The Porous Coated Discovery™ Elbow is intended for cemented use in patients with the following conditions: 1. Non-Inflammatory degenerative joint disease including osteoarthritis, and avascular necrosis. 2. Rheumatoid arthritis. 3. Revision where other devices or treatments have failed. 4. Correction of severe functional deformity. 5. Treatment of acute or chronic fractures with humeral epicondyle involvement, which are unmanageable using other treatments.
Device Story
Porous Coated Discovery™ Elbow is a total elbow prosthesis consisting of humeral and ulnar components. Humeral and ulnar components feature porous coating to facilitate enhanced fixation. Device is intended for cemented use in clinical settings by orthopedic surgeons. It replaces damaged joint surfaces to restore function in patients with degenerative disease, rheumatoid arthritis, or complex fractures. Output is a restored joint articulation. Benefit includes improved joint function and pain relief for patients with severe elbow pathology.
Clinical Evidence
No clinical data provided as a basis of substantial equivalence. Mechanical bench testing was performed to demonstrate device performance.
Technological Characteristics
Total elbow prosthesis comprising humeral and ulnar components. Features porous coating for enhanced fixation. Designed for cemented use. Materials, design, and sizing are similar or identical to the predicate device.
Indications for Use
Indicated for patients requiring total elbow arthroplasty due to non-inflammatory degenerative joint disease (osteoarthritis, avascular necrosis), rheumatoid arthritis, revision of failed prior treatments, severe functional deformity, or acute/chronic fractures with humeral epicondyle involvement unmanageable by other means.
Regulatory Classification
Identification
An elbow joint metal/polymer constrained cemented prosthesis is a device intended to be implanted to replace an elbow joint. It is made of alloys, such as cobalt-chromium-molybdenum, or of these alloys and of an ultra-high molecular weight polyethylene bushing. The device prevents dislocation in more than one anatomic plane and consists of two components that are linked together. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).
Special Controls
*Classification.* Class II. The special controls for this device are:(1) FDA's:
(i) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Devices—Part I: Evaluation and Testing,’ ”
(ii) “510(k) Sterility Review Guidance of 2/12/90 (K90-1),”
(iii) “Guidance Document for Testing Orthopedic Implants with Modified Metallic Surfaces Apposing Bone or Bone Cement,”
(iv) “Guidance Document for the Preparation of Premarket Notification (510(k)) Application for Orthopedic Devices,”
(v) “Guidance Document for Testing Non-articulating, ‘Mechanically Locked’ Modular Implant Components,”
(2) International Organization for Standardization's (ISO):
(i) ISO 5832-3:1996 “Implants for Surgery—Metallic Materials—Part 3: Wrought Titanium 6-Aluminum 4-Vandium Alloy,”
(ii) ISO 5832-4:1996 “Implants for Surgery—Metallic Materials—Part 4: Cobalt-Chromium-Molybdenum Casting Alloy,”
(iii) ISO 5832-12:1996 “Implants for Surgery—Metallic Materials—Part 12: Wrought Cobalt-Chromium-Molybdenum Alloy,”
(iv) ISO 5833:1992 “Implants for Surgery—Acrylic Resin Cements,”
(v) ISO 5834-2:1998 “Implants for Surgery—Ultra High Molecular Weight Polyethylene—Part 2: Moulded Forms,”
(vi) ISO 6018:1987 “Orthopaedic Implants—General Requirements for Marking, Packaging, and Labeling,”
(vii) ISO 9001:1994 “Quality Systems—Model for Quality Assurance in Design/Development, Production, Installation, and Servicing,” and
(viii) ISO 14630:1997 “Non-active Surgical Implants—General Requirements,”
(3) American Society for Testing and Materials':
(i) F 75-92 “Specification for Cast Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implant Material,”
(ii) F 648-98 “Specification for Ultra-High-Molecular-Weight Polyethylene Powder and Fabricated Form for Surgical Implants,”
(iii) F 799-96 “Specification for Cobalt-28 Chromium-6 Molybdenum Alloy Forgings for Surgical Implants,”
(iv) F 981-93 “Practice for Assessment of Compatibility of Biomaterials (Nonporous) for Surgical Implant with Respect to Effect of Material on Muscle and Bone,”
(v) F 1044-95 “Test Method for Shear Testing of Porous Metal Coatings,”
(vi) F 1108-97 “Specification for Titanium-6 Aluminum-4 Vanadium Alloy Castings for Surgical Implants,”
(vii) F 1147-95 “Test Method for Tension Testing of Porous Metal Coatings, ” and
(viii) F 1537-94 “Specification for Wrought Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implants.”
{0}------------------------------------------------
SEP - 6 2005
Kos1975
Image /page/0/Picture/2 description: The image shows the word "BIOMET" in a bold, sans-serif font. The letters are all capitalized and outlined in black. The word appears to be a logo or brand name.
## 510(k) Summary
Applicant/Sponsor: Biomet Manufacturing Corp.
Patricia Sandborn Beres Contact Person: Senior Regulatory Specialist
Proprietary Name: Porous Coated Discovery™ Elbow
Common Name: Constrained Elbow Joint
Classification Name: Elbow joint/polymer constrained cemented prosthesis (21 CFR 888.3150)
Legally Marketed Device To Which Substantial Equivalence Is Claimed: Biomet Discovery™ Elbow (K013042)
Device Description: The Porous Coated Discovery™ Elbow is a total elbow prosthesis Device Desonption: The Porocol component. Placing the humeral articulation comprised of an unfidrar components of components. The humeral components. The humeral through the unfar anticulation into the une and ulnar components are porous coated to provide enhanced fixation.
