VITEK 2 GRAM POSITIVE ERTAPENEM FOR STREPTOCOCCUS PNEUMONIAE

K051849 · bioMerieux, Inc. · LON · Aug 16, 2005 · Microbiology

Device Facts

Record IDK051849
Device NameVITEK 2 GRAM POSITIVE ERTAPENEM FOR STREPTOCOCCUS PNEUMONIAE
ApplicantbioMerieux, Inc.
Product CodeLON · Microbiology
Decision DateAug 16, 2005
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.1645
Device ClassClass 2
AttributesAI/ML

AI Performance

OutputAlgorithmAcceptanceObservedDev DSDev ReadersTest DSTest Readers
Antimicrobial susceptibility of Streptococcus pneumoniae to ertapenem94.9% Category AgreementClinical study at three sites using fresh clinical isolates, stock isolates, and a challenge set

Indications for Use

The VITEK®2 Antimicrobial Susceptibility Test is intended to be used with the VITEK®2 System for the automated quantitative or qualitative susceptibility testing of isolated colonies for most clinically significant aerobic gram-negative bacilli, Staphylococcus spp., Enterococcus spp., Streptococcus agalactiae, and S. pneumoniae. The VITEK®2 Gram Positive Susceptibility Card is intended for use with the VITEK®2 system in clinical laboratories as an in vitro test to determine the susceptibility of Streptococcus pneumoniae to antimicrobial agents when used as instructed in the Online Product Information.

Device Story

VITEK®2 system performs automated antimicrobial susceptibility testing (AST) on isolated colonies. Device uses 64-microwell cards containing premeasured antibiotics and culture medium. Organism suspension (McFarland 0.5) is prepared manually or automatically, vacuum-filled into cards, and sealed. Instrument incubates cards at 35.5°C while performing optical scanning every 15 minutes. Transmittance optics measure organism growth by monitoring light blockage. Discriminate analysis algorithm processes growth characteristics to calculate MIC and determine category interpretation (S, I, R). Results assist clinicians in selecting appropriate antibiotic therapy for S. pneumoniae infections.

Clinical Evidence

External evaluation conducted using fresh and stock clinical isolates and stock challenge strains. Performance compared against CLSI (NCCLS) microbroth dilution reference method. Results demonstrated 94.9% overall Category Agreement. Reproducibility and Quality Control testing yielded acceptable results.

Technological Characteristics

Miniaturized AST card containing antimicrobial medium. Sensing principle: optical monitoring of bacterial growth in wells over 18 hours. Connectivity: integrated with VITEK 2 and VITEK 2 Compact instrument platforms. Software-based automated analysis of growth kinetics to determine MIC.

Indications for Use

Indicated for antimicrobial susceptibility testing of Streptococcus pneumoniae isolates. Intended for use in clinical laboratory settings to determine in vitro susceptibility to antimicrobial agents.

Regulatory Classification

Identification

A fully automated short-term incubation cycle antimicrobial susceptibility system is a device that incorporates concentrations of antimicrobial agents into a system for the purpose of determining in vitro susceptibility of bacterial pathogens isolated from clinical specimens. Test results obtained from short-term (less than 16 hours) incubation are used to determine the antimicrobial agent of choice to treat bacterial diseases.

Special Controls

*Classification.* Class II (special controls). The special control for this device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA.”

Predicate Devices

Submission Summary (Full Text)

