55 SV, MULTIFOCAL AND TORIC (OCUFILCON C) SOFT (HYDROPHILIC) CONTACT LENS FOR DAILY WEAR
K051184 · United Contact Lens, Inc. · LPL · Jul 1, 2005 · Ophthalmic
Device Facts
Record ID
K051184
Device Name
55 SV, MULTIFOCAL AND TORIC (OCUFILCON C) SOFT (HYDROPHILIC) CONTACT LENS FOR DAILY WEAR
Applicant
United Contact Lens, Inc.
Product Code
LPL · Ophthalmic
Decision Date
Jul 1, 2005
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.5925
Device Class
Class 2
Indications for Use
The 55 SV (ocufilcon C) Soft (Hydrophilic) Contact Lens for Daily Wear is indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia) in aphakic or not-aphakic persons with non-diseased eyes that may exhibit refractive and/or corneal astigmatism up to 2.00 diopters that does not interfere with visual acuity. The 55 Multifocal (ocufilcon C) Soft (Hydrophilic) Contact Lens for Daily Wear is indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia) and presbyopia in aphakic and not-aphakic persons with non-diseased eyes that may exhibit refractive and/or corneal astigmatism up to 2.00 diopters that does not interfere with visual acuity. The 55 Toric (ocutilcon C) Soft (Hydrophilic) Contact Lens for Daily Wear is indicated for daily wear for the correction of refractive ametropia (myopia, hyperopia and astigmatism) in aphakic and/or not aphakic persons with non diseased eyes. The lenses may be disinfected using chemical or hydrogen peroxide disinfecting systems. Eye care practitioners may prescribe the lenses for daily wear and/or frequent replacement. When prescribed for a Frequent Replacement Program, the lenses may be disinfected using chemical or hydrogen peroxide disinfecting systems.
Device Story
Hemispherical flexible hydrophilic contact lenses; cover cornea and adjacent sclera. Available in single vision, double slab-off back surface (toric), and aspheric (multifocal) designs. Prescribed by eye care practitioners for daily wear or frequent replacement. Patients use lenses to correct refractive errors; disinfection via chemical or hydrogen peroxide systems. Benefits include vision correction for ametropia and presbyopia.
Clinical Evidence
Bench testing only. Pre-clinical testing included systemic injection, primary ocular irritation, cytotoxicity, and physicochemical testing. Results confirmed non-toxicity, non-irritation, and stability of physical properties. Shelf life testing confirmed sterility and physical property stability.
Technological Characteristics
Material: ocufilcon C (HEMA and methacrylic acid cross-linked with ethyleneglycol dimethacrylate, 55% water). Tint: Reactive Blue 19. Designs: single vision, double slab-off (toric), aspheric (multifocal). Manufacturing: molded. Sterilization: not specified.
Indications for Use
Indicated for daily wear correction of myopia, hyperopia, astigmatism, and presbyopia in aphakic or non-aphakic persons with non-diseased eyes.
Regulatory Classification
Identification
A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.
Predicate Devices
55 F (ocufilcon C) Soft (Hydrophilic) Single Vision Contact Lens (P920008/K941607)
55 Multifocal (ocufilcon C) Soft (Hydrophilic) Contact Lens (P920008/K941607)
{0}------------------------------------------------
JUL 1 - 2005
K051184
Image /page/0/Picture/2 description: The image shows the words "United Contact Lens" in a stylized font. Below the words "United Contact Lens" is a black box with the words "SOFT CONTACT LENSES" in white font. The words "United Contact Lens" are stacked on top of each other, with "United" on top, "Contact" in the middle, and "Lens" on the bottom.
