TERASON ULTRASOUND SYSTEM WITH CONTINUOUS WAVE SOPPLER
Applicant
Teratech Corp.
Product Code
IYN · Radiology
Decision Date
Jun 3, 2005
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1550
Device Class
Class 2
Attributes
Pediatric, 3rd-Party Reviewed
Indications for Use
The TERASON™ Ultrasound System with Continuous Wave (CW) Doppler is intended for diagnostic ultrasound imaging or fluid flow analysis of the human body; specific indications for use a tabulated in Section 4.3 of this submission.
Device Story
The TERASON Ultrasound System (Models t3000 and Echo) is a diagnostic ultrasound platform providing B-mode, M-mode, PWD, CWD, Color Doppler, and harmonic imaging. It processes ultrasound echoes to generate real-time images and flow analysis. The system features 128 transmit/receive channels (Echo model features 128 receive channels). It supports ECG triggering, tissue Doppler, and LV dyssynchrony assessment (Echo model). Operated by clinicians in hospitals, clinics, or military settings, the system provides visual output for diagnostic assessment and procedural guidance (e.g., needle/catheter placement). The device aids clinical decision-making by providing anatomical and hemodynamic information, potentially improving diagnostic accuracy and procedural safety.
Clinical Evidence
Bench testing only. No clinical data provided. Substantial equivalence is supported by performance testing and comparison of technical specifications and intended uses to the predicate device.
Technological Characteristics
Diagnostic ultrasound system with 128 transmit/receive channels. Modes: B, M, PWD, CWD, Color Doppler, Harmonic Imaging. Features include ECG trigger, tissue Doppler, and LV dyssynchrony assessment. Connectivity and software architecture are consistent with diagnostic ultrasound standards. Sterilization/biocompatibility requirements apply to transducers.
Indications for Use
Indicated for diagnostic ultrasound imaging and fluid flow analysis in fetal, abdominal, intra-operative, pediatric, small organ, neonatal/adult cephalic, trans-rectal, trans-vaginal, musculoskeletal, cardiac, and peripheral vessel applications. Includes ultrasound guidance for needle/catheter placement, cryosurgery, brachytherapy, and infertility monitoring. Used in hospital, clinic, and military field settings.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
K051334
# 510(k) Summary TERATECH Corporation TERASON™ Ultrasound System with Continuous Wave (CW) Doppler
- 1. SPONSOR
2.
TERATECH Corporation 77-79 Terrace Hall Rd. Burlington, MA 01803
| Contact Person: | Charles F. Hottinger, Ph.D., RAC,<br>Regulatory Affairs Consultant | | | |
|--------------------|--------------------------------------------------------------------|--|--|--|
| Telephone: | 206-780-7945 | | | |
| Date Prepared: | May 11, 2005 | | | |
| DEVICE NAME | | | | |
| Proprietary Name: | TERASON™ Ultrasound System with Continuous<br>Wave (CW) Doppler | | | |
| Common/Usual Name: | Diagnostic Ultrasound System | | | |
- Classification Name: Diagnostic Ultrasound Transducer (21 CFR 892.1570, 90-ITX) Ultrasonic Pulsed Echo Imaging System (21 CFR 892.1560, 90-IYO) Diagnostic Ultrasonic Transducer (21 CFR 892.1570, 90-ITX)
- 3. PREDICATE DEVICES
Toshiba Powervision 6000 (K991710)
- 4. INTENDED USE
The TERASON™ Ultrasound System with Continuous Wave (CW) Doppler is intended for diagnostic ultrasound imaging or fluid flow analysis of the human body; specific indications for use a tabulated in Section 4.3 of this
{1}------------------------------------------------
submission.
#### 5. DEVICE DESCRIPTION
The TERASON™ Ultrasound System with Continuous Wave (CW) Doppler introduces two new versions of the TERATECH Model 2000 Imaging System.
| | TERATECH | TERASON ™ | TERASON ™ |
|--------------------------------------------|----------|-----------|-----------|
| | 2000 | t3000 | Echo |
| Transmit: Receive<br>Beam forming Channels | 128:64 | 128:64 | 128:128 |
| CW Doppler | No | Yes | Yes |
| ECG Trigger | No | Yes | Yes |
| Tissue Doppler | No | No | Yes |
| LV Dyssynchrony<br>Assessment | No | No | Yes |
#### BASIS FOR SUBSTANTIAL EQUIVALENCE 6.
The TERASON™ Ultrasound System is substantially equivalentto the Toshiba Powervision 6000, which is currently in commercial distribution in the United States. Since it is identical in modes of operation, and intended for the same clinical applications, as described in this submission.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image is a seal for the Department of Health & Human Services - USA. The seal is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract image of an eagle.
