K051335 · Exactech, Inc. · JDI · Jun 21, 2005 · Orthopedic
Device Facts
Record ID
K051335
Device Name
ACUMATCH 12/14 PRESS-FIT FEMORAL STEMS
Applicant
Exactech, Inc.
Product Code
JDI · Orthopedic
Decision Date
Jun 21, 2005
Decision
SESE
Submission Type
Special
Regulation
21 CFR 888.3350
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
All Exactech Hip Systems are indicated for use in skeletally mature individuals undergoing primary surgery for hip replacement due to osteoarthritis, rheumatoid arthritis, osteonecrosis, post-traumatic degenerative problems of the hip, and for treatment of proximal femoral fractures where prosthetic replacement is determined by the surgeon as the preferred treatment. Components of Exactech Hip Systems are also potentially indicated for ankylosing spondylitis, congenital hip dysplasia, revision of failed previous reconstructions where sufficient bone stock is present, and to restore mobility resulting from previous fusion. AcuMatch® P-Series and AcuMatch® L-Series press-fit femoral stems are intended for press-fit fixation. Press-fit components without hydroxyapatite (HA) coating may also be used with bone cement at the discretion of the surgeon.
Device Story
AcuMatch 12/14 Press-Fit Femoral Stems are orthopedic implants used in hip replacement surgery. The device consists of a titanium alloy femoral stem with a trapezoidal cross-section and distal taper. The P-Series features a plasma-spray surface enhancement, while the L-Series has a corundum finish; both offer an optional hydroxyapatite (HA) coating. The stems are designed for press-fit fixation, though non-HA coated versions may be used with bone cement at the surgeon's discretion. The device is implanted by an orthopedic surgeon in a clinical/hospital setting. It serves as a structural replacement for the femoral head and neck, restoring joint function and mobility for patients suffering from degenerative hip disease or fractures. Modifications from the predicate include a threaded insertion hole (P-Series), expanded taper threadform tolerances, and a collared P-Series option.
Clinical Evidence
No clinical data provided. Substantial equivalence is supported by bench testing and engineering analyses.
Technological Characteristics
Materials: Titanium alloy (ASTM F1472). Design: Trapezoidal cross-section, distal taper. Surface: Plasma-spray (P-Series) or corundum finish (L-Series), with optional hydroxyapatite (HA) coating. Fixation: Press-fit or cemented (non-HA versions).
Indications for Use
Indicated for skeletally mature individuals undergoing primary hip replacement for osteoarthritis, rheumatoid arthritis, osteonecrosis, post-traumatic degenerative hip conditions, or proximal femoral fractures. Also indicated for ankylosing spondylitis, congenital hip dysplasia, and revision of failed reconstructions with sufficient bone stock, or to restore mobility after previous fusion.
Regulatory Classification
Identification
A hip joint metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and an acetabular resurfacing component made of ultra-high molecular weight polyethylene and is limited to those prostheses intended for use with bone cement (§ 888.3027).
{0}------------------------------------------------
**Exactech**
JUN 2 1 2005
FL 32653
FAX 352-378-2617
# Exactech® AcuMatch 12/14 Press-Fit Femoral Stems
#### 510(k) Summary of Safety and Effectiveness Special 510(k)
#### Trade or proprietary or model name(s):
AcuMatch 12/14 P-Series Press-Fit Plasma Femoral Stems AcuMatch 12/14 L-Series Press-Fit Femoral Stems
# Information on devices to which substantial equivalence is claimed:
| 510(k) | Trade or Proprietary or Model Name | Manufacturer |
|-------------------|-----------------------------------------------------------------------------------------------------------|---------------|
| Number<br>K041906 | AcuMatch 12/14 P-Series Plasma Press-Fit Femoral Stems<br>AcuMatch 12/14 L-Series Press-Fit Femoral Stems | Exactech, Inc |
#### INDICATIONS
All Exactech Hip Systems are indicated for use in skeletally mature individuals undergoing primary surgery for hip replacement due to osteoarthritis, rheumatoid arthritis, osteonecrosis, post-traumatic degenerative problems of the hip, and for treatment of proximal femoral fractures where prosthetic replacement is determined by the surgeon as the preferred treatment. Components of Exactech Hip Systems are also potentially indicated for ankylosing spondylitis, congenital hip dysplasia, revision of failed previous reconstructions where sufficient bone stock is present, and to restore mobility resulting from previous fusion.
AcuMatch® P-Series and AcuMatch® L-Series press-fit femoral stems are intended for press-fit fixation.
Press-fit components without hydroxyapatite (HA) coating may also be used with bone cement at the discretion of the surgeon.
#### Special 510(k) Modifications
The AcuMatch 12/14 Press-Fit femoral stems were modified from the predicate as follows:
- The geometry of the insertion hole feature was modified from an oblong slot to an . oblong slot with threads (P-Series only).
