The Uritek-720+ Urine Analyzer is intended for use with Teco Urine Reagent Strips for Urinalysis such as URS-10, used in the determination of glucose, protein, pH, bilirubin, blood, ketone, urobilinogen, nitrite, specific gravity, and leukocytes in urine.
Device Story
Uritek-720+ Urine Analyzer is an automated urinalysis system; processes urine reagent strips (e.g., URS-10). Input: colorimetric changes on reagent strips; captured via Contact Image Sensor (CIS). Operation: user dips strip, places on test table; device scans reflectance of red, green, blue light; computes reflectance ratio using internal algorithm. Output: semi-quantitative/qualitative results for 10 analytes. Used in clinical settings; operated by healthcare personnel. Benefits: standardized, automated interpretation of urine chemistry compared to visual reading; reduces subjectivity; improves consistency in clinical decision-making regarding patient metabolic or renal status.
Clinical Evidence
No clinical data. Bench testing only. Precision/reproducibility evaluated using 20 replicates of three control levels across two instruments and two lots, showing high agreement (95-100%). Method comparison performed on 100 specimens using Teco strips on Uritek-720+ versus Bayer Multistix 10SC on Clinitek-200+, confirming substantial equivalence.
Technological Characteristics
Automated urinalysis system using reflectance photometry. Sensing: Contact Image Sensor (CIS). Energy: Electrical. Form factor: Benchtop (260 x 320 x 178 mm; 2.4 kg). Connectivity: Manual or barcode patient ID entry. Calibration: Self-calibrating on power-up using internal white reference. Software: Embedded system for signal processing and reflectance ratio computation.
Indications for Use
Indicated for use with Teco Urine Reagent Strips (e.g., URS-10) for the semi-quantitative or qualitative determination of glucose, protein, pH, bilirubin, blood, ketone, urobilinogen, nitrite, specific gravity, and leukocytes in urine samples.
Regulatory Classification
Identification
An automated urinalysis system is a device intended to measure certain of the physical properties and chemical constituents of urine by procedures that duplicate manual urinalysis systems. This device is used in conjunction with certain materials to measure a variety of urinary analytes.
Submission Summary (Full Text)
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE
A. 510(k) Number:
k051526
B. Purpose for Submission:
New submission for the Uritek-720+ Urine Analyzer for the use with Teco Diagnostic’s previously cleared urine strips URS-10 k970250.
C. Measurand:
Glucose, Blood, Leukocytes, Specific Gravity, pH, Nitrite, Protein, Ketone, Urobilinogen and Bilirubin in urine
D. Type of Test:
Qualitative /Semi-Quantitative
E. Applicant:
TECO Diagnostics
F. Proprietary and Established Names:
Uritek-720+ Urine Analyzer
G. Regulatory Information:
1. Regulation section:
21 CFR § 862.1340 Urinary glucose (nonquantitative) test system
21 CFR § 864.6550 Occult blood test
21 CFR § 864.7675 Leukocyte peroxidase test
21 CFR § 862.2800 Refractometer for clinical use
21 CFR § 862.1550 Urinary pH (nonquantitative) test system
21 CFR § 862.1510 Nitrite (nonquantitative) test system
21 CFR § 862.1645 Urinary protein or albumin (nonquantitative) test system
21 CFR § 862.1435 Ketones (nonquantitative) test system
21 CFR § 862.1785 Urinary urobilinogen (nonquantitative) test system
21 CFR § 862.1115 Urinary bilirubin and its conjugates (nonquantitative) test system
{1}
21 CFR § 862.2900 Automated urinalysis system
2. Classification:
Class II; Urinary glucose (nonquantitative) test system and Occult blood test
Class I; Leukocyte peroxidase test, Refractometer for clinical use, Urinary pH (nonquantitative) test system, Nitrite (nonquantitative) test system, Urinary protein or albumin (nonquantitative) test system, Ketones (nonquantitative) test system, Urinary urobilinogen (nonquantitative) test system and Urinary bilirubin and its conjugates (nonquantitative) test system and Automated urinalysis system
3. Product code:
JIL Enzymatic Method, Glucose (urinary, non-quant.)
JIO Blood, Occult, Colorimetric, in urine
LJX Test, Urine Leukocyte
JRE Refractometer for clinical use
CEN Dye-Indicator, pH (urinary, non-Quant.)
JMT Diazo (colorimetric), Nitrite (urinary, non-quant.)
JIR Indicator method, Protein or Albumin (urinary, non-quant.)
JIN Nitroprusside, Ketones (urinary, non-quant.)
CDM Diazonium Colorimetry, Urobilinogen (urinary, non-quant.)
JJB Azo-dye, colorimetric, Bilirubin & its conjugates (urinary, non-quant.)