19111 - 615 Avenue NE, Suite 5. Arlington, Washington 98223 Office: 425-743-7343 Fax: 425-743-8795 Toll Free: 800-446-1666
### 510(k) Summary
#### Submitter Information:
United Contact Lens Inc. 19111 61st Ave. NE #5 Arlington, WA 98223 Registration No. 2918644
| Contact Person: | Garold L. Edwards, O.D., F.A.A.O. |
|-----------------|-----------------------------------|
| | Regulatory Consultant |
| Telephone: | (408) 221 - 3860 |
| Fax: | (425) 743 - 8795 |
May 4, 2005 Date Prepared:
### Device Name:
| Common Name: | Soft (Hydrophilic) Contact Lens |
|--------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade/Proprietary Names: | 55 SV (ocufilcon C) Soft<br>(Hydrophilic) Contact Lens for Daily Wear<br><br>55 Multifocal (ocufilcon C) Soft (Hydrophilic)<br>Contact Lens for Daily Wear<br><br>55 Toric (ocufilcon C) Soft (Hydrophilic)<br>Contact Lens for Daily Wear |
| Classification Name: | Soft (Hydrophilic) Contact Lens |
| Device Classification: | Class II (21 CFR 886.5925) |
### Predicate Devices:
The 55 F (ocufilcon C) Soft (Hydrophilic) Single Vision Contact Lens, the 55 Multifocal (ocufilcon C) Soft (Hydrophilic) Contact Lens and the 55 Toric Contact Lens were selected as the predicate devices manufactured in the same facility, using the same formulation, under the same quality system packaging and sterilization processes as the subject devices.
{1}------------------------------------------------
# Description of Devices:
The 55 SV, 55 Multifocal, and the 55 Toric (ocufilcon C) Soft (Hydrophilic) Daily Wear Contact Lenses are hemispherical flexible shells which cover the cornea and a portion of the adjacent sclera. The 55 Contact Lens is available in a single vision lens design, the 55 Toric Contact Lens is available in a double slab-off back surface design and the 55 Multifocal Contact Lens is available in an aspheric lens design. The lens material (ocufilcon C) is a hydrophilic polymer of 2-hydroxyethyl methacrylate (HEMA) and methacrylic acid cross-linked with ethyleneglycol dimethacrylate (45.0%) and water (55.0%). Lenses are tinted using the color additive Reactive Blue 19.
| PARAMETER | 55 SV, 55 Toric and 55<br>Multifocal Soft<br>(hydrophilic) Contact<br>Lenses for Daily Wear | 55 F, Toric 55 and Multifocal<br>(ocufilcon C) Soft<br>(hydrophilic) Contact Lenses<br>for Daily Wear |
|--------------------------------|------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------|
| Submission number | N/A | P920008/K941607 |
| Material | ocufilcon C | ocufilcon C |
| Material classification | Hydrophilic Lens Group 4 | Hydrophilic Lens Group 4 |
| Indication for use | myopia, hyperopia,<br>astigmatism and<br>presbyopia | myopia, hyperopia,<br>astigmatism and presbyopia |
| Water content | 54.9% | 54.4% |
| Visible light<br>transmittance | 98.9% | 98.3% |
| Dk (35° C) | $18.77 x 10^{-11}$ | $18.78 x 10^{-11}$ |
| Powers | +12.00 D to -20.00 D;<br>Continuous add power to<br>+3.00 (55 Multifocal only);<br>Cylinder powers -0.50 D to<br>-2.50 D (55 Toric only) | +20.00D to -20.00 D<br>Continuous add power to<br>+3.00 (55 Multifocal only);<br>Cylinder powers -0.50 D to<br>6.00D (Toric 55 only) |
| Color | Blue visibility,<br>Reactive Blue #19 | clear |
| Refractive index | 1.415 | 1.414 |
| Method of<br>manufacture | Molded | Lathed |
# Comparison to Predicate Device
{2}------------------------------------------------
### Indications for Use:
The 55 SV (ocufilcon C) Soft (Hydrophilic) Contact Lens is indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia) in aphakic or notaphakic persons with non-diseased eyes that may exhibit refractive and/or corneal astigmatism up to 2.00 diopters that does not interfere with visual acuity.
The 55 Multifocal (ocufilcon C) Soft (Hydrophilic) Contact Lens is indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia) and presbyopia in aphakic or not-aphakic persons with non-diseased eyes that may exhibit refractive and/or corneal astigmatism up to 2.00 diopters that does not interfere with visual acuity.
The 55 Toric (ocufilcon C) Soft (Hydrophilic) Contact Lens is inclicated for daily wear for the correction of refractive ametropia (myopia, hyperopia and astigmatism) in aphakic and/or not aphakic persons with non- diseased eyes.
