SANARUS CASSI II ROTATIONAL CORE BIOPSY SYSTEM, MODEL CS2000
Applicant
Sanarus Medical, Inc.
Product Code
KNW · Gastroenterology, Urology
Decision Date
Jul 27, 2005
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1075
Device Class
Class 2
Indications for Use
The device is indicated for use in obtaining biopsies from soft tissues such as liver, kidney, prostate, spleen, lymph nodes and various soft tissue tumors. It is not intended for use in bone. The device is also indicated to provide breast tissue samples for diagnostic sampling of breast abnormalities. It is designed to provide breast tissue for histologic examination with partial or complete removal of the imaged abnormality. The extent of histologic abnormality cannot be reliably determined from its mammographic appearance. Therefore, the extent of removal of the imaged evidence of an abnormality does not predict the extent of removal of a histologic abnormality (e.g., malignancy). When the sampled abnormality is not histologically benign, it is essential that the tissue margins be examined for completeness of removal using standard surgical procedures.
Device Story
System comprises control unit (handle), biopsy needle unit (sticking needle, cutting cannula), and sample collection tray. Sticking needle uses cold temperatures at tip to engage tissue; coaxially mounted cutting cannula cores tissue specimen. Available in multiple gauge sizes/lengths. Used for soft tissue and breast abnormality sampling. Provides tissue for histologic examination; allows partial or complete removal of imaged abnormalities. Physician-operated. Output is physical tissue sample for pathology. Benefits include diagnostic tissue acquisition for malignancy assessment.
Clinical Evidence
Bench testing only. Performance testing confirmed sample quality is equivalent to the predicate device.
Technological Characteristics
System includes control unit, biopsy needle unit, and sample collection tray. Features cryo-assisted tissue engagement (cold tip) and coaxial cutting cannula. Available in various gauge sizes and lengths. Sterilization methods equivalent to predicate.
Indications for Use
Indicated for soft tissue biopsy (liver, kidney, prostate, spleen, lymph nodes, tumors) and diagnostic breast tissue sampling. Contraindicated for use in bone.
Regulatory Classification
Identification
A gastroenterology-urology biopsy instrument is a device used to remove, by cutting or aspiration, a specimen of tissue for microscopic examination. This generic type of device includes the biopsy punch, gastrointestinal mechanical biopsy instrument, suction biopsy instrument, gastro-urology biopsy needle and needle set, and nonelectric biopsy forceps. This section does not apply to biopsy instruments that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Predicate Devices
Sanarus Cassi™ Rotational Core Biopsy System (K042136)
Submission Summary (Full Text)
{0}------------------------------------------------
K051581 pg, 1 of 2
Sanarus Cassi II Rotational Core Biopsy System
JUL 27 2005
Caye 102
# Section 2: Summary of Safety and Effectiveness
### Contact Information (1)
Trena Depel Director, Requlatory & Clinical Affairs Telephone: (925) 460-5731 FAX: (925) 460-0688 E-mail: tdepel@sanarus.com
# Company Information (2)
Sanarus Medical, Inc. 4696 Willow Road Pleasanton, CA 94588 Telephone: (925) 460-6080 FAX: (925) 460-6084
#### (3) Device Name
Sanarus Cassi™ II Rotational Core Biopsy System
#### (4) Device Description
The Sanarus Cassi II Rotational Core Biopsy System consists of a fully integrated control unit (Handle), Biopsy Needle Unit (comprised of a sticking needle and cutting cannula) and sample collection tray. The sticking needle is operated by the control unit and uses cold temperatures at its tip to engage the tissue to be sampled. The cutting cannula is coaxially mounted around the sticking needle and is used to core the tissue specimen. The cutting cannula will be available in several gauge sizes and lengths.
## Indications for Use (5)
The device is indicated for use in obtaining biopsies from soft tissues such as liver, kidney, prostate, spleen, lymph nodes and various soft tissue tumors. It is not intended for use in bone.
The device is also indicated to provide breast tissue samples for diagnostic sampling of breast abnormalities. It is designed to provide breast tissue for histologic examination with partial or complete removal of the imaged abnormality. The extent of histologic abnormality cannot be reliably determined from its mammographic appearance. Therefore, the extent of removal of the imaged evidence of an abnormality does not predict the extent
{1}------------------------------------------------
of removal of a histologic abnormality (e.g., malignancy). When the sampled abnormality is not histologically benign, it is essential that the tissue margins be examined for completeness of removal using standard surgical procedures.