JUN 3 2005 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
TERATECH Corporation % Mr. Mark Job Responsible Third Party Regulatory Technology Services LLC 1394 25th Street NW BUFFALO MN 55313
Re: K051334
K031334
Trade Name: TERASON Ultrasound System with Continuous Wave Doppler Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulation Number: 21 CFR 892.1560 Regulation Name: Ultrasonic pulsed echo imaging system Regulation Number: 21 CFR 892.1570 Regulation Name: Diagnostic ultrasonic transducer Regulatory Class: II Product Code: IYN, IYO, and ITX Dated: May 19, 2005 Received: May 23, 2005
## Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your Soctor o re(i) proxing is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device merslate collinered pror to May 20, 1978, its classified in accordance with the provisions of the Amendinents, or to devices and Cosmetic Act (Act). You may, therefore, market the device, subject to Federal Food, Drug, and Oosmetic Fee et. The general controls provisions of the Act include the general controls provisions of the Frica - Fift going of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
This determination of substantial equivalence applies to the following transducers intended for I mis deternmination of subbanial vith Continuous Wave Doppler, as described in your premarket notification:
Transducer Model Number
4C2 4V2
{3}------------------------------------------------
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), ( If your device is classified (see above) into entrols. Existing major regulations affecting your device. FDA it may be subject to such additional controls: "Listing negation of 898. In addition, FDA
can be found in the Code of Federal Regulations, Title 21, Parts 800 to 800 to 100 can be found in the Code of I cuctar Rogalations, Friderice in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean Please be advised that FDA s issualite of a substitive with other requirements of the Act
that FDA has made a determination that your device also anaios. You must that FDA has made a delefinination and your arres by other Federal agencies. You must or any Federal statutes and regulations administered by Septiration and listing (21)
comply with all the Act's requirements, including, but not limited on viraments as set comply with all the Act s requirements, mortalians practice requirements as set
CFR Part 807); labeling (21 CFR Part 801); god if caplicable, the electronic CFR Part 807); labeling (21 CFA 1 at 601); good manaxt 820); and if applicable, the electronic
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicab forth in the quality systems (Q5) regulation (Sections 531-542 of the Act); 21 CFR 1000-1050.
product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-
This determination of substantial equivalence is granted on the prior to shipping of This determination of substantial cquivalence special report. This seport should contain complete the first device, you submit a postclearance spocal reportan ine devices, requested
information, including acoustic output measurements based on production for Masufaqurers information, including accusur output measurements 30, 1997 "Information for Manufacturers" in Appendix G, (enclosed) of the Celler's beptems and Translucers." If the special Seeking Marketing Clearance of Diagnosite United Oritus output greater than approved report is incomplete of contails macceptable values (orgi, accuration of possible as a result may be considered adulterated or misbranded.
The special report should reference the manufacturer's 510(k) number. It should be clearly and The special report should reference are marked be submitted in duplicate to:
> Food and Drug Administration Center for Devices and Radiological Health Document Mail Center (HFZ-401) 9200 Corporate Boulevard Rockville, Maryland 20850
This letter will allow you to begin marketing your device as described in your premarket I his letter will anow you to begin marketing your device of your device to a legally marketed nothication. The FDA midnig of substantial equivalier of thus permits your device to proceed to market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific advice ion your act (240) 276-0120. Also, please note the regulation entitled, Contact the Office of Compunation at (210FR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small other general information on your response Assistance at its toll-free number (800) 638-204 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html
{4}------------------------------------------------
Page 3 – Mr. Job
If you have any questions regarding the content of this letter, please contact Rodrigo C. Perez at (301) 594-1212.