The lenses may be disinfected using chemical or hydrogen peroxide disinfecting systems. Eye care practitioners may prescribe the lenses for daily wear and/or frequent replacement. When prescribed for a Frequent Replacement Program, the lenses may be disinfected using chemical or hydrogen peroxide disinfecting systems.
### Description of Safety and Substantial Equivalence:
A series of pre-clinical tests were performed to demonstrate the safety and effectiveness of the 55 SV, the 55 Multifocal, and the 55 Toric (ocufilcon C) Soft (Hydrophilic) Contact Lenses for Daily Wear, and to establish substantial equivalence to the predicate devices.
Results of Systemic Injection, Primary Ocular Irritation and Cytotoxicity Tests show the lenses to be non-toxic and non-irritating. The 55 SV lenses were extracted and evaluated for presence of residue. Results showed no evidence of unsafe amounts of residue in the extracts. Physicochemical testing of the 55 lenses demonstrated equivalency to the predicate devices.
## Conclusion:
Information submitted in the 510(k) establishes that the 55 SV, the 55 Multifocal and the 55 Toric Contact Lenses (ocufilcon C) have comparable physicochemical properties to the predicate devices and do not raise questions of safety and effectiveness. Shelf life testing has shown the lenses remain sterile and that the physical lens properties are stable. Therefore, the devices are substantially equivalent to the predicate devices.
{3}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image contains the text 'Public Health Service'. The text is in a simple, bold font and is horizontally aligned. The background is plain and white, which makes the text stand out.
Image /page/3/Picture/2 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" arranged around the top half of the circle. Inside the circle is a stylized image of an eagle or bird-like figure, with three curved lines representing the body and wings.
JUL 1 - 2005
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
United Contact Lens, Inc. c/o Garold L. Edwards, O.D., F.A.A.O Regulatory Consultant 2091 Upper Scenic Drive Felton, CA 95018
Re: K051184
Trade/Device Name:
55 SV (ocufilcon C) Soft (hydrophilic) Contact Lens for Daily Wear 55 Multifocal (ocufilcon C) Soft (hydrophilic) Contact Lens for Daily Wear 55 Toric (ocufilcon C) Soft (hydrophilic) Contact Lens for Daily Wear Regulation Number: 21 CFR 886.5925 Regulation Name: Soft (hydrophilic) Contact Lens Regulatory Class: Class II Product Code: LPL Dated: May 4, 2005 Received: May 9, 2005
Dear Mr. Edwards:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
#### Page 2 - Garold L. Edwards, O.D., F.A.A.O
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 827-8910. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
David M. Whipple
David M. Whipple Acting Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
### INDICATIONS STATEMENT
#### Device Names:
55 SV (ocufilcon C) Soft (Hydrophilic) Contact Lens for Daily Wear
55 Multifocal (ocufilcon C) Soft (Hydrophilic) Contact Lens for Daily Wear
55 Toric (ocufilcon C) Soft (Hydrophilic) Contact Lens for Daily Wear
### Indications for Use:
The 55 SV (ocufilcon C) Soft (Hydrophilic) Contact Lens for Daily Wear is indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia) in aphakic or not-aphakic persons with non-diseased eyes that may exhibit refractive and/or corneal astigmatism up to 2.00 diopters that does not interfere with visual acuity.
The 55 Multifocal (ocufilcon C) Soft (Hydrophilic) Contact Lens for Daily Wear is indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia) and presbyopia in aphakic and not-aphakic persons with non-diseased eyes that may exhibit refractive and/or corneal astigmatism up to 2.00 diopters that does not interfere with visual acuity.
The 55 Toric (ocutilcon C) Soft (Hydrophilic) Contact Lens for Daily Wear is indicated for daily wear for the correction of refractive ametropia (myopia, hyperopia and astigmatism) in aphakic and/or not aphakic persons with non diseased eyes.
The lenses may be disinfected using chemical or hydrogen peroxide disinfecting systems. Eye care practitioners may prescribe the lenses for daily wear and/or frequent replacement. When prescribed for a Frequent Replacement Program, the lenses may be disinfected using chemical or hydrogen peroxide disinfecting systems.
Prescription Use X
Over-the-Counter Use_
06 001
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH. O
OR
n of Ophthalmic Ear, and Throat Devises
K051184 510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.