SPYSCOPE ACCESS AND DELIVERY CATHETER

K051504 · Boston Scientific Corp · FED · Jun 16, 2005 · Gastroenterology, Urology

Device Facts

Record IDK051504
Device NameSPYSCOPE ACCESS AND DELIVERY CATHETER
ApplicantBoston Scientific Corp
Product CodeFED · Gastroenterology, Urology
Decision DateJun 16, 2005
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2
AttributesTherapeutic, 3rd-Party Reviewed

Indications for Use

The SpyScope Access and Delivery Catheter will be intended to guide both optical and accessory type devices for diagnostic and therapeutic applications during endoscopic procedures in the biliary system including the hepatic ducts.

Device Story

The SpyScope Access and Delivery Catheter is a sterile, single-use device consisting of a flexible delivery catheter and a handle. It is used during endoscopic retrograde cholangiopancreatography (ERCP) procedures. The device is introduced through a duodenoscope (minimum 4.2mm working channel) to reach the biliary system. The distal tip features four-way articulation controlled by the handle. The device serves as a conduit to guide visualization probes (e.g., SpyGlass Direct Visualization Probe) and various accessory devices such as biopsy forceps, cytology brushes, and stone retrieval baskets. It is operated by a physician in a clinical setting. By providing a stable, steerable pathway, it facilitates targeted diagnostic and therapeutic interventions within the biliary tree.

Clinical Evidence

Bench testing only. No clinical data provided.

Technological Characteristics

Sterile, single-use catheter. Features: 4-way articulating distal tip, control knobs, locking mechanism, and working channel. Compatible with duodenoscopes having >=4.2mm working channel. Mechanical design.

Indications for Use

Indicated for patients undergoing endoscopic procedures in the biliary system, including hepatic ducts, requiring guidance of optical or accessory devices for diagnostic or therapeutic applications.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls: 1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following: 1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures. 2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use. 3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use. 2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes. 3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches 4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error. 5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program. (6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms. (7) Labeling must include: (i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified; (ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA; (iii) Identification of compatible devices; (iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification; (v) Reprocessing instructions for reusable components; (vi) A shelf life for any sterile components; (vii) A description of the device-specific use training program; (viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and (ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance. (8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include: (i) Device motion accuracy and repeatability; (ii) System testing; (iii) Instrument reliability; (iv) Crosstalk; (v) Table motion control; (vi) Thermal effects on tissue; (vii) User-device interface performance; (viii) Workspace access testing; and (ix) Performance testing with compatible devices. (9) Software verification, validation, and hazard analysis must be performed. (10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed. (11) Performance data must demonstrate the sterility of all patient-contacting device components. (12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life. (13) Performance data must validate the reprocessing instructions for the reusable components of the device. (14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible. (15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic. (16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information: (i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and (ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

Predicate Devices

Submission Summary (Full Text)

{0} DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center – WO66-G609 Silver Spring, MD 20993-0002 Boston Scientific Corporation % Mr. Neil E. Devine Sr. Staff Engineer Intertek Testing Services NA, Inc. 70 Codman Hill Road Boxborough, MA 01719 Re: K051504 Trade/Device Name: SpyScope™ Access and Delivery Catheter Regulation Number: 21 CFR§ 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: FED, GCJ, FBN Dated (Date on orig SE ltr): June 2, 2005 Received (Date on orig SE ltr): June 7, 2005 Dear Mr. Devine, This letter corrects our substantially equivalent letter of June 16, 2005. We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be {1} found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the *Federal Register*. Please be advised that FDA’s issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act’s requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH’s Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Benjamin R. Fisher -S Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2} SECTION 3 INDICATIONS FOR USE # Indications for Use 510(k) Number (if known): To be determined Device Name: SpyScope Access and Delivery Catheter Indications For Use: The SpyScope Access and Delivery Catheter is intended to guide both optical and accessory devices for diagnostic and therapeutic applications during endoscopic procedures in the biliary system including the hepatic ducts. Prescription Use ☑ (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use ☐ (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Davige Brogdon (Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number K051504 Page 1 of __ Premarket Notification, SpyScope Access and Delivery Catheter, May 16, 2005 Proprietary and Confidential Information of Boston Scientific Corporation {3} Boston Scientific JUN 16 2005 K051504 SECTION 10 510(K) SUMMARY # 510(K) SUMMARY ## 1. Submitter: Boston Scientific Corporation One Boston Scientific Place Natick, MA 01760 Owner/operator #9912058 Contact: Jennifer Johnson Sr. Regulatory Specialist Boston Scientific Marlborough 100 Boston Scientific Way Marlborough, MA 01752 Phone: 508-683-4178, Fax: 508-683-5939 Date Prepared: May 16, 2005 ## 2. Device: Trade Name: SpyScope Access and Delivery Catheter Common Name: Catheter Classification Name: Endoscopes and Accessories ## 3. Predicate Device: Olympus SwingTip Cannula, K011149 Olympus CHF Type BP30 Choledochofiberscope, K944473 ## 4. Device Description: The SpyScope Access and Delivery Catheter is a sterile, single-use device comprised of two main components: a flexible delivery catheter and a handle. The device is intended to be used to guide both the SpyGlass Direct Visualization Probe (K050403) or other visualization devices and accessory devices, (such as biopsy forceps, cytology brushes, stone retrieval baskets, etc.) during endoscopic retrograde cholangiopancreatography (ERCP) procedures. The SpyScope Access and Delivery Catheter is introduced to the desired anatomical location through a duodenoscope with a minimum working channel diameter of 4.2mm. The distal tip of the SpyScope Access and Delivery Catheter is designed to articulate in four directions. ## 5. Intended Use: The SpyScope Access and Delivery Catheter will be intended to guide both optical and accessory type devices for diagnostic and therapeutic applications during endoscopic procedures in the biliary system including the hepatic ducts. {4} # Boston Scientific ## 6. Technological Characteristics: Technological similarities between the SpyScope Access and Delivery Catheter and the Olympus SwingTip Cannula include the articulating tip and between the SpyScope Access and Delivery Catheter and the Olympus Choledochofiberscope includes the working channel, control knobs, and locking mechanism to hold tip in position. In instances where the technological characteristics may differ, it has been demonstrated that there are no new questions raised regarding safety and effectiveness of the SpyScope Access and Delivery Catheter. ## 7. Performance Data: Bench testing was conducted to evaluate the design features of the SpyScope Access and Delivery Catheter and establish specifications of the descriptive characteristics for the proposed device. These specifications were used to demonstrate substantial equivalence of the proposed device to the predicate devices. ## 8. Conclusion: BSC has demonstrated that the SpyScope Access and Delivery Catheter is substantially equivalent to Olympus’s currently marketed SwingTip Cannula and Olympus’s currently marketed CHF Type BP30 Choledochofiberscope.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...