SAF-T-SYRINGE/SAF-T-NEEDLE

K051762 · Safety Medical International, Incorporated · MEG · Sep 23, 2005 · General Hospital

Device Facts

Record IDK051762
Device NameSAF-T-SYRINGE/SAF-T-NEEDLE
ApplicantSafety Medical International, Incorporated
Product CodeMEG · General Hospital
Decision DateSep 23, 2005
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.5860
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Saf-T-Syringe's primary intended use is to provide a safe, accurate and reliable method of injecting medication into a patient. The syringe is available in a 3cc, 5cc and 10cc with various needle sizes. Because the contaminated needle automatically withdraws into the needle plunger the syringe is protected from accidental needle sticks

Device Story

Saf-T-Syringe is a sterile, disposable, one-handed, passively activated, automatic retraction piston syringe; available in 3cc, 5cc, and 10cc volumes with interchangeable 18G-25G needles. Device functions by automatically withdrawing the contaminated needle into the plunger after injection, rendering the syringe inoperable for reuse. Used by clinicians to administer subcutaneous and intramuscular injections; primary benefit is the reduction of accidental needle stick injuries. No external power or complex processing required; operation is purely mechanical.

Clinical Evidence

No clinical data. Bench testing only. Biocompatibility testing performed per ISO 10993-1, including cytotoxicity, sensitization, and irritation. Sterilization validated per ANSI/AAMI/ISO 11135.

Technological Characteristics

Piston syringe with automatic needle retraction mechanism. Materials: biocompatible per ISO 10993. Sterilization: Ethylene Oxide per ANSI/AAMI/ISO 11135. Sizes: 3cc, 5cc, 10cc. Needle gauges: 18G-25G. Mechanical operation; no software or electronic components.

Indications for Use

Indicated for subcutaneous and intramuscular medication injection in patients requiring such therapy. Available in 3cc, 5cc, and 10cc sizes with 18G to 25G needles. Designed to prevent accidental needle stick injuries via automatic needle retraction.

