WIESLAB CAP PR-3 ANCA
Device Facts
| Record ID | K051458 |
|---|---|
| Device Name | WIESLAB CAP PR-3 ANCA |
| Applicant | Eurodiagnostica |
| Product Code | MOB · Immunology |
| Decision Date | Nov 22, 2005 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 866.5660 |
| Device Class | Class 2 |
| Attributes | Real-World Evidence |
Real-World Evidence
| Submission | Device | Sponsor | RWD Sources | RWE Use Summary | Key Tags |
|---|---|---|---|---|---|
| K051458 · Nov 22, 2005 | WIESLAB CAP PR-3 ANCA | Eurodiagnostica | Retrospective clinical serum samples | Retrospective clinical serum samples were used to evaluate the clinical sensitivity and specificity of the device against established clinical diagnoses (WG, MP, SLE, RA, GBM) and to compare performance against existing diagnostic methods (IIF-ANCA and alternative ELISA). | Retrospective clinical samples; Clinical sensitivity; Clinical specificity; Diagnostic aid |
Clinical Evidence
| Study Design | Population | Comparator | Key Endpoints |
|---|---|---|---|
| Retrospective analysis of frozen clinical sera | Patients with Wegener's granulomatosis (WG), microscopic polyangiitis (MP), systemic lupus erythematosus (SLE), rheumatoid arthritis (RA), glomerular basement membrane (GBM) disease, and blood donors; Sample Size: 295 | Not applicable for this study | Clinical sensitivity and specificity for detection of IgG antibodies to PR3 |
| Retrospective comparative study | Patients with systemic vasculitis; Sample Size: 78 | IIF-ANCA | Relative sensitivity and specificity |
| Retrospective comparative study | Clinical serum samples; Sample Size: 180 | Alternative ELISA | Relative sensitivity, specificity, and accuracy |
Indications for Use
The Wieslab™ Cap PR3-ANCA is an enzyme-linked immunosorbent assay (ELISA) for detection and semi-quantitation of IgG antibodies to proteinase 3 (PR3) in human sera. The assay is used to detect antibodies in a single serum specimen. The results of the assay are to be used as an aid to the diagnosis of Wegener's granulomatosis. The assay is intended for use in patients with signs and symptoms consistent with WG. It is not intended for screening a healthy population. The analysis should be performed by trained laboratory professionals.
Device Story
Wieslab™ Cap PR3-ANCA is an ELISA test kit for detecting IgG antibodies to proteinase 3 (PR3) in human serum. Microtiter wells are coated with monoclonal antibody to PR3 and PR3 antigen. Patient serum is incubated in wells; specific antibodies bind to antigen. After washing, alkaline phosphatase-labeled anti-human IgG conjugate is added. Substrate solution is added; color intensity (optical density) correlates to bound antibody concentration. Performed by trained laboratory professionals in clinical settings. Results are calculated against a calibrator curve in arbitrary units. Output aids physicians in diagnosing Wegener's granulomatosis in symptomatic patients.
Clinical Evidence
Clinical performance evaluated on 363 characterized sera. Sensitivity for Wegener’s granulomatosis (WG) was 96.0% (95% CI: 83.3-99.5%). Specificity in healthy controls was 100% (95% CI: 97.0-100%), SLE 100% (95% CI: 92.0-100%), and RA 100% (95% CI: 80.5-100%). Method comparison with predicate (n=180) showed 98.3% overall agreement. Analytical precision (intra-assay CV 3-13%, inter-assay CV 4-14%) and linearity (10-320 U/mL) were established.
Technological Characteristics
ELISA-based immunological test system. Components: microtiter strips coated with monoclonal antibody to PR3 and PR3 antigen. Detection: alkaline phosphatase-labeled anti-human IgG conjugate and substrate solution. Measurement: optical density (OD) via spectrophotometry. Quantitative output in arbitrary units.
Indications for Use
Indicated for patients with signs and symptoms consistent with Wegener’s granulomatosis (WG) to aid in diagnosis. Not intended for screening healthy populations.
Regulatory Classification
Identification
A multiple autoantibodies immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the autoantibodies (antibodies produced against the body's own tissues) in serum and other body fluids. Measurement of multiple autoantibodies aids in the diagnosis of autoimmune disorders (disease produced when the body's own tissues are injured by autoantibodies).
Predicate Devices
- Wieslisa® PR-3 ANCA Kit (k974167)