PROTON PHC 8731 10 SECOND DIGITAL THERMOMETER; PROTON PHC 109 EAR THERMOMETER; PROTON 889 EAR THERMOMETER

K051566 · Proton Healthcare , Ltd. · FLL · Jan 17, 2006 · General Hospital

Device Facts

Record IDK051566
Device NamePROTON PHC 8731 10 SECOND DIGITAL THERMOMETER; PROTON PHC 109 EAR THERMOMETER; PROTON 889 EAR THERMOMETER
ApplicantProton Healthcare , Ltd.
Product CodeFLL · General Hospital
Decision DateJan 17, 2006
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.2910
Device ClassClass 2
AttributesPediatric

Indications for Use

The Proton Thermometer models PHC 8731, PHC 109 and PHC 889 are intended for measuring body temperature of all ages. The Proton Thermometers are reusable and intended for use in the home setting.

Device Story

Proton PHC 8731 (digital), PHC 109 (ear), and PHC 889 (ear) thermometers measure human body temperature. Devices are reusable and intended for home use by patients or caregivers. Thermometers utilize internal sensors to detect thermal energy; convert signals into digital temperature readings; display results on integrated screen. Output provides immediate temperature data for health monitoring. No complex clinical workflow or remote processing required.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

Clinical electronic thermometers; digital sensing technology; reusable form factor; intended for home use. No specific materials, energy sources, or software algorithm classes detailed in the provided documentation.

Indications for Use

Indicated for measuring body temperature in patients of all ages in a home setting.

Regulatory Classification

Identification

A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.

Special Controls

(1) Device is not a clinical thermometer with telethermographic functions; (2) Device is not a clinical thermometer with continuous temperature measurement functions; and (3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the seal of the Department of Health and Human Services (HHS). The seal features an abstract image of an eagle with three lines forming its wings and head. The words "DEPARTMENT OF HEALTH AND HUMAN SERVICES . USA" are arranged in a circular pattern around the eagle. #### Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 ### JAN 1 7 2006 Proton HealthCare Limited C/O Mr. Kevin Walls Principal Consultant Regulatory Insight, Incorporated 13 Red Fax Lane Littleton, Colorado 80127 Re: K051566 Trade/Device Name: Proton PHC 8731 10 Second Digital Thermometer Proton PHC 109 Ear Thermometer Proton PHC 889 Ear Thermometer Regulation Number: 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: II Product Code: FLL Dated: January 2, 2005 Received: January 4, 2005 Dear Mr. Walls: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into cither class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ #### Page 2 - Mr. Walls Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/indcx.html. Sincerely yours, Suette y. Michien Omd Chih-Lin, Ph.D. Chiú Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known): K051566 Device Name: - 1. Proton PHC 8731 10 Second Digital Thermometer - 2. Proton PHC 109 Ear Thermometer - 3. Proton PHC 889 Ear Thermometer Indications for Use: The Proton Thermometer models PHC 8731, PHC 109 and PHC 889 are intended for measuring body temperature of all ages. The Proton Thermometers are reusable and intended for use in the home setting. Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use X (21 CFR 801 Subpart C) ## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Arthur D. Little , ජූලි විද්‍යා පිළිබඳව, Ga .. Con Control, Davital Devic Page 1 of 1 1 - 1 -
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...