#### (6) Name of Predicate or Legally Marketed Device
Sanarus Cassi™ Rotational Core Biopsy System
#### (7) Substantial Equivalence
The Sanarus Cassi II Rotational Core Biopsy System is substantially equivalent to the Cassi Rotational Core Biopsy System that was determined to be substantially equivalent on Sep 8 2004 (reference K042136).
The Sanarus Cassi II Rotational Core Biopsy System has the same indications for use and technological characteristics as the predicate device. The patient contact components and component materials for obtaining core biopsy samples in both the new and predicate device are the same. The packaging materials, packaging configurations, sterilization methods and sterility assurance level are also equivalent.
Based on the indications for use, technological characteristics and performance testing results, the Sanarus Cassi II Rotational Core Biopsy System does not raise significant new questions of safety and effectiveness.
#### (8) Performance Testing Summary
Performance testing confirms that the quality of samples obtained with the Sanarus Cassi II Rotational Core Biopsy System is equivalent to the predicate device.
{2}------------------------------------------------
Image /page/2/Picture/2 description: The image shows a black and white logo. The logo features a stylized bird in flight, with three lines representing its wings. The bird is positioned within a circle, and there is some text visible around the circumference of the circle, although the text is not clear enough to read.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20050
JUL 2 7 2005
Ms. Trena Depel Director, Regulatory and Clinical Affairs Sanarus Medical, Inc. 4696 Willow Road PLEASANTON CA 94588
Re: K051581 Trade/Device Name: Sanarus Cassi™ II Rotational Core Biopsy System Regulation Number: 21 CFR §876.1075 Regulation Name: Gastroenterology-urology biopsy instrument Regulatory Class: II
Product Code: KNW Dated: June 14, 2005 Received: June 15, 2005
Dear Ms. Depel:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced we ha re reviewed the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to the cholors of to regary ment date of the Medical Device Amendments, or to devices that have been May 20, 1710, and classince with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the do not require approvisions of the Act. The general controls provisions of the Act do froo, subjoct to at to for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket It your all to is vascent in to such additional controls. Existing major regulations affecting your Apployali, It the you be begiller Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that I reast of advisou that i Dr. in that your device complies with other requirements of the Act or any I Drives and regulations administered by other Federal agencies. You must comply with all the I cacrates and reguirements, including, but not limited to registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) (21 CFR Part 801); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
This letter will allow you to begin marketing your device as described in your section 5 [0(k) I his letter will anow you to begin marketing your and equivalence of your device to a legally premarket notification. THE PDA multing of bassantial sept revice and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Patt 801), please If you desire specific advice for your correst on the following numbers, based on the regulation number at the top of this letter:
| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|-----------------|----------------------------------|--------------|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification the Antification the Also, please note the regulation entitled, "Misetanaling on your responsibilities under the Act from the 190 807.97). You may obtain outci general miornation on your Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address
http://www.fda.gov/cdrlv/industry/support/index.html.
Sincerely yours,
Nancy C. Hogdon
Nancv C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications For Use
510(k) Number: K051581
Device Name: Sanarus Cassi™ II Rotational Core Biopsy System
Indications for Use: The device is indicated for use in obtaining biopsies from soft tissues such as liver, kidney, prostate, spleen, lymph nodes and various soft tissue tumors. It is not intended for use in bone.
The device is also indicated to provide breast tissue samples for diagnostic sampling of breast abnormalities. It is designed to provide breast tissue for histologic examination with partial or complete removal of the imaged abnormality. The extent of histologic abnormality cannot be reliably determined from its mammographic appearance. Therefore, the extent of removal of the imaged evidence of an abnormality does not predict the extent of removal of a histologic abnormality (e.g., malignancy). When the sampled abnormality is not histologically benign, it is essential that the tissue margins be examined for completeness of removal using standard surgical procedures.
Concurrence of CDRH, Office of Device Evaluation (ODE):
David br. Leggrm
(Division Sign-Off)
Division of Reproductive, Abdominal and Radiological De 510(k) Number
Prescription Use: × (Per 21 CFR 801.109)
Sanarus Medical CONFIDENTIAL
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.