Sincerely yours,
Rat C. Pelly
ﮨﮯ
Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure(s)
{5}------------------------------------------------
### DIAGNOSTIC ULTRASOUND INDICATIONS FOR USE FORM
| System: | TERASON™ Ultrasound System | | | | | | | |
|-----------------------------|--------------------------------------------------------------------------------------------------|-------------------|------|------|-----|----------------|-----------------|--------|
| Transducer: | (see comments) | | | | | | | |
| | Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | | | | | | |
| Clinical Application | | Mode of Operation | | | | | | |
| General<br>(Track I Only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppa | Comb.<br>Modesb | Otherc |
| Ophthalmic | Ophthalmic | | | | | | | |
| Fetal<br>Imaging<br>& Other | Fetalh | P1,4 | P2,4 | P2,4 | | P2,4 | P2,4 | P2,4 |
| | Abdominali: | P1,4 | P2,4 | P2,4 | N | P2,4 | P2,4 | P2,4 |
| | Intra-operative (Spec.)j,k | P4 | P4 | P4 | | P4 | P4 | P4 |
| | Intra-operative (Neuro) | P5 | P5 | P5 | | P5 | P5 | P5 |
| | Laparoscopic | P6 | P6 | P6 | | P6 | P6 | P6 |
| | Pediatricd: | P1,4 | P2,4 | P2,4 | N | P2,4 | P2,4 | P2,4 |
| | Small Organ (Thyroid, Breast, Testes, etc.)c,d: | P2,4 | P2,4 | P2,4 | | P2,4 | P2,4 | P2,4 |
| | Neonatal Cephalice: | P1,4 | P2,4 | P2,4 | N | P2,4 | P2,4 | P2,4 |
| | Adult Cephalice: | P1,4 | P2,4 | P2,4 | N | P2,4 | P2,4 | P2,4 |
| | Trans-rectalf: | P2,4 | P3,4 | P3,4 | | P3,4 | P3,4 | P3,4 |
| | Trans-vaginalg: | P2,4 | P3,4 | P3,4 | | P3,4 | P3,4 | P3,4 |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | |
| | Musculo-skel. (Convent.)d: | P2,4 | P2,4 | P2,4 | | P2,4 | P2,4 | P2,4 |
| | Musculo-skel. (Superfic)e: | P2,4 | P2,4 | P2,4 | | P2,4 | P2,4 | P2,4 |
| | Intra-luminal | | | | | | | |
| | Other (Specify) | | | | | | | |
| Cardiac | Cardiac Adult | P1 | P2 | P2 | N | P2 | P2 | P2 |
| | Cardiac Pediatric | P1 | P2 | P2 | N | P2 | P2 | P2 |
| | Trans-esoph. (Cardiac) | | | | | | | |
| | Other (Specify) | | | | | | | |
| Peripheral<br>Vessel | Peripheral vesseld: | P1,4 | P2,4 | P2,4 | | P2,4 | P2,4 | P2,4 |
| | Other (Specify) | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
Includes Color Doppler (CD), Directional Power Doppler (DPD), and (non-directional) Power Doppler.
" Molades Oblor Boppicil (OD), B+DPD, B+PD.
° Harmonic Imaging (HI)
4 Includes ultrasound guidance for placement of needles, catheters.
* Abdominal organs and peripheral vessel.
I Includes ultrasound guidance for placement of needles, catheters, cryosurgery, and brachytherapy
9 Includes ultrasound guidance of transvaginal biopsy, infertility monitoring of follicle development.
Includes guidance of amniocentesis, infertility monitoring of follicle development. 1
System uses previously cleared under K992505 with 3 MHz Mindel L3 (Linear).
2 System uses previously cleared under K0121919
3 System uses previously cleared under K0121919
3 System uses previously cleared under K012191
3 System uses previously cleared under K010883.
* System uses previously cleared under K010003.
* System uses previously cleared under K030199.
System uses previously cleared under K040840.
System uses previously cleared under K040040.
Includes uses in military field settings in addition to hospital/clinic settings.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of Center for Devices and Radiological Health, Office of Device Evaluation Prescription Use (Per 21 CFR 801.109)
Ruta Phillips
(Division Sign-Off and Ra 510k) Nur
{6}------------------------------------------------
| System: | TERASON TM Ultrasound System |
|---------|------------------------------|
|---------|------------------------------|
Transducer: 4C2
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | | Mode of Operation | | | | | | |
|-----------------------------|--------------------------------------------------|-------------------|------|------|-----|----------------|-----------------|--------|
| General<br>(Track I Only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppa | Comb.<br>Modesb | Otherc |
| Ophthalmic | Ophthalmic | | | | | | | |
| | Fetald | P1,4 | P2,4 | P2,4 | | P2,4 | P2,4 | P2,4 |
| | Abdominald: | P1,4 | P2,4 | P2,4 | N | P2,4 | P2,4 | P2,4 |
| | Intra-operative (Spec.)d,e | | | | | | | |
| | Intra-operative (Neuro) | | | | | | | |
| | Laparoscopic | | | | | | | |
| Fetal<br>Imaging<br>& Other | Pediatricd: | P1,4 | P2,4 | P2,4 | | P2,4 | P2,4 | P2,4 |
| | Small Organ (Thyroid,<br>Breast, Testes, etc.)d: | | | | | | | |
| | Neonatal Cephalicd: | | | | | | | |
| | Adult Cephalicd: | | | | | | | |
| | Trans-rectalf: | | | | | | | |
| | Trans-vaginalg: | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | |
| | Musculo-skel. (Convent.)d: | | | | | | | |
| | Musculo-skel. (Superfic)d: | | | | | | | |
| | Intra-luminal | | | | | | | |
| | Other (Specify) | | | | | | | |
| Cardiac | Cardiac Adult | N1 | N2 | N2 | N | N2 | N2 | N2 |
| | Cardiac Pediatric | N1 | N2 | N2 | N | N2 | N2 | N2 |
| | Trans-esoph. (Cardiac) | | | | | | | |
| | Other (Specify) | | | | | | | |
| Peripheral<br>Vessel | Peripheral vesseld: | | | | | | | |
| | Other (Specify) | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
' Includes Color Doppler (CD), Directional Power Doppler (DPD), and (non-directional) Power Doppler.
