GF-UE160-AL5 ULTRASOUND ENDOSCOPE

K051541 · Olympus America, Inc. · FDF · Jul 1, 2005 · Gastroenterology, Urology

Device Facts

Record IDK051541
Device NameGF-UE160-AL5 ULTRASOUND ENDOSCOPE
ApplicantOlympus America, Inc.
Product CodeFDF · Gastroenterology, Urology
Decision DateJul 1, 2005
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2
Attributes3rd-Party Reviewed

Indications for Use

The Olympus GF-UE160-AL5 Ultrasonic Endoscope is intended to be used for endoscopic ultrasonic inaging of the gastrointestinal wall, bile and pancreatic ducts and surrounding organs. It is to be used with the Aloka SSD-5000 (K033311) Diagnostic Ultrasound system and various other video and light source accessories.

Device Story

The Olympus GF-UE160-AL5 is an ultrasonic gastro-video endoscope designed for use with the Aloka SSD-5000 diagnostic ultrasound system. It captures ultrasound signals from the gastrointestinal tract, bile ducts, and pancreatic ducts; these signals are processed by the Aloka system to produce real-time diagnostic images. The device features a 100° field of view, 55° forward-oblique direction of view, and 3-100mm depth of field. It is operated by physicians in clinical settings (e.g., endoscopy suites, OR) to visualize internal structures and surrounding organs. The output is displayed on a video monitor for the clinician to perform diagnostic imaging or fluid flow analysis, aiding in clinical decision-making regarding gastrointestinal and abdominal pathology. The device utilizes balloon or de-aerated water immersion methods for acoustic coupling.

Clinical Evidence

Bench testing only. No clinical data provided. Substantial equivalence is based on design comparison and performance specifications consistent with predicate devices.

Technological Characteristics

Ultrasonic gastro-video endoscope; 13.8mm distal end diameter; 11.8mm insertion tube diameter; 1250mm working length; 2.2mm instrument channel. Imaging modes: B-mode, M-mode, PWD, Color Doppler, Amplitude Doppler. Connectivity: interfaces with Aloka SSD-5000 ultrasound system, video center, and light source. Coupling: balloon or de-aerated water immersion.

