K051869 · Natus Medical, Inc. · LBI · Oct 7, 2005 · General Hospital
Device Facts
Record ID
K051869
Device Name
NEOBLUE COZY LED PHOTOTHERAPY SYSTEM
Applicant
Natus Medical, Inc.
Product Code
LBI · General Hospital
Decision Date
Oct 7, 2005
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5700
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The neoBLUE cozy phototherapy light is intended for the treatment for neonatal hyperbilirubinemia. It is designed to provide phototherapy treatment from underneath the baby. The neoBLUE cozy system must be used within a patient enclosure, such as a bassinet, an open crib, a warming table or an incubator. The neoBLUE cozy system can be used in a clinical setting or in the home.
Device Story
Portable phototherapy system delivering high-intensity blue light (400-550 nm) via LEDs to treat neonatal hyperbilirubinemia; device placed under infant within bassinet, crib, warming table, or incubator. System comprises light source, mattress, disposable cover, bumper, and power supply. Operated by clinicians or caregivers in clinical or home settings. Blue light degrades bilirubin; LEDs emit no significant UV radiation. Output intensity adjustable via internal potentiometer; thermal protection circuit automatically disables LEDs if overheating occurs. Healthcare providers use device to provide phototherapy; can be used with overhead units for increased coverage. Benefits include effective bilirubin degradation without UV exposure risk.
Clinical Evidence
Bench testing only. Results of prototype testing to specifications were as expected; no new safety or effectiveness issues identified.
Technological Characteristics
Blue LED light source (400-550 nm, peak 450-470 nm); intensity 30-35 µW/cm²/nm. Materials: polycarbonate cover, polyurethane base, nonwoven spunlaced polyester mattress cover, foam bumper with cotton fabric. Electrical safety: EN 60601-1, UL 2601-1, CSA C22.2 601.1, EN 60601-2-50. Ingress protection: IPX4. Thermal safety: cooling fan and thermal protection circuit. Portable form factor.
Indications for Use
Indicated for treatment of neonatal hyperbilirubinemia in neonates.
Regulatory Classification
Identification
A neonatal phototherapy unit is a device used to treat or prevent hyperbilirubinemia (elevated serum bilirubin level). The device consists of one or more lamps that emit a specific spectral band of light, under which an infant is placed for therapy. This generic type of device may include supports for the patient and equipment and component parts.
{0}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. Inside the circle is a stylized image of three human profiles facing to the right, with a flowing design below them.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
NOV - 7 2005
Ms. Luanne Ng Regulatory/Clinical Affairs Manager Natus Medical, Incorporated 1501 Industrial Road San Carlos. California 94070
Re: K051869
Trade/Device Name: NEOBLUE COZY LED PHOTOTHERAP Y SYSTEM Regulation Number: 21 CFR 880.5700 · Regulation Name: Neonatal phototherapy unit Regulatory Class: II Product Code: LBI Dated: October 21, 2005 Received: October 24, 2005
Dear Ms. Ng:
This letter corrects our substantially equivalent letter of October 7, 2005.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Page 2 - Ms. Ng
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to continue marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Sajette Y. Michael Omd
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# INDICATIONS FOR USE STATEMENT
510(k) Number (if known):
Device Name: neoBLUE cozy™ LED Phototherapy System
INDICATIONS FOR USE: The neoBLUE cozy phototherapy light is intended for the treatment for neonatal hyperbilirubinemia. It is designed to provide phototherapy treatment from underneath the baby. The neoBLUE cozy system must be used within a patient enclosure, such as a bassinet, an open crib, a warming table or an incubator. The neoBLUE cozy system can be used in a clinical setting or in the home.
Prescription Use × AND/OR Over-The-Counter Use (Part 21 CFR 801 Supbart B) (21 CFR 801 Subpart C)
(PLEASE DONOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Anten Dm
Division of Anesthesiology, General Hospital, Infection Contion Dental Devices
3 10(k) Number ..