KQO Automated urinalysis system
4. Panel:
75 Chemistry
H. Intended Use:
1. Intended use(s):
See indications for use.
2. Indication(s) for use:
The Uritek-720+ Urine Analyzer is intended for use with Teco Urine Reagent Strips for Urinalysis such as URS-10, used in the determination of glucose, protein, pH, bilirubin, blood, ketone, urobilinogen, nitrite, specific gravity and leukocytes in urine.
3. Special conditions for use statement(s):
For prescription use
{2}
4. Special instrument requirements:
Uritek-720+ Urine Analyzer
I. Device Description:
The Uritek-720+ Urine Analyzer combines a contact Image Sensor (CIS), electronics and computer technology with other technologies. The controlling system processes the reflectance ratios of the lights and converts them into electrical signals for result interpretation.
J. Substantial Equivalence Information:
1. Predicate device name(s):
Bayer, Clinitek 200+ Urine Chemistry Analyzer
2. Predicate 510(k) number(s):
k0926359
3. Comparison with predicate:
| Comparison | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Intended use | Used in determination of glucose, protein, pH, bilirubin, blood, ketone, urobilinogen, nitrite, specific gravity and leukocytes in urine. | Same |
| Basic Operating Principle | Reflectance | Same |
| Testing options | Single-step or continuous testing | Same |
| Test steps | Dip and place urine strip onto test table | Same |
| Environment Requirement | 0-40 °C Relative humidity ≤ 85% | 18-30 °C Relative humidity 20% - 85% |
| Dimension | 260 x 320 x 178 mm | 49.3 x 35.4 x 19.7 cm |
{3}
4
| Comparison | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Weight | 2.4 kg | 13.6 kg |
K. Standard/Guidance Document Referenced (if applicable):
None Referenced
L. Test Principle:
The instrument uses three electric signals of red, green and blue (conversion of reflected light) to scan the difference and depth of color change. The controlling system processes the electrical signal and computes the reflectance ratio of testing color according to the following equation:
$$
\mathrm{R}\% = \frac{\mathrm{Tm} \times \mathrm{Cr}}{\mathrm{Tr} \times \mathrm{Cm}} \times 100\%
$$
R -- Reflectance ratio
Tr -- Reflectance intensity of reference light of test sector
Cr -- Reflectance intensity of reference light of blank sector
Tm -- Reflectance intensity of test sector of test light
Cm -- Reflectance intensity of blank sector of test light
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
The repeatability study was performed on 20 replicates of three levels of controls using two instruments and two lots. The 2x2 contingency table was applied to evaluate the total agreement, in terms of percentage, on each parameter. Cutoff values were used to determine the difference of positive and negative samples. Results are summarized below:
Control I
| Sample | Glucose | | Bilirubin | | Ketone | | SG | | Blood | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Cutoff | 759.8 mg/dL | | 4.6 mg/dL | | 59.4 mg/dL | | 1.022 | | 129.9 cells/μl | |
| | Pos | Neg | Pos | Neg | Pos | Neg | Pos | Neg | Pos | Neg |
| Positive | 20 | 0 | 20 | 0 | 20 | 0 | 19 | 0 | 20 | 0 |
| Negative | 0 | 20 | 0 | 20 | 0 | 20 | 1 | 20 | 0 | 20 |
| Agreement % | 100% | 100% | 100% | 100% | 100% | 100% | 95% | 100% | 100% | 100% |
{4}
Control I
| Sample | pH | | Protein | | Urobilinogen | | Nitrite | | Leukocyte | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Cutoff | 7.7 | | 178.7 mg/dL | | 4.9 mg/dL | | 0.8 | | 315.4 cells/μl | |
| | Pos | Neg | Pos | Neg | Pos | Neg | Pos | Neg | Pos | Neg |
| Positive | 20 | 0 | 20 | 0 | 20 | 0 | 20 | 0 | 20 | 0 |
| Negative | 0 | 20 | 0 | 20 | 0 | 20 | 0 | 20 | 0 | 20 |
| Agreement % | 100% | 100% | 100% | 100% | 100% | 100% | 100% | 100% | 100% | 100% |
Control II
| Sample | Glucose | | Bilirubin | | Ketone | | SG | | Blood | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Cutoff | 417.2 mg/dL | | 1.9 mg/dL | | 27.9 mg/dL | | 1.013 | | 39.3 cells/μl | |
| | Pos | Neg | Pos | Neg | Pos | Neg | Pos | Neg | Pos | Neg |
| Positive | 20 | 0 | 20 | 1 | 20 | 0 | 20 | 1 | 20 | 1 |
| Negative | 0 | 20 | 0 | 19 | 0 | 20 | 0 | 19 | 0 | 19 |
| Agreement % | 100% | 100% | 100% | 95% | 100% | 100% | 100% | 95% | 100% | 95% |
Control II
