K051983 · Axya Medical, Inc. · MAI · Aug 5, 2005 · Orthopedic
Device Facts
Record ID
K051983
Device Name
MODEL 3000 AXYALOOP BIO-ABSORBABLE BONE ANCHOR
Applicant
Axya Medical, Inc.
Product Code
MAI · Orthopedic
Decision Date
Aug 5, 2005
Decision
SESE
Submission Type
Special
Regulation
21 CFR 888.3030
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Axya Model 3000 AxyaLoop™ Bio-Absorbable Bone Anchor is indicated for securing suture to bone. This device is intended for use in the following applications: Shoulder: Rotator cuff, Bankart, and SLAP lesion repair, Biceps tenodesis, Acromio-Clavicular separation and Deltoid repair, Capsular shift and Capsulolabral reconstruction Foot/Ankle: Lateral and Medial stabilization, Achilles tendon and Metatarsal ligament repair, Hallux Valgus and Midfoot reconstruction Knee: Medial collateral and Lateral collateral ligament repair, Patellar tendon and Posterior oblique ligament repair, Iliotibial band tenodesis Hand/Wrist: Scapholunate ligament, Radial collateral ligament and Ulnar collateral ligament reconstruction Elbow: Biceps tendon reattachment, Tennis elbow repair, Ulnar and Radial collateral ligament reconstruction Pelvis: Bladder neck suspension for urinary incontinence due to urethral hypermobility or intrinsic sphincter deficiency in females
Device Story
Bio-absorbable bone anchor system; 3.0 mm diameter corkscrew thread design; pre-threaded with various suture materials (polypropylene, nylon, polyester, polyethylene). Used in open or arthroscopic soft tissue-to-bone attachment procedures. System includes anchor, thread tap, and placement tool (driver). Operated by surgeons in clinical settings. Anchor provides mechanical fixation of suture to bone; facilitates soft tissue repair. Benefits include secure tissue reattachment and bio-absorbable material properties.
Clinical Evidence
Bench testing only. In vitro studies conducted to evaluate anchor pull-out strength per FDA Guidance Document for Testing Bone Anchor Devices (April 20, 1996). Results demonstrate pull-out force and failure modes comparable to predicate devices.
Technological Characteristics
Bio-absorbable bone anchor; 3.0 mm diameter; corkscrew thread geometry. System includes manual surgical instruments (tap, driver). Sterile, single-patient-use. No software or electronic components.
Indications for Use
Indicated for securing suture to bone in shoulder, foot/ankle, knee, hand/wrist, elbow, and pelvic surgical procedures (including bladder neck suspension for female urinary incontinence).
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
Predicate Devices
Axya Model 5000 AxyaLoop™ Bio-Absorbable Bone Anchor (K041698)
Arthrex Inc. BioFASTak™ Suture Anchor Model AR-1324B (K000506)
Submission Summary (Full Text)
{0}------------------------------------------------
K051983 9'/3
# AUG 5 - 2005
## 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS (Per 21 CFR 807.92)
## General Company Information
- Axya Medical, Inc. Name: Howard Schrayer Contact: Regulatory Affairs Consultant
- 100 Cummings Center Address: Suite 444C Beverly, MA 01915
- (978) 232 9997 Telephone: (978) 232 - 9998 Fax:
- July 21, 2005 Date Prepared
#### General Device Information
| Product Name: | Model 3000 AxyaLoop™ Bio-Absorbable Bone Anchor |
|-----------------|-----------------------------------------------------------------------------|
| Classification: | "Biodegradable soft tissue fixation fastener"<br>Product code: MAI Class II |
#### Predicate Devices
Axya Model 5000 AxyaLoop™ Bio-Absorbable Bone Anchor [510(k) Number K041698]
Arthrex Inc. BioFASTak™ Suture Anchor Model AR-1324B [501(k) Number K000506]
#### Description
The device is designed with a corkscrew style thread 3.0 mm in diameter, for use in a range of soft tissue to bone attachment procedures. The Axya Bone Anchor will be made available as a system together with a thread tap and a placement tool (driver). These accessories are the same types of instruments included in procedure sets for currently marketed bone anchor systems. Axya Medical believes that the accessory instruments are Class I Manual Surgical Instruments and are exempt from the premarket Notification regulations. The Axya Model 3000 Bio-Absorbable Bone Anchor is prethreaded with polypropylene monofilament, nylon, braided polyester, and braided polyethylene suture material.