{3}------------------------------------------------
#### 8 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
#### Submitter
Natus Medical Inc. 1501 Industrial Road San Carlos, CA 94070 Phone: 650-802-6400 FAX: 650-802-0401
Contact: Luanne Ng Regulatory/Clinical Affairs Manager Date Prepared: July 7, 2005 Device Names Classification name Unit, Neonatal Phototherapy Common name Phototherapy Light Trade Name Natus® ncoBLUE cozy systemPhototherapy Unit
### Predicate Devices
Medela Bilibed, K962612 Natus Blue Light Phototherapy Unit, K022196
### Device Description
The neoBLUE cozy system is a portable phototherapy light that delivers a narrow band of high-intensity blue light via blue light emitting diodes (LEDs) to provide treatment for neonatal hyperbilirubinemia. The neoBLUE cozy system is designed to vrovide phototherapy treatment from underneath the baby. The neoBLUE cozy system must be used within a patient enclosure, such as a bassinett, an open crib, a warming tat hlast or incubator.
Blue LEDs emit light in the range of 400 -- 550 nm (peak wavelength 450-470 nm). This range corresponds to the spectral absorption of light by bilirubin, and is thus considered to be the most effective for the degradation of bilirubin. Blue LEDs do not emit significant energy in the ultraviolet (UV) region of the spectrum, so there is no concern about UV exposure to the infant. As with all phototherapy lights, protective eyeshades must be used to protect the infant's eyes from excessive light exposure.
LEDs have minimal light output degradation over their lifetime with proper use. Nevertheless, the biomedical engineer can adjust the output of the LEDs using the
{4}------------------------------------------------
potentiometer located behind the filter cover at the flat end of the light enclosure. The system is expected to operate at intensities above 30 uW/cm2/nm for approximately 3.000 hours.
The neoBLUE cozy system consists of five components: the neoBLUE cozy light source (light), the mattress, the disposable mattress cover, the bumper accessory, and the power supply.
### Intended Use
The neoBLUE cozy system is a portable phototherapy light that delivers a narrow band of high-intensity blue light via blue light emitting diodes (LEDs) to provide treatment for neonatal hyperbilirubinemia. The neoBLUE cozy system is designed to provide phototherapy treatment from underneath the baby. The neoBLUE cozy system must be used within a patient enclosure, such as a bassinett, an open crib, a warming table or an incubator. The neoBLUE cozy system can be used in a clinical setting or in the home.
Blue LEDs emit light in the range of 400 – 550 nm (peak wavelength 450-470 nm). This range corresponds to the spectral absorption of light by bilirubin, and is thus considered to be the most effective for the degradation of bilirubin. Blue LEDs do not emit significant energy in the ultraviolet (UV) region of the spectrum, so there is no concern about UV exposure to the infant. As with all phototherapy lights, protective eyeshades. must be used to protect the infant's eyes from excessive light exposure
The neoBLUE cozy system can be used in conjunction with an overhead phototherapy light for more patient surface area coverage. A simple blanket may be placed over the patient during phototherapy treatment with the neoBLUE cozy system to praced or the surroundings from stray blue light.
## Comparison with Predicate Device
The neoBLUE cozy system, the Medela BiliBed and the Natus Blue Light Phototherapy Unit have the same intended use of treatment of neonatal hyperbilirubinemia and use the same operating principle of delivery of light to degrade bilirubin. See the Comparison Table below for details.
