DENTRONIX DDS 7000 RAPID DRY HEAT STERILIZER STERILIZATION SYSTEM
Device Facts
| Record ID | K051660 |
|---|---|
| Device Name | DENTRONIX DDS 7000 RAPID DRY HEAT STERILIZER STERILIZATION SYSTEM |
| Applicant | Coltene/Whaledent, Inc. |
| Product Code | KMH · General Hospital |
| Decision Date | Aug 4, 2005 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 880.6870 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Indications for Use
The DDS 7000 is a countertop convective dry heat sterilizer designed for use in Healthcare facilities for the sterilization of un-bagged dental instruments that can withstand exposure temperatures up to 216°C (420° F). There is a single software-controlled cycle that includes a warm-up and cool down phase in addition to the 3-minute sterilization (exposure) phase. Sterilize (exposure) starts when the RTD/control temperature reaches the operating temperature set point of 190°C (374° F). Complete cycle times average 34-44 minutes and are load dependent. The DDS 7000 achieves optimum results by uniform load distribution utilizing a dedicated rack and tray system as detailed below. 1) Maximum recommended loads for the DDS 7000-115 sterilizer: a) Maximum load for a four (4)-rack configuration is four (4) plier racks each holding nine (9) orthodontic pliers each for a total of thirty-six (36) pliers. Other maximum load configuration options are: 32 pliers on eight mini vertical racks; or 20 pliers and hand instruments on four Combo Racks -(OR)- b) Maximum load for a full tray configuration is one (1) Horizontal Tray each holding up to six (6) orthodontic pliers or up to (12) single-handle hand instruments. Other maximum load combinations include: (5) orthodontic pliers with up to (2) single-handle hand instruments; or (4) orthodontic pliers with up to (4) single-handle hand instruments; or (3) orthodontic pliers with up to (6) single-handle hand instruments; or (2) Orthodontic pliers with up to (8) single-handle hand instruments; or (1) Orthodontic pliers with up to (10) single-handle hand instruments. -(OR)- c) Maximum load for the one (1) half tray and two (2) racks is one (1) Half Tray holding three (3) Orthodontic pliers or six (6) single-handle hand instruments in combination with two (2) plier racks capable of holding nine (9) pliers each for a total of eighteen (18) pliers on plier racks. Other maximum load combinations include: (2) orthodontic pliers with up to (2) single-handle hand instruments in combination with two (2) plier racks capable of holding nine (9) pliers each for a total of eighteen (18) pliers on plier racks; or (1) orthodontic pliers with up to (4) single-handle hand instruments in combination with two (2) plier racks capable of holding nine (9) pliers each for a total of eighteen (18) pliers on plier racks.
Device Story
Countertop convective dry heat batch sterilizer; processes un-bagged dental instruments. Input: manual loading of instruments into dedicated rack/tray systems. Operation: electric heating element with mechanical convection; software-controlled cycle (warm-up, 3-minute sterilization at 190°C, cool down). Output: sterilized instruments; diagnostic error codes; printable cycle logs via PC COM port. Used in dental clinics/healthcare facilities. Features HEPA filter (99.7% efficiency for 0.3-micron particles) for cooling air. Software monitors cycle parameters, provides audible alarms, and manages safety interlocks. Benefits: rapid sterilization and immediate instrument availability; uniform load distribution ensures efficacy.
Clinical Evidence
Bench testing only. Performance validated using Bacillus atrophaeus spore strips and inoculated tools. Testing across full, partial, and empty chamber loads demonstrated a 12-log reduction of spores, achieving a 10^-6 sterility assurance level (SAL).
Technological Characteristics
Convective dry heat sterilizer; electric heating element; mechanical convection. Includes HEPA filter (99.7% efficiency at 0.3 microns) per ANSI/AAMI ST50:2004. Features RTD temperature monitoring, LED display, and software-controlled cycle management. Connectivity: PC COM port for log printing. Power: 115V/230V. Safety: door interlock.
Indications for Use
Indicated for sterilization of un-bagged dental instruments in healthcare facilities; instruments must withstand exposure temperatures up to 216°C (420°F).
Regulatory Classification
Identification
A dry-heat sterilizer is a device that is intended for use by a health care provider to sterilize medical products by means of dry heat.
Predicate Devices
- Dentronix DDS 5000 Dry Heat Sterilization System (K880322)