# VNUS RFS, MODEL RFS-6-XX (XX = 10, 12, 13, & 15); VNUS RFS FLEX, MODEL RFX-6-XX (XX = 10, 20, 30 & 40) (K052003)

_Vnus Medical Technologies, Inc. · GEI · Oct 11, 2005 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K052003

## Device Facts

- **Applicant:** Vnus Medical Technologies, Inc.
- **Product Code:** [GEI](/productcode/GEI.md)
- **Decision Date:** Oct 11, 2005
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4400
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The VNUS RFS and RFSFlex Devices are intended for use in vessel and tissue coagulation including: Treatment of incompetent (i.e., refluxing) perforator and tributary veins .

## Device Story

VNUS RFS and RFSFlex are sterile, single-use, bipolar electrosurgical instruments; designed for vessel and tissue coagulation. Device delivers bipolar RF energy to treatment site; relays temperature and feedback to compatible VNUS RF Generator. Used by physicians in clinical settings for treatment of incompetent perforator and tributary veins. Bipolar configuration eliminates need for grounding pads; minimizes damage to adjacent tissue by confining current path. Efficiency allows lower voltage/power settings compared to monopolar systems. Physician selects device configuration and length based on vessel access method and anatomy. Output is thermal coagulation of target tissue; facilitates vein closure.

## Clinical Evidence

Bench testing only. In-vitro testing demonstrated safety and effectiveness for intended function. Biocompatibility of materials confirmed.

## Technological Characteristics

Bipolar electrosurgical instrument; RF energy delivery; sterile, single-use, disposable. Available in 2 bipolar configurations and multiple lengths. Compatible with VNUS RF Generator and Instrument Cable. No grounding pads required.

## Regulatory Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

## Predicate Devices

- VNUS Closure System ([K982816](/device/K982816.md), [K003092](/device/K003092.md), and [K030557](/device/K030557.md))
- VNUS Vessel and Tissue Coagulation System ([K033547](/device/K033547.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

## VNUS® RFS and RFSFlex Devices

K052003

# Summary of Safety and Effectiveness

- Determination of Substantial Equivalence A. VNUS RFS and RFSFlex devices
#### B. Common Name

Bipolar Electrosurgical Instrument

#### C. Predicate Device(s)

VNUS® Closure® System (K982816, K003092, and K030557); and VNUS® Vessel and Tissue Coagulation System (proprietary name: TBD) (K033547)

#### D. Device Description

The VNUS RFS family of devices is a bipolar, high frequency electrosurgical system designed for use in general surgical procedures where blood vessel and tissue coagulation is desired. These devices are compatible with separately cleared Radiofrequency (RF) Generator, and the Instrument Cable manufactured by VNUS Medical Technologies.

The VNUS RFS family of devices are sterile disposable devices intended for a singleuse only. The device's function is to deliver bipolar RF energy to the desired treatment site and relay temperature and other feedback to the RF Generator. The disposable device is available in 2 bipolar configurations and multiple lengths for selection by the physician based on preference for method of vessel access, location and length of the vessel to be treated. This submission reflects a name change, minor modification of the indication statement (more specific inclusion of perforator and tributary veins), and minor design changes.

#### ய் Intended Use

The VNUS RFS and RFSFlex Devices are intended for use in vessel and tissue coagulation including:

- Treatment of incompetent (i.e., refluxing) perforator and tributary veins .
#### Intended Use of Predicate Devices F.

The specified predicate devices are indicated for "coagulation of blood vessels in patients with superficial vein reflux" (VNUS Closure System) and "vessel and tissue coagulation" (VNUS Vessel and Tissue Coagulation System).

#### G. Technological Comparison

RF energy has been widely used in electrosurgical equipment for many years. The safety and efficacy of such devices has been well established for a variety of intended uses. The use of bipolar RF energy delivery has potential advantages over monopolar systems. No grounding pads are required, and the potential for damage to adjacent tissue is minimized, as the patient is no longer the return path for electrical current. The efficiency of bioolar RF energy delivery allows systems such as the VNUS RF Generator to be used at lower voltage and power settings as compared with monopolar systems.

{1}------------------------------------------------

Page 2 of (2)

The technological characteristics and principals of operation of the VNUS RFS family of devices are substantially equivalent to the noted predicate devices relv on the delivery of RF energy to achieve their intended use.

### Discussion of Clinical/Non-Clinical Tests and Conclusions H.

## Performance

Results of in-vitro testing demonstrate that the VNUS RFS and RFSFlex devices are safe and effective for their intended function.

## Biocompatibility

The materials used in the VNUS RFS and RFSFlex Devices have been shown to be biocompatible.

### l. Summary of Safety and Effectiveness

Based upon the intended use, design, materials, function, comparison with currently marketed devices and the non-clinical testing it is concluded that the VNUS RFS and RFSFlex Devices are substantially equivalent to the noted predicate devices. The RF ablation/coagulation of blood vessels (i.e., VNUS Closure) has a well-established history of safe an efficacious use in over 100,000 procedures.

Signature

Brady Esch . Directory, R&D VNUS Medical Technologies, Inc.

{2}------------------------------------------------

Public Health Service

0C1 1 1 2005

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

VNUS Medical Technologies, Inc. c/o Sam Nanavati 750 Chesapeake Drive Redwood City, California 94063

Re: K052003

Trade/Device Name: VNUS® RFS and RFS Flex Devices Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: II Product Code: GEI Dated: July 22, 2005 Received: July 25, 2005

Dear Mr. Nanavati:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adultcration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination docs not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the clectronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

## Page 2- Sam Nanavati

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours.

Harlow Annett
to!

Mark N. Melkerson Acting Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Device Name: Derices VNUS® RFS and RFSFlex

510(k) Number (if known):

): K052003

Indications for Use:

The VNUS RFS and RFSFlex devices are intended for use in vessel and tissue coagulation including:

Treatment of incompetent (i.e., refluxing) perforator and tributary veins .

Over the Counter Use: (Per 21 CFR 801.109)

or

Prescription Use:

(Please do not write below this line - continue on another Page if needed)

CONCURRENCE OF CDRH, OFFICE OF DEVICE EVALUATION (ODE)

Caitano Pereira

(Division Sign-Off) Division of General, Restorative, and Neurological Devices

510(k) Number K052005

. - -

Premarket Notification [510(K)] Submission

Page 42

---

**Source:** [https://fda-staging.innolitics.com/device/K052003](https://fda-staging.innolitics.com/device/K052003)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).