| Feature | Natus neoBLUE<br>cozy LED<br>Phototherapy<br>System | Medela BiliBed<br>(K962612) | Natus Blue Light<br>Phototherapy Unit<br>(K022196) |
|-------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------|----------------------------------------------------------------------------------------|
| Intended Use | For the treatment of<br>neonatal<br>hyperbilirubinemia | For the treatment of<br>neonatal<br>hyperbilirubinemia | For the treatment of<br>neonatal<br>hyperbilirubinemia |
| Treatment Method | Underbaby<br>phototherapy | Underbaby<br>phototherapy | Overhead<br>phototherapy |
| Targeted Population | Neonates | Neonates | Neonates |
| Sites of Use | Clinical setting,<br>home-use | Clinical setting,<br>home-use | Clinical setting |
| Feature | Natus neoBLUE<br>cozy LED<br>Phototherapy<br>System | Medela BiliBed<br>(K962612) | Natus Blue Light<br>Phototherapy Unit<br>(K022196) |
| Specifications | | | |
| Type of Device | Free standing device | Free standing device | Free standing device |
| Type of Light | Blue light (LED) | Blue light<br>fluorescent | Blue light (LED) |
| Intensity | 30-35 µW/cm²/nm | Approx. 40-60 µW<br>/cm²/nm or more | 30 – 35 µW/cm²/nm |
| Height | <6.4 cm (2.5 in)<br>patient surface,<br>≤12.7 cm (5.0 in)<br>rest of device | 13.0 cm (5.1 in),<br>including baby<br>support | Not Applicable |
| Width and Length | 30.5 cm (12.0 in)<br>width x 64.8 cm<br>(25.5 in) length | 32.6 cm (12.8 in)<br>width x 63 cm<br>(24.8 in) length | Not Applicable |
| Weight | <4.3 kg (9.5 lbs) | 5 kg (11 lbs) | Not Applicable |
| Treatment area | Minimum 613 cm²<br>(95 in²) | Not Published | 1250 cm² (200 in²) |
| Materials | | | |
| Device | Polycarbonate cover<br>& polyurethane base | Aluminum frame,<br>plastic base | Not Applicable |
| Mattress | Polyurethane cover<br>with polyolefin<br>bubble cushioning | Polyurethane plastic | Not Applicable |
| Mattress Cover | Nonwoven<br>spunlaced polyester | Bilicombi – 100%<br>cotton or 100%<br>nonwoven<br>spunlaced fabric | Not Applicable |
| Patient side<br>cushioning/Jacket | Foam bumper<br>encased by 100%<br>cotton fabric | Bilicombi – 100%<br>cotton or 100%<br>nonwoven<br>spunlaced fabric | Not Applicable |
| Miscellaneous | | | |
| Shape | Oval with flat end at<br>foot area | Rectangle | Not Applicable |
| Portable | Yes with carrying<br>case | Yes with carrying<br>case | Not Applicable |
| Patient side<br>cushioning/Jacket | Foam bumper<br>encased by 100%<br>cotton fabric | Bilicombi – 100%<br>cotton or 100%<br>nonwoven<br>spunlaced fabric | Not Applicable |
| Feature | Natus neoBLUE<br>cozy LED<br>Phototherapy<br>System | Medela BiliBed<br>(K962612) | Natus Blue Light<br>Phototherapy Unit<br>(K022196) |
| Standards and Safety | | | |
| Electrical Safety | EN 60606-1<br>UL 2601-1<br>CSA C22.2 601.1<br>EN 60601-2-50 | UL 2601-1<br>IEC 529 | EN 60606-1<br>UL 2601-1<br>CSA C22.2 601.1<br>EN 60601-2-50 |
| Mechanical Safety | Disposable cover<br>acts as a diffuser to<br>minimize accidental<br>viewing of single<br>point light source | Not Applicable | Plastic diffuser<br>minimizes<br>accidental viewing<br>of single point light<br>source |
| Thermal Safety | Fan to cool<br>circuitry, minimize<br>device heating.<br>Thermal protection<br>circuit - turns off<br>LEDs if device gets<br>too warm. | Fan to cool | Fan to cool<br>circuitry, minimize<br>device heating |
| Radiation Safety | LED light source<br>emits no significant<br>ultraviolet light | Fluorescent light<br>sources may<br>produce minimal<br>ultraviolet light | LED light source<br>emits no significant<br>ultraviolet light |
| Ingress of Liquids | IPX4 | IPX1 | Not Applicable |
| Human Factors | | | |
| Controls and<br>Indicators | On/Off switch.<br>Overheat indicator<br>light | On/Off switch | On/Off switch.<br>High/Low intensity<br>switch. Positioning<br>light. |
| Compatibility with<br>environment or<br>other devices | Used inside<br>bassinet, open crib,<br>warmer table,<br>incubator | Used inside bassinet<br>and/or under radiant<br>warmer | Used in conjunction<br>with a bassinet,<br>open crib, warmer<br>table, incubator |
051860
2.55
{5}------------------------------------------------
{6}------------------------------------------------
## Summary of Nonclinical Testing
This submission includes the results of testing of prototype devices to specifications. The results were as expected and no new issues of safety and effectiveness were raised as a result of the nonclinical testing .
52869
4 15
{7}------------------------------------------------
### Conclusion
Based on the data and the information presented in this submission, the neoBLUE cozy LED Phototherapy System is substaintially equivalent to the currently marketed Medela BiliBed and the Natus Blue Light Phototherapy Unit (for the blue LED light only).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.