{0} 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE A. 510(k) Number: k051849 B. Purpose for Submission: Addition of ertapenem to the Vitek®2 Antimicrobial Susceptibility Test (AST) System C. Measurand: Ertapenem ≤0.5 – ≥ 8 μg/ml D. Type of Test: Qualitative AST growth based detection E. Applicant: bioMerieux, Inc. F. Proprietary and Established Names: VITEK®2 Gram Positive Susceptibility Card G. Regulatory Information: 1. Regulation section: 21 CFR 866.1645 Fully Automated Short-Term Incubation Cycle Antimicrobial Susceptibility System 2. Classification: II 3. Product code: LON System, Test, Automated, Antimicrobial Susceptibility, Short Incubation 4. Panel: 83 Microbiology H. Intended Use: 1. Intended use(s): The VITEK®2 Antimicrobial Susceptibility Test is intended to be used with the VITEK®2 System for the automated quantitative or qualitative susceptibility testing of isolated colonies for most clinically significant aerobic gram-negative bacilli, Staphylococcus spp., Enterococcus spp., Streptococcus agalactiae, and S. pneumoniae. {1} The VITEK®2 Gram Positive Susceptibility Card is intended for use with the VITEK®2 system in clinical laboratories as an in vitro test to determine the susceptibility of *Streptococcus pneumoniae* to antimicrobial agents when used as instructed in the Online Product Information. 2. Indication(s) for use: This submission is for the addition of the antibiotic ertapenem at concentrations of 0.5, 1, and 4 ug/ml for a calling range of ≤ 0.5 – ≥ 8 ug/ml for testing *S. pneumoniae* only. 3. Special conditions for use statement(s): For prescription use 4. Special instrument requirements: N/A I. Device Description: Each VITEK®2 test card contains 64 microwells. A control well, that contains only microbiological culture medium is resident on all cards, with the remaining wells containing premeasured amounts of a specific antibiotic combined with culture medium. A suspension of organism is made in 0.45-0.5% sterile saline from a pure culture and standardized to a McFarland 0.5 standard using the DensiChek. The desired card(s) are placed in the cassette along with an empty tube for the susceptibility card. The cassette is placed in the VITEK®2 instrument where a susceptibility test will be automatically diluted from the ID suspension by the VITEK®2. The cards are then automatically vacuum filled; the tubes are cut and the cards sealed prior to proceeding to the Incubator Loading Station. Cards are then transferred from the cassette into the carousel for incubation (35.5° C) and optical scanning during testing. Readings are performed every 15 minutes. In addition to the automatic dilution, there is also a manual inoculation dilution procedure described in the package insert. J. Substantial Equivalence Information: 1. Predicate device name(s): VITEK®2 Gram Positive Susceptibility Card for Gatifloxacin 2. Predicate 510(k) number(s): k032314 3. Comparison with predicate: {2} | Similarities | | | | --- | --- | --- | | Item | Device | Predicate | | Intended Use | Determine antimicrobial susceptibility to antimicrobial agents | Same | | Test Organism | Gram Positive Cocci | Same | | Test Card | VITEK®2 card format with base broth | Same | | Instrument | VITEK®2 System | Same | | Differences | | | | --- | --- | --- | | Item | Device | Predicate | | Antibiotic | Ertapenem at specific concentrations | Gatifloxacin at specific concentrations | | Reading algorithm | Unique for ertapenem | Unique for gatifloxacin | K. Standard/Guidance Document Referenced (if applicable): “Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test Systems; Guidance for Industry and FDA”; CLSI M7 (M100-S15) “Methods for Dilution Antimicrobial Susceptibility Tests for Bacteria That Grow Aerobically; Approved Standard.” L. Test Principle: Optics systems use visible light to directly measure organism growth. These transmittance optics are based on an initial light reading of a well before significant growth has begun. Periodic light transmittance samplings of the same well measure organism growth by how much light is prevented from going through the well. An interpretive call is made between 4 and 16 hours for a “rapid” read but may be extended to 18 hours in some instances. The VITEK®2 Susceptibility Card test is based on the microdilution minimum inhibitory concentration technique with concentrations equivalent to standard method concentrations. Several parameters based on the growth characteristics observed are used to provide appropriate input for the MIC calculations. Discriminate analysis is used to develop the