NON-STERILE POWDER FREE MULTICOLOR - RED, BLUE, PURPLE AND YELLOW LATEX EXAMINATION GLOVES WITH PROTEIN CONTENT LABELING
Applicant
Sgmp Co., Ltd.
Product Code
LYY · General Hospital
Decision Date
Sep 1, 2005
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6250
Device Class
Class 1
Indications for Use
A medical glove to be worn on the hand of the health care and similar personnel to prevent contamination between health care personnel and patient.
Device Story
Non-sterile, powder-free, multicolored latex examination gloves. Designed for use by healthcare personnel to provide a barrier against contamination between the examiner and the patient. Manufactured from latex; meets ASTM D3578-01aE2 standards for physical dimensions, tensile strength, and elongation. Device is disposable; intended for single use. Provides a protein labeling claim of 50 micrograms or less per square decimeter. Used in clinical or similar settings to maintain hygiene and prevent cross-contamination.
Clinical Evidence
Bench testing only. Testing included physical property evaluation (tensile strength, elongation), water-tightness (1,000 ml leak test), residual powder content (ASTM D6124-00), and residual protein levels. Results confirmed compliance with ASTM D3578-01aE2 standards and FDA requirements for powder-free latex examination gloves.
Technological Characteristics
Latex material; powder-free; multicolored. Dimensions and physical properties (tensile strength, elongation) conform to ASTM D3578-01aE2. Residual powder content < 2 mg/glove; residual protein level < 50 µg/dm2. Non-sterile.
Indications for Use
Indicated for use as a disposable medical glove worn on the examiner's hand or finger to prevent contamination between patient and examiner. Suitable for healthcare and similar personnel.
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows a handwritten string of characters, "K052016". The characters are written in a dark ink on a white background. The characters are all connected, and the string is slightly curved.
APPENDIX L
### 510(k) SUMMARY SUMMARY OF THE SAFETY AND EFFECTIVENESS FOR POWDER-FREE MULTICOLOR LATEX EXAMINATION GLOVES
Submitted For : SGMP Company Limited, 181 Moo 6, Tambol Kanpaengpetch, Rattaphum, Songkhla 90180, Thailand.
- Submitted By: Tucker & Associates Official Correspondent for SGMP Co Ltd Janna P. Tucker, President - CEO 198 Avenue de la D'emcrald, Sparks, NV 89434-9550 Phone No : 775-342-2612 Fax No : 775-342-2613 E-mail: Tuckerjan@aol.com
Equivalent Predicate Device: POWDER FREE BLUE LATEX EXAM GLOVES which was granted a 510(k) # K011377 and POWDER-FREE PINK LATEX EXAM GLOVES which was also granted a 510 (k) # K011371 as shown in APPENDIX M
This summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990.
### Device Information:
Trade Name - NON-STERILF POWDER FREE MULTICOLOR LATEX EXAMINATION GLOVES
Common Name - Exam gloves
Classification Name - Patient examination glove (per 21 CFR 880.6250)
Classification Information - Class I Latex patient examination glove 80LYY, powder free and meeting all the requirements of ASTM-D3578-01aE2 Standard Specification for U.atex Examination Gloves for Medical Application.
### Device Description:
Class I latex patient examination gloves 80LYY, powder free and mecting all the requirements of ASTM-D3578-01aE2 Standard Specification for Latex Examination Glowes for Medical Application.
### Intended Use of Device:
A medical glove to be worn on the hand of the health care and similar personnel to prevent contamination between health care personnel and patient.
{1}------------------------------------------------
Technological Characteristics of Device:
## 1. Dimension
| DIMENSION | ASTM D3578-01aE2 | SGMP |
|--------------------|---------------------------------|--------------|
| X-Small | 70 mm +/- 10 mm | 70 – 80 mm |
| Small | 80 mm +/- 10 mm | 80 – 85 mm |
| Medium | 95 mm +/- 10 mm | 90 – 97 mm |
| Large | 111 mm +/- 10 mm | 105 - 111 mm |
| Length | 230 mm minimum<br>for all sizes | 242 mm |
| Thickness - Finger | 0.08 mm min | 0.13 mm min |
| Palm | 0.08 mm min | 0.1 mm min |
## 2. Physical Properties (ASTM-D3578-00aE2 Standard Specification for Latex Exam Gloves) on Lot# 4144
| | TENSILE STRENGTH | | ULTIMATE ELONGATION | |
|--------------|------------------|------|---------------------|------|
| | ASTM-D3578-00aE2 | SGMP | ASTM-D378-00aE2 | SGMP |
| | Mpa | Mpa | % | % |
| Before Aging | 18.0 | | 650 | |
| X-Small | | 23.5 | | 890 |
| Small | | 26.8 | | 800 |
| Medium | | 22.5 | | 850 |
| Large | | 29.2 | | 870 |
| After Aging | 14.0 | | 500 | |
| X-Small | | 22.8 | | 780 |
| Small | | 27.5 | | 820 |
| Medium | | 28.3 | | 750 |
| Large | | 26.8 | | 820 |
| | | | | |
{2}------------------------------------------------
### 3. Water Tight Test
| BATCH # | SIZE | SAMPLE SIZE | LEAK STATUS | NUMBER<br>LEAKED |
|-----------------|---------|-------------|-------------|------------------|
| UN-AGED<br>4144 | X-Small | 125 | No | 0 |
| 4144 | Small | 125 | Yes | 1 |
| 4144 | Medium | 125 | Yes | 2 |
| 4144 | Large | 125 | No | 0 |
| AGED<br>4144 | X-Small | 125 | No | 0 |
| 4144 | Small | 125 | Yes | 1 |
| 4144 | Medium | 125 | Yes | 1 |
| 4144 | Large | 125 | Yes | 2 |
Using the FDA specified 1,000 ml water leak test, 125 pieccs of each size of the gloves were tested and our results are as given below.