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows handwritten text that appears to be a combination of letters and numbers. The text reads 'K051983' followed by '1 2/3'. The handwriting is somewhat stylized, with some characters slightly distorted or connected.
The Model 3000 Bio-Absorbable Bone Anchor is designed for use in both standard open surgical procedures and in minimally invasive (arthroscopic) surgical procedures.
## Intended Use (Indications)
The Axya Model 3000 AxyaLoop™ Bio-Absorbable Bone Anchor is indicated for securing suture to bone. This device is intended for use in the following applications:
Shoulder: Rotator cuff, Bankart, and SLAP lesion repair, Biceps tenodesis, Acromio-Clavicular separation and Deltoid repair, Capsular shift and Capsulolabral reconstruction
Foot/Ankle: Lateral and Medial stabilization, Achilles tendon and Metatarsal ligament repair, Hallux Valgus and Midfoot reconstruction
Knee: Medial collateral and Lateral collateral ligament repair, Patellar tendon and Posterior oblique ligament repair, Iliotibial band tenodesis
Hand/Wrist: Scapholunate ligament, Radial collateral ligament and Ulnar collateral ligament reconstruction
Elbow: Biceps tendon reattachment, Tennis elbow repair, Ulnar and Radial collateral ligament reconstruction
Pelvis: Bladder neck suspension for urinary incontinence due to urethral hypermobility or intrinsic sphincter deficiency in females
### Substantial Equivalence
This submission supports the position that the Axya Model 3000 Bio-Absorbable Bone Anchor is substantially equivalent to a number of previously cleared devices, including the Axya Model 5000 AxyaLoop Bio-Absorbable Bone Anchor [510(k) Number K041698] and the Arthrex Inc. BioFASTak Model AR-1324B Suture Anchor.
The 510(k) Notice contains summaries of in vitro studies that were conducted to evaluate the anchor pull-out strength as specified in the FDA Guidance Document for Testing Bone Anchor Devices (dated April 20, 1996).
The data presented demonstrate that the anchor pull-out force of the Axya Model 3000 AxyaLoop Bio-Absorbable Bone Anchor compared favorably with the predicate device of similar corkscrew geometry. The failure mode observed for the Axya anchor was predominately the same as that of the predicate anchor.
The single-patient-use components of the bone anchor system are provided sterile. The suture material and bone anchors are sterilized using a process equivalent to the process used by the original suture manufacturer.
{2}------------------------------------------------
K051983 p³/₃
### Conclusions
Axya Medical, Inc. believes that the information provided establishes that similar legally Axya Mculoar, me. bonovoo that the me clinical applications as the Axya Model 3000 Biomanotod have boom and . The materials from which the Axya device is fabricated have an Absorbable Bono I theneve the nical applications, and devices produced by Axya have been tested in accordance with applicable FDA guidelines.
{3}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
AUG 5 - 2005
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. Howard L. Schrayer Regulatory Affairs Consultant . . . . Axya Medical, Inc. 100 Cummings Center, Suite 444C Beverly, Massachusetts 01915
↓
Re: K051983 K051983
Trade/Device Name: Model 3000 AxyaLoop™ Bio-Absorbable Bone Anchor Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: II Product Code: MAI Dated: July 21, 2005 Received: July 25, 2005
Dear Mr. Schrayer:
We have reviewed your Section 510(k) premarket notification of intent to market the indication We have reviewed your Section 510(x) prematics is substantially equivalent (for the indications
referenced above and have determined the devices marketed in interstate referenced above and nave determined the arreed predicate devices marketed in interstate for use stated in the enclosure) to tegally manced produced Device Amendments, or to
commerce prior to May 28, 1976, the enactment at the Federal Food. Drug commerce prior to May 28, 1970, the chaculters with the provisions of the Federal Food, Drug, devices that have been reclassified in accordance with as proval application (PMA).