algorithm that determines the susceptibility result for all antimicrobials on the VITEK®2 system. The MIC result must be linked to organism identification in order to determine a category interpretation. A category interpretation will be reported along with a MIC. M. Performance Characteristics (if/when applicable): 1. Analytical performance: a. Precision/Reproducibility: Ten gram-positive on-scale organisms were tested three times at three sites to determine within site and site to site reproducibility demonstrating >95% reproducibility. This testing was performed using both the manual dilution of {3} the inoculum and also the automatic dilution method. b. Linearity/assay reportable range: Not applicable c. Traceability, Stability, Expected values (controls, calibrators, or methods): The recommended QC isolates were tested on every test occasion with the reference method and the VITEK®2. The reference method QC results were in range for every day tested. The VITEK®2 was tested a sufficient number of times to demonstrate that the system can produce QC results in the recommended range. Quality Control was performed during the studies using both the auto-dilution and the manual method of diluting the organisms. Results demonstrated that methods were comparable with the same mode. Quality Control Table | ORGANISM | VITEK®2 Conc. (ug/mL) | VITEK®2 Auto | VITEK®2 Manual | Reference Conc. (ug/mL) | Reference | | --- | --- | --- | --- | --- | --- | | | | | | | | | S. pneumoniae ATCC 49619 Expected Range: 0.03 – 0.25 μg/ml | ≤0.5 | 65 | 64 | 0.0625 | 17 | | | 1 | | 1 | 0.125 | 48 | | | 2 | | | 0.25 | | | | 4 | | | 0.5 | | | | 8 | | | 0.8 | | | | | | | | | Inoculum density control was monitored using the DensiChek instrument. This was standardized weekly with all results recorded and in the expected range. Verification was performed during internal testing. d. Detection limit: Not Applicable e. Analytical specificity: Not Applicable f. Assay cut-off: Not Applicable 2. Comparison studies: a. Method comparison with predicate device: A clinical study was conducted at three sites using the VITEK®2 gram positive cards with ertapenem and the broth microdilution method using {4} Mueller Hinton (MH) broth with lysed horse blood prepared as recommended by CLSI. Inoculum was prepared with direct colony suspension. The testing included both fresh clinical isolates and stock isolates along with a challenge set with known results. Two methods of inoculation (manual and automated) were evaluated. Clinical testing was performed using the automated method of inoculation and the challenge set was tested using both the manual and the automated method. The test device had a growth rate of >95%. All isolates grew in the VITEK®2 cards in less than 16 hours. Essential agreement was not calculated because the VITEK®2 card contained <5 dilutions of ertapenem. A comparison was provided to the reference method with the following agreement. Summary Table for S. pneumoniae | | CA Tot | CA N | CA % | #R | Min | maj | vmj | | --- | --- | --- | --- | --- | --- | --- | --- | | Clinical | 311 | 297 | 95.5 | 0 | 14 | 0 | 0 | | Challenge | 80 | 74 | 92.5 | 4 | 5 | 1 | 0 | | Combined | 391 | 371 | 94.9 | 4 | 19 | 1 | 0 | CA-Category Agreement R-resistant isolates maj-major discrepancies vmj-very major discrepancies min- minor discrepancies CA is when the interpretation of the reference method agrees exactly with the interpretation of the VITEK®2 results. Manual Dilution: The challenge set of organisms was also tested at one site using the manual method of inoculation with the following performance that demonstrated that there was little or no difference between the two inoculation methods. Summary Table for S. pneumoniae | | CA Tot | CA N | CA % | #R | Min | maj | vmj | | --- | --- | --- | --- | --- | --- | --- | --- | | Challenge | 80 | 73 | 91.3 | 4 | 7 | 0 | 0 | b. Matrix comparison: Not Applicable 3. Clinical studies: a. Clinical Sensitivity: Not Applicable b. Clinical specificity: Not Applicable {5} c. Other clinical supportive data (when a. and b. are not applicable): Not Applicable 4. Clinical cut-off: Not Applicable 5. Expected values/Reference range: S. pneumoniae ≤ 1 (S), 2 (I), ≥4 (R) N. Proposed Labeling: The Interpretative criteria, QC isolates and the expected ranges are the same as recommended by FDA and CLSI. All values will be included in the package insert. O. Conclusion: The submitted information in this premarket notification is complete and supports substantial equivalence decision. 6
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...