The CONSERVE® Total Femoral Head is indicated for use in total hip arthroplasty for reduction or relief of pain and/or improved hip function in skeletally mature patients with the following conditions: 1. non-inflammatory degenerative joint disease such as osteoarthritis, avascular necrosis, ankylosis, protrusio acetabuli, and painful hip dysplasia; 2. inflammatory degenerative joint disease such as rheumatoid arthritis; 3. correction of functional deformity; and, 4. revision procedures where other treatments or devices have failed The CONSERVE® Total Femoral Head is intended for single patient use only.
Device Story
CONSERVE® Total Femoral Head is a cobalt-chromium-molybdenum orthopedic implant used in total hip arthroplasty. It serves as a replacement femoral head articulating with an acetabular component. The device features a superfinished articulating surface and a WMT12/14 taper connection compatible with existing femoral stems. It is intended for use by orthopedic surgeons in clinical settings to reduce pain and improve hip function. The device is provided sterile for single-patient use. Clinical benefit is derived from the restoration of joint function and relief of pain associated with degenerative or inflammatory joint conditions.
Clinical Evidence
Bench testing only. In vitro hip wear simulator testing compared the CONSERVE® Total Femoral Head to the Metal TRANSCEND® Femoral Head (both 54mm diameter) over 5 million cycles using orbital bearing machines. Results showed the CONSERVE® device had an average of 10 times lower volumetric run-in wear rate and 3.2 times lower cumulative volumetric wear than the predicate. Note: In vitro simulator results do not quantitatively predict clinical wear performance.
Indicated for skeletally mature patients undergoing total hip arthroplasty for non-inflammatory degenerative joint disease (osteoarthritis, avascular necrosis, ankylosis, protrusio acetabuli, painful hip dysplasia), inflammatory degenerative joint disease (rheumatoid arthritis), correction of functional deformity, or revision procedures.
Regulatory Classification
Identification
A hip joint metal/metal semi-constrained, with a cemented acetabular component, prosthesis is a two-part device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that consist of a femoral and an acetabular component, both made of alloys, such as cobalt-chromium-molybdenum. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).
Predicate Devices
Metal TRANSCEND® Femoral Heads (Larger Sizes)
Submission Summary (Full Text)
{0}------------------------------------------------
K051348 (pg 1 of 3)
AUG ] 9 2005
## 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS
In accordance with the Food and Drug Administration Rule to implement provisions of the Safe in accordance with the 100d and in conformance with 21 CRF 807, this information serves as a Medical Devices 110 01 199 cities of the CONSERVE® Total Femoral Head.
| Submitted By: | Wright Medical Technology, Inc. |
|-------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date: | August 3, 2005 |
| Contact Person: | Theresa Leister<br>Regulatory Affairs Specialist |
| Proprietary Name: | CONSERVE® Total Femoral Head |
| Common Name: | Femoral Head |
| Classification Name and Reference: | 21 CFR 888.3320 Hip joint metal/ metal semi-<br>constrained, with a cemented acetabular component<br>prosthesis - Class III<br>21 CFR 888.3330 Hip joint metal/ metal semi-<br>constrained, with an uncemented acetabular<br>component prosthesis – Class III |
| Device Product Code and Panel Code: | Orthopedics/87/JDL<br>Orthopedics/87/KWA |
## DEVICE INFORMATION
## A. INTENDED USE
The CONSERVE® Total Femoral Head is indicated for use in total hip arthroplasty for reduction or relief of pain and/or improved hip function in skeletally mature patients with the following conditions:
- 1. non-inflammatory degenerative joint disease such as osteoarthritis, avascular necrosis, ankylosis, protrusio acetabuli, and painful hip dysplasia;
- 2. inflammatory degenerative joint disease such as rheumatoid arthritis;
- 3. correction of functional deformity; and,
- 4. revision procedures where other treatments or devices have failed
The CONSERVE® Total Femoral Head is intended for single patient use only.
{1}------------------------------------------------
KCS1348 (pg 2 of 3)
### B. DEVICE DESCRIPTION
#### Design Features
The design features of the CONSERVE® Total Femoral Head are summarized below:
- Manufactured from wrought cobalt-chromium-molybdenum .
- The articulating surface of the implants will be superfinished .
- Available sizes: 36mm, 38mm, 42mm, 42mm, 46mm, 48mm, 50mm, 52mm, 52mm, . 54mm, 56mm
- The taper connection is identical to previously cleared taper connections and is . intended to be used with our existing femoral stems manufactured with WMT12/14 taper.
#### Wear Claims
The following marketing claims will be made for the CONSERVE® Total Femoral Head:
- The articulation of the CONSERVE® Total Femoral Head Bearing has an average of . 10 times lower volumetric run-in wear rate than the currently available Metal TRANSCEND® Femoral Head Bearings. *
- The articulation of the CONSERVE® Total Femoral Head Bearing has an average of . 3.2 times lower cumulative volumetric wear than the currently available Metal TRANSCEND® Femoral Head Bearings. *
#### * Wear Test Information
- (a) The CONSERVE® Total Femoral Head and the Metal TRANSCEND® Femoral Head are manufactured by Wright Medical Technology, Inc.
- (b) The test device was the CONSERVE® Total Femoral Head and the control device was the Metal TRANSCEND® Femoral Head
- (c) Both the test and control devices were 54mm diameter femoral heads.