- The tolerance of the 12/14 taper threadform geometry was expanded. .
- Collared 12/14 P-Series Press-Fit Stem option was added. .
Rev. 05/19/05
{1}------------------------------------------------
Exactech
Image /page/1/Picture/1 description: The image shows an address written in cursive and print. The address is "Kari 1335 2320 NW 66TH COURT GAINESVILLE, FL 32653". The first line is written in cursive, while the rest of the address is printed.
FAX 352-378-2617
## Exactech® AcuMatch 12/14 Press-Fit Femoral Stems
#### 510(k) Summary of Safety and Effectiveness Special 510(k)
## AcuMatch 12/14 Press-Fit Femoral Stems
AcuMatch 12/14 Press-Fit femoral stems are composed of titanium alloy (ASTM F1472), have a trapezoidal cross-sectional geometry and distal taper.
- 12/14 P-Series model has a plasma-spray surface enhancement. ●
- 12/14 L-Series model has a corundum finish. .
Both models have a hydroxyapatite (HA) coating option. The femoral stems are intended Doth models have a try are models without the HA coating may be used with bone cement at the discretion of the surgeon.
#### Conclusion:
Conclusion." results one clightering - 12:14 Press-Fit Femoral Stem components would be adequate for anticipated in vivo use. This includes empirical testing and engineering analyses. Based on successful results we conclude that the proposed devices are substantially equivalent to Exactech's predicate femoral stems.
Rev. 05/19/05
Section 4 Page 3 of 3
{2}------------------------------------------------
Image /page/2/Picture/2 description: The image shows the seal of the Department of Health and Human Services (HHS) of the United States. The seal features the department's name in a circular arrangement around a stylized eagle emblem. The eagle is depicted with three lines forming its head and wings, giving it a modern and abstract appearance.
JUN 2 1 2005
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Ms. Maritza Elias Regulatory Representative Exactech, Inc. 2320 N.W. 66th Court Gainesville, Florida 32653
Re: K051335
Trade/Device Name: Exactech Acumatch 12/14 Press-Fit Femoral Stems Regulation Number: 21 CFR 888.3353, 888.3350 Regulation Name: Hip joint metal/cemented/polymer semi-constrained cemented or nonregaration Names - Hip joint metal/polymer semiconstrained cemented prosthesis Regulatory Class: II Product Code: JDI, LWJ, MEH, LZO Dated: May 17, 2005 Received: May 23, 2005
Dear Ms. Elias:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for doo based or to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The I ou mayy atch s provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it rr your de rise to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean r that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set
{3}------------------------------------------------
## Page 2 – Ms. Maritza Elias
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic forth in the quality systems (Sections 531-542 of the Act); 21 CFR 1000-1050. product radiation control provisions (seting your device as described in your Section 510(k) This letter will anow you to oegin mains of substantial equivalence of your device to a legally premarket notheadon. The PDF mining sitems of your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific acriously at (240) 276-0120. Also, please note the regulation entitled, Comaci the Office of Comphanes and in (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small other general mionnation en Journer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Miriam C. Provost
Miriam C. Provost, Ph.D. Acting Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Exactech®, Inc.
# Exactech AcuMatch 12/14 Press-Fit Femoral Stems
## Indications for Use
K051335 510(k) Number:
#### INDICATIONS
INDICATIONS
All Exactech Hip Systems are indicated for use in skeletally mature individuals All Exactech Hip Systems are more to ass them to osteon thritis, the matoid
undergoing primary surgery for hip replacement due to steem of the hin, and for undergoing primary surgely for mp replacement we problems of the hip, and for arthritis, osteonecrosis, post-tradinatio copines where prosthetic replacement is determined by treatment of proximal remoral fractures when provins of Exactech Hip Systems are also
the surgeon as the preferred treatment. Components hip dysplassia, revision of the surgeon as the preletted treatinent. Configure on tall hip dysplass, revision of
potentially indicated for ankylosing spondylitis, congent, and to restore potentially indicated for allRylosing spondyinia, occaged is present, and to restore mobility resulting from previous fusion.
AcuMatch® P-Series and AcuMatch® L-Series press-fit femoral stems are intended for press-fit fixation.
Press-fit components without hydroxyapatite (HA) coating may also be used with bone cement at the discretion of the surgeon.
Prescription Use
X
Over the Counter Use _
Please do not write below this line - use another page if needed.
or
Concurrence of CDRH, Office of Device Evaluation (ODE)
Miriam C. Provost
(Division Sign-Off) Division of General, Restorative and Neurological Devices
510(k) Number_________________________________________________________________________________________________________________________________________________________________
Section 3 Page 1 of 1
05/18/05
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.