Regulatory Classification

Identification

A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # SECTION 5: 510(k) RELEASABLE SUMMARY #### 510(k) Summary (As Required by 21 section 807.92 (c)) - 1. Submitter Name: Safety Medical International, Incorporated - 2. Address: 2055 Sprint Blvd Apopka, FL 32703 - 3. Phone: (407) 880-2301 - 4. Fax: (407) 880-3357 - 5. Contact Person: David L. Zarrilli, Chief Executive Officer - 6. Date Summary Prepared: June 29, 2005 - 7. Device Trade or Proprietary Name: Saf-T-Syringe - 8. Device Common or usual name: Retractable Needle Safety Syringe - 9. Device Classification Name: Piston Syringe, AntiStick, with Hypodermic Needle - 10. Substantial Equivalency is claimed against the following devices: POP-N-LOCK SYRINGE ( currently marketed as the VANISHPOINT™ SYRINGE) from RETRACTABLE TECHNOLOGIES, INC., 510k #k946219, BD INTEGRA™ Syringe, from BECTON DICKINSON MEDICAL SURGICAL, 510(k) #k011103. ### 11. Description of the Device: The Saf-T-Syringe is a 3cc, 5cc, and 10cc, sterile, non-toxic, non-pyrogenic, disposable, one-hand operated, passively activated, automatic retraction, plunger type, Anti-Stick syringe with a unique, dedicated luer-lock hypodermic needle, as detailed within this submission and in relevant patents. The syringe includes an interchangeable needle; 18 gauge to 25 gauge and lengths of 5/8", 1" and 1-1/2". The primary intended use of the device is to administer safe and accurate subcutaneous and intramuscular injections. Its secondary intended use is to retract and contain the contaminated needle after injection which renders the syringe inoperable for reuse, while not introducing any additional steps to the one-handed operating method, for the purpose of aiding in the prevention of accidental needle stick injuries. ### 12. Intended Use of the Device: The Saf-T-Syringe's primary intended use is to provide a safe, accurate and reliable method of injecting medication into a patient. The syringe is available in a 3cc, 5cc and 10cc with various needle sizes. Because the contaminated needle automatically withdraws into the needle plunger the syringe is protected from accidental needle sticks {1}------------------------------------------------ K051762 ## 13. Safety and Effectiveness of the Device: This device is safe and effective as the predicated devices cited above The Saf-T-Syringe is sterilized using the Ethylene Oxide Sterilization method. The Sal-1 -Oyninge is assured using this sterilization method, which is The stenlity of the dones to assistance with ANSI/AAMI/ISO 11135-1994, "Medical Devices validation and Routine Control of Ethylene Oxide Sterilization" and BS EN 1174; Validation and Nouthe Souther of Ecripes – Part 1: Estimation of the Parts 1-3(1950-1991) Stoninzation of Microorganisms Part 2. Guidance; and Fopulation of Mioroorganisms on Falidation of Microbiological Techniques." The r art of Surate to the ilized to provide a Sterility Assurance Level (SAL) of 108 The Saf-T-Syringe is classified as an Externally Communicating Device, Blood The Our T Oynings to Use (< 24 hours). The device's blood contacting materials were tested in accordance with the tests recommended in the FDA General word toolou in toosel im #G95-1 (5/1/95): "Use of International Standard ISO 1 rogram Memoranualm » 000 + (31.000) Devices Part 1: Evaluation and Testing" 10000 - Biological Evaluation), 5, 10, 11 (1993-2004) "Biological Evaluation and 100 10000 Part 1: Evaluation and Testing; Part 4 (AMD 1): Selection of of Mcdical Devioco - Part 1: Ed. Part 5: Test for In Vitro Cytotoxicity; Part 10: Tests for Internation with Diyed Type Hypersensitivity; and Part 11: Tests for Teste for imation and Diage of the testing demonstrate that the blood contacting materials are biocompatible. ## 14. Summary Comparing Technological Characteristics With Predicate Devices: Safety Medical International, Incorporated makes a Substantial Equivalence Oalery Modioan Internance Retractable Technologies, Inc., Pop-n-Lock Syringe (currently marketed as the Vanishpoint™ syringe), 510(k) #k946219, and to (ourrently marketed Syringe (currently marketed as the Integra™ syringe) 510(k) #011103. All three syringes are similar, and in some cases the Synngo) o re(n) ir o roarts, design, material, operating procedure, and intended sumo, while rogarde to to to to VanishPoint™, and the Integra™, all consist of a syringe barrel, syringe plunger, single lumen hypodermic needle and needle hub, and all have a retracting mechanism on the syringe plunger tip and the needle stem. {2}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image is a black and white circular seal for the Department of Health and Human Services - USA. The seal features the department's emblem, which is a stylized depiction of an eagle with its wings spread. The emblem is positioned to the right side of the seal. The text "DEPARTMENT OF HEALTH AND HUMAN SERVICES - USA" is arranged around the upper half of the circle. Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 SEP 2 3 2005 Mr. David L. Zarrilli Chief Executive Officer Safety Medical International, Inc. 2055 Sprint Boulevard. Apopka, Florida 32703 Re: K051762 Trade/Device Name: SAF-T-SYRINGE/SAF-T-NEEDEL Regulation Number: 21 CFR 880.5860 Regulation Name: Piston Syringe Regulatory Class: II Product Code: MEG, FMF Dated: June 29, 2005 Received: June 30, 2005 Dear Mr. Zarrilli: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ #### Page 2 - Mr. Zarrilli Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits vour device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Chiu Lin, Ph.D. Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # SECTION 4: STATEMENT OF INDICATIONS OF USE # INDICATIONS FOR USE 510(k) Number (if known): 051762 Device Name: Saf-T-Syringe Indications for Use: The Saf-T-Syringe's primary intended use is to provide a safe, accurate and reliable method of injecting medication into a patient. The syringe is available in a 3cc, 5cc and 10cc with various needle sizes. The needle automatically withdraws and prevents accidental needle sticks. 3cc, 5cc, and 10cc Syringes are available for: OR Over-the-Counter Use Prescription Use × (Part 21 CFR 801, Subpart D) (Part 21 CFR 801, Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrance of CDRH. Office of Device Evaluation (ODE) Oxtin Vma (Division Sign-Off) Division of Anesthesiology, General Hospital, Infection Control, Dental Devices 510(k) Number KuS1762
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...