indicated for intrauterine insemination in patients requiring assisted reproduction procedures.
Regulatory Classification
Identification
Assisted reproduction catheters are devices used in in vitro fertilization (IVF), gamete intrafallopian transfer (GIFT), or other assisted reproduction procedures to introduce or remove gametes, zygote(s), preembryo(s), and/or embryo(s) into or from the body. This generic type of device may include catheters, cannulae, introducers, dilators, sheaths, stylets, and component parts.
Special Controls
*Classification.* Class II (special controls) (mouse embryo assay information, endotoxin testing, sterilization validation, design specifications, labeling requirements, biocompatibility testing, and clinical testing).
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows a logo for InnTec, Inc. The logo features a stylized letter 'C' that wraps around a lowercase 'i'. The 'i' has a circular dot above it. Below the symbol is the text "InnTec, Inc".
## 510(k) Summary
Company Name: InnTec, Inc. 401 E. Edgewater St. Portage, WI 53901
Contact: Michael Kvalo, PE
Office-401 E. Edgewater St. Portage, WI. 53901 Phone 608-444-4544 Fax 608-846-6071
Phone: 608 444-4544 Fax: 608 846-6071
Summary Date: October 10, 2005
Shipping & Receiving-121 Evco Circle DeForest, WI. 53532
Trade Name: IUI Catheter
Common Name: Assisted Reproduction Catheter
Classification Name: 21 CFR 884.6110; Product Code: MQF
Predicate Device:
| 510(k) | Manufacturer | Product Code | Class | Trade Name |
|---------|--------------------------|--------------|-------|--------------|
| K921518 | Lifetek<br>Medical, Inc. | MFD | II | IUI Catheter |
របា Catheters 1.0
Oocyte
Needle
Sets
Retrieval
### Description of Device
The InnTec IUI Catheter has been specifically designed for Intrauterine Inseminations. The over all Catheter length is 17.5cm, and tapers from 7 Fr. at the hub down to 5 Fr at the distal tip which is 5cm long. The tip is closed and smoothly rounded, eyelets are located at 180° from one another. Catheter internal volume is .07cc. Centimeter marks are printed at 6, 7, 8, 9, & 10cm from the tip, the ink is non-toxic and will not rub off.
2.0 Embryo Transfer
The InnTec, Inc. Intrauterine Insemination (IUI) Catheter is used for
Intended Use
intrauterine insemination.
Catheters
Custom Product
Design &
Manufacturing
#### 3.0 Technology
The technology of the device is the same as the predicate.
#### 4.0 Conclusions
The intended use, technology, materials and manufacturing processes of the Inn Tec, Inc. IUI Catheter are the same as the predicate device. No new questions of safety or effectiveness are raised.
File name: IUI 510(k) K052059 Reply to Questions 10-24-2005
the submit the state the big the first to the state of the many of a contractional with well not
{1}------------------------------------------------
510(k) Number:
K052059 InnTec, Inc. - IUI Catheter
Standard Organization No: Standard Identification No: CDRH Internal Reference No: ANSI/AAMI/ISO 10993 Part 7 10
| Declaration of Conformity Elements: | | | |
|-------------------------------------------------------|-----|----|----|
| Any Adaptations Applied | yes | no | X |
| Any Requirements Not Applicable | yes | no | X |
| Any Deviations Applied | yes | X | no |
| Any Differences in Device Tested and Finished Product | yes | no | X |
| *Is There a Third Party or Test Lab Involved | yes | X | no |
Was there another standard used in the review of this submission? X yes no If another standard was used, please fill out an additional form.
... . . . . . . . . . . . . . . . . . . . .
* This is not the third party that reviews 510ks
{2}------------------------------------------------
# Standards Data Form for Abbreviated 510(k)s
510(k) Number:
K052059 InnTec, Inc. - IUI Catheter
Standard Organization No: Standard Identification No: CDRH Internal Reference No: ANSI/AAMI/ISO 11135 7
| Declaration of Conformity Elements: | | | | |
|-------------------------------------------------------|-----|---|----|---|
| Any Adaptations Applied | yes | | no | X |
| Any Requirements Not Applicable | yes | X | no | |
| Any Deviations Applied | yes | | no | X |
| Any Differences in Device Tested and Finished Product | yes | | no | X |
| *Is There a Third Party or Test Lab Involved | yes | X | no | |
X Was there another standard used in the review of this submission? yes no
... .............
If another standard was used, please fill out an additional form.
* This is not the third party that reviews 510ks
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the Department of Health and Human Services (HHS). The logo features a stylized caduceus, a symbol often associated with medicine and healthcare, consisting of a staff with two snakes coiled around it. The text "DEPARTMENT OF HEALTH AND HUMAN SERVICES . USA" is arranged in a circular fashion around the caduceus.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
InnTec. Inc. % Mr. Gary Syring Quality & Regulatory Associates, LLC 800 Levanger Lane STOUGHTON WI 53589
K052059 Re.:
NOV 2 2 2005
Trade/Device Name: IUI Catheter Regulation Number: 21 CFR 884.6110 Regulation Name: Assisted reproduction catheters Regulatory Class: II Product Code: MQF Dated: October 24, 2005 Received: October 31, 2005
Dear Mr. Syring:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) This letter will anow you to ogin maning of substantial equivalence of your device to a legally premaired notheation: "The PDF mainly of your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific advice to: your as the following numbers, based on the regulation number at the top of this letter:
| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|-----------------|----------------------------------|--------------|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Also, prease note the regulation on your responsibilities under the Act from the 607.77). Tou may ooum outer general and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Nancy C. Hogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Indications for Use
510(k) Number (if known): Kos 2059
Device Name: IUI Catheter
Indications for Use:
The InnTec, Inc. Intrauterine Insemination (IUI) Catheter is used for intrauterine insemination.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
David A. Grayson
.. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
(Division Sign-Off) Abdominal, Division of Reproductive, and Radiological Devi 510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.