K051753 · Oncology Systems, Inc. · JAQ · Mar 29, 2006 · Radiology
Device Facts
Record ID
K051753
Device Name
ACCU-SOURCE
Applicant
Oncology Systems, Inc.
Product Code
JAQ · Radiology
Decision Date
Mar 29, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.5700
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The ACCU-SOURCE™ remote controlled high dose rate afterloading brachytherapy system is intended for the treatment of cancer and other lesions by intracavitary, interstitial, intralumenal, intraoperative, endobronchial and surface applicator irradiation treatments.
Device Story
Accu-Source™ is a remote afterloading system for High Dose Rate (HDR) Brachytherapy. It utilizes a radioactive source (Ir192 or Yb169) housed in a tungsten-shielded safe. The system includes an operator console and a remote afterloading device. Encoders and motors, controlled by software, drive the source from the safe through channels to the patient to deliver a predetermined radiation dose. The system features 24 channels, supports up to 1200 dwell points, and includes an integrated radiation monitor and emergency manual retraction. It is operated by clinical staff in a controlled environment. The output is the precise positioning of the radioactive source to treat target tissue. The system includes battery backup and an uninterruptible power supply to ensure treatment data is maintained during power failures. Clinical benefit is the targeted delivery of radiation for cancer treatment while minimizing radiation exposure to staff through remote operation.
Indicated for treatment of cancer and other lesions via intracavitary, interstitial, intralumenal, intraoperative, endobronchial, and surface applicator irradiation. Prescription use only.
Regulatory Classification
Identification
A remote controlled radionuclide applicator system is an electromechanical or pneumatic device intended to enable an operator to apply, by remote control, a radionuclide source into the body or to the surface of the body for radiation therapy. This generic type of device may include patient and equipment supports, component parts, treatment planning computer programs, and accessories.
Predicate Devices
Nucletron-Oldeft Corporation Microselectron-HDR Version 2 (K953946)
Submission Summary (Full Text)
{0}------------------------------------------------
MAR 2 9 2006
K051753
### 510(k) Summary
Submitter: Oncology Systems, Inc.
Address: 206 N. Randolph Street, Suite 300 Champaign, IL 61820
Phone number: (217) 355-4460
Fax number: (217) 355-4470
Contact Person: Jennifer Williams
Date Prepared: May 27, 2005
Trade Name: Accu-Source™
Common Name: Remote Control High Dose Rate Afterloading Brachytherapy Device
Classification name: Remote Controlled Radionuclide System, 90 JAQ, Regulation 21 CFR క్ర892.5700.
Substantial Equivalence claimed to:
- 1. Nucletron-Oldeft Corporation K953946 Microselectron-HDR Version 2
{1}------------------------------------------------
Kos1753
### Description:
The Accu-Source™ is a remote afterloading system for High Dose Rate Brachytherapy. The system is designed to deliver a predetermined dose of radiation to organs and tissue via a radioactive source.
The ACCU-SOURCE™ system includes an operator console and a remote afterloading device. The ACCU-SOURCE™ system has a tungsten shielded storage safe which houses a radioactive source when the system is not in use. There are encoders and motors used in conjunction with electronic circuitry that drive the source and keep track of its position. A battery is used for back-up power and an uninterruptible power supply is also integrated into the ACCU-SOURCE™ system to charge the battery backup subsystem.
### Intended Use:
The ACCU-SOURCE™ remote controlled high dose rate afterloading brachytherapy system is intended for the treatment of cancer and other lesions by intracavitary, interstitial, intralumenal, intraoperative, endobronchial and surface applicator irradiation treatments.
Technical Specifications:
Comparison Table Substantial Equivalent Device
The ACCU-SOURCE™ system is comparable to:
- Nucletron MicroSelectron HDR Version 2 .