and Cosmetic Act (Act) that do not require approval assesses approvisions of the Act . The and Cosmetic Act (Act) that to not require appro the general controls provisions of the Act. The You may, therefore, market the device, sucject to use going mishranding and general controls provisions of the Act merade requirement of the security misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is classified (see aove) mio entire) and in trepulations affecting your device can
may be subject to such additional controls. Existing major regulations FDA may be subject to such additional controls: Extroing may or 898 In addition, FDA may
be found in the Code of Federal Regulations, Title 21, Parts 800 Register be found in the Couc of Pederal Tegarations)
publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean Please be advised that FDA s issualice of a substitution with other requirements of the Act
that FDA has made a determination that your device Forel agemes You must that FDA has made a determination that your are res or other Federal agencies. You must
or any Federal statutes and regulations administered by other not listing or any Federal statures and regulations administration and listing (21
comply with all the Act's requirements, including, but not of seguirements as set comply with all the Act s requirements, menames and starting practice requirements as set
CFR Part 807); labeling (21 CFR Part 801); great 800); and if enplicable, the electr CFR Part 807); labeling (21 CFR 1 at 801); good manata 820); and if applicable, the electronic
forth in the quality systems (QS) regulation (21 CFR 1000-1050 forth in the quality systems (QS) regulation (Sections 531-542 of the Act); 21 CFR 1000-1050.
product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-
{4}------------------------------------------------
# Page 2 - Mr. Howard L. Schrayer
This letter will allow you to begin marketing your device as described in your Section 510(k) This letter will anow you to begin markstill equivalence of your device to a legally
premarket notification. The FDA finding of substantial equivales and thus, normits vour prematket notification. The PDA miding of substanted by and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please extitled If you desire specific advice for your car (240) 276-__. Also, please note the regulation entitled, comact the Office of Complanoo at (210) 276 - 17 (21CFR Part 807.97). You may obtain
"Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtai "Misbranding by reference to premance nouthead.org the Act from the Division of Small
other general information on your responsibilities under the Act from the 1900 C28, 000 other general information on your responsion.iss and of its toll-fire number (800) 638-2041 or Manufacturers, International and Consumer Plastianse and Constitutions try/support/index.html.
Sincerely yours,
Stipt Rlrodu
Mark N. Melkerson Acting Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
### INDICATIONS FOR USE
510(k) Number (if known):
Device Name: Axya, Model 3000 AxyaLoop™ Bio-Absorbable Bone Anchor
Indications For Use:
The Axya Model 3000 AxyaLoop™ Bio-Absorbable Bone Anchor is indicated for securing The Axya Model 5000 / kyazoop - dee in the following applications:
Shoulder: Rotator cuff, Bankart, and SLAP lesion repair, Biceps tenodesis, Acromo-Shoulder: Rotator cun, Dankart, and SEAR shift and Capsulolabral reconstruction
Foot/Ankle: Lateral and Medial stabilization, Achilles tendon and Metatarsal ligament repair, Hallux Valgus and Midfoot reconstruction
Knee: Medial collateral and Lateral collateral ligament repair, Patellar tendon and Posterior oblique ligament repair, Iliotibial band tenodesis
Hand/Wrist: Scapholunate ligament, Radial collateral ligament and Ulnar collateral ligament reconstruction
Elbow: Biceps tendon reattachment, Tennis elbow repair, Ulnar and Radial collateral ligament reconstruction
Pelvis: Bladder neck suspension for urinary incontinence due to urethral hypermobility or intrinsic sphincter deficiency in females
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use_ (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Division of General, Restorative, and Neurological Devices
510(k) Number K051983
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.