0 B+M; B+PWD; B+CD; B+DPD; B+PD.
° Harmonic Imaging (HI)
4 Includes ultrasound guidance for placement of needles, catheters.
Abdominal organs and peripheral vessel.
Includes ultrasound guidance for placement of needles, catheters, cryosurgery, and brachytherapy
9 Includes ultrasound guidance of transvaginal biopsy, infertility monitoring of follicle development.
" Includes guidance of amniocentesis, infertility monitoring of follicle development.
System uses previously cleared under K992505 with 3 MHz Model L3 (Linear).
3 System uses previously cleared under K992505
2 System uses previously cleared under K012191.
System uses previously cleared under K012103183.
* System uses previously cleared under K030191.
System uses previously cleared under K040840.
· System uses previously cleared under K043278.
Includes uses in military field settings in addition to hospital/clinic settings. (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of Center for Devices and Radiological Health, Office of Device Evaluation Prescription Use (Per 21 CFR 801.109)
Rach Phillips
(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number _
{7}------------------------------------------------
TERASON ™ Ultrasound System System:
Transducer: 4V2
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | | | | | | | |
|-----------------------------|--------------------------------------------------|------|------|------|-----|----------------|----------------|--------|
| General<br>(Track I Only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppa | Comb<br>Modesb | Otherc |
| Ophthalmic | Ophthalmic | | | | | | | |
| | Fetalh | P1,4 | P2,4 | P2,4 | | P2,4 | P2,4 | P2,4 |
| | Abdominald: | P1,4 | P2,4 | P2,4 | N | P2,4 | P2,4 | P2,4 |
| | Intra-operative (Spec.)a,e | N4 | N4 | N4 | | N4 | N4 | N4 |
| | Intra-operative (Neuro) | | | | | | | |
| | Laparoscopic | | | | | | | |
| Fetal<br>Imaging<br>& Other | Pediatricc: | P1,4 | P2,4 | P2,4 | N | P2,4 | P2,4 | P2,4 |
| | Small Organ (Thyroid,<br>Breast, Testes, etc.)d: | | | | | | | |
| | Neonatal Cephalicc: | P1 | P2 | P2 | N | P2 | P2 | P2 |
| | Adult Cephalicc: | P1 | P2 | P2 | N | P2 | P2 | P2 |
| | Trans-rectalf: | | | | | | | |
| | Trans-vaginalf: | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | |
| | Musculo-skel. (Convent.)d: | | | | | | | |
| | Musculo-skel. (Superfic)d: | | | | | | | |
| | Intra-luminal | | | | | | | |
| | Other (Specify) | | | | | | | |
| Cardiac | Cardiac Adult | P1 | P2 | P2 | N | P2 | P2 | P2 |
| | Cardiac Pediatric | P1 | P2 | P2 | N | P2 | P2 | P2 |
| | Trans-esoph. (Cardiac) | | | | | | | |
| | Other (Specify) | | | | | | | |
| Peripheral<br>Vessel | Peripheral vesseld: | | | | | | | |
| | Other (Specify) | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
Includes Color Doppler (CD), Directional Power Doppler (DPD), and (non-directional) Power Doppler.
B+M; B+PWD; B+CD; B+DPD; B+PD.
°Harmonic Imaging (HI)
4 Includes ultrasound guidance for placement of needles, catheters.
Abdominal organs and peripheral vessel.
Includes ultrasound guidance for placement of needles, catheters, cryosurgery, and brachytherapy
4 Includes ultrasound guidance of transvaginal biopsy, infertility monitoring of follicle development.
" Includes guidance of amniocentesis, infertility monitoring of follicle development.
System uses previously cleared under K992505 with 3 MHz Model L3 (Linear).
3 System uses previously cleared under K992505 with 3 MHz Model L3 (Linear).
2 System uses previously cleared under K012191.
3 System uses previously cleared under K010883.
"System uses previously cleared under K010003.
* System uses previously cleared under K030191.
System uses previously cleared under K0408441.
* System uses previously cleared under K040844.
* System uses previously cleared under K040830.
* System uses previously cleared under K043278.
Includes uses in military field settings in addition to hospital/clinic settings.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of Center for Devices and Radiological Health, Office of Device Evaluation Prescription Use (Per 21 CFR 801.109)
Rata Phillips
(Division Sign-Off) Division of Reproductive, Abdominal and Radiological Devices 510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.