| FEATURE<br>Device | Nucletron<br>Micro-Selectron | Oncology Systems, Inc.<br>ACCU-SOURCETM | |
|------------------------------------------------------------------------------|------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Intended Use | Remote<br>Afterloading<br>Brachytherapy unit<br>for interstitial,<br>intracavity,<br>intralumenal,<br>including bronchial,<br>intraoperative, and<br>surface applicator<br>treatments | The ACCU-SOURCE remote<br>control high dose rate<br>brachytherapy system is<br>intended for the treatment of<br>cancer and other lesions by<br>intracavity, interstitial,<br>intralumenal, intraoperative,<br>endobronchial and surface<br>applicator irradiation<br>treatments |
| | Base Area | | |
| | Height | Variable | 101cm |
| | Weight | 150 lbs | 180 lbs |
| | Transportable | V2 only | Yes |
| | Power Supply | 110 VAC | 110 VAC/220 VAC |
| | Mobile | Yes | Yes |
| | Nucletron<br>Micro-Selectron | Oncology Systems, Inc.<br>ACCU-SOURCE TM | |
| HDR | Yes | Yes | |
| Number of<br>Channels | 18 | 24 | |
| Shielding | Tungsten | Tungsten | |
| Maximum Activity of<br>Shielding | 10 Ci | 12 Ci | |
| Maximum<br>Treatment Activity | 10 Ci | 12 Ci | |
| Maximum exposure<br>rate a XX distance<br>containing the<br>maximum activity | At 10cm =<br>>2mR/h | Inner vault: > 100mR/h at<br>20cm<br>Outer vault: >2mR/h at 10cm | |
| Dwell positions per<br>channel | 48 | Up to 100 dwell points per<br>channel not to exceed 1200<br>dwell points (Ex. 24 channels<br>w/ 50 dwell points each or 12<br>channels w/100 dwell points<br>each) | |
| Total Channels | 18 | 24 | |
| Contains radiation<br>monitor | Optional | Yes, Included | |
| Device Control OS<br>Software | Windows | Windows XP | |
| Treatment Data<br>maintained during<br>power failure | Yes | Yes | |
| Simulator (Dummy)<br>Source | Yes | Yes | |
| Verification of<br>Channel Length by<br>direct measurement | By direct<br>measurements | Yes<br>By direct measurements | |
| Source Positioning | Distal to proximal | Distal to proximal | |
| Maximum Source<br>Position error over<br>treatment length (+/-<br>.5mm) | 1mm | .5 mm | |
| Emergency<br>container for the<br>source included | Optional | Yes, Included | |
| Response to<br>Emergency signal | Automatic<br>Retraction of<br>Source | Automatic Retraction of<br>Source | |
| Emergency manual | Yes | Yes | |
| | | Nucletron<br>Micro-Selectron | Oncology Systems, Inc.<br>ACCU-SOURCE TM |
| Source | Isotope | Ir192 | Ir192 |
| | | | |
| | Maximum Activity | 10 Ci | 12 Ci for Ir192 |
| | | | |
| | Maximum<br>Treatment Activity | 10 Ci | 12 Ci for Ir192 |
| | | | |
| | Capsule<br>Dimensions | 4.5mm length by<br>.9mm width | 4.5mm length by .9mm width<br>for Ir192 and 6.9 mm length by<br>1.17 mm width for Yb169 |
| | | | |
| | Active dimensions | 3.6mm length by<br>.65mm width | 3.6mm length by .65mm width<br>for Ir192 and 5.4mm length by<br>.8mm width for Yb169 |
| | Source extension<br>length | 150cm | 160cm |
| Operator Console | Operating console<br>with personal<br>computer | Yes | Yes |
| | Keyswitch Control | Yes | Yes |
| | Operating Voltage | 110/220 VAC | 110/220 VAC |
{2}------------------------------------------------
# 【 【 】 【 】 【 】 【 】 【 】 【 】 【 】 【 】 【 】 【 】 【 】 【 】 【 】 【 】 【 】 【 】 【 】 【 】 【 】 【 】 【 】 【 】 【 】 【 】 【 】 【 】 【 】 【 】 【 】 【 】 【 】 【 】 【 】 【 】 【 】 【 】 【 】 【 】 【 】 【 】 【 】 【 】 【
.
{3}------------------------------------------------
## KOSI753
the state of the state of the states and
.
:
{4}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an emblem featuring a stylized depiction of three wavy lines, resembling a bird or a symbol representing health and human services.
#### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Ms. Jennifer Williams Official Correspondent Oncology Systems, Inc. 206 N. Randolph Street, Suite 300 CHAMPAIN IL 61820
Re: K051753
MAR 2 9 2006
Trade/Device Name: ACCU-SOURCE Regulation Number: 21 CFR 892.5700 Regulation Name: Remote controlled radionuclide applicator system Regulatory Class: II Product Code: JAQ
Dated: February 16, 2006 Received: February 21, 2006
Dear Ms. Williams:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements. including, but not limited to registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); 200d manufacturing practice requirements as set forth in the quality systems (05) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
{5}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|-----------------|----------------------------------|--------------|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology) | 240-276-0120 |
| Other | 240-276-0100 | |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industrv/support/index.html
Sincerely yours.
T. Nancy C. Brigdon
Naney C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices ()fice of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
Device Name: Accu-Source™
Indications for Use:
The Accu-Source™ remote control high dose rate afterloading brachytherapy system is indicated for the treatment of cancer and other lesions by intracavitary, interstitial, intralumenal, intraoperative, endobronchial and surface applicator irradiation treatments.
| (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) |
|--------------------------------------------------------------------------|
| Concurrence of CDRH, Office of Device Evaluation (ODE) |
| Prescription Use<br>(Per 21 CFR 801.109) | OR | Over-the-Counter Use |
|------------------------------------------|----|----------------------|
|------------------------------------------|----|----------------------|
| | <img alt="Signature" src="signature.png"/> |
|------------------------------------------------------------------|--------------------------------------------|
| (Division Sign-Off) | |
| Division of Reproductive, Abdominal,<br>and Radiological Devices | |
| 510(k) Number | K051753 |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.