Intended Use: The Porous Coated Discovery™ Elbow is intended for cemented use in patients with the following conditions:
- patients with the following conditions:
1. Non-Inflammatory degenerative joint disease including osteoarthritis, and avascular necrosis.
- 2. Rheumatoid arthritis.
- 2. Revision where other devices or treatments have failed.
- 4. Correction of severe functional deformity.
- 4. Concection of Sciere functional doferming.
5. Treatment of acute or chronic fractures with humeral epicondyle involvement, which is are unmanageable using other treatments.
Summary of Technologies: The technological characteristics (materials, design, sizing Summal y of Technologics: The toonnological entiled in are similar to or identical to the predicate devices.
Non-Clinical Testing: Mechanical testing was provided to demonstrate that devices ability to perform.
Clinical Testing: None provided as a basis of substantial equivalence.
All trademarks are property of Biomet, Inc.
MAILING ADDRESS P.O. Box 587 Warsaw, IN 46581-0587
E
SHIPPING ADDRESS 56 E. Bell Drive Warsaw, IN 46582
574.267.6639
FAX 574.267.8137
E-MAIL biomet@biomet.com
{1}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH AND HUMAN SERVICES . USA" around the perimeter. Inside the circle is a stylized image of an eagle with three stripes representing the agency's mission to protect the health of all Americans and provide essential human services.
SEP - 6 2005
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Ms. Patricia Sandborn Beres Senior Regulatory Specialist Biomet Manufacturing Corp. P. O. Box 587 Warsaw, Indiana 46581-0578
Re: K051975
K051973
Trade/Device Name: Porous Coated Discovery™ Elbow Praders on Number: 21 CFR 888.3150 Regulation Number: 21 CFR 868.5136
Regulation Name: Elbow joint metal/polymer constrained cemented prosthesis Regulatory Class: II Product Code: JDC Dated: August 17, 2005 Received: August 19, 2005
Dear Ms. Beres:
We have reviewed your Section 510(k) premarket notification of intent to market the device We have reviewed your Section > I0(x) premainer is substantially equivalent (for the indications
referenced above and have determined the device is subserverketed in intersta referenced above and have delemined the devices marketed in interstate
for use stated in the enclosure) to legally marketed president Device Amendments for use stated in the enclosure) to regary manced promoted in orice Amendments, or to
commerce prior to May 28, 1976, the enactment of the Federal Food Drug commerce prior to May 28, 1970, the chaculture with the provisions of the Federal Food. Drug, devices that have been recrassified in accordance was are approval application (PMA).
and Cosmetic Act (Act) that do not require approval assessed approvisions of the Act . T and Cosmetic Act (Act) that do not require approval controls provisions of the Act. The
You may, therefore, market the device, subject to the general construction of listin You may, therefore, market the ucvice, subject to the gents for annual registration, listing of
general controls provisions of the Act include requirements misbranding and general controls provisions of the recentered to quarterolibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is classified (see above) mio ennier enably in the minent of the same of the same of the addition. FDA may may be subject to such additional controls. Existing in harts 800 to 898. In addition, FDA may
be found in the Code of Federal Regulations, Title 21, Parts 800 Locister be found in the Code of Poderal Regaring your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean Please be advised that FDA s issualice or a subscription with other requirements of the Act
that FDA has made a determination that your device aronaise. You must that FDA has made a decemination that your are ros reserve agencies. You must and list or any Federal statutes and regulations administered of registration and listing (21 l
eomply with all the Act's requirements, including, but not limited to: requirements as comply with all the Act S requirements, moradial wantics requirements as set CH Part 807); labeling (21 CFR Part 800); good and 11 applicable, the electronic
forth in the quality systems (QS) regulation (21 CFR 1000 1050 forth in the quality systems (Q5) regalation (Sections 531-542 of the Act), 21 CFR 1000-1050.
product radiation control provisions (Sections 531-542 of the Act), 21 CFR 1000
{2}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k)
The same of the same of a the same of the same of your device to a legal This letter will allow you to begin marketing your article quivalence of your device to a legally
premarket notification. The FDA finding of substantial equivalence of your premarket notification. The PDA initime of substantial by and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please
rs the latest of the many of 10, 2019 - All any desse mate the regulation e If you desire specific advice for your ac not on one of the regulation entitled, the regulation entitled, contact the Office of Comphalled at (217) 276 - 8 at 807.97). You may obtain
"Misbranding by reference to premarket notification" (21 CFR Part 807.97). You may obtain "Misbranding by reference to premarket inclifetes under the Act from the Division of Small other general information on your responsion its toll-free number (800) 638-2041 or
Manufacturers, International and Consumer Assistance and Consumer (index) html Manufacturers, International and Consumer Abolokation on on corner in the manufactures of the
Sincerely yours,
signature
r Mark N. Melkerson Acting Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known):
Device Name: Porous Coated Discovery™ Elbow
Indications For Use: The Porous Coated Discovery™ Elbow is intended for cemented mdications + or with the following conditions:
- use in patients with the following conditions.
1. Non-Inflammatory degenerative joint disease including osteoarthritis, and avascular necrosis.
- 2. Rheumatoid arthritis.
- 2. Revision where other devices or treatments have failed.
- 4. Correction of severe functional deformity.
- 4. Correction of severe inncilonal delomity.
5. Treatment of acute or chronic fractures with humeral epicondyle involvement, which are unmanageable using other treatments.
Prescription Use
(Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use_ (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
| Concurrence of CDRH, Office of Device Evaluation (ODE) | |
|------------------------------------------------------------|-----------|
| (Division Sign-Off) | Page 1 of |
| Division of General, Restorative, and Neurological Devices | |
| 510(k) Number | K081975 |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.