The above figures are within the ASTM D3578-00aE2 requirements for latex exam gloves of 2.5% AQL.
#### Biocompatibility 4.
The bio-compatibility test results are as per attached in APPENDIX K and show that the gloves passed the tests for examination gloves.
### 5. Total Residual Powder Content & Presence of Cornstarch
| TEST | FDA<br>REQUIREMENT | INTERNAL SGMP's |
|---------------------------------------------|--------------------|--------------------------------------------------|
| Residual Powder Content<br>(ASTM D 6124-00) | 2 mg/glove max | Ranger: 0.4 - 1.0 mg/glove<br>Mean: 0.8 mg/glove |
| Presence of Cornstarch | Negative | Negative |
{3}------------------------------------------------
## 6. Residual Protein Level.
| Acres . F FERENCED<br>-----------------------------------<br>A B-16-449, Tre- A------------------------------------------------------------------------------------------------------------------------------------------------------------<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>1111 17 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 1<br>A. A. A.<br><br>American All Art 1911<br>Advertising the country of the comments of the comments of the comments of the comments of the comments of the commended to the commended to the commended to the commended to<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>A belowed an experience of the same of the control of the<br>- L- STORES<br>Acres of Canadian Accessional Property<br>1<br>1<br>1<br>ALA VE VERSELLER ATTE<br>----------------------------------------------------------------------------<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>Commender Accession of Children Comments of Children Comments<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | 1<br>1<br>1<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>Career States of Children<br>Acres of the last to the<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>American American Partis Canada Andrews Comments of<br>Amade A. H. F. F. F. F. F. F. F. F. F. F. F. F. F. F. F. F. F. F. F. F. F. F. F. F. F. F. F. F. F. F. F. F. F. F. F. F. F. F. F. F. F. F. F. F. F. F. F. F. F. F. F. F. F. F.<br>Commend Course of Concess of Concess of |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>इ<br>1<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>and management to the Phy of the same to a montant and and and and and and<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | -------------------------------------------------------------------------------------------------<br>Advertising a libert<br>Concession and Concession Lines Annual Party of<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | Mark<br>16, 1994, 1994, 1999, 1999, 1999, 1999, 1999, 1999, 1999, 1999, 1999, 1999, 1999, 1999, 1999, 1999, 1999, 1999, 1999, 1999, 1999, 1999, 1999, 1999, 1999, 1999, 1999, 1999, 19<br>ﻟﻠﻠﻠ<br>1<br>3<br>1<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>STATISTIC COLUMN OF<br>1 |
### Conclusion:-
The data presented indicate that the Non-sterile Powder Free Multicolor Latex Examination Gloves.
- 1. meets/exceeds ASTM- D3578-00aE2 Standard Specifications For Latex Examination Glove,
- 2. meets FDA pinhole requirements,
- 3. meets FDA claim criterion of a powder free glove.
- 4. meets the protein labelling claims level at <50 µg/ dm2
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the seal of the Department of Health and Human Services, USA. The seal features an abstract eagle design with three stylized lines representing the eagle's body and wings. The text "DEPARTMENT OF HEALTH AND HUMAN SERVICES, USA" is arranged in a circular pattern around the eagle emblem.
### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
SEP 0 1 2005
SGMP Company Limited C/O Janna P. Tucker Tucker & Associates 198 Avenue De La D'Emerald Sparks, Nevada 89434
Re: K052016
Trade/Device Name: Non-Sterile, Latex, Powder-Free Patient Examination Glove (Multicolored), with Protein Labeling Claim (50 Micrograms or Less) Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LYY Dated: August 9, 2005 Received: August 25, 2005
Dear Ms. Tucker:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting ( vour device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 - Ms. Tucker
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements moan that I Drima muse statutes and regulations administered by other Federal agencies. or the Fer of all the Act's requirements, including, but not limited to: registration r od intilety (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice und listing (21 er read vo the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) I mo reter notification. The FDA finding of substantial equivalence of your device to a premaired predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), 11 you desire specific at Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the maj overs Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Syndie Y. Michie Dmd.
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# Indications for Use
Applicant: SGMP Company Ltd.
510(k) Number (if known): K052016
Device Name: _Non-Sterile, Latex, Powder-Free Patient Examination Gloves (Multicolored), with Protein Labeling Claim (50 Micrograms or less)
Indications For Use:
The Non-Sterile, Latex, Powder-Free Patient Examination Gloves (Multicolored), with Protein Labeling Claim (50 Micrograms or less) is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Sheila f. Murphy, M.D. 8/31/15
(Division Sign-Off)
Division of Anesthesiology. General Hospital Infection Control, Dental Devices
510(k) Number: K052016
Page 1 of ____________________________________________________________________________________________________________________________________________________________________
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.