| Sample | pH | | Protein | | Urobilinogen | | Nitrite | | Leukocyte | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Cutoff | 7.3 | | 67.4 mg/dL | | 0.5 mg/dL | | 0.8 | | 98.6cells/μl | |
| | Pos | Neg | Pos | Neg | Pos | Neg | Pos | Neg | Pos | Neg |
| Positive | 20 | 0 | 19 | 0 | 20 | 0 | 20 | 0 | 19 | 0 |
| Negative | 0 | 20 | 1 | 20 | 0 | 20 | 0 | 20 | 1 | 20 |
| Agreement % | 100% | 100% | 95% | 100% | 100% | 100% | 100% | 100% | 95% | 100% |
Control III
| Sample | Glucose | | Bilirubin | | Ketone | | SG | | Blood | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Cutoff | 60 mg/dL | | 0.6 mg/dL | | 2.6 mg/dL | | 1.007 | | 8.8cells/μl | |
| | Pos | Neg | Pos | Neg | Pos | Neg | Pos | Neg | Pos | Neg |
| Positive | 20 | 0 | 20 | 0 | 20 | 0 | 20 | 0 | 20 | 0 |
| Negative | 0 | 20 | 0 | 20 | 0 | 20 | 0 | 20 | 0 | 20 |
| Agreement % | 100% | 100% | 100% | 100% | 100% | 100% | 100% | 100% | 100% | 100% |
Control III
| Sample | pH | | Protein | | Urobilinogen | | Nitrite | | Leukocyte | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Cutoff | 6.2 | | 11 mg/dL | | 0.5 mg/dL | | 0.8 | | 9.7cells/μl | |
| | Pos | Neg | Pos | Neg | Pos | Neg | Pos | Neg | Pos | Neg |
| Positive | 20 | 0 | 20 | 0 | 20 | 0 | 20 | 0 | 20 | 0 |
| Negative | 0 | 20 | 0 | 20 | 0 | 20 | 0 | 20 | 0 | 20 |
| Agreement % | 100% | 100% | 100% | 100% | 100% | 100% | 100% | 100% | 100% | 100% |
b. Linearity/assay reportable range:
See k970250
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
{5}
See k970250
d. Detection limit:
See k970250
e. Analytical specificity:
See k970250
f. Assay cut-off:
See k970250
2. Comparison studies:
a. Method comparison with predicate device:
100 specimens were tested using the Teco strips (Lot 3: 04454 Exp.: 10/07) on Uritek-720+ (S/N: 6015-2184) verses the Bayer Multistix 10SC reagent strips (Lot#: 2K02C Exp.: 04/06) on the Clinitek-200+ (S/N: 1034). The samples included every range measurement for every analyte. The Percent agreement for each analyte was determined by the number of positive and negative results. The comparison data showed substantial equivalence to the predicate device.
b. Matrix comparison:
Not applicable
3. Clinical studies:
a. Clinical Sensitivity:
Not applicable
b. Clinical specificity:
Not applicable
c. Other clinical supportive data (when a. and b. are not applicable):
Not applicable
4. Clinical cut-off:
Not applicable
{6}
5. Expected values/Reference range:
See k970250
N. Instrument Name:
URITEK-720+ Urine Analyzer
O. System Descriptions:
1. Modes of Operation:
Single-step or continuous testing
2. Software:
FDA has reviewed applicant’s Hazard Analysis and software development processes for this line of product types:
Yes ☐ X or No ☐
3. Specimen Identification:
Manual or Barcode patient ID entry
4. Specimen Sampling and Handling:
Manual Dip and place urine strip onto test table
5. Calibration:
Uritek-720+ is self-calibrated every time the power is turned on. The calibration is performed twice when the strip bed moves in-out two times. As a preliminary step, the sponsor recommends that users should ensure the position of strip bed and clean the calibration white dot.
6. Quality Control:
The sponsor recommends the following to their users:
1. Positive and negative controls are always recommended for testing purpose.
2. Use commercially available controls: High Abnormal, Abnormal, and Normal.
3. Prepare the controls accordingly to the instruction insert provided.
4. Test the urine strips with the controls. Perform the control testing according to the sample test instructions.
5. Record the results and compare to the ranges provided. Remark: The control testing is
{7}
only requested on every 100 tests or after turn on the power.
6. Retest the controls with different lot of strips, if test results are in doubt. If the test results are consistently in doubt, please contact TECO technical support for help.
P. Other Supportive Instrument Performance Characteristics Data Not Covered In The "Performance Characteristics" Section above:
Software documentation provided demonstrates the Uritek-720+ Urine Analyzer was designed and manufactured under well developed software lifecycle processes.
Q. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
R. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
8
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.