Indications for Use

Indicated for endoscopic ultrasound imaging of the gastrointestinal wall, bile and pancreatic ducts, and surrounding abdominal organs in patients requiring diagnostic ultrasound imaging or fluid flow analysis. Used in abdominal, intraoperative, and transesophageal applications. Not indicated for ophthalmic applications.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls: 1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following: 1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures. 2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use. 3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use. 2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes. 3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches 4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error. 5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program. (6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms. (7) Labeling must include: (i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified; (ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA; (iii) Identification of compatible devices; (iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification; (v) Reprocessing instructions for reusable components; (vi) A shelf life for any sterile components; (vii) A description of the device-specific use training program; (viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and (ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance. (8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include: (i) Device motion accuracy and repeatability; (ii) System testing; (iii) Instrument reliability; (iv) Crosstalk; (v) Table motion control; (vi) Thermal effects on tissue; (vii) User-device interface performance; (viii) Workspace access testing; and (ix) Performance testing with compatible devices. (9) Software verification, validation, and hazard analysis must be performed. (10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed. (11) Performance data must demonstrate the sterility of all patient-contacting device components. (12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life. (13) Performance data must validate the reprocessing instructions for the reusable components of the device. (14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible. (15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic. (16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information: (i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and (ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Cos154// Date: Date: __________________________________________________________________________________________________________________________________________________________________ .............................................................................................................................................................................. 510(K) Summary Olympus GF-UE160-AL5 Endoscope used with Aloka Model SSD- 5000 Diagnostic Ultrasound System This summary statement complies with 21 CFR, section 807.92 as amended March 14, 1995. This premarket notification has been submitted by Olympus American, Inc. and covers the Olympus GF-UE160-AL5 endoscope used with the Aloka SSD-5000 diagnostic ultrasound system. ## 1. GENERAL INFORMATION Applicant: | Submission Correspondent: | Laura Storms-Tyler<br>Director, Regulatory Affairs and Quality Assurance<br>Olympus America Inc.<br>Two Corporate Center Drive, Melville, NY 11747-9058<br>Phone: 631-844-5688<br>Fax: 631-844-5554 | |---------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Official Correspondent: | Laura Storms-Tyler<br>Director, Regulatory Affairs and Quality Assurance<br>Olympus America Inc.<br>Two Corporate Center Drive, Melville, NY 11747-9058<br>Phone: 631-844-5688<br>Fax: 631-844-5554<br>Establishment Registration No.: 2429304 | #### 2. DEVICE IDENTIFICATION Trade/Common Name: The proprietary name is the Olympus GF-UE160-AL5 Gastro-video Endoscope used with the Aloka SSD-5000 Diagnostic Ultrasound System. The common name for this type of device is an Ultrasound Gastro-video Scope. Regulation Number/Name: The items in this submission are covered under the following classification: | 90 IYN | Ultrasonic Pulsed Doppler Imaging System | 21 CFR 892.1550 | |--------|------------------------------------------|-----------------| | 90 ITX | Diagnostic Ultrasound Transducer | 21 CFR 892.1570 | | 90 IYO | Ultrasonic Pulsed Echo Imaging System. | 21 CFR 892.1560 | | 70 FDF | Endoscope and accessories | 21 CFR 876.1500 | The above have been classified as regulatory Class II. {1}------------------------------------------------ ## 510(K) Summary #### 3. Predicate Devices: The Olympus GF-UE160-AL5 Endoscope is substantially equivalent to the Olympus GF-UC140P-AL5 Endoscope (K011314) and the GF-UM130 Endoscope (K971660). Both predicates share common indications and some features. ## 4. Device Description: The Olympus GF-UC140P-A15 is an electronic radial scan ultrasound endoscope providing a 360° view angle. #### 5. Intended Use: The Olympus GF-UE160-AL5 Ultrasonic Endoscope is intended to be used for endoscopic ultrasonic inaging of the gastrointestinal wall, bile and pancreatic ducts and surrounding organs. It is to be used with the Aloka SSD-5000 (K033311) Diagnostic Ultrasound system and various other video and light source accessories. #### 6. Comparison of Technological Characteristics: | Below is the comparison table between the subject device and its predicate devices | | | | |------------------------------------------------------------------------------------|--------------------------------------------------------|-----------------------------|-----------------------| | Specifications | Subject Device<br>GF-UE160-A1.5 | Predicate Devices | | | | | GF-UC140P-A1.5<br>(K011314) | GF-UM130<br>(K971660) | | Optical Characteristics: | | | | | Field of View: | 100° (video) | Same | Same | | Direction of View: | 55° forward-oblique | Same | Same | | Depth of Field: | 3~100mm | Same | Same | | Outer Diameter of Distal End | ⌀ 13.8mm | ⌀ 14.2mm | ⌀12.7mm | | Outer Diameter of Insertion Tube | ⌀ 11.8mm | Same | ⌀ 10.5mm | | Angulations: | | | | | Up: | 130° | Same | Same | | Down: | 90° | Same | Same | | Left: | 90° | Same | Same | | Right: | 90° | Same | Same | | Working Length | 1250mm | Same | Same | | Instrument Channel | ⌀ 2.2mm | ⌀ 2.8mm | ⌀ 2.2mm | | Contact Method | Ballon Method/De-<br>Aerated Water Immersion<br>Method | Same | Same | n the subject, douing and its prodicate #### 8. Conclusion: When the GF-UE 160-AL5 is compared to its predicates, the device does not incorporate any significant changes in intended use, method of operation, material or design that could affect the safety and effectiveness. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a stylized eagle with three stripes forming its body, symbolizing health, human services, and well-being. Encircling the eagle is the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 JUL 1 - 2005 Olympus America, Inc. % Mr. Neil E. Devine, Jr. Responsible Third Party Official Intertek Testing Services NA, Inc. 70 Codman Hill Road BOXBOROUGH MA 01719 Re: K051541 Trade Name: Olympus GF-UE160-AL5 Endoscope used with the Aloka SSD-5000 Diagnostic Ultrasound System Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulation Number: 21 CFR 892.1570 Regulation Name: Diagnostic ultrasonic transducer Regulatory Class: II Product Code: FDF and ITX Dated: June 9, 2005 Received: June 10, 2005 Dear Mr. Devine: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. This determination of substantial equivalence applies to the following transducers intended for use with the Aloka SSD-5000 Diagnostic Ultrasound System, as described in your premarket notification: Transducer Model Number #### GF-UE160-AL5 {3}------------------------------------------------ Page 2 - Mr. Devine If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This determination of substantial equivalence is granted on the condition that prior to shipping the first device, you submit a postclearance special report. This report should contain complete information, including acoustic output measurements based on production line devices, requested in Appendix G, (enclosed) of the Center's September 30, 1997 "Information for Manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound Systems and Transducers." If the special report is incomplete or contains unacceptable values (e.g., acoustic output greater than approved levels), then the 510(k) clearance may not apply to the production units which as a result may be considered adulterated or misbranded. The special report should reference the manufacturer's 510(k) number. It should be clearly and prominently marked "ADD-TO-FILE" and should be submitted in duplicate to: > Food and Drug Administration Center for Devices and Radiological Health Document Mail Center (HFZ-401) 9200 Corporate Boulevard Rockville, Maryland 20850 This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html {4}------------------------------------------------ Page 3 – Mr. Devine If you have any questions regarding the content of this letter, please contact Rodrigo C. Perez at (301) 594-1212. Sincerely yours, Nancy C Brogdon Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure(s) {5}------------------------------------------------ # Indications for use statement | 510(K) Number (if known): | | |---------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Device Name: | Olympus GF-UE160-AL5 Endoscope used with the Aloka SSD-5000 Diagnostic Ultrasound System | | Indications For Use: | The Olympus GF-UE160-AL5 is an ultrasonic gastro video endoscope to be used with an Aloka diagnostic ultrasound system, video system center, light source, video monitor, endo-therapy accessories for endoscopic ultrasound imaging of the gastrointestinal wall, bile and pancreatic ducts and surrounding organs.<br>The Aloka SSD-5000 (K033311) is an all-digital diagnostic ultrasonic scanner with a digital beam former supporting gray scale, spectral Doppler and Color Flow imaging. Depending on the probe, the Aloka SSD-5000 may be used for diagnostic ultrasound imaging in Cardiac, Gynecological, Neurological, Obstetrical, Neonatal, Pediatric, Perinatal, Radiological, Vascular, Urological, Trauma and Surgical applications. The Aloka SSD-5000 is a Track 3 system and is not indicated for ophthalmic applications. | く Prescription use (Pet 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR *)! Subpart C) # (PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED) .............................................................................................................................................................................. Concurrence of CDRH, Office of Device Evaluation (ODE) Nancy Brogdon (Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices and Radiological Devices 510(k) Number _ Page 1 of 1 {6}------------------------------------------------ #### 4.3.1 # Diagnostic Ultrasound Indications for Use Form SSD-5000 Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | Clinical Application | Modes of operation | | | | | | | | | | |----------------------------------|--------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------| | | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify) | | Ophthalmic | | | | | | | | | | | | Fetal | | P | P | P | | P | P | | See Below | | | Abdominal | | P | P | P | | P | P | | See Below | | | Intraoperative (specify) | | P | P | P | | P | P | | See Below | | | Intraoperative Neurological | | P | P | P | | P | P | | See Below | | | Pediatric | | P | P | P | | P | P | | See Below | | | Small Organ (specify) | | P | P | P | | P | P | | See Below | | | Neonatal Cephalic | | P | P | P | | P | P | | See Below | | | Adult Cephalic | | P | P | P | | P | P | | See Below | | | Cardiac | | P | P | P | | P | P | | See Below | | | Transesophageal | | P | P | P | | P | P | | See Below | | | Transrectal | | P | P | P | | P | P | | See Below | | | Transvaginal | | P | P | P | | P | P | | See Below | | | Transurethral | | | | | | | | | | | | Intravascular | | | | | | | | | | | | Peripheral Vascular | | P | P | P | | P | P | | See Below | | | Laparoscopic | | P | P | P | | P | P | | See Below | | | Musculo-skeletal<br>Conventional | | P | P | P | | P | P | | See Below | | | Musculo-skeletal Superficial | | P | P | P | | P | P | | See Below | | | Other | | | | | | | | | | | N= new indication; P= previously cleared by FDA; E= added under Appendix E Additional Comments: Mixed mode operation includes B/M, B/Bflow/PWD. Intra-operative applications: include liver, pancreas and gall bladder. Small parts applications include breast, testes and thyroid. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Preseription Use (Per 21 CFR 801.109) Prescription Use (Division Sign/Off) Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number K051541 {7}------------------------------------------------ ### Diagnostic Ultrasound Indications for Use Form GF-UE160-AL5 Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | Modes of operation | | | | | | | | | | |----------------------------------|--------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------| | Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify) | | Ophthalmic | | | | | | | | | | | | Fetal | | | | | | | | | | | | Abdominal | | N | N | N | | N | N | | See Below | | | Intraoperative (specify) | | N | N | N | | N | N | | See Below | | | Intraoperative Neurological | | | | | | | | | | | | Pediatric | | | | | | | | | | | | Small Organ (specify) | | | | | | | | | | | | Neonatal Cephalic | | | | | | | | | | | | Adult Cephalic | | | | | | | | | | | | Cardiac | | | | | | | | | | | | Transesophageal | | N | N | N | | N | N | | See Below | Non-<br>Cardiac | | Transrectal | | | | | | | | | | | | Transvaginal | | | | | | | | | | | | Transurethral | | | | | | | | | | | | Intravascular | | | | | | | | | | | | Peripheral Vascular | | | | | | | | | | | | Laparoscopic | | | | | | | | | | | | Musculo-skeletal<br>Conventional | | | | | | | | | | | | Musculo-skeletal Superficial | | | | | | | | | | | | Other | | | | | | | | | | | N= new indication; P= previously cleared by FDA; E= added under Appendix E Additional Comments: Mixed mode operation includes B/M, B/PWD, B/Bflow/PWD., Intraoperative : Gastro-intestinal tract & surrounding abdominal organs (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109) Nancy C. Hodgon (Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number K051541 21
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...