- (d) The articulating surface of both the test and control device was mirror finished.
- (e) The hip simulators used were orbital bearing hip wear test machines manufactured by Shore Western Manufacturing, Inc.
- (f) Both the test and control devices were tested for 5 million cycles.
- (g) The material of the device was a CoCr alloy and the bearing size was 54mm.
- (h) The lubricant used was 90% alpha calf serum with 0.2% sodium azide, 20mM EDTA and distilled water.
- (i) The average amount of volumetric wear at 5 million cycles for the test device was 1.02mm3 less than that of the control device.
- (j) Both the test and control devices were sterilized.
- (k) The results of in vitro hip wear simulator tests have not been shown to quantitatively predict clinical wear performance.
Image /page/1/Picture/25 description: The image shows the logo for Wright Medical Technology, Inc. The logo features a stylized "W" symbol above the word "WRIGHT" in a bold, sans-serif font. Below "WRIGHT" are the words "MEDICAL TECHNOLOGY, INC." in a smaller, sans-serif font.
Image /page/1/Picture/26 description: The image shows a black and white drawing of a globe. The globe is tilted on its axis and is sitting on a stand. The continents of Asia and Australia are visible on the globe. The drawing is simple and has a vintage look.
{2}------------------------------------------------
Special 510(k)
CONSERVE® Total Femoral Head
510(K) SUMMARY
Page 3 of 3
K051348(pg 3 of 3)
# C. SUBSTANTIAL EQUIVALENCE INFORMATION
The intended use and type of interface of the CONSERVE® Total Femoral Head are The intended use and type of internation Femoral Heads (Larger Sizes). The design features and material are substantially equivalent to the Metal TRANSCEND® Femoral Heads (Larger Sizes). The safety and effectiveness of this device are adequately supported by the substantial equivalence information, materials data, and testing results, provided within the Premarket Notification.
Image /page/2/Picture/4 description: The image shows a black and white drawing of a globe. The globe is tilted on its axis and is sitting on a stand. The continents are visible on the globe, but they are not labeled. The globe is surrounded by a ring, which may represent the equator or another line of latitude. The image is simple and straightforward, and it is likely intended to be used for educational purposes.
510(K) SUMMARY- Amended 8-3-05
WRIGHT
MEDICAL TECHNOLOGY, INC.
{3}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Image /page/3/Picture/2 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it, representing medicine and healing. The text "DEPARTMENT OF HEALTH AND HUMAN SERVICES - USA" is arranged in a circular pattern around the caduceus symbol.
# AUG 1 9 2005
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Ms. Theresa Leister Regulatory Affairs Specialist Wright Medical Technology, Inc. . . . . . 5677 Airline Road Arlington, Tennessee 38002
Re: K051348
Trade/Device Name: NSERVE® Total Femoral Head Regulation Number: 21CFR 888.3320, 21 CFR 888.3330 Regulation Name: Hip joint metal/metal semi-constrained, with a cemented acetabular component prosthesis, Hip joint metal/metal semi-constrained, with an uncemented acetabular component, prosthesis Regulatory Class: III
Product Code: JDL, KWA Dated: August 1, 2005 Received: August 2, 2005
Dear Ms. Leister:
We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your bedeemined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the encreated of the enactment date of the Medical Device Amendments, or to conninered processions) in accordance with the provisions of the Federal Food, Drug, de necs mat nave been require approval of a premarket approval application (PMA). and Cosmetic Prov (110) market the device, subject to the general controls provisions of the Act. The I va mayy are or visions of the Act include requirements for annual registration, listing of general controls profice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it rr your device is elabilition (tional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act than v Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
Page 2 – Ms. Theresa Leister
This letter will allow you to begin marketing your device as described in your Section 510(k) I his letter will anow you to begin manxoting your and equivalence of your device to a legally premarket nothleadon: The PDA midning of castification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific acvice for your accel (240) 276-0210. Also, please note the regulation entitled, comact the Office of Compullion in (21 CFR Part 807.97). You may obtain Misblanding by reference to premaintonions in the Act from the Division of Small other general information on your responsive Assistance at its toll-free number (800) 638-2041 or Manufacturers, International and Oolbanhoon of the world the more of the many of the states html.
Sincerely yours,
Z
Michael W. Mollen
Mark N. Melkerson Acting Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Indications for Use
510(k) Number (if known):
Device Name: CONSERVE® Total Femoral Head
Indications For Use:
The CONSERVE® Total Femoral Head is indicated for use in total hip arthroplasty for reduction The CONSERVE® Total Fead is marcally mature patients with the following conditions:
- non-inflammatory degenerative joint disease such as osteoarthritis, avascular necrosis, 1. ankylosis, protrusio acetabuli, and painful hip dysplasia;
- allkylosis, protrasto accuse joint disease such as rheumatoid arthritis;
- 3. correction of functional deformity; and,
- 5. conceedures where other treatments or devices have failed
The CONSERVE® Total Femoral Head is intended for single patient use only.
X Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
2-1 - 1 - 1 - - -
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
signature
(Division Sign-Off) Division of General, Restorative, and Neurological Devices
**510(k) Number** k051348
Page 1 of ____________________________________________________________________________________________